K Number
K190069

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-05-14

(119 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SixFix™ Hexapod and Deformity Analysis and Correction Software (DACS) are intended to be used for posttraumatic joint contracture which has resulted in loss of range of motion; fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction; open and closed fracture fixation; pseudo-arthrosis of long bones; limb lengthening by epiphyseal or metaphyseal distraction; correction of bony or soft tissue deformities; correction of bony or soft tissue defects; joint arthrodesis; infected fractures or nonunions.

Device Description

The SixFix™ Hexapod Fixator is a multilateral circular external fixation system. The system includes the following external fixator elements: rings, footplates, arches, struts, threaded rods, wires, external fixation accessories, and software. All of the elements are provided non-sterile and are for single use only.

The system is designed such that gradually adjusting the lengths of the struts in relation to one another alters the orientation of the rings and, consequently, the bone segments connected to the rings by half-pins and wires during the treatment period in order to achieve the patient's treatment goals.

The Deformity Analysis and Correction Software is an optional software component and is used to assist the physician in calculating the lengths of the struts connecting the rings to manipulate the bone fragments. The software receives inputs from the physician and allows the physician to visualize the moving bone position. The program computes the strut lengths necessary to implement any desired translation and/or rotation required by the surgeon.

AI/ML Overview

The provided text describes the SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software (DACS) but does not detail specific acceptance criteria or a study that rigorously proves the device meets those criteria with quantitative performance metrics.

Instead, the document highlights:

  • Substantial Equivalence: The primary assertion is that the SixFix™ Hexapod Fixator and DACS are "substantially equivalent" to predicate devices (Taylor Spatial Frame, Deformity Analysis and Correction Software (DASC), and Ilizarov External Fixation Wires). This means it performs as well as or better than legally marketed predicate devices without raising new questions of safety or effectiveness.
  • Performance Data Overview: It mentions that "Static and dynamic mechanical testing were performed" and "Software verification and validation testing were conducted." However, it does not provide the results of these tests in terms of specific acceptance criteria.

Therefore, an answer using the requested format would mostly consist of "Not Provided" or "N/A" for many fields, as the document does not contain the level of detail asked for in the prompt regarding a specific study proving quantifiable acceptance criteria.

Here's an attempt to answer based on the given text, with the understanding that much information is missing:


1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Mechanical PerformanceThe testing confirmed that the subject SixFix™ Hexapod Fixator is substantially equivalent to the predicate device. (No specific quantitative criteria or performance metrics are provided).
Software FunctionalityThe functions performed by the SixFix™ DACS match the functions performed by the predicate DACS (K180539). Software validation confirmed that the DACS should perform as intended. (No specific quantitative criteria or performance metrics are provided for software accuracy or performance).
Safety and EffectivenessThe design characteristics of the subject system do not raise any new types of questions of safety or effectiveness. (General claim based on comparison to predicates, no specific criteria).

2. Sample sized used for the test set and the data provenance
Not provided in the document. The text refers to "static and dynamic mechanical testing" and "software verification and validation testing" but does not specify sample sizes for these tests or the provenance of any data used (e.g., country of origin, retrospective/prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable/Not provided. The document focuses on substantial equivalence based on mechanical and software testing, not on clinical performance evaluated against expert ground truth for interpretation or diagnosis.

4. Adjudication method for the test set
Not applicable/Not provided. No clinical test set or expert adjudication process is described in the provided text.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable/Not provided. This document does not describe an MRMC study. The DACS is software to assist the physician in calculating strut lengths and visualizing bone position, not for diagnostic interpretation by human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
While the DACS is a standalone software component in terms of its execution, the performance demonstration focuses on its functional equivalence to a predicate software and its role in assisting a physician. No specific "standalone performance" study (e.g., accuracy of calculations compared to a gold standard without human input as part of the evaluation) is detailed with quantifiable results. The text states, "The DACS matches the functions performed by the predicate DACS."

7. The type of ground truth used
For the mechanical testing, the ground truth would likely be established engineering standards or validated predicate device performance data.
For the software testing, the ground truth would be the expected output or behavior according to the software requirements specification and comparison to the predicate software's known functionality.
Specific details are not provided.

8. The sample size for the training set
Not applicable/Not provided. The document describes software verification and validation, not a machine learning model that would require a "training set" in the conventional sense.

9. How the ground truth for the training set was established
Not applicable/Not provided. As no training set for a machine learning model is mentioned, this information is not relevant to the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA part includes the acronym in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 5, 2022

Arrowhead DE, LLC Patrick Mullaney President/CEO 328 Poplar View Lane East, Suite 2 Collierville, Tennessee 38017

Re: K190069

Trade/Device Name: SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software (DACS) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT, HTY, OSN

Dear Patrick Mullaney:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated May 14, 2019. Specifically, FDA is updating this SE Letter to address the inconsistent indications for use in the FDA Form 3881 as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Ting Song, Ph.D., OHT6: Office of Orthopedic Devices, 301-796-7677, Ting.Song@fda.hhs.gov.

