RENOVIS SURGICAL HIP JOINT REPLACEMENT PROSTHESIS

K112897 · Renovis Surgical Technologies, LLC · OQG · Jun 13, 2012 · Orthopedic

Device Facts

Record IDK112897
Device NameRENOVIS SURGICAL HIP JOINT REPLACEMENT PROSTHESIS
ApplicantRenovis Surgical Technologies, LLC
Product CodeOQG · Orthopedic
Decision DateJun 13, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Renovis A400 Surgical Hip Joint Replacement Prosthesis is indicated for patients suffering from the following: 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 2. Rheumatoid arthritis; 3. Correction of functional deformity; 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and 5. Revision procedures where other treatment or devices have failed. The Renovis A400 Surgical Hip Joint Replacement Prosthesis is intended for cementless applications.

Device Story

Renovis A400 is a cementless total hip replacement system; components include titanium alloy femoral stem and acetabular cup, CoCr or Biolox delta ceramic femoral head, and mechanically annealed HXL polyethylene acetabular liners (with/without alpha-tocopherol). System replaces both femur and acetabulum in patients with degenerative joint disease, arthritis, or fractures. Used in surgical settings by orthopedic surgeons. Components provided sterile for single use. Design relies on standard mechanical fixation; no software or active components. Clinical benefit derived from restoring joint function and mobility in patients with failed prior treatments or severe joint pathology.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by comprehensive bench performance testing, including distal/proximal stem fatigue, snap-lock testing, range of motion analysis, hip simulator wear testing (normal and abrasive), wear debris characterization, porous coating integrity testing, and extensive polyethylene material characterization. Biocompatibility testing performed per ISO 10993.

Technological Characteristics

Cementless hip prosthesis. Materials: Titanium alloy (stem/cup), CoCr/Biolox delta ceramic (head), HXL polyethylene (liner). Testing standards: ISO 7206 (fatigue), ASTM F1820 (snap-lock), ISO 14242 (wear), ASTM F1877 (debris), ASTM F1978/F1160/F1044/F1147/F1854 (porous coating), ISO 10993 (biocompatibility). Mechanical fixation. No software or electronics.

