GMK Sphere Revision

K211664 · Medacta International S.A. · JWH · Jul 29, 2021 · Orthopedic

Device Facts

Record IDK211664
Device NameGMK Sphere Revision
ApplicantMedacta International S.A.
Product CodeJWH · Orthopedic
Decision DateJul 29, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Evolis®/GMK® knee prosthesis is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. This knee replacement system is indicated in the following cases: - Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis. - Avascular necrosis of femoral condyle. - Post traumatic loss of joint configuration. - Primary implantation failure. Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement. The screwed tibial augments are for screwed fixation to the tibial baseplate. In case a semi-constrained liner is used, an extension stem must be implanted both on the tibial and on the femoral components. In case a GMK Revision or GMK Sphere Revision tibial tray is used, an extension stem must be implanted.

Device Story

GMK Sphere Revision is a line extension of the Medacta GMK knee prosthesis system; designed for cemented total knee arthroplasty. System includes femoral components, tibial trays, distal/posterior wedges, and offset connectors. Implants are manufactured from Co-Cr-Mo alloy (ISO 5832-4) or Ti6Al4V (ISO 5832-3), with optional TiNbN coating. Device is used by orthopedic surgeons in clinical settings to replace damaged knee joints. Components are implanted to restore joint function and alleviate pain caused by arthritis, necrosis, or previous implant failure. Surgeons select appropriate sizes and configurations based on patient anatomy and bone quality. The device provides a mechanical replacement for the patellofemorotibial joint, potentially improving patient mobility and quality of life.

Clinical Evidence

No clinical studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including range of motion, dynamic endurance of femoral condyles and extension stems, tibial augmentation connection strength, wear behavior, and ion release testing for TiNbN coating.

Technological Characteristics

Materials: Co-Cr-Mo alloy (ISO 5832-4) for femoral/tibial components; Ti6Al4V (ISO 5832-3) for wedges/connectors. Optional TiNbN coating. Fixation: Cemented. Design: Semi-constrained knee prosthesis. Components: Femoral components (sizes 2-8), tibial trays, distal/posterior wedges, offset connectors. Sterilization: Not specified (standard industry practices implied).

