K Number
K221535

Validate with FDA (Live)

Date Cleared
2022-07-22

(56 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Align Lumbar Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.

The Align Lumbar Interbody Fusion System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radios.

The Align Lumbar Interbody Fusion System is intended to be used with supplemental spinal fixation system cleared for use in the lumbosacral spine.

Device Description

The Align Lumbar Interbody Fusion System consists of several models of intervertebral body fusion devices. The Align ALIF unitary cages and Align fully round ALIF (FRA) cages are intended for anterior approaches. The Align AOLIF cages are intended for anterolateral approaches. The Align Lateral cages are intended for lateral approaches. The Align TLIF cages and Align Oblique TLIF cages are intended for transforaminal approaches. The Align PLIF cages are intended for posterior approaches.

Align Lumbar Interbody Fusion System devices are made from titanium alloy (Ti-6A1-4V ELI) per ASTM F3001 Standard Specification for Additive Manufacturing Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) with Powder Bed Fusion. Align Lumbar Interbody Fusion System devices are also made from polyetheretherketone (Zeniva ZA-500 PEEK®) per ASTM F2026 Standard Specification for Polyetheretherketone (PEEK) Polymers for Surgical Implant Applications. The PEEK devices contain radiopaque markers constructed from Tantalum per ASTM F560 Standard Specification for Unalloyed Tantalum for Surgical Implant Applications to assist with intraoperative placement. The radiopaque markers are positioned so the implant can be visualized from an Anterior-Posterior (AP) and Lateral x-ray view.

Reusable instruments to support implantation of the subject device are provided with non-sterile implants in sterilization trays. Implants are also available sterile packaged.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device, the "Align Lumbar Interbody Fusion System." Within this specific type of regulatory submission to the FDA, detailed information regarding acceptance criteria and the study that proves the device meets those criteria, especially in the context of AI/ML performance, is generally not present.

The 510(k) pathway focuses on demonstrating "substantial equivalence" to a legally marketed predicate device, primarily through non-clinical testing (bench testing, biocompatibility) and comparisons of technological characteristics. Clinical data is often not required for 510(k) submissions unless the device has new indications for use, different technological characteristics that raise new safety or effectiveness questions, or if the predicate device itself required clinical data.

Based on the provided document, I can primarily address the information regarding non-clinical testing acceptance criteria and studies. The questions related to AI/ML specific performance, such as MRMC studies, expert ground truth adjudication, and training/test set details, are not applicable to this specific submission as there is no mention of AI/ML components in the "Align Lumbar Interbody Fusion System."

Here's an analysis of what is available in the document regarding acceptance criteria and studies:

1. A table of acceptance criteria and the reported device performance:

The document broadly states that the device met acceptance criteria, especially for bench testing. However, it does not provide a specific table detailing numerical acceptance criteria and corresponding reported performance values. This level of detail is typically found in the full submission, not in the publicly available 510(k) summary.

Acceptance CriteriaReported Device Performance
Biocompatibility: Complies with ISO 10993-1, ISO 10993-5, ISO 10993-11 standards.Met: Device was subjected to and complied with the listed ISO 10993 standards for biological evaluation.
Mechanical Strength (Bench Testing): Design features meet required mechanical strength criteria for intended use. Specifically, static compression, dynamic compression, static compression-shear, dynamic compression-shear (per ASTM F2077), and subsidence testing (per ASTM F2267) standards. Performance equivalence to predicate device.Met: "the design features met the required mechanical strength criteria for their intended use." "Performance equivalence was shown through the verification comparison to the predicate device." "The results on the non-clinical testing demonstrated that the subject device met the acceptance criteria of the standard and is substantially equivalent to the predicate devices."
Sterilization & Packaging Validation: Validated.Met: "The sterilization method and packaging of the subject device were validated and remain unchanged."

