SPAR-K Instruments (for use with Gemini SL Total Knee System)

K191755 · Waldemar Link GmbH & Co. KG · JWH · Jul 18, 2019 · Orthopedic

Device Facts

Record IDK191755
Device NameSPAR-K Instruments (for use with Gemini SL Total Knee System)
ApplicantWaldemar Link GmbH & Co. KG
Product CodeJWH · Orthopedic
Decision DateJul 18, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3560
Device ClassClass 2

Indications for Use

The LINK GEMINI SL Total Knee System is indicated for patients suffering from disability due to: - Degenerative, post-traumatic or rheumatoid arthritis; - Avascular necrosis of the femoral condyle; - Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; - Moderate valgus, varus or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts. The device is intended for cemented use. Only cementless labeled modular stems are indicated for uncemented use.

Device Story

SPAR-K instruments are manual, reusable orthopedic surgical tools designed to aid implantation of the Gemini SL Total Knee System. These instruments serve as a line extension to previously cleared instruments (K182872), incorporating minor design modifications to resection guides for improved simplicity and ease of use. The instruments are used by orthopedic surgeons in a clinical/OR setting to perform femoral, tibial, and patellar bone resections. They support both tibia-first and femur-first surgical workflows. The modifications do not alter the fundamental surgical technique, intended use, or operating principle. By facilitating accurate bone preparation, the instruments assist the surgeon in implanting the knee prosthesis, aiming to restore joint function and alleviate patient disability.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Manual, reusable orthopedic surgical instruments (femoral, tibial, and patellar resection guides). Materials and operating principles are identical to the predicate system. No energy source or software components. Designed for use with the Gemini SL Total Knee System.

Indications for Use

Indicated for patients with disability due to degenerative, post-traumatic, or rheumatoid arthritis; avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (patellofemoral erosion, dysfunction, or prior patellectomy); or moderate valgus, varus, or flexion deformities. Also indicated for salvage of failed surgical attempts.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue. July 18, 2019 Waldemar Link GmbH & Co. KG % Terry Powell Regulatory Affairs Program Director LinkBio Corp. 69 King Street Dover, New Jersey 07801 Re: K191755 Trade/Device Name: SPAR-K Instruments (for use with Gemini SL Total Knee System) Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: July 1, 2019 Received: July 1, 2019 Dear Terry Powell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For CAPT Raquel Peat, PhD, MPH, USPHS Director OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K191755 Device Name SPAR-K Instruments (for use with Gemini SL Total Knee System) #### Indications for Use (Describe) The LINK GEMINI SL Total Knee System is indicated for patients suffering from disability due to: - · Degenerative, post-traumatic or rheumatoid arthritis; - · Avascular necrosis of the femoral condyle; · Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; · Moderate valgus, varus or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts. The device is intended for cemented use. Only cementless labeled modular stems are indicated for uncemented use | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | 510(k)<br>Submitter: | Waldemar Link GmbH & Co. KG<br>Barkhausenweg 10<br>D-22339 Hamburg, Germany<br>Phone: +49-40-539950<br>Facility Registration:3003386935 (Oststraße 4-10)<br>Facility Registration: 3007118403 (Harckesheyde 95) | |-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact<br>Person: | Terry Sheridan Powell<br>Regulatory Affairs Program Director<br>LinkBio Corp.<br>69 King Street<br>Dover, NJ 07801<br>973-625-1333 x112 | | Date<br>Prepared: | June 27, 2019 | | Proprietary<br>Name: | SPAR-K Instruments (for LINK GEMINI SL Total Knee System) | | Common<br>Name: | Total Knee Prosthesis (Accessory Instruments) | | Classification<br>Name: | Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented,<br>Polymer/Metal/Polymer (888.3560, JWH, Class II) | | Predicate<br>Device(s): | LINK GEMINI SL Total Knee System Gemini SL Instruments (510(k) #K182872) | | Device | Description: The SPAR-K instruments are a line extension to the instrument system cleared in<br>510(k) #K182872 with the Gemini SL Total Knee System. The SPAR-K<br>instruments are manual orthopedic surgical reusable instruments offered to aid the<br>implantation of the Gemini SL Total Knee System (K182872). The SPAR-K<br>Instruments incorporate design changes for simplicity of use. The modifications do<br>not significantly alter the surgical workflow or technique. Both the original and<br>modified (SPAR-K) instruments accommodate tibia first or femur first workflows<br>according to surgeon preference. The modifications do not change the intended<br>use, or involve any change in technology. The Class II accessory instruments<br>within the SPAR-K Instrument system that are the subjects of this 510(k) are the<br>femoral, tibial, and patellar resection guides. | # Intended The LINK GEMINI SL Total Knee System is indicated for patients suffering from Use: disability due to: {4}------------------------------------------------ - Degenerative, post-traumatic or rheumatoid arthritis: . - Avascular necrosis of the femoral condyle; - . Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy: - Moderate valgus, varus or flexion deformities. . This device may also be indicated in the salvage of previously failed surgical attempts. The device is intended for cemented use. Only cementless labeled modular stems are indicated for uncemented use. ## Technological Characteristics and Substantial Equivalence The modified Class II accessory instruments within the SPAR-K Instrument system that are the subjects of this 510(k) have the same intended use, operating principle, basic device designs and purposes, and materials as the unmodified instruments. The modified cutting quides create the same bone cuts, but feature minor design changes for simplicity and ease of use. A comparison of designs and features supported the substantial equivalence of the modified to the original instruments. Non-clinical and clinical performance testing were not required to demonstrate substantial equivalence. ## Conclusion The subject SPAR-K Instruments that are the subjects of this 510(k) (Class II accessory instruments) are Substantially Equivalent to the predicate instruments identified in this premarket notification.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...