Indications for Use

Indicated for patients with severely painful/disabled joints due to arthritis (traumatic, rheumatoid, or polyarthritis), avascular necrosis of the femoral condyle, post-traumatic joint configuration loss, or primary implantation failure, provided there is sufficient sound bone for component seating.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ July 29, 2021 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is the Department of Health & Human Services logo. The right side of the logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Medacta International S.A. % Chris Lussier Senior Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis, Tennessee 38118 Re: K211664 Trade/Device Name: GMK Sphere Revision Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: May 28, 2021 Received: June 1, 2021 Dear Chris Lussier: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Ting Song, Ph.D., R.A.C. For Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K211664 Device Name GMK Sphere Revision ### Indications for Use (Describe) The Evolis®/GMK® knee prosthesis is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. This knee replacement system is indicated in the following cases: - · Severely painful and/or disabled joint as a result of arthritis, traumator arthritis, theumatord arthritis. - Avascular necrosis of femoral condyle. - · Post traumatic loss of joint configuration. - · Primary implantation failure. Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement. The screwed tibial augments are for screwed fixation to the tibial baseplate. In case a semi-constrained liner is used, an extension stem must be implanted both on the tibial and on the femoral components. In case a GMK Revision or GMK Sphere Revision tibial tray is used, an extension stem must be implanted. × | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary #### I. Submitter Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66 Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director of Quality and Regulatory, Medacta USA Date Prepared: May 27, 2021 Date Revised: July 28, 2021 #### II. Device | Device Proprietary Name: | GMK Sphere Revision | |--------------------------|---------------------------------------------------------------------------------------------| | Common or Usual Name: | Total Knee Prosthesis | | Classification Name: | Knee joint patellofemorotibial polymer/metal/polymer<br>semiconstrained cemented prosthesis | | Primary Product Code: | JWH | | Regulation Number: | 21 CFR 888.3560 | | Device Classification | II | #### III. Predicate Device Substantial equivalence is claimed to the following devices: Primary predicate device - > GMK Sphere, K121416, Medacta International SA Secondary predicate devices - > GMK Revision, K102437, Medacta International SA - A GMK Revision Extension, K123721, Medacta International SA - GMK Total Knee System TiNbN Coating, K202684, Medacta International SA A - A GMK Revision & Hinge - TiNbN Coating, K210010, Medacta International SA #### IV. Device Description The GMK Sphere Revision is a Medacta GMK line extension to provide a larger product offering. The subject devices are designed for cemented use in total knee arthroplasty procedures. The GMK Sphere Revision system includes: {4}------------------------------------------------ K211664 Page 2 of 4 - Femoral components, left and right, sizes from 2 to 8, with or without TiNbN coating; ● - . Tibial trays, left and right, sizes T3I4 and T4I3, with or without TiNbN coating; - Distal wedges, 4 sizes (2, 3-4, 5-6, 7-8) with thicknesses 4, 8, 12, 16 and 20 mm; ● - Posterior wedges, 6 sizes (2, 3-4, 5, 6, 7, 8) with thicknesses 4, 8 and 12 mm; ● - Offset connectors from 2 to 5 mm. ● The GMK Sphere Revision implants, both femoral components and tibial trays, are manufactured from cobalt-chromium-molybdenum alloy (Co-Cr-Mo) according to ISO 5832-4 and they are available with or without Titanium Niobium Nitride (TiNbN) coating. The GMK Sphere Revision wedges and offset connectors, are manufactured from Ti6A14V per ISO 5832-3. #### V. Indications for Use The Evolis®/GMK® knee prosthesis is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. This knee replacement system is indicated in the following cases: - Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polvarthritis. - . Avascular necrosis of femoral condyle. - Post traumatic loss of joint configuration. - . Primary implantation failure. Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement. The screwed tibial augments are for screwed fixation to the tibial baseplate. In case a semiconstrained liner is used, an extension stem must be implanted both on the femoral components. In case a GMK Revision or GMK Sphere Revision tibial tray is used, an extension stem must be implanted. #### VI. Comparison of Technological Characteristics The GMK Sphere Revision implants and the predicate devices share the following characteristics: - indications for use; - shape and design (except for ML dimension and posterior condyle height of the femoral ● components and tibial travs holes): - fixation; ● - materials: - coating (when applicable); ● - biocompatibility: - device usage: - sterility; and - packaging. . The GMK Sphere Revision implants differ from the predicate devices as follow: - sizes: - ML dimension and posterior condyle height of the femoral components; and ● {5}------------------------------------------------ - tibial trays holes. ● ## Discussion Medacta International SA has not made any change to the indications for use, general design and shape, materials and coating, device usage, biocompatibility, sterility, shelf life, and packaging of the subject devices respect to the predicate devices. Based on the comparison of technological characteristics and performance data provided within this submission, the data supports the substantial equivalence of the GMK Sphere Revision implants to the identified predicate devices. #### Performance Data VII. Based on the risk analysis, performance testing were conducted to written protocols. The following tests and rationales are provided in support of the substantial equivalence determination: ### Non-Clinical Studies - PERFORMANCE TESTING - GMK Sphere Revision ROM and mobility of the articulating surface (per ISO 21536 and o ASTM F2083) - GMK Sphere Revision Dynamic Endurance test of the Posterior Femoral Condyle (per ISO O 7207-1:07, ISO 7207-2:11, ISO 5832-4:14, ASTM F 1814-15, ASTM F 2083-12, ASTM F 1800-19, ASTM F 3161-16) - o GMK Sphere Revision - Dynamic Endurance test in combination with Extension Stem (per ASTM F 1814-15, ASTM F 897-02, ASTM F 1800-19, ASTM F F2009-20, ISO 7207-1:07, ISO7207-2:11, ISO 5832-4:14, ISO 5832-3:16, ISO 14243-1:09, ISO 5834-2:19) - GMK Sphere Revision Tibial augmentation screwed connection (per PI-53: 2010-11 and O ASTM F2009-20) - GMK Sphere Revision femoral components Wear behavior (per ISO 14243-1) O - TiNbN Coating Excessive Ions Release ( per ISO 21534. ISO 14577, ISO 5832-2, ISO 14243-O 1, EN ISO 10993, EN ISO 10993-1, ISO 10993-18, and EN ISO 10993-6) - PYROGENICITY - Bacterial endotoxin test (LAL test) according to European Pharmacopoeia \$2.6.14 (which is o equivalent to USP chapter <85>) - Pyrogen test according to USP chapter <151> for pyrogenicity determination о - The subject devices are not labeled as non-pyrogenic or pyrogen free. O - BIOCOMBATIBILITY evaluation - SHELF-LIFE evaluation {6}------------------------------------------------ Medacta International SA ## Clinical Studies: - No clinical studies were conducted. ● ### VIII. Conclusion The information provided above supports that the GMK Sphere Revision implants are substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...