K Number
K223548

Validate with FDA (Live)

Date Cleared
2023-01-20

(56 days)

Product Code
Regulation Number
888.3565
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GMK knee cemented prosthesis is designed for cemented application in total knee arthroplasty if there is evidence of sufficient sound bone to seat and support the components.

The use of the cementless GMK femur is limited to cases in which the surgeon considers the bone quality to be sufficient for cementless applications.

The GMK 3D Metal Tibial Baseplate is indicated for cemented application if there is evidence of sufficient sound bone to seat and support the components.

This knee replacement system is indicated in the following cases:

  • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis
  • Collagen disorders, and avascular necrosis of the femoral condyle
  • Post traumatic loss of joint configuration
  • Primary implantation failure

Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement.

The screwed tibial augments are for screwed fixation to the tibial baseplate. In case a semiconstrained liner is used, an extension stem must be implanted both on the tibial and on the femoral components. In case a GMK Revision or GMK Sphere Revision tibial tray is used, an extension stem must be implanted.

It is not possible to implant tibial wedges and extension stems with the GMK 3D Metal Tibial Baseplate.

Limitations for use for GMK Sphere/GMK SpheriKA used with kinematic alignment

GMK Sphere and GMK SpheriKA can be implanted in kinematic alignment. In this case, this knee replacement system is indicated for:

  • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis
  • Collagen disorders, and/or avascular necrosis of the femoral condyle
  • Moderate valgus, varus, or flexion deformities.
Device Description

The GMK Sphere & GMK SpheriKA Cementless is a line extension to the GMK Total Knee System and includes the following devices:

  • GMK Sphere Cementless Femoral Components, Left and Right, Sizes from 1 to 7 and from 1+ to 6+ (intermediate sizes);
  • GMK SpheriKA Cementless Femoral Components. Left and Right, Sizes from 1 to 7 and from 1+ to 6+ (intermediate sizes);
  • GMK SpheriKA ST Cementless Femoral Components, Left and Right, Sizes from 1 to 7 and from 1+ to 6+ (intermediate sizes).

The subject devices are marketed as individually packaged femoral components, designed for cementless use in total knee arthroplasty procedures where the surgeon considers the bone quality to be sufficient for cementless applications.

The GMK Sphere & GMK SpheriKA Cementless implants are manufactured from cobaltchromium-molybdenum alloy (Co-Cr-Mo) according to ISO 5832-4:2014 Implants for Surgery -Metallic Materials-Part 4: Cobalt-Chromium-Molybdenum Casting Alloy with double-layer coating composed of Titanium plasma spray according to ISO 13179-1 Implants for surgery--Plasma-sprayed unalloyed titanium coatings on metallic surgical implants -- Part 1: General requirements and Hydroxyapatite according to ISO 13779-2 Implants for surgery - Hydroxyapatite - Part 2: Thermally sprayed coatings of hydroxyapatite on the internal surface pockets.

AI/ML Overview

This document refers to the review of a 510(k) premarket notification for the "GMK Sphere & GMK SpheriKA Cementless" knee prosthesis. Based on the provided text, there is no acceptance criteria and study that proves the device meets the acceptance criteria in the traditional sense of evaluating an algorithm's performance against predefined metrics (e.g., sensitivity, specificity for diagnostic devices).

This submission is for a medical device (knee prosthesis), not an AI/ML powered device, and the "performance data" provided refers to engineering and biocompatibility testing, not clinical performance metrics or studies involving human readers or ground truth as would be relevant for AI.

Here's a breakdown of why the requested information cannot be provided from the given text:

  1. A table of acceptance criteria and the reported device performance: Not applicable. The document describes a comparison to predicate devices, not the setting of specific performance targets (like accuracy, precision, etc.) for an AI/ML algorithm. The "performance data" listed are non-clinical engineering tests (mechanical properties, microscopy, pyrogenicity, biocompatibility, shelf-life) which are not presented with acceptance criteria and results in the format requested.

  2. Sample size used for the test set and the data provenance: Not applicable. There is no "test set" in the context of an algorithm's performance evaluation. The "tests" performed are on the physical device components.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth, in the context of AI/ML, refers to an independently verified correct answer for a given input. This concept is not relevant to the described premarket notification for a knee prosthesis.

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. There is no adjudication process described as it pertains to medical image interpretation or similar tasks for an algorithm.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document explicitly states: "No clinical studies were conducted." The device is a physical knee implant, not an AI-assisted diagnostic tool.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. There is no algorithm.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. "Ground truth" as an expert-derived annotation or definitive diagnosis is not relevant to the evaluation of a physical knee implant.

  8. The sample size for the training set: Not applicable. There is no AI/ML algorithm requiring a training set.

  9. How the ground truth for the training set was established: Not applicable, as there is no AI/ML algorithm or training set.