Sincerely,

Ting Song -S

Ting Song, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 14, 2019

Arrowhead DE, LLC Patrick Mullaney President/CEO 328 Poplar View Lane East, Suite 2 Collierville, Tennessee 38017

Re: K190069

Trade/Device Name: SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software (DACS) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories Regulatory Class: Class II Product Code: KTT, HTY, OSN Dated: April 17, 2019 Received: April 18, 2019

Dear Patrick Mullaney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{2}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Digitally signed by Peter G. Peter G. Allen -S Date: 2019.05.14 Allen -S 10:08:37 -04'00'

FOR CAPT Raquel Peat, PhD, MPH, USPHS Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K190069

Device Name

SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software (DACS)

Indications for Use (Describe)

The SixFix™ Hexapod and Deformity Analysis and Correction Software (DACS) are intended to be used for posttraumatic joint contracture which has resulted in loss of range of motion; fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction; open and closed fracture fixation; pseudo-arthrosis of long bones; limb lengthening by epiphyseal distraction; correction of bony or soft tissue deformities; correction of bony or soft tissue defects; joint arthrodesis; infected fractures or nonunions.

Type of Use (Select one or both, as applicable):

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary SixFix™ Hexapod Fixator and DACS K190069

l. Submitter:

Arrowhead Medical Device Technologies, LLC 328 Poplar View Lane East, Suite 2 Collierville, TN 38017

Contact Person: Thomas J. Twardzik Vice President, Marketing and Operations Office: (901) 853-4366 Fax: (206) 222-9173 Email: INFO@ArrowheadDevices.com

Date of Summary: May 7, 2019

II: Device

Proprietary Name: SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software (DACS)

Common Name: Orthopaedic Software for Treatment and Instrumentation Regulatory Class: Class II

Regulation:

21 CFR 888.3030 Single/multiple component metallic bone fixation appliances and accessories

21 CFR 888.3040 Smooth or threaded metallic bone fixation fastener Device Product Codes: KTT, HTY, and OSN Panel: Orthopedic

III. Predicate Devices
--------------------------
DeviceManufacturer510(k) No.ClearanceDate
Primary Predicate
Taylor Spatial FrameSmith & NephewK09304709/27/2010
K97074805/09/1997
Secondary Predicates
Deformity Analysis andCorrection Software(DASC) andInstrumentationArrowhead MedicalDeviceTechnologies, LLCK18053908/13/2018
Ilizarov ExternalFixation WiresSmith & NephewK87096103/19/1987
K96280808/19/1996

{5}------------------------------------------------

510(k) Summary SixFix™ Hexapod Fixator and DACS K190069

IV. Device Description

The SixFix™ Hexapod Fixator is a multilateral circular external fixation system. The system includes the following external fixator elements: rings, footplates, arches, struts, threaded rods, wires, external fixation accessories, and software. All of the elements are provided non-sterile and are for single use only.

The system is designed such that gradually adjusting the lengths of the struts in relation to one another alters the orientation of the rings and, consequently, the bone segments connected to the rings by half-pins and wires during the treatment period in order to achieve the patient's treatment goals.

The Deformity Analysis and Correction Software is an optional software component and is used to assist the physician in calculating the lengths of the struts connecting the rings to manipulate the bone fragments. The software receives inputs from the physician and allows the physician to visualize the moving bone position. The program computes the strut lengths necessary to implement any desired translation and/or rotation required by the surgeon.

V. Intended Use

The SixFix™ Hexapod and Deformity Analysis and Correction Software (DACS) are intended to be used for post-traumatic joint contracture which has resulted in loss of range of motion; fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction; open and closed fracture fixation; pseudo-arthrosis of long bones; limb lengthening by epiphyseal or metaphyseal distraction; correction of bony or soft tissue deformities; correction of bony or soft tissue defects; joint arthrodesis; infected fractures or nonunions.

VI. Comparison of Technological Characteristics with the Predicate Devices

The SixFix™ Hexapod Fixator is technologically substantially equivalent to predicate devices in terms of intended use, material, design, mechanical performance and safety. Testing and analyses confirmed that the Hexapod™ Fixator is substantially equivalent when compared to the predicate device. The design characteristics of the subject system do not raise any new types of questions of safety or effectiveness. From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate device.

The subject device and predicate devices are optional software programs that are used with a spatial frame external fixation system. The functions performed by the SixFix™ DACS match the functions performed by the predicate DACS (K180539). Software validation confirmed that the DACS should perform as intended.

{6}------------------------------------------------

510(k) Summary SixFix™ Hexapod Fixator and DACS K190069

VII. Performance Data

Static and dynamic mechanical testing were performed in support of the submission. The testing confirmed that the subject SixFix™ Hexapod Fixator is substantially equivalent to the predicate device.

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a Moderate Level of Concern. Documentation for Moderate Level of Concern was provided as follows.

  • Device Hazard Analysis
  • Software Requirements Specification ●
  • Architectural Design Charts
  • Software Design Specification Traceability Analysis
  • Software Development Environment Description
  • Software Verification Plan and Protocol
  • Variety Testing Software Verification Protocol
  • Cybersecurity Risk Analysis Report
  • Off-the-Shelf Software Analysis
  • Software Release Record
  • Unresolved Anomaly Report
  • Verification and Validation Documentation

From the evidence submitted in this 510(k) for the software, the Deformity Analysis and Correction Software (DACS) demonstrates that the device is as safe, as effective, and performs as well as or better than the legally marketed device predicates. The design characteristics of the subject system software do not raise any new types of questions of safety or effectiveness.

VIII. Conclusions

A review of the device indications, material composition, external element design, and technological characteristics confirmed that the SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software are substantially equivalent to the predicate device. While the SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software are not identical to the predicate device, comparisons of the subject and predicate device confirmed that any differences between the subject device and predicate do not render the device NSE as there is not a new intended use; and any differences in technological characteristics do not raise different questions of safety and effectiveness than the predicate device. Therefore, it is concluded that the SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software is substantially equivalent to the predicate devices.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.