K230961 · Shanghai Sanyou Medical Co, Ltd. · NKB · Oct 10, 2023 · Orthopedic
Device Facts
Record ID
K230961
Device Name
Zeus Spinal System
Applicant
Shanghai Sanyou Medical Co, Ltd.
Product Code
NKB · Orthopedic
Decision Date
Oct 10, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Zeus Spinal System is intended for posterior, non-cervical fixation as an adjunct to fusion for skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studes); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis and/or lordosis); tumor, pseuudarthrosis; and/or failed previous fusion. In addition, when used as a pedicle screw fixation system, the Zeus Spinal System is intended for skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral, L5-S1 vertebra, who are receiving fusion by autogenous bone graft only, who are having the device attached to the lumbar and sacral spine (levels may be from L3 to the sacrum/ilium), who are having the device removed after the attainment of a solid fusion
Device Story
Zeus Spinal System is a posterior, non-cervical pedicle screw fixation system used as an adjunct to spinal fusion. The system comprises various titanium alloy or cobalt chrome alloy components, including polyaxial/monoaxial screws, reduction screws, rods, crosslinks, and connectors. Surgeons assemble these components intraoperatively to create a rigid construct tailored to the patient's anatomy. The device provides mechanical stabilization of the spine to facilitate fusion. It is intended for use in clinical settings by surgeons. The system is a modification of the Adena-Zina Spinal System, expanding the range of available screw geometries and rod sizes to accommodate diverse surgical needs.
Clinical Evidence
Bench testing only. Testing included static axial compression, dynamic axial compression, and static torsion per ASTM F1717, and screw dissociation testing per ASTM F1798.
Indicated for skeletally mature patients requiring posterior, non-cervical spinal fixation as an adjunct to fusion. Conditions include degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, scoliosis, kyphosis, lordosis, tumor, pseudarthrosis, and failed previous fusion. Specific indication for severe spondylolisthesis (Grades 3-4) at L5-S1 using autogenous bone graft, with fixation from L3 to sacrum/ilium and subsequent device removal after fusion.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
DePuy Spine, Inc. EXPEDIUM® Spine System (K102249)
Related Devices
K152781 — Adena-Zina System · Shanghai Sanyou Medical Co, Ltd. · Feb 12, 2016
K231931 — Adena-Zina Spinal System · Shanghai Sanyou Medical Co, Ltd. · Jul 28, 2023
K103237 — LANX SPINAL SYSTEM · Lanx, Inc. · Mar 25, 2011
K062912 — SCIENT'X SPINAL SYSTEM · Scient'X · Feb 5, 2007
K170964 — Zenius Spinal System · Medyssey USA, Inc. · Apr 26, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION" in a smaller font.
October 10, 2023
Shanghai Sanyou Medical Co, LTD % Christine Scifert Partner MRC Global, LLC 9085 E. Mineral Cir., Suite 110 Centennial, Colorado 80112
Re: K230961
Trade/Device Name: Zeus Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: September 11, 2023 Received: September 11, 2023
Dear Christine Scifert:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software
Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Image /page/2/Picture/3 description: The image contains the text "Colin O'neill -S" in black font on the left side of the image. On the right side of the image, the letters "FDA" are written in a light blue font. The letters "FDA" are much larger than the text on the left side of the image.
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230961
Device Name Zeus Spinal System
#### Indications for Use (Describe)
The Zeus Spinal System is intended for posterior, non-cervical fixation as an adjunct to fusion for skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studes); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis and/or lordosis); tumor, pseuudarthrosis; and/or failed previous fusion.
