Intended Use

Smith & Nephew Total Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Total Knee Systems and their cleared Indications for Use.

Indications for Cruciate Retaining Cemented Knee Replacement:

  1. Rheumatoid arthritis.
  2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
  3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
    Cruciate Retaining Cemented Knee components are indicated for use with cement and are single use devices.

Indications for Cruciate Retaining Cementless Knee Replacement:

  1. Rheumatoid arthritis.
  2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
  3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
    Cruciate Retaining Cementless Knee components are indicated for use without cement and are single use devices.

Indications for Posterior Stabilized Cemented Knee Replacement:

  1. Rheumatoid arthritis.
  2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
  3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
  4. Primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
    Posterior Stabilized Cemented Knee components are indicated for use with cement and are single use devices.

Indications for Posterior Stabilized Cementless Knee Replacement:

  1. Rheumatoid arthritis.
  2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
  3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
  4. Primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
    Posterior Stabilized Cementless Knee components are indicated for use without cement and are single use devices.

Indications for Constrained Total Knees:

  1. Rheumatoid arthritis.
  2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
  3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
  4. Constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are incompetent.
    Constrained Total Knee components are indicated for use with cement and are single use devices.

Indications for Hinged Total Knees:

  1. Rheumatoid arthritis.
  2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
  3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
  4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
  5. Hinge knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are absent or incompetent.
    Hinged Total Knee components are indicated for use with cement and are single use devices.
Device Description

Subject of this Traditional 510(k) Premarket Notification are the Smith & Nephew, Inc. Total Knee System Instruments. The subject devices are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Total Knee Systems and their cleared Indications for Use. Smith & Nephew Total Knee System Instruments can be organized into instrument families which are categorized as follows: Trials, Cutting Instruments and Cutting Guides; Cutting Blocks, Alignment and Sizing Instruments, Impactors and Handles, Clamps, Extraction, Torque, Instrument Guides, and Covers and Protectors.

AI/ML Overview

This document is a 510(k) Summary of Safety and Effectiveness for the Smith & Nephew Total Knee System Instruments. It explicitly states that these instruments are "accessory devices" used to assist in the implantation of Smith & Nephew Total Knee Systems.

The document does not contain any information regarding acceptance criteria or a study proving the device meets those criteria, nor does it provide details about any AI-related performance metrics.

The content is focused on:

  • Device classification and regulatory information.
  • Listing predicate devices (previously cleared knee systems).
  • Device description (types of instruments).
  • Intended use and Indications for Use for various knee replacement types (Cruciate Retaining, Posterior Stabilized, Constrained, Hinged).
  • FDA clearance letter.

Therefore, I cannot provide the requested information. The document is for mechanical surgical instruments, not an AI/Software as a Medical Device (SaMD) that would typically involve performance studies with ground truth and expert reviews for diagnostic accuracy.

{0}------------------------------------------------

Summary of Safety and Effectiveness Total Knee System Instruments Smith & Nephew, Inc.

K12/393

Date of Summary: May 8, 2011

AUG 7 2012

Contact Person and Address Gino Rouss, MS Group Manager, Regulatory Affairs Smith & Nephew, Inc. Orthopaedic Division 7135 Goodlett Farms Parkway Memphis, Tennessee 38016 Tel: (901) 399-6707 FAX: (901) 566-7080

Name of Device: Total Knee System Instruments Common Name: Orthopaedic Surgical Instrumentation Device Classification Name and Reference:

CFR NumberDescriptionClassification
888.3510Knee joint femorotibial metal/polymer constrained cemented prosthesisII
888.3520Knee joint femorotibial metal/polymer non-constrained cemented prosthesisII
888.3530Knee joint femorotibial metal/polymer semi-constrained cemented prosthesisII
888.3540Knee joint patellofemoral polymer/metal semi-constrained cemented prosthesisII
888.3560Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesisII
888.3565Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesisII

Device Class: Class II Panel Code: Orthopaedics/87

{1}------------------------------------------------

Summary of Safety and Effectiveness Total Knee System Instruments Smith & Nephew, Inc.

Predicate Devices:

Smith & Nephew Inc. Cemented Total Knee Systems

Description510(k)Clearance Date
GENESIS II TOTAL KNEE SYSTEMK9519878/22/1995
GENESIS POROUS TIBIAL TRAYS & REVISIONKNEE SYSTEMK9532742/5/1996
PROFIX POSTERIOR STABILIZED KNEE SYSTEMK9549091/31/1996
GENESIS II POROUS PLUS HA KNEE SYSTEMK03268310/15/2003
GENESIS II DEEP FLEXION CRUCIATE RETAININGARTICULAR INSERTK0418253/11/2005
HIGH PERFORMANCE KNEEK0425153/14/2005
PROFIX FLEX CRUCIATE RETAINING ARTICULARINSERTK0512297/20/2005
JOURNEY CR KNEE SYSTEMK1014998/26/2010
GENESIS II ZIRCONIUM FEMORALCOMPONENT/PROFIX ZIRCONIUM FEMORALCOMPONENTK96255712/5/1996
GENESIS II POSTERIOR STABILIZED HIGH FLEXIONINSERTK0322958/21/2003
PROFIX CONFORMING PLUS TIBIAL INSERTK9462367/6/1995

