K120832 · X-Spine Systems, Inc. · MNH · Oct 23, 2012 · Orthopedic
Device Facts
Record ID
K120832
Device Name
FORTEX PEDICLE SCREW SYSTEM
Applicant
X-Spine Systems, Inc.
Product Code
MNH · Orthopedic
Decision Date
Oct 23, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fortex Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. In addition, the Fortex Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
Device Story
Fortex Pedicle Screw System is a spinal fixation construct consisting of screws, rods, and cross-bar connectors. Used by surgeons in clinical settings to provide supplemental stabilization and immobilization of spinal segments as an adjunct to fusion. Modified version introduces cobalt chrome alloy rods (ASTM F1537) alongside existing titanium alloy components (ASTM F136). Components are assembled into a construct to support bone graft fusion; titanium and cobalt chrome may be used together, but not with stainless steel. Device provides mechanical support to stabilize the spine during the healing process, potentially benefiting patients by correcting deformities or stabilizing instabilities.
Clinical Evidence
Bench testing only. Static and dynamic axial compression, bending, and torsion tests were performed in accordance with ASTM F1717 to compare the modified system with the predicate. Results demonstrated that the modified device functioned as intended and possessed mechanical properties comparable to the predicate system.
Technological Characteristics
Construct components include screws, rods, and cross-bar connectors. Materials: medical grade titanium alloy (ASTM F136) and medical grade cobalt chrome alloy (ASTM F1537). Mechanical testing performed per ASTM F1717. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring spinal fusion for severe spondylolisthesis (Grade 3-4, L5-S1) or acute/chronic instabilities/deformities (degenerative spondylolisthesis with neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, pseudarthrosis) of the thoracic, lumbar, and sacral spine.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
K132101 — ANAX 5.5 SPINAL SYSTEM · U&I Corporation · Nov 25, 2013
K103237 — LANX SPINAL SYSTEM · Lanx, Inc. · Mar 25, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
K120832
## 510(k) Summary
### ADMINISTRATIVE INFORMATION
Manufacturer Name:
X-spine Systems, Inc. 452 Alexandersville Rd. Miamisburg, OH 45342
Telephone (937) 847-8400 FAX (937) 847-8410
Fortex Pedicle Screw System
Pedicle Screw Spinal System
Official Contact:
David Kirschman, MD Chief Medical Officer
Initial Date Prepared:
March 16, 2012
#### DEVICE NAME
Trade/Proprietary Name: Common Name(s):
Classification Name(s):
Device Class: Classification(s): Product Codes(s): Orthosis, Spondylolisthesis Spinal Fixation Orthosis, Pedicle Spinal Fixation Class II \$888.3070 MNH, MNI
#### ESTABLISHMENT REGISTRATION NUMBER
X-spine Systems, Inc. has submitted an Establishment Registration to FDA. The Establishment Registration number is 3005031160. The owner/operator number for X-spine Systems, Inc. is 9063903.
#### INTENDED USE
The Fortex Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition, the Fortex Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
{1}------------------------------------------------
page 2 of 2
#### DEVICE DESCRIPTION AND TECHNOLOGICAL CHARACTERISTICS
The purpose of this 510(k) submission is to add rods made of a different material to the Fortex Pedicle Screw System. The modified system has the same intended use and fundamental scientific technology as the previously-cleared Fortex Pedicle Screw System.
Both the modified and predicate Fortex Pedicle Screw System consist of various screws, rods, and cross bar connectors that are used to build a construct to provide supplemental stabilization of spinal segments to support fusion.
The components of the predicate Fortex.Pedicle Screw System are fabricated from medical grade titanium alloy per ASTM F136. The modified system includes components fabricated from medical grade cobalt chrome alloy per ASTM F1537, in addition to the titanium components. As documented within this premarket notification, the addition of cobalt chrome alloy rods does not raise new issues of safety or effectiveness. Titanium and cobalt chrome components may be used together within the same construct. These components should never be used with stainless steel implant components.
### PERFORMANCE DATA
Non-clinical performance testing was conducted to characterize the modified Fortex Pedicle Screw System. Static and dynamic axial compression bending and static torsion tests in accordance with ASTM F1717 were performed on the modified and predicate systems, and the results compared. The modified device functioned as intended and the observed test results demonstrate substantial equivalence to the predicate devices.
### EQUIVALENCE TO MARKETED PRODUCT
The modified Fortex Pedicle Screw System has the same intended use, indications, technological characteristics, and principles of operation as the predicate Fortex Pedicle Screw System, and the material change is to a type of material that has been used in other legally marketed devices within the same classification regulation for the same intended use such as the K2M, Inc. Range Spinal System (K080792). The modifications to the Fortex Pedicle Screw System do not raise new issues of safety or effectiveness. Also, mechanical testing demonstrated comparable mechanical properties to the predicate devices. Thus, the modified Fortex Pedicle Screw System is substantially equivalent to the predicate devices.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, rendered in a bold, black color.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
X-spine Systems, Incorporated % David Kirschman, M.D. Chief Medical Officer 452 Alexandersville Road Miamisburg, Ohio 45342
OCT 23 2012
Re: K120832
Trade/Device Name: Fortex Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI Dated: October 5, 2012 Received: October 9, 2012
Dear Dr. Kirschman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - David Kirschman, M.D
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Daniel. Krone
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Fortex Pedicle Screw System
### Indications for Use
## 510(k) Number (if known): K20832
Device Name: Fortex Pedicle Screw System
Indications for Use:
The Fortex Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition, the Fortex Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
Prescription Use ਮ (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Divigion Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
Page 14 of 39
K120832 510(k) Number_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.