Indications for Use

Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or proximal femur fractures (non-union, femoral neck, trochanteric) requiring hip replacement, including revision procedures. Intended for cementless applications.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # RENOVIS #### RENOVIS Surgical Technologies, Inc. #### 510(k)# K112897 ### 510(k) Summary - Submitter's Name: Renovis Surgical Technologies, Inc. - Address: 1901 W. Lugonia Ave, Ste 340 Redlands, CA 92374 - Contact Person: Anthony DeBenedictis Vice President of Quality Assurance Phone: 909.557.2360 Date Prepared: May 31, 2012 | Classification Name: | Prosthesis, Hip Revision System | |----------------------|-------------------------------------------------| | Common Name: | Surgical Hip Joint Replacement System | | Model Number: | A400 | | Product Code: | OQG, OQI, LPH, LZO | | C.F.R. Section: | 888.3353-Hip joint | | | metal/ceramic/polymer semi-constrained | | | cemented or nonporous uncemented<br>prosthesis. | | | 888.3358-Hip joint | | | metal/polymer/ | metal semi-constrained porouscoated uncemented prosthesis. | Recommended Classification: ...................................... | Class II | |--------------------------------------------------------------------|--------------------| | Classification Panel: ...................................... | Orthopedics | | Reason for Submission: ...................................... | Traditional 510(k) | ### Device Description: The Renovis Surgical Technologies, Inc. A400 Surgical Hip Replacement System (Renovis A400 Hip System) is a family of components designed in a variety of standard clinically usable sizes for cementless total hip replacements (i.e. replacements where both the femur and acetabulum are replaced). {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the text 'K112897' at the top. Below that is the word 'RENOVIS' in large, bold, sans-serif font. To the left of the word 'RENOVIS' is a circular graphic with three overlapping shapes. The system consists of four main components available in a variety of clinically usable sizes. A) Femoral Stem (titanium alloy) B) Femoral Head (CoCr and Biolox delta ceramic) C) Acetabular Cup (titanium alloy) D) Acetabular Liners (Mechanically annealed HXL polyethylene with and without alphatocopherol) The Renovis A400 Hip System implant components are provided sterile and are for single use. ### Indications for Use The Renovis A400 Surgical Hip Replacement System is indicated for patients suffering from: 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 2. Rheumatoid arthritis; 3. Correction of functional deformity; 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and 5. Revision procedures where other treatment or devices have failed. The Renovis A400 Surgical Hip Joint Replacement Prosthesis is intended for cementless applications. ### Predicate Devices Predicate List | 510(k) Number | Manufacturer | |---------------|----------------------| | K030055 | Biomet Manufacturing | | K032396 | Biomet Manufacturing | | K051411 | Biomet Manufacturing | | K061312 | Biomet Manufacturing | | K070364 | Biomet Manufacturing | | K073102 | Biomet Manufacturing | 2 Page {2}------------------------------------------------ ### ×112897 ## RENOVIS | K091956 | DJO Surgical | |---------|----------------------| | K094035 | StelKast, Inc. | | K043537 | Biomet Manufacturing | | K093235 | Biomet Manufacturing | | K981211 | Sulzer Orthopedics | ### Substantial Equivalence The Renovis A400 Surgical Hip Replacement System claims substantial equivalence to commercially available hip prosthesis systems and related hip system components manufactured by Biomet Manufacturing Corporation (Biomet). The Renovis A400 Hip System's intended use and indications for use are identical to the Biomet hip prosthesis system and associated components. The Renovis A400 Hip System implanted components: femoral head; acetabular cup; acetabular liners, with vitamin E and without vitamin E are equivalent and raise no new question of safety or technology to those devices manufactured by Biomet. The materials used for all Renovis A400 Hip System components are similar to those used in the components of the Biomet hip prosthesis system predicate devices. All materials used in the Renovis A400 Surgical Hip Replacement System are biocompatible and are well known and commonly used in the orthopedic implant industry. ### Performance Testing Comprehensive bench performance testing conducted on the hip system components: femoral stems, femoral heads, and acetabular liners demonstrated that the A400 Hip System design is safe for its intended use and is substantially equivalent to the predicate devices. The following bench/ performance testing was conducted on the A400 Hip System: A) A400 Hip System – The following device tests were performed on the complete A400 Hip system and demonstrate the system is safe for its intended uses and is substantially equivalent to the predicate devices: - Distal Stem