Smith & Nephew Inc. Cementless Total Knee Systems

Description510(k)Clearance Date
GENESIS II POROUS PLUS HA KNEE SYSTEMK03268310/15/2003
PROFIX TOTAL KNEE SYSTEMK0306235/22/2003
GENESIS II TOTAL KNEE SYSTEMK0306125/27/2003
LEGION POROUS + HA TIBIAL BASEPLATESK1008975/13/2010
LEGION POROUS PRIMARYK07332512/20/2007
LEGION POROUS PLUS HA PRIMARY FEMORALCOMPONENTSK09154312/21/2009
PROFIX FLEX CRUCIATE RETAINING ARTICULARINSERTK0512297/20/2005
GENESIS II POSTERIOR STABILIZED HIGH FLEXIONINSERTK0322958/21/2003
GENESIS II DEEP FLEXION CRUCIATE RETAININGARTICULAR INSERTK0418253/11/2005

{2}------------------------------------------------

Summary of Safety and Effectiveness Total Knee System Instruments Smith & Nephew, Inc.

Smith & Nephew Inc. Constrained Total Knee Systems

Description510(k)Clearance Date
GENESIS II CONSTRAINED KNEE SYSTEMK9621378/2/1996
REVISION KNEE SYSTEMK0434402/18/2005
LEGION COBALT CHROME REVISION KNEESYSTEMK0607425/3/2006
CROSSLINKED POLYETHYLENE ARTICULARINSERTSK0710719/19/2007
LEGION STEM WITH HOLESK07253112/6/2007
PROFIX PLUS TIBIAL INSERT, PROFIX PS PLUSTIBIAL INSERT, PROFIX P/S TIBIAL INSERTK9632551/2/1997
REVISION KNEE SYSTEMK0411067/1/2004

Smith & Nephew Inc. Hinged Total Knee System

Description510(k)Clearance Date
LEGION HINGE KNEE SYSTEMK0811117/23/2008

Device Description

Subject of this Traditional 510(k) Premarket Notification are the Smith & Nephew, Inc. Total Knee System Instruments. The subject devices are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Total Knee Systems and their cleared Indications for Use. Smith & Nephew Total Knee System Instruments can be organized into instrument families which are categorized as follows: Trials, Cutting Instruments and Cutting Guides; Cutting Blocks, Alignment and Sizing Instruments, Impactors and Handles, Clamps, Extraction, Torque, Instrument Guides, and Covers and Protectors.

Intended Use / Indications for Use

Smith & Nephew Total Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Total Knee Systems and their cleared Indications for Use.

Indications for Cruciate Retaining Cemented Knee Replacement

    1. Rheumatoid arthritis.
  • Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 2.
    1. Failed osteotomies, unicompartmental replacement, or total knee replacement.

Cruciate Retaining Cemented Knee components are indicated for use with cement and are single use devices.

{3}------------------------------------------------

Indications for Cruciate Retaining Cementless Knee Replacement

  • Rheumatoid arthritis. ﺎ ﻳﺘﻢ ﺍﻟﻤﺴﺘﻘﺒﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤ
    1. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
  • Failed osteotomies, unicompartmental replacement, or total knee replacement. 3.

Cruciate Retaining Cementless Knee components are indicated for use without cement and are single use devices.

Indications for Posterior Stabilized Cemented Knee Replacement

    1. Rheumatoid arthritis.
    1. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
  • Failed osteotomies, unicompartmental replacement, or total knee replacement. 3.
    1. Primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.

Posterior Stabilized Cemented Knee components are indicated for use with cement and are single use devices.

Indications for Posterior Stabilized Cementless Knee Replacement

    1. Rheumatoid arthritis.
    1. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
    1. Failed osteotomies, unicompartmental replacement, or total knee replacement.
    1. Primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.

Posterior Stabilized Cementless Knee components are indicated for use without cement and afe single use devices.

Indications for Constrained Total Knees

  • Rheumatoid arthritis. 1.
  • Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 2.
  • Failed osteotomies, unicompartmental replacement, or total knee replacement. 3.
  • Constrained knee systems are designed for use in patients in primary and revision surgery, 4. where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are incompetent.

