K Number
K963288

Validate with FDA (Live)

Device Name
LM BONE ANCHOR
Date Cleared
1996-11-13

(84 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Bladder neck suspension for female urinary incontinence due to urethral hypermobility. Fixation in pubic bone for the purpose of bladder neck suspension using USP #2 Class I, non-absorbable, synthetic suture (e.g., polypropylene). A minimum of two LM Anchors should be used in this (modified MMK, Raz, or Stamey, i.e., Benderev) procedure.

Device Description

Made from surgical grade stainless steel (316L) or titanium (6A1-4V ELI), the LM Anchor Products are designed with a crown and a center pin through which suture is passed to provide a means for soft tissue to bone attachment.

AI/ML Overview

This 510(k) summary describes a device, the LM Anchor Products, not an AI/ML powered device. Therefore, many of the requested categories related to AI performance, such as sample sizes for test and training sets, expert qualifications, and MRMC studies, are not applicable.

However, I can extract the available performance information from the provided text for the medical device.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Substantial equivalence in mean pullout strength to Mitek anchors.The mean pullout strength of the LM Anchor Products was substantially equivalent to the mean pullout strength of the Mitek anchors.

2. Sample size used for the test set and the data provenance

  • Sample Size: Not explicitly stated as a number of devices or subjects. The document mentions "cadaver specimen," implying human cadaveric tissue was used.
  • Data Provenance: "Pre-clinical testing in cadaver specimen." This indicates a laboratory study using cadavers, not live human subjects. The country of origin is not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is a mechanical performance test, not a diagnostic or interpretative one that would require expert ground truth establishment in the traditional sense. The "ground truth" here is the measured pullout strength.

4. Adjudication method for the test set

Not applicable. This is a direct measurement of pullout strength. No adjudication method typically applies to such biomechanical testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a non-AI mechanical device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is a non-AI mechanical device. The test measures the device's inherent mechanical property.

7. The type of ground truth used

The "ground truth" in this context is the measured pullout strength of both the LM Anchor Products and the predicate Mitek anchors, obtained through pre-clinical cadaveric testing.

8. The sample size for the training set

Not applicable. This is a hardware device; there is no "training set" in the AI/ML sense.

9. How the ground truth for the training set was established

Not applicable. There is no training set. The "ground truth" (measured pullout strength) for the performance evaluation was established through direct experimental measurement on cadaveric specimens.

{0}------------------------------------------------

NOV 1 3 1995

510(K) SUMMARY

长963288

Device Sponsor:Li Medical, 4 Armstrong Road, Shelton, CT 06484,203-944-2800
Contact:Rhodemann Li, Vice President
Date:October 30, 1996
Classification Name:Staple, Fixation, Bone
Common Name:Bone anchor
Proprietary Name:LM Anchor Products or other proprietary names
Predicate Devices:Mitek Surgical Products GII QuickAnchor (K904436)Li Medical Bone Anchor (K951233), (K951472),(K953764), (K960448), (K960439), (K960825)
Device Description:Made from surgical grade stainless steel (316L) or titanium(6A1-4V ELI), the LM Anchor Products are designed witha crown and a center pin through which suture is passed toprovide a means for soft tissue to bone attachment.
Intended Use:(UltraFix RC,UltraFix Mini,UltraFix)Bladder neck suspension for female urinary incontinencedue to urethral hypermobility. Fixation in pubic bone forthe purpose of bladder neck suspension using USP #2 ClassI, non-absorbable, synthetic suture (e.g., polypropylene). Aminimum of two LM Anchors should be used in this(modified MMK, Raz, or Stamey, i.e., Benderev)procedure.
Technical Comparison:The LM Anchor Products are similar to the Mitek anchorsin its cylindrical shape however, the bony purchase isaccomplished by the LM Anchor Products throughcantilevered beams versus nitinol arcs with the Mitekanchors.
Performance Data:Pre-clinical testing in cadaver specimen showed thatthe mean pullout strength of the LM Anchor Products weresubstantially equivalent to the mean pullout strength ofthe Mitek anchors

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.