In summary, the provided FDA 510(k) clearance letter and summary describe the regulatory review for a physical orthopedic implant (a knee prosthesis), not an AI/ML-powered medical device. Therefore, the questions related to acceptance criteria and study design for AI/ML performance evaluation are not applicable to this document. The "Performance Data" section details non-clinical laboratory testing for material properties, sterility, and biocompatibility to demonstrate substantial equivalence to predicate devices, which is a different type of evaluation from AI/ML performance studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 20, 2023

Medacta International S.A. % Chris Lussier Senior Director, Quality, Regulatory and Clinical Research Medacta USA 3973 Delp Street Memphis, Tennessee 38118

Re: K223548

Trade/Device Name: GMK Sphere & GMK SpheriKA Cementless Regulation Number: 21 CFR 888.3565 Regulation Name: Knee Joint Patellofemorotibial Metal/Polymer Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: MBH Dated: November 23, 2022 Received: November 25, 2022

Dear Chris Lussier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Ting Song -S

Ting Song, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223548

Device Name

GMK Sphere & GMK SpheriKA Cementless

Indications for Use (Describe)

The GMK knee cemented prosthesis is designed for cemented application in total knee arthroplasty if there is evidence of sufficient sound bone to seat and support the components.

The use of the cementless GMK femur is limited to cases in which the surgeon considers the bone quality to be sufficient for cementless applications.

The GMK 3D Metal Tibial Baseplate is indicated for cemented application if there is evidence of sufficient sound bone to seat and support the components.

  • This knee replacement system is indicated in the following cases:
  • Severely painful and/or disabled joint as a result of arthritis, theumatoid arthritis or polyarthritis
  • · Collagen disorders, and avascular necrosis of the femoral condyle
  • Post traumatic loss of joint configuration
  • · Primary implantation failure

Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement. The screwed tibial augments are for screwed fixation to the tibial baseplate. In case a semi-constrained liner is used, an extension stem must be implanted both on the tibial and on the femoral components. In case a GMK Revision or GMK Sphere Revision tibial tray is used, an extension stem must be implanted.

It is not possible to implant tibial wedges and extension stems with the GMK 3D Metal Tibial Baseplate.

Limitations for use for GMK Sphere/GMK SpheriKA used with kinematic alignment

GMK Sphere and GMK SpheriKA can be implanted in kinematic alignment. In this case, this knee replacement system is indicated for:

  • Severely painful and/or disabled joint as a result of arthritis, traumatoid arthritis or polyarthritis
  • · Collagen disorders, and/or avascular necrosis of the femoral condyle
  • Moderate valgus, varus, or flexion deformities.

Type of Use (Select one or both, as applicable)

Requisition Use (Part 31 CFR 301 Subpart D)Other Than Requisition Use (31 CFR 301 Subpart C)
------------------------------------------------------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. Submitter

Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66

Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director of Quality and Regulatory, Medacta USA Date Prepared: November 23, 2022 Date Revised: January 13, 2023

II. Device

Device Proprietary Name:GMK Sphere & GMK SpheriKA Cementless
Common or Usual Name:Total Knee Prosthesis
Classification Name:Knee joint patellofemorotibial metal/polymer porous-coateduncemented prosthesis
Primary Product Code:MBH
Regulation Number:21 CFR 888.3565
Device ClassificationII

III. Predicate Device

Substantial equivalence is claimed to the following devices:

Primary predicate device

  • LEGION Porous Plus HA Primary Femoral Components, K091543, Simth & Nephew, . Inc.
    Secondary predicate devices

  • . GMK Sphere, K121416, Medacta International SA (also referred to as the GMK Total Knee System)

  • . GMK Sphere Extensions, K140826, Medacta International SA (also referred to as the GMK Total Knee System)

  • GMK SpheriKA, K211004, Medacta International SA ●

  • Vanguard Complete Knee System, K113550, Biomet Manufacturing Corp. ●

Reference devices

  • Versafit Cup Double Mobility Cementless Acetabular shells, K083116, Medacta . International SA

{4}------------------------------------------------

  • Evolis Total Knee System. Cementless femur porous Ti coated, K081023. Medacta . International SA.

Device Description IV.

The GMK Sphere & GMK SpheriKA Cementless is a line extension to the GMK Total Knee System and includes the following devices:

  • GMK Sphere Cementless Femoral Components, Left and Right, Sizes from 1 to 7 and from . 1+ to 6+ (intermediate sizes);
  • . GMK SpheriKA Cementless Femoral Components. Left and Right, Sizes from 1 to 7 and from 1+ to 6+ (intermediate sizes);
  • GMK SpheriKA ST Cementless Femoral Components, Left and Right, Sizes from 1 to 7 ● and from 1+ to 6+ (intermediate sizes).