2. Sample size used for the test set and the data provenance:

  • Test Set (for physical device testing): Not explicitly stated in terms of a "sample size" like one would for a clinical study. For bench testing, typically a certain number of samples per configuration are tested (e.g., 6 or 12 per test type). This detail is not provided in a summary.
  • Data Provenance: The studies are laboratory-based bench tests and biocompatibility tests, not clinical data from patients. Therefore, terms like "country of origin" or "retrospective/prospective" are not applicable in this context.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Not Applicable: This submission relies on engineering and material science standards (ASTM, ISO), not expert clinical reviews for "ground truth". The "ground truth" for these tests is generally defined by the parameters and methodologies outlined in the referenced standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable: Adjudication methods like 2+1 or 3+1 are used in clinical studies, particularly for imaging interpretation, to resolve discrepancies between readers. This is not relevant for bench testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No: The document states, "No clinical data was provided to demonstrate substantial equivalence." This implies no human reader studies of any kind were performed for this submission, negating the possibility of an MRMC study related to AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable: This device is an implantable medical device (intervertebral body fusion system), not a software algorithm or AI-powered system. No standalone algorithm performance was assessed.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Bench Testing Standards & Specifications: For the non-clinical tests, the "ground truth" is adherence to the specified engineering standards (e.g., ASTM F2077, ASTM F2267) and biocompatibility standards (ISO 10993 series), and comparison to the performance of the predicate device. The goal is to show that the device performs mechanically as intended and is biocompatible, similar to existing devices.

8. The sample size for the training set:

  • Not Applicable: As this is a physical medical device and not an AI/ML system, there is no "training set."

9. How the ground truth for the training set was established:

  • Not Applicable: No training set exists for this device.

In summary: The provided document describes a 510(k) submission for a physical implantable device. The "acceptance criteria" and "proof" primarily stem from meeting established international and national standards for biocompatibility and mechanical performance, and demonstrating equivalence to predicate devices through laboratory bench testing. The questions related to AI/ML performance, clinical studies, and expert ground truth are not applicable to the information contained within this specific 510(k) summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 22, 2022

Acuity Surgical Devices, LLC Charlie Forton Director of Engineering and Regulatory Affairs 8710 N. Royal Lane Irving, Texas 75063

Re: K221535

Trade/Device Name: Align Lumbar Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: May 26, 2022 Received: May 27, 2022

Dear Charlie Forton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K221535

Device Name Align Lumbar Interbody Fusion System

Indications for Use (Describe)

The Align Lumbar Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.

The Align Lumbar Interbody Fusion System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radios.

The Align Lumbar Interbody Fusion System is intended to be used with supplemental spinal fixation system cleared for use in the lumbosacral spine.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SECTION 5: 510(k) SUMMARY

5.1 SUBMITTER

Acuity Surgical Devices LLC 8710 N. Royal Lane Irving, TX 75063 USA

Contact Person: Bryan Cowan Job Title: President Phone: 844-228-4890 Fax: 866-616-2789 Email: bcowan@acuitysurgical.com

Correspondent: Charlie Forton Job Title: Director of Engineering and Regulatory Affairs Phone: 512-585-3537 Fax: 866-616-2789 Email: cforton(@acuitysurgical.com

Date Prepared: 05/19/2022

5.2 DEVICE

Trade Name: Align Lumbar Interbody Fusion System

Common or Usual Name: Intervertebral body fusion device

Classification Name: Intervertebral fusion device with bone graft, lumbar

Regulation: 21 CFR 888.3080

Regulatory Class: II

Product Code: MAX

5.3 PREDICATE DEVICE

Predicate Device: Aurora Spine Interbody Fusion System (K133967)

Additional Predicate A-Link Z (K201671) Devices: TxTiHA IBF System & AxTiHA Stand-Alone ALIF System (K201614)

{4}------------------------------------------------

Subject DevicePrimary PredicateAdditional Predicate Devices
Align LumbarAnterior CagesAurora Spine InterbodyFusion System, K133967A-Link Z, TxTiHA IBF System & AxTiHAK201671 Stand-Alone ALIF System, K201614
Align LumbarAnterolateral CageAurora Spine InterbodyFusion System, K133967A-Link Z,K201671
Align Lumbar LateralAurora Spine InterbodyFusion System, K133967A-Link Z,K201671
Align LumbarPosterior CageAurora Spine InterbodyFusion System, K133967
Align LumbarTransforaminal andObliqueTransforaminal CageAurora Spine InterbodyFusion System, K133967TxTiHA IBF System & AxTiHA Stand-Alone ALIFSystem, K201614

5.4 DEVICE DESCRIPTION

The Align Lumbar Interbody Fusion System consists of several models of intervertebral body fusion devices. The Align ALIF unitary cages and Align fully round ALIF (FRA) cages are intended for anterior approaches. The Align AOLIF cages are intended for anterolateral approaches. The Align Lateral cages are intended for lateral approaches. The Align TLIF cages and Align Oblique TLIF cages are intended for transforaminal approaches. The Align PLIF cages are intended for posterior approaches.

