AVS® AL and ALign PEEK Spacers, AVS® PL and UniLIF PEEK Spacers, AVS® TL PEEK Spacer, AVS® Navigator PEEK Spacer, AVS® ARIA PEEK Spacer, AccuLIF TL and PL Cage, AVS® Anchor-L Spacer, Aero-AL Lumbar Cage System
Applicant
Stryker Corporation
Product Code
OVD · Orthopedic
Decision Date
Jan 26, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Stryker Spine AVS® AL and AVS® ALign PEEK Spacers are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy. The AVS® AL and AVS® ALign PEEK Spacers are to be implanted via anterior or anterolateral approach. The AVS® AL and AVS® ALign PEEK Spacers are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine.
Device Story
Intervertebral body fusion devices (spacers/cages) designed to facilitate spinal fusion; implanted via anterior, anterolateral, posterior, or posterolateral approaches. Devices feature hollow geometries with graft windows for bone graft containment; serrated surfaces for fixation; radiopaque tantalum markers for visualization. Materials include PEEK OPTIMA LT1, titanium alloy (Ti-6Al-4V), stainless steel, and silicone rubber. Used in clinical settings by surgeons to treat degenerative disc disease; intended to be used with supplemental internal spinal fixation (e.g., pedicle screw/rod systems) to provide stability during fusion. Some models (AccuLIF, Aero-AL, Anchor-L) include integrated fixation mechanisms. Output is structural support for the intervertebral space; clinical benefit is stabilization of the spinal segment to promote arthrodesis.
Clinical Evidence
No new clinical or bench testing was performed. Substantial equivalence is supported by published clinical literature demonstrating that the use of allogenic bone graft with these interbody fusion devices does not adversely affect performance or represent a new worst-case scenario.
Technological Characteristics
Materials: PEEK OPTIMA LT1, titanium alloy (Ti-6Al-4V), stainless steel (316 LVM), silicone rubber (MED-4870), and tantalum markers. Design: Hollow cages with graft windows, serrated superior/inferior surfaces, various lordotic angles/heights. AccuLIF models are expandable. Aero-AL features a titanium jacket. Sterilization: Not specified. Connectivity: None.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, including those with up to Grade I spondylolisthesis or retrolisthesis, who have failed six months of nonoperative therapy. Used as an adjunct to fusion with autograft and/or allogenic bone graft (cancellous/corticocancellous).
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K240830 — Reef TO/TA System; Regatta Lateral System; Explorer TO System; WaveForm C Interbody System; WaveForm TO Interbody System; WaveForm TA Interbody System; FORZA XP Expandable Spacer System; Shoreline ACS Interbody System; Shoreline RT Interbody System; Meridian Interbody System; WaveForm A Interbody System · Orthofix Medical, Inc. · Jun 20, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 26, 2015
Stryker Corporation Garry T. Hayeck, Ph.D. Senior Regulatory Affairs Specialist Stryker Spine 2 Pearl Court Allendale, New Jersey 07401
Re: K143163
Trade/Device Name: AVS® AL and AVS® ALign PEEK Spacers, AVS® PL and AVS® UniLIF PEEK Spacers, AVS® TL PEEK Spacer, AVS® Navigator PEEK Spacer, AVS® ARIA PEEK Spacer, AccuLIF TL and PL Cage, AVS® Anchor-L Spacer, and Aero™-AL Lumbar Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD, MAX Dated: October 31, 2014 Received: November 3, 2014
Dear Dr. Hayeck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
Page 2 - Garry T. Hayeck, Ph.D.
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K143163
Device Name AVS® AL and ALign PEEK Spacers
#### Indications for Use (Describe)
The Stryker Spine AVS® AL and AVS® ALign PEEK Spacers are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy.
The AVS® AL and AVS® ALign PEEK Spacers are to be implanted via anterior or anterolateral approach.
The AVS® AL and AVS® ALign PEEK Spacers are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known) K143163
Device Name AVS® PL and UniLIF PEEK Spacers
#### Indications for Use (Describe)
The Stryker Spine AVS® PL and AVS® UniLIF™ PEEK Spacers are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy.
The AVS® PL PEEK Spacers and AVS® UniLIF™ PEEK Spacers are to be implanted via posterior approach.
The AVS® PL PEEK Spacers and AVS® UniLIF™ PEEK Spacers are intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems).
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
{4}------------------------------------------------
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known) K143163
Device Name AVS® TL PEEK Spacer
#### Indications for Use (Describe)
The Stryker Spine AVS® TL PEEK Spacers are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy.
The AVS® TL Peek Spacers are to be implanted via posterior approach.
The AVS® TL PEEK Spacers are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems).
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{6}------------------------------------------------
510(k) Number (if known) K143163
Device Name AVS® Navigator PEEK Spacer
#### Indications for Use (Describe)
The Stryker Spine AVS® Navigator PEEK Spacers are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy.
The AVS® Navigator PEEK Spacers are to be implanted via a posterolateral approach.
The AVS® Navigator PEEK Spacers are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
{7}------------------------------------------------
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{8}------------------------------------------------
510(k) Number (if known) K143163
Device Name AVS® ARIA PEEK Spacer
#### Indications for Use (Describe)
The Stryker Spine AVS® ARIA PEEK Spacers are intervertebral body fusion devices indicated for use with autograff and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy.
The AVS® ARIA PEEK Spacers are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine.
## Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{9}------------------------------------------------
510(k) Number (if known) K143163
Device Name AccuLIF TL and PL Cage
#### Indications for Use (Describe)
The AccuLIF TL and PL Cage are indicated for intervertebral body fusion with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have completed six months of non-operative treatment. The AccuLIF TL and PL Cages are always to be used with supplemental internal spinal fixation. Additionally, the AccuLIF TL and PL Cages are to be used with autograft and/or allogenic bone graft comprised of cancellous bone graft when the subject device is used as an adjunct to fusion.
## Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{10}------------------------------------------------
510(k) Number (if known) K143163
Device Name AVS® Anchor-L Spacer
#### Indications for Use (Describe)
The Stryker Spine AVS® Anchor-L is an intervertebral body fusion device indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy.
The AVS® Anchor-L Lumbar Cage system is to be implanted via an open, anterior approach.
The AVS® Anchor-L Lumbar Cage system may be used as a stand-alone device or in conjunction with supplemental fixation. When used as a stand-alone device, the AVS® Anchor-L Lumbar Cage must be used with the internal screw and plate fixation provided by AVS® Anchor-L Fixation Screws and Locking Plate. If AVS® Anchor-L is used with less than three or none of the provided screws, then additional supplemental fixation that has been cleared by the FDA for use in the lumbar spine must be used to augment stability. The accompanying Locking Plate must be vice is used with any number of screws.
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
{11}------------------------------------------------
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{12}------------------------------------------------
510(k) Number (if known) K143163
Device Name Aero™-AL Lumbar Cage System
#### Indications for Use (Describe)
The Stryker Spine Aero™-AL is an intervertebral body fusion device indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy.
The Aero™-AL Lumbar Cage System is to be implanted via an anterior approach.
The Aero™-AL Lumbar Cage System is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (e.g., posterior pedicle screw and rod systems) in addition to the included fixation anchors.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
ال Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
{13}------------------------------------------------
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{14}------------------------------------------------
| 510(k) Summary: Expanded Indications for Use | | |
|---------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Stryker Spine Lumbar Intervertebral Body Fusion Devices | | |
| Submitter | Stryker Spine<br>2 Pearl Court<br>Allendale, NJ 07401 | |
| Contact Person | Garry T. Hayeck, Ph.D.<br>Senior Regulatory Affairs Specialist<br>Phone: 201-760-8043<br>Fax: 201-962-4043<br>E-mail: garry.hayeck@stryker.com | |
| Date Prepared | January 22, 2015 | |
| Trade Name | 1. AVS® AL and AVS® ALign PEEK Spacers<br>2. AVS® PL and AVS® UniLIF PEEK Spacers<br>3. AVS® TL PEEK Spacer<br>4. AVS® Navigator PEEK Spacer<br>5. AVS® ARIA PEEK Spacer<br>6. AccuLIF TL and PL Cage<br>7. AVS® Anchor-L Spacer<br>8. AeroTM-AL Lumbar Cage System | |
| Common Name | 1. AVS® AL and AVS® ALign PEEK Spacers<br>Intervertebral fusion device with bone graft, lumbar<br>2. AVS® PL and AVS® UniLIF PEEK Spacers<br>Intervertebral fusion device with bone graft, lumbar<br>3. AVS® TL PEEK Spacer<br>Intervertebral fusion device with bone graft, lumbar<br>4. AVS® Navigator PEEK Spacer<br>Intervertebral fusion device with bone graft, lumbar<br>5. AVS® ARIA PEEK Spacer<br>Intervertebral fusion device with bone graft, lumbar<br>6. AccuLIF TL and PL Cage<br>Intervertebral fusion device with bone graft, lumbar<br>7. AVS® Anchor-L Spacer<br>Intervertebral fusion device with bone graft, lumbar<br>8. AeroTM-AL Lumbar Cage System<br>Intervertebral fusion device with integrated fixation, lumbar | |
| Proposed Class | 1. AVS® AL and AVS® ALign PEEK Spacers<br>Class II<br>2. AVS® PL and AVS® UniLIF PEEK Spacers<br>Class II<br>3. AVS® TL PEEK Spacer<br>Class II<br>4. AVS® Navigator PEEK Spacer<br>Class II | |
| | | 5. AVS® ARIA PEEK Spacer |
| | | Class II |
| | | 6. AccuLIF TL and PL Cage |
| | | Class II |
| | | 7. AVS® Anchor-L Spacer |
| | | Class II |
| | | 8. Aero™-AL Lumbar Cage System |
| | | Class II |
| Classification Name,<br>Codification | 1. | AVS® AL and AVS® ALign PEEK Spacers<br>Intervertebral body fusion device , 21 CFR § 888.3080 |
| | 2. | AVS® PL and AVS® UniLIF PEEK Spacers<br>Intervertebral body fusion device , 21 CFR § 888.3080 |
| | 3. | AVS® TL PEEK Spacer<br>Intervertebral body fusion device , 21 CFR § 888.3080 |
| | 4. | AVS® Navigator PEEK Spacer<br>Intervertebral body fusion device , 21 CFR § 888.3080 |
| | 5. | AVS® ARIA PEEK Spacer<br>Intervertebral body fusion device , 21 CFR § 888.3080 |
| | 6. | AccuLIF TL and PL Cage<br>Intervertebral body fusion device , 21 CFR § 888.3080 |
| | 7. | AVS® Anchor-L Spacer<br>Intervertebral body fusion device , 21 CFR § 888.3080 |
| | 8. | Aero™-AL Lumbar Cage System<br>Intervertebral body fusion device , 21 CFR § 888.3080 |
| Product Codes | 1. | AVS® AL and AVS® ALign PEEK Spacers<br>MAX |
| | 2. | AVS® PL and AVS® UniLIF PEEK Spacers<br>MAX |
| | 3. | AVS® TL PEEK Spacer<br>MAX |
| | 4. | AVS® Navigator PEEK Spacer<br>MAX |
| | 5. | AVS® ARIA PEEK Spacer<br>MAX |
| | 6. | AccuLIF TL and PL…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.