The UNITY Sacroiliac Joint Fixation System is indicated for use in skeletally mature patients for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.
The UNITY Sacroiliac Joint Fixation System consists of screws designed to enhance sacroiliac joint fusion. The UNITY Sacroiliac Joint Fixation System is offered in various diameters, lengths, and three screw types in cannulated form to accommodate patient anatomy. The three design types of the subject device are: 1. Standard Thread Screw (with and without slots) 2. Lag Screw (with and without slots) and 3. Washer Screw (with slots)
This appears to be an FDA 510(k) clearance letter and a 510(k) summary for the UNITY Sacroiliac Joint Fixation System. This type of documentation permits a device to be legally marketed, but it typically does not contain detailed information about the acceptance criteria and study results in the way you've requested for proving a device meets acceptance criteria, especially regarding AI or diagnostic performance studies.
The document states: "The UNITY Sacroiliac Joint Fixation does not create a new worst case for device performance; additional testing is not needed." This is a key statement indicating that the device is cleared based on substantial equivalence to an existing predicate device, not on new clinical performance data demonstrating its effectiveness in a traditional clinical trial or AI performance study.
Therefore, I cannot provide the requested information from this document. The questions you've asked (sample sizes for test/training sets, expert consensus, MRMC studies, standalone performance, etc.) are common for AI/diagnostic device clearances, but this particular device is a fixation system (essentially, screws for the sacroiliac joint), which falls under a different regulatory pathway.
To directly answer your points based on the provided text:
- A table of acceptance criteria and the reported device performance: Not present. The device is cleared based on equivalence to a predicate, not performance against specific criteria.
- Sample sized used for the test set and the data provenance: Not applicable. No new test set data for performance was generated as per the document.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
- Adjudication method for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI device.
- The type of ground truth used: Not applicable.
- The sample size for the training set: Not applicable.
- How the ground truth for the training set was established: Not applicable.
{0}------------------------------------------------
October 13, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the agency's acronym in a blue square, followed by the full name "U.S. Food & Drug Administration" in blue text.
Dio Medical Corporation Milan George VP of R&D 2100 Campus Lane Suite 100 East Norriton, Pennsylvania 19403
Re: K222448
Trade/Device Name: UNITY Sacroiliac Joint Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR Dated: August 12, 2022 Received: August 15, 2022
Dear Milan George:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug Administration | Form Approved: OMB No. 0910-0120Expiration Date: 06/30/2023See PRA Statement below. |
|---|---|
| ------------------------------------------------------------------------- | --------------------------------------------------------------------------------------------- |
Indications for Use
| 510(k) Number (if known) | K222448 |
|---|---|
| Device Name | UNITY Sacroiliac Joint Fixation System |
| Indications for Use (Describe) | The UNITY Sacroiliac Joint Fixation System is indicated for use in skeletally mature patients for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis. |
| Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to
review instructions, search existing data sources, gather and maintain the data needed and complete and review the
collection of information. Send comments regarding this burden estimate or any other aspect of this information
collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA)
Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a
collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (6/20) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 |
|---|---|---|
| ---------------------- | ------------- | ---------------------------------------- |
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
510(k) SUMMARY
Dio Medical UNITY Sacroiliac Joint Fixation System
| Sponsor: | Manufacturer: | Dio Medical Corp.2100 Campus Lane, Suite 100East Norriton, PA 19403 |
|---|---|---|
| Official Contact:Email:Phone: | Milan Georgemgeorge@dio-us.com1-877-394-5407 ext.102 | |
| Date Prepared: | August 12, 2022 | |
| Device Name:Common Name: | UNITY Sacroiliac Joint Fixation SystemSacroiliac Joint Fixation Screw | |
| ClassificationName: | Smooth or threaded metallic bone fixation fastener | |
| ClassificationNumber: | 21 CFR 888.3040 | |
| Product Code/Classification: | OUR, class II | |
| Description: | The UNITY Sacroiliac Joint Fixation System consists of screwsdesigned to enhance sacroiliac joint fusion. The UNITY SacroiliacJoint Fixation System is offered in various diameters, lengths, andthree screw types in cannulated form to accommodate patientanatomy. The three design types of the subject device are:1. Standard Thread Screw (with and without slots)2. Lag Screw (with and without slots) and3. Washer Screw (with slots) | |
| Indicationsfor Use: | The UNITY Sacroiliac Joint Fixation System is indicated for use inskeletally mature patients for sacroiliac joint fusion for conditionsincluding sacroiliac joint disruptions and degenerative sacroiliitis. | |
| PredicateDevice: | Huvexel Co. Ltd. - UNITY Sacroiliac Joint Fixation System(K173201) | |
| SubstantialEquivalence: | The UNITY Sacroiliac Joint Fixation System has the same intendeduse, indications for use, and principles of operation as the primarypredicate device. It has similar technological characteristics as itspredicate device. There are no technological differences betweenthe subject device and its predicate device resulting in no newissues of safety or effectiveness. Thus, the UNITY Sacroiliac JointFixation System is substantially equivalent to the predicate. | |
| PerformanceData: | The UNITY Sacroiliac Joint Fixation does not create a new worstcase for device performance; additional testing is not needed. | |
| Conclusion: | Based upon a comparison of technological characteristics,intended use, design features, and mechanical performance, theDio Medical UNITY Sacroiliac Joint Fixation System does not raiseany new safety or efficacy concerns and has demonstrated |
{4}------------------------------------------------
substantial equivalence to the identified predicates.
N/A