Fatigue Testing per ISO 7206-4:2002/ Guidance for Industry and FDA 1. Staff - Non-clinical Information for Femoral Stem Prostheses, September 1997 - Proximal Stem Fatigue Testing per ISO 7206-6:1992/ Guidance for Industry and FDA 2. Staff – Non-clinical Information for Femoral Stem Prostheses, September 1997 - Push-out Snap-Lock Testing of Acetabular Liners and Shells per ASTM F1820-97 3. {3}------------------------------------------------ K"2897 RENOVIS - Lever-out and Torque-out Snap-Lock Testing of Acetabular Liners and Shells 4. - 5. CoCr Head Attachment Testing per ISO 7206-10:2003 - BIOLOX delta Ceramic Heads Attachment Testing per FDA Guidance Document for 6. the Preparation of Premarket Notifications for Ceramic Ball Hip Systems - Range of Motion Analysis 7.. - Hip Simulator Wear Testing under Normal Conditions per ISO 14242-1 8. - Hip Simulator Wear Testing under Abrasive Wear Conditions per ISO 14242-1 ക് - Wear Debris Characterization per ASTM F1877-05 10. B) Femoral Stem and Acetabular Cup Porous Surface/Coating- the following tests were performed to evaluate the integrity of the porous surface/coating: - 1. Abrasion resistance per ASTM F1978 - 2. Shear fatigue strength per ASTM F1160 - 3. Static shear strength per ASTM F1044 - 4. Static tensile strength per ASTM F1147 - 5. Characterization per ASTM F1854 C) Acetabular Liner – The following poly characterization testing, both chemical and mechanical, was performed on the HXL polyethylene materials: - 1. Heat of fusion , melting point, and percent crystallinity per ASTM F2625/D3418 - 2. Small punch test properties (peak load, ultimate load, ultimate displacement, work to failure) per ASTM F2183 - 3. Oxidation measurement per ASTM F2102 - 4. Trans-vinylene vield per ASTM F2381 - 5. Crosslink density (Swell Ratio Testing) per ASTM F2214 - 6. Tensile properties (yield strength, ultimate strength, elongation at break, elastic modulus) per ASTM D638 - 7. Izod impact strength per ASTM D256 /F648 - 8. GCMS and LCMS extract analyses - 9. Vitamin E content - 10. Fatigue crack propagation per ASTM E647 - 11. Compressive modulus per ASTM D695 - 12. Uniformity of radiation dose - 13. Free radical content (Electron Spin Resonance) - 14. Consolidation verification - 15. Density per ASTM D1505/D792 {4}------------------------------------------------ K112897 ## SRENOVIS All poly characterization test results demonstrate that the A400 Hip System design is safe for its intended use and is substantially equivalent to the predicate devices. D) Polyethylene Biocompatibility - The following tests in accordance with ISO 10993 (permanent implant panel) were performed on the polyethylene material to demonstrate biocompatibility: - 1. Systemic Toxicity: Materials Mediated Rabbit Pyrogen Test per ISO 10993-11 and -12 - 2. Systemic Toxicity: Acute Systemic Toxicity per ISO 10993-11 and -12 - 3. In Vitro Cytotoxicity Study Using the ISO Elution Method per ISO 10993-5 - 4. Genotoxicity: Chromosomal Aberration per ISO 10993-1, -3,-5, and -12. - 5. Genotoxicity: In-Vivo Mouse Micronucleus Assay per ISO 10993-3 and -12 - 6. Genotoxicity: Bacterial Reverse Mutation per ISO 10993-3. - 7. Irritation and Skin Sensitization: Intracutaneous Study per ISO 10993-2 and -10. - 8. ISO Muscle Implantation Study in Rabbits- 2 Weeks per ISO 10993-2 and -6 - 9. ISO Muscle Implantation Study in Rabbits- 13 Weeks per ISO 10993-2 and -6 - 10. Irritation and Skin Sensitization: Guinea Pig Maximization Sensitization Test per ISO 10993-2 and -10 - 11. Systemic Toxicity: Sub-Chronic (14 Day) Intravenous Toxicity Study per ISO 10993-11 and -12 {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle faces right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUN 1 3 2012 Renovis Surgical Technologies, LLC c/o David Teicher, DuVal & Associates 825 Nicollet Mall Minneapolis, Minnesota 55402 Re: K112897 Trade/Device Name: Renovis A400 Hip Joint Replacement Prosthesis Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: OQG, OQI, LPH, LZO Dated: May 4, 2012 Received: May 7, 2012 Dear Mr. Teicher: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {6}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ### X112897 ### RENOVIS Surgical Technologies, Inc. ### Indications for Use 510(k) Number (if known): K112897 Device Name: Renovis A400 Surgical Hip Joint Replacement Prosthesis Indications for Use: The Renovis A400 Surgical Hip Joint Replacement Prosthesis is indicated for patients suffering from the following: 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 2. Rheumatoid arthritis; 3. Correction of functional deformity; 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and 5. Revision procedures where other treatment or devices have failed. The Renovis A400 Surgical Hip Joint Replacement Prosthesis is intended for cementless applications. Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of (Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K112.897
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...