LIFE SPINE SACROILIAC JOINT FIXATION SYSTEM

K141246 · Life Spine, Inc. · OUR · Feb 22, 2015 · Orthopedic

Device Facts

Record IDK141246
Device NameLIFE SPINE SACROILIAC JOINT FIXATION SYSTEM
ApplicantLife Spine, Inc.
Product CodeOUR · Orthopedic
Decision DateFeb 22, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Life Spine Sacroiliac Joint Fixation System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.

Device Story

System consists of fully threaded and partially threaded cannulated screws used for sacroiliac joint fusion. Device implanted by surgeons in clinical setting to stabilize joint and promote fusion. Screws provide mechanical fixation; titanium alloy construction ensures biocompatibility and structural integrity. Output is physical stabilization of joint; clinical decision-making based on patient diagnosis of joint disruption or degenerative sacroiliitis. Benefits include joint stabilization and potential pain relief associated with fusion.

Clinical Evidence

Bench testing only. Performance testing conducted per ASTM F543 and ASTM F2191, including static and dynamic cantilever bending, screw pullout, and insertion tests. No clinical data presented.

Technological Characteristics

Titanium alloy 6AL-4V-ELI (ASTM F136). Components include fully threaded and partially threaded cannulated screws. Single-use, non-sterile. Mechanical fixation via bone screws.

Indications for Use

Indicated for patients requiring sacroiliac joint fusion due to sacroiliac joint disruptions or degenerative sacroiliitis.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 22, 2015 Life Spine, Incorporated Mr. Randy Lewis General Manager 2401 West Hassell Road, Suite 1535 Hoffman Estates, Illinois 60169 Re: K141246 Trade/Device Name: Sacroiliac Joint Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: OUR, HWC Dated: January 5, 2015 Received: January 7, 2015 Dear Mr. Lewis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ Page 2 - Mr. Randy Lewis forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Lori A. Wiggins -S - for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K141246 Device Name Sacroiliac Joint Fixation System Indications for Use (Describe) The Life Spine Sacroiliac Joint Fixation System is intended for sacroiliac joint fusion for conditions including sacrolliac joint disruptions and degenerative sacroiliitis. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary Life Spine Sacroiliac Joint Fixation System | Submitted By: | Life Spine, Inc.<br>13951 S Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 | |-------------------|-------------------------------------------------------------------------------------------------------------------------| | 510(k) Contact: | Randy Lewis<br>Life Spine<br>13951 S Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 | | Date Prepared: | February 17th, 2015 | | Trade Name: | Life Spine Sacroiliac Joint Fixation System | | Common Name: | Smooth or Threaded Metallic Bone Fixation Fastener | | Classification: | HWC, CFR 888.3040, Class II<br>OUR, CFR 888.3040, Class II | | Predicate Device: | Globus SI-LOK Sacroiliac Joint Fixation System (K112028) | #### Device Description: The Life Spine Sacroiliac Joint Fixation System consists of fully threaded screws and partially threaded cannulated screws in various diameters and lengths to enhance sacroiliac joint fusion. All components are fabricated and manufactured from titanium alloy 6AL-4V-ELI per ASTM F-136. ### Intended Use of the Device: The Life Spine Sacroiliac Joint Fixation System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis. {4}------------------------------------------------ ### Technological Characteristics: The Life Spine Sacroiliac Joint Fixation System is substantially equivalent to the predicate systems in terms of design, materials, indications for use and sizing. ### Material: The Life Spine Sacroiliac Joint Fixation System is 6AL-4V-ELI titanium manufactured according to ASTM F136. The device is comprised of a variety of non-sterile titanium, single use components. ### Performance Data: Performance testing was conducted in accordance with ASTM F543 and ASTM F2191 including static and dynamic cantilever bending, screw pullout and insertion tests. Engineering analysis is presented to demonstrate the substantial equivalence of the Life Spine Sacroiliac Joint Fixation System to the predicate device. #### Conclusion: The information presented demonstrates the substantial equivalency of the Life Spine Sacroiliac Joint Fixation System.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...