The subject devices are marketed as individually packaged femoral components, designed for cementless use in total knee arthroplasty procedures where the surgeon considers the bone quality to be sufficient for cementless applications.

The GMK Sphere & GMK SpheriKA Cementless implants are manufactured from cobaltchromium-molybdenum alloy (Co-Cr-Mo) according to ISO 5832-4:2014 Implants for Surgery -Metallic Materials-Part 4: Cobalt-Chromium-Molybdenum Casting Alloy with double-layer coating composed of Titanium plasma spray according to ISO 13179-1 Implants for surgery--Plasma-sprayed unalloyed titanium coatings on metallic surgical implants -- Part 1: General requirements and Hydroxyapatite according to ISO 13779-2 Implants for surgery - Hydroxyapatite - Part 2: Thermally sprayed coatings of hydroxyapatite on the internal surface pockets.

V. Indications for Use

The GMK knee cemented prosthesis is designed for cemented application in total knee arthroplasty if there is evidence of sufficient sound bone to seat and support the components.

The use of the cementless GMK femur is limited to cases in which the surgeon considers the bone quality to be sufficient for cementless applications.

The GMK 3D Metal Tibial Baseplate is indicated for cemented application if there is evidence of sufficient sound bone to seat and support the components.

This knee replacement system is indicated in the following cases:

  • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid . arthritis or polyarthritis
  • Collagen disorders, and avascular necrosis of the femoral condyle .
  • Post traumatic loss of joint configuration ●
  • Primary implantation failure

Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement.

The screwed tibial augments are for screwed fixation to the tibial baseplate. In case a semiconstrained liner is used, an extension stem must be implanted both on the tibial and on the femoral

{5}------------------------------------------------

components. In case a GMK Revision or GMK Sphere Revision tibial tray is used, an extension stem must be implanted.

It is not possible to implant tibial wedges and extension stems with the GMK 3D Metal Tibial Baseplate.

Limitations for use for GMK Sphere/GMK SpheriKA used with kinematic alignment

GMK Sphere and GMK SpheriKA can be implanted in kinematic alignment. In this case, this knee replacement system is indicated for:

  • . Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis
  • . Collagen disorders, and/or avascular necrosis of the femoral condyle
  • Moderate valgus, varus, or flexion deformities.

Comparison of Technological Characteristics VI.

The subject GMK Sphere & GMK SpheriKA Cementless and the predicate devices (K121416, K140826. K211004) share the following characteristics:

  • sizes and design:
  • substrate material;
  • biocompatibility:
  • device usage:
  • sterility: ●
  • shelf life: and ●
  • packaging.

The subject GMK Sphere & GMK SpheriKA Cementless differ from the predicate devices (K121416, K140826, K211004) with respect to:

  • . fixation: and
  • . coating.

The presence of a double-layer coating of the subject devices with respect to the predicate is justified by the different fixation method: cementless for the subject devices and cemented for the predicate (K121416, K140826, K211004). This difference does not raise any new issue related to subject devices' safety and effectiveness since it is shared with the primary predicate LEGION Porous Plus HA Primary Femoral Components (K091543) and with the additional predicate Vanguard Complete Knee System (K113550) and is supported by performance data.

VII. Performance Data

Based on the risk analysis, testing was conducted according to written protocols. The following tests and rationales are being provided in support of a substantial equivalence determination:

Non-Clinical Studies:

  • . PERFORMANCE TESTING

{6}------------------------------------------------

  • GMK Sphere porous coated femoral components mechanical properties o
  • о GMK SpheriKA porous coated femoral components mechanical properties
  • Scanning Electron Microscopy analysis of Medacta GMK Sphere femoral components o
  • Cross Sectioned Area of the coated implant surfaces of the Medacta GMK Sphere O femoral component
  • o XRD analyses comparing the coating features deposed in the Medacta GMK Sphere femoral component and on planar samples made of CoCr
  • PYROGENICITY
    • Bacterial Endotoxin Test (LAL test) was conducted according to European O Pharmacopoeia $2.6.14 (which is equivalent to USP chapter <85>)
    • Pyrogen test according to USP chapter <151> for pyrogenicity determination O
    • O The subject devices are not labeled as non-pyrogenic or pyrogen free
  • BIOCOMPATIBILITY assessment as per ISO 10993 series and FDA Biocompatibility Guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"
  • SHELF-LIFE evaluation

Clinical Studies

  • . No clinical studies were conducted.

VIII. Conclusion

Based on the above information, the GMK Sphere & GMK SpheriKA Cementless implants can be considered substantially equivalent to the identified predicate devices.

Substantial equivalence has been demonstrated through a comparison of intended use, design, and technological characteristics, as well as performance evaluations.

N/A