UROTAC BONE ANCHOR INSERTER

K962372 · Influence, Inc. · HXJ · Aug 19, 1996 · Orthopedic

Device Facts

Record IDK962372
Device NameUROTAC BONE ANCHOR INSERTER
ApplicantInfluence, Inc.
Product CodeHXJ · Orthopedic
Decision DateAug 19, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.4540
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The UroTac Bone Anchor System is intended for soft tissue fixation to the pubic bone by means of bone anchors threaded with suture. The UroTac Bone Anchor System is indicated for the treatment of stress type (female) urinary incontinence resulting from urethral hypermobility.

Device Story

System consists of Nitinol bone anchor and stainless steel inserter; used for soft tissue-to-bone fixation during bladder neck suspension procedures. Anchor inserted into pubic bone medulla; locks upon application of pull force to attached USP Class I suture. Operated by surgeons in clinical/OR setting. Provides mechanical fixation to treat female stress urinary incontinence caused by urethral hypermobility. Benefits include secure soft tissue attachment to bone to restore urethral support.

Clinical Evidence

Bench testing only. Mean failure load tested per FDA Guidance Document for Testing Bone Anchor Devices (Draft Nov 1, 1993). Results demonstrate fixation strength substantially equivalent to Mitek GII predicate.

Technological Characteristics

Anchor material: Nitinol. Inserter material: Stainless Steel (ASTM F899-94). Mechanical fixation principle: anchor locks into pubic bone medulla via pull force on suture. Non-powered, manual surgical instrument.

Indications for Use

Indicated for female patients with stress urinary incontinence resulting from urethral hypermobility requiring soft tissue fixation to the pubic bone.

Regulatory Classification

Identification

An orthopedic manual surgical instrument is a nonpowered hand-held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery. This generic type of device includes the cerclage applier, awl, bender, drill brace, broach, burr, corkscrew, countersink, pin crimper, wire cutter, prosthesis driver, extractor, file, fork, needle holder, impactor, bending or contouring instrument, compression instrument, passer, socket positioner, probe, femoral neck punch, socket pusher, reamer, rongeur, scissors, screwdriver, bone skid, staple driver, bone screw starter, surgical stripper, tamp, bone tap, trephine, wire twister, and wrench.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} AUG 19 1996 Influence, Inc. 637 Steiner Street San Francisco, California 94117 Tel. 415-552-4303/Fax. 415-552-4309 K962372 510(k) SUMMARY Prepared: June 18, 1996. This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Section 807.92. ## SUBMITTER | NAME | Influence, Inc. | | --- | --- | | ADDRESS | 637 Steiner Street San Francisco, California 94117 | | TEL.#/FAX# | 415-552-4303 / 415-552-4309 | | CONTACT | Peter A. Bick, MD President | NAME OF DEVICE Consisting of: - **UroTac Bone Anchor System** TRADE NAME: UroTac Bone Anchor COMMON NAME: Device for soft tissue to bone attachment. CLASSIFICATION NAME: Staple, Fixation, Bone TRADE NAME: UroTac Bone Anchor Inserter COMMON NAME: Instrument used to insert bone anchor in desired situs. CLASSIFICATION NAME: Surgical Instrument - Staple Driver ## PREDICATE DEVICE - MITEK GII Bone Anchor (and accessories) (K920213) - MICROVASIVE (VESICA) Bone Anchor (K932925) ## DESCRIPTION OF THE DEVICE ### FUNCTION Fixation of non-absorbable suture below bone surface for soft tissue fixation. UroTac Bone Anchor System WDC - 65853/2 - 0292817 02 {1} # DEVICE DESIGN Suture anchor constructed of Nitinol designed to lock into medulla of the pubic bone when pull force is applied to the suture attached to the anchor. # MATERIAL Anchor - Nitinol Inserter - Stainless Steel (ASTM F899-94). # INTENDED USE The UroTac Bone Anchor System is intended for soft tissue fixation to the pubic bone by means of bone anchors threaded with suture. The UroTac Bone Anchor System is indicated for the treatment of stress type (female) urinary incontinence resulting from urethral hypermobility. # COMPARISON TO PREDICATE DEVICES The predicate device is the Mitek GII Anchor. Equivalent features include: use in connection with bladder neck suspension procedure to accomplish soft tissue fixation to the pubic bone; insertion of anchor into pubic bone; use of an inserter to set the anchor into the pubic bone; use of Nitinol as an essential material; use of USP Class I suture; and initial fixation strengths that are substantially equivalent. Microvasive’s Vesica Bone Anchor is a predicate device by virtue of the similarity of certain aspects of the bladder neck suspension procedure. # DESCRIPTION OF NON CLINICAL TESTS The mean failure load of the UroTac Bone Anchor, when pulled from pubic bone, has been tested in a manner consistent with the Guidance Document for Testing Bone Anchor Devices (Draft November 1, 1993). # CONCLUSIONS FROM TESTS The mean fixation strength of the device was demonstrated to be substantially equivalent to that of the Mitek predicate device. The tests did not yield any data suggesting safety or effectiveness issues which are other than those presented by the Mitek predicate device or the Microvasive bladder neck suspension procedure using the Vesica Bone Anchor. UroTac Bone Anchor System 01DC - 65853/2 - 0292817 02
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...