Constrained Total Knee components are indicated for use with cement and are single use devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Smith & Nephew, Inc. % Mr. Gino Rouss Group Manager, Regulatory Affairs 1450 E Brooks Road Memphis, Tennessee 38116

Re: K121393

Trade/Device Name: Smith & Nephew Inc. Total Knee System Instrumentation Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: MBH, JWH, HRY, HSX, KRO, KRQ, KRR, NPJ Dated: May 8, 2012 Received: May 9, 2012

Dear Mr. Rouss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

7 2012 AUG

{5}------------------------------------------------

Page 2 - Mr. Gino Rouss

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark A. Milliken

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known):

Device Name: Smith & Nephew Inc. Hinged Total Knee Systems

Indications for Use:

Smith & Nephew Hinge Total Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Hinge Total Knee Systems and their cleared' Indications for Use.

Indications for Hinged Total Knees

    1. Rheumatoid arthritis.
    1. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
  • Failed osteotomies, unicompartmental replacement, or total knee replacement. 3.
  • Posterior stabilized knee systems are designed for use in patients in primary and revision বা surgery where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
  • Hinge knee systems are designed for use in patients in primary and revision surgery, where 5. the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are absent or incompetent.

Hinged Total Knee components are indicated for use with cement and are single use devices.

Over-The-Counter Use x AND/OR Prescription Use (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Andut

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K121393

Page of

{7}------------------------------------------------

510(k) Number (if known):

Device Name: Smith & Nephew Inc. Constrained Total Knee Systems

Indications for Use:

inditorient to revew Constrained Total Knee Instruments are accessory devices and are intended to Smith & Rophen and and of Smith & Nephew Constrained Total Knee Systems and their cleared Indications for Use.

Indications for Constrained Total Knees

  • Rheumatoid arthritis. 1.
  • Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 2.
  • Failed osteotomies, unicompartmental replacement, or total knee replacement. 3.
  • r uned restored thee systems are designed for use in patients in primary and revision surgery, 4. where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are incompetent.

Constrained Total Knee components are indicated for use with cement and are single use devices.

Prescription Use(Part 21 CFR 801 Subpart D)X AND/OR
Over-The-Counter Use(21 CFR 807 Subpart C)

(Please do not WRITE BELOW THIS LINE -- Continue on ANOTHER PAGE IF NEÉDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number_________________________________________________________________________________________________________________________________________________________________

Page_ ੀ

{8}------------------------------------------------

510(k) Number (if known):

Device Name: Smith & Nephew Inc. Posterior Stabilized Cementless Knee Systems

Indications for Use:

Smith & Nephew Posterior Stabilized Cementless Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Posterior Stabilized Cementless Knee Systems and their cleared Indications for Use.

Indications for Posterior Stabilized Cementless Knee Replacement

    1. Rheumatoid arthritis.
  • Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 2.
    1. Failed osteotomies, unicompartmental replacement, or total knee replacement.
    1. Primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.

Posterior Stabilized Cementless Knee components are indicated for use without cement and are single use devices.

Prescription Use x (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Aisul

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K121393

Page _________________________________________________________________________________________________________________________________________________________________________ of

{9}------------------------------------------------

510(k) Number (if known):

Device Name: Smith & Nephew Inc. Posterior Stabilized Cemented Knee Systems

Indications for Use:

Smith & Nephew Posterior Stabilized Cemented Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Posterior Stabilized Cemented Knee Systems and their cleared Indications for Use.

Indications for Posterior Stabilized Cemented Knee Replacement

  • Rheumatoid arthritis. 1.
  • Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 2.
    1. Failed osteotomies, unicompartmental replacement, or total knee replacement.
    1. Primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intáct.

Posterior Stabilized Cemented Knee components are indicated for use with cement and are single use devices.

AND/OR

Prescription Use x (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

As

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K121393

Page of

{10}------------------------------------------------

510(k) Number (if known):

Device Name: Smith & Nephew Inc. Cruciate Retaining Cementless Knee Systems

Indications for Use:

Smith & Nephew Cruciate Retaining Cementless Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Cruciate Retaining Cementless Knee Systems and their cleared Indications for Use.

Indications for Cruciate Retaining Cementless Knee Replacement

    1. Rheumatoid arthritis.
  • Post-traumatic arthritis, osteoarthritis, or degenerative arthritis. 2.
    1. Failed osteotomies, unicompartmental replacement, or total knee replacement.

AND/OR

Cruciate Retaining Cementless Knee components are indicated for use without cement and are single use devices.

Prescription Use × (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K121393

Page of

{11}------------------------------------------------

510(k) Number (if known):

Device Name: Smith & Nephew Inc. Cruciate Retaining Cemented Knee Systems

Indications for Use:

Smith & Nephew Cruciate Retaining Cemented Knee Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Cruciate Retaining Cemented Knee Systems and their cleared Indications for Use.

Indications for Cruciate Retaining Cemented Knee Replacement

    1. Rheumatoid arthritis.
    1. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
    1. Failed osteotomies, unicompartmental replacement, or total knee replacement.

Cruciate Retaining Cemented Knee components are indicated for use with cement and are single use devices.

Prescription Use(Part 21 CFR 801 Subpart D)X
AND/OR Over-The-Counter Use(21 CFR 807 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE – Continue on ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K121393