ENOS Software Guided External Fixation System

K253764 · Disior, Ltd. · OSN · Feb 26, 2026 · Orthopedic

Device Facts

Record IDK253764
Device NameENOS Software Guided External Fixation System
ApplicantDisior, Ltd.
Product CodeOSN · Orthopedic
Decision DateFeb 26, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The ENOS Software Guided External Fixation System application is used with the Paragon 28 Monkey Rings™ External Fixation System for the treatment of traumatic or reconstructive deformities within the indications for use of the Monkey Rings External Fixation System. It is used to generate a prescription of strut adjustments to provide to the patient.

Device Story

Cloud-based software assisting orthopedic surgeons in planning patient-specific deformity corrections for the Monkey Rings Circular External Fixation System. Surgeon inputs anatomical alignment values, hardware configuration parameters, and treatment objectives; software algorithm processes these inputs alongside mechanical hardware characteristics to calculate required strut lengths. Output is a patient-specific strut adjustment schedule (prescription) provided to the patient for manual hardware adjustment. Used in clinical settings by orthopedic surgeons to visualize bone position and determine necessary mechanical corrections. Benefits include precise, patient-specific deformity correction planning.

Clinical Evidence

No clinical data. Substantial equivalence supported by software verification and validation (unit, integration, and system levels) and non-clinical bench testing using mathematically randomized input sets covering various anatomical sites and hardware configurations.

Technological Characteristics

Cloud-based medical device software. Operates via algorithmic processing of manual surgeon inputs (anatomical alignment, hardware parameters) to compute mechanical strut adjustments. Designed for use with Monkey Rings Circular External Fixation System. Software verification conducted per FDA guidance.

Indications for Use

Indicated for patients requiring treatment of traumatic or reconstructive deformities using the Paragon 28 Monkey Rings External Fixation System.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION February 26, 2026 Disior, Ltd. % Edward Wells-Spicer Senior Regulatory Affairs Specialist Paragon 28, Inc. 14445 Grasslands Dr. Englewood, Colorado 80112 Re: K253764 Trade/Device Name: ENOS Software Guided External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: OSN Dated: November 25, 2025 Received: November 25, 2025 Dear Edward Wells-Spicer: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253764 - Edward Wells-Spicer Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K253764 - Edward Wells-Spicer Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali -S Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} Page 1 of 1 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253764 | ? | | Please provide the device trade name(s). | | ? | | ENOS Software Guided External Fixation System | | | | Please provide your Indications for Use below. | | ? | | The ENOS Software Guided External Fixation System application is used with the Paragon 28 Monkey Rings™ External Fixation System for the treatment of traumatic or reconstructive deformities within the indications for use of the Monkey Rings External Fixation System. It is used to generate a prescription of strut adjustments to provide to the patient. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} # 510(K) SUMMARY Manufacturer: Disior Ltd HTC Helsinki, Building Pinta, Tammasaarenkatu 3 Helsinki Uusimaa, FL, 00180, Finland Contact: Aarno Jussila Director of Smart 28 - Quality Phone: +358 40 673 4939 Email: aarno.jussila@zimmerbiomet.com Prepared By: Edward Wells-Spicer Senior Regulatory Affairs Specialist Paragon 28, Inc. 14445 Grasslands Dr., Englewood, CO, 80112 Phone: 720-994-5481 Date Prepared: February 25, 2025 Device Trade Name: ENOS Software Guided External Fixation System Device Class and Common Name: Class II, Software for diagnosis/treatment Classification: 888.3030 - Single/multiple component metallic bone fixation appliances and accessories Product Codes: OSN Indications for Use: The ENOS Software Guided External Fixation System application is used with the Paragon 28 Monkey Rings™ External Fixation System for the treatment of traumatic or reconstructive deformities within the indications for use of the Monkey Rings External Fixation System. It is used to generate a prescription of strut adjustments to provide to the patient. Device Description: The ENOS Software Guided External Fixation System (also known as Smart Monkey Rings) is a cloud-based medical device software system that assists orthopedic surgeons in planning patient-specific deformity corrections for use with the Monkey Rings Circular External Fixation System. The surgeon enters anatomical alignment values, hardware configuration parameters, and treatment objectives, and the software's algorithm processes these inputs together with the mechanical characteristics of the Monkey Rings hardware to generate a patient-specific strut adjustment schedule. Traditional 510(k) - K253764 {5} Traditional 510(k) - K253764 Page 2 of 3 Primary Predicate: SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software (DACS) (K190069) Additional Predicate: Orthohub External Fixator Software (K140550) Substantial Equivalence: The subject device and the primary predicate device have similar intended uses. Both the ENOS Software Guided External Fixation System and the Sixfix Hexapod primary predicate are intended to be used for traumatic (subject device) and post-traumatic injuries (primary predicate device). The subject device and the secondary predicate device are both intended for the treatment of traumatic or reconstructive deformities. The subject device and secondary predicate device have identical intended uses, except for the secondary predicate specifying reconstructive tibia deformities while the subject device is intended for reconstructive deformities within the Monkey Rings External Fixation System (K212895). The subject device, primary predicate device and the secondary predicate device generate a prescription of strut adjustments. The subject device and primary predicate have manual input methods to assist the surgeon calculating the lengths of the struts connecting the rings to manipulate the bone fragments. Also, for the subject and primary predicate, the software receives inputs from the physician and allows the physician to visualize the moving bone position. The subject and primary predicate program computes the strut lengths necessary to implement adjustments required by the surgeon. The output of the subject device is a prescription of strut adjustments, which is identical to the primary and secondary predicates. A difference between the subject device and secondary predicate is the use of a web-based software run in the cloud for the subject device and the use of a macintosh computer for the secondary predicate. Additionally, the secondary predicate utilizes x-rays in addition to manual surgeon input for calculations of recommended adjustments while the subject device is manual input only. The differences between the subject device and predicates do not introduce new issues of safety and effectiveness. Performance Testing: All necessary testing has been performed on the ENOS Software Guided External Fixation System to assure substantial equivalence to its predicate and demonstrate the subject device performs as intended. Software Verification and Validation Software verification and validation were carried out based on the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", at the unit, integration, and system levels to determine substantial equivalence to the predicate device. Non-Clinical Bench Testing Beyond Formal Verification Testing, incremental testing of software inputs against expected frame data was also performed. Given the number of anatomical sides, sites, ring combinations, and ring sizes possible, a mathematically randomized set of inputs over defined limits was selected to be used to validate that the software was developed to satisfy its intended purpose. {6} Conclusions: Clinical data are not needed to support the safety and effectiveness of the subject device. The subject device possesses similar intended uses, technological characteristics, and principles of operation as the primary predicate and secondary predicate devices. The subject device does not raise new issues of safety or effectiveness. The subject device is substantially equivalent to the legally marketed predicate devices. Traditional 510(k) - K253764 Page 3 of 3
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...