In addition, when used as a pedicle screw fixation system, the Zeus Spinal System is intended for skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral, L5-S1 vertebra, who are receiving fusion by autogenous bone graft only, who are having the device attached to the lumbar and sacral spine (levels may be from L3 to the sacrum/ilium), who are having the device removed after the attainment of a solid fusion
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary Zeus Spinal System 4 April 2023
| Company: | Manufacturing Facility and Headquarters:<br>Shanghai Sanyou Medical Co, LTD<br>385 Huirong Road, Jiading District,<br>Shanghai 201815<br>Phone: 021-58389980 |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Manufacturing Facility:<br>Shanghai Sanyou Medical Co, LTD<br>385 Huirong Road, Jiading District,<br>Shanghai 201815<br>Phone: 021-58389980 |
| Company Contact: | David Fan<br>VP Marketing<br>david.fan@sanyou-medical.com<br>Phone: +86 21 58389980<br>Fax: +86 21 59990826 |
| Official Correspondent: | Christine Scifert - MRC Global, LLC<br>Christine.scifert@askmrcglobal.com<br>901-831-8053 |
| Trade Name: | Zeus Spinal System |
| Common Name: | Thoracolumbosacral pedicle screw system |
| Classification: | Class II |
| Regulation Number: | 21 CFR 888.3070 (Thoracolumbosacral pedicle screw system) |
| Panel: | Orthopedic |
| Product Code: | NKB |
### Device Description:
The Zeus Spinal System is a modification of the previously cleared Adena-Zina Spinal System which consists of a variety of shapes and sizes of screws and Ø5.35mm rods that can be rigidly locked in a variety of configurations, with each construct being tailor-made for the individual case.
{5}------------------------------------------------
Fixation is provided via a posterior approach. The components are made from titanium alloy or cobalt chrome alloy. Components of the system include the originally cleared (K152781) straight and pre-bent rods, fixed and fixed angle reduction screws, multi-axial reduction screws, T-links, domino connectors, sacro-illiac connectors, as well as the Duetto dual headed screws and connectors, along with associated set screws (K212066)
The subject submission seeks to add new Ø5.5mm pre-bend rods, single head and dual head polyaxial (multi-axial) screws, dual head reduction screws, polyaxial reduction screws (fixed polyaxial fixed screws, dual head monoaxial (fixed) screws, cannulated reduction screws (multiaxial and monoaxial), coronal and sagittal uniplanar screws, preferable angle screws, iliac screws, monoaxial Screw, associated set screws, connector and Ø5.5mm crosslinks, as well as additional sizes of existing components and cobalt chrome tulips for associated screws.
### Indications for Use:
The Zeus Spinal System is intended for posterior, non - cervical fixation as an adjunct to fusion for skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion. In addition, when used as a pedicle screw fixation system, the Zeus Spinal System is intended for skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral, L5-S1 vertebra, who are receiving fusion by autogenous bone graft only, who are having the device attached to the lumbar and sacral spine (levels may be from L3 to the sacrum/ilium), who are having the device removed after the attainment of a solid fusion.
#### Substantial Equivalence:
The subject Zeus Spinal System is substantially equivalent to the following predicate devices: Primary Predicate:
- Shanghai Sanyou Adena-Zina System – K152781
### Secondary Predicates:
- Shanghai Sanyou Adena-Zina System K212066 ●
- . Medtronic CD Horizon Spinal System– K201407
- DePuy Spine, Inc. EXPEDIUM® Spine System – K102249
There are insignificant differences between the subject Zeus Spinal System and the predicate devices. The Indications for Use and Materials for predicate devices are all inclusive of the subject device, with the subject device materials and indications being identical to the primary predicate Adena-Zina System (K152781). While Zeus Spinal System does include size offering/geometry outside of the predicate offerings, performance bench testing had been performed and shows that the subject Zeus Spinal System performs equivalent to or better than the primary predicate device. Thus, it can be concluded that the subject does not raise new questions about safety and effectiveness.
{6}------------------------------------------------
## Performance Testing:
Bench performance testing was performed on the subject Zeus Spinal System implants including static axial compression, dynamic axial compression, and static torsion per ASTM F1717 Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model. Additionally, screw dissociation testing was performed per ASTM F1798 Standard Test Method for Evaluating the Static and Fatigue Propertied of Interconnection Mechanism and Subassemblies Used in Spinal Arthrodesis Implants
#### Conclusion:
Based on the test results and the comparison to the predicate devices, the subject device is determined to be substantially equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.