Align Lumbar Interbody Fusion System devices are made from titanium alloy (Ti-6A1-4V ELI) per ASTM F3001 Standard Specification for Additive Manufacturing Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) with Powder Bed Fusion. Align Lumbar Interbody Fusion System devices are also made from polyetheretherketone (Zeniva ZA-500 PEEK®) per ASTM F2026 Standard Specification for Polyetheretherketone (PEEK) Polymers for Surgical Implant Applications. The PEEK devices contain radiopaque markers constructed from Tantalum per ASTM F560 Standard Specification for Unalloyed Tantalum for Surgical Implant Applications to assist with intraoperative placement. The radiopaque markers are positioned so the implant can be visualized from an Anterior-Posterior (AP) and Lateral x-ray view.

Reusable instruments to support implantation of the subject device are provided with non-sterile implants in sterilization trays. Implants are also available sterile packaged.

{5}------------------------------------------------

ર્સ્ટ INDICATIONS FOR USE

The Align Lumbar Interbody Fusion System cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft consisting of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion.

The Align Lumbar Interbody Fusion System cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.

The Align Lumbar Interbody Fusion System cages are intended to be used with supplemental spinal fixation system cleared for use in the lumbosacral spine.

ર્સ્વ COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Acuity Surgical Devices LLC believes that the Align Lumbar Interbody Fusion System is substantially equivalent to the predicate and reference devices based on information summarized here:

For the Align Lumbar Interbody Fusion System, all devices have similar indications for use as the devices cleared in K133967, K201671, and K201614.

The Align Lumbar anterior cages have a similar surgical approach as the devices cleared in K133967, K201671, and K201614. The device's function, type of bone graft, need for supplemental fixation, and sterilization methods of the subject device and primary predicate are the same. The subject devices have the same design materials as the devices cleared in K201671. When compared to the primary predicate device, the main differences are the extended range of cage configurations and available sizes. The dimensions of the subject devices are similar to the devices cleared in K133967, K201671, and K201614.

The Align Lumbar anterolateral cages have a similar surgical approach as the devices cleared in K133967 and K201671. The subject device has the same implant shape with graft window and threaded instrument attachment and uses the same sterilization methods as the devices cleared in K133967 and K201671. The design material and implantation anatomical site is the same for the subject device and the device cleared in K201671. When compared to the predicate and reference devices, the main differences are the extended range of dimensions and size offerings.

{6}------------------------------------------------

The Align Lumbar lateral cages have a similar surgical approach as the devices cleared in K133967 and K201671. The subject devices have the same design materials as the devices cleared in K201671. The subject device has the same implant shape with graft window and threaded instrument attachment and uses the same sterilization methods as the devices cleared in K133967 and K201671. The main differences are the extended range of cage configurations and size offerings when compared to the predicate and reference devices. The dimensions of the subject devices are similar to the devices cleared in K133967 and K201671.

The Align Lumbar transforaminal and oblique transforaminal cages have the same function, implant shape and design with graft window and threaded instrument attachment, requirement for supplement fixation, and sterilization methods as the devices cleared in K133967. The design material is similar for the subject device and primary predicate device. The subject device has an extended range of cage configurations and available sizes compared to the primary predicate device, K133967.

The Align Lumbar posterior cages have the same function, implant shape and design with graft window and threaded instrument attachment, requirement for supplement fixation, and sterilization methods as the devices cleared in K133967 and K201614. The design material is similar for the subject device and primary predicate device. The subject device has an extended range of cage configurations and available sizes compared to the devices cleared in K133967 and K201614.

In order to ensure that the different technological characteristics do not affect the safety and effectiveness of the subject device, both mechanical testing and worstcase analysis were conducted on new implant sizes and configurations. The sterilization method and packaging of the subject device were validated and remain unchanged. Based on the testing conducted, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device.

5.7 SUMMARY OF NON-CLINICAL TESTING

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility Testing

The subject device was subjected to biocompatibility testing in compliance to ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity, and ISO 10993-11 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.

{7}------------------------------------------------

Bench Testing

The subject device underwent verification evaluation (static compression, dynamic compression, static compression-shear, dynamic compression-shear per ASTM F2077 Test Methods For Intervertebral Body Fusion Devices, and subsidence testing per ASTM F2267 Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression) to ensure that the design features met the required mechanical strength criteria for their intended use.

Performance equivalence was shown through the verification comparison to the predicate device.

The results on the non-clinical testing demonstrated that the subject device met the acceptance criteria of the standard and is substantially equivalent to the predicate devices.

5.8 CLINICAL TESTING

No clinical data was provided to demonstrate substantial equivalence.

રું. વ CONCLUSIONS

Based on the testing conducted, including mechanical performance testing, design validation analysis, biocompatibility testing, and sterilization and packaging validation, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices.

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.