Stitch Cerclage – Suture Tapes

K243905 · GM Dos Reis Industria e Comercio Ltda. · JDQ · Aug 14, 2025 · Orthopedic

Device Facts

Record IDK243905
Device NameStitch Cerclage – Suture Tapes
ApplicantGM Dos Reis Industria e Comercio Ltda.
Product CodeJDQ · Orthopedic
Decision DateAug 14, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3010
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Stitch Cerclage - Suture Tapes are indicated in orthopedic surgeries as bone fixation cerclage, specifically: - Sternotomy indications including the "rewiring" of osteomized sternums; - Repair of long bone fractures due to trauma or reconstruction.

Device Story

Stitch Cerclage - Suture Tapes are orthopedic implants used for internal fixation of discontinued bone structures. The system consists of UHMWPE suture tapes and AISI300 steel needles (ASTM F899) mounted on an ABS loader. Used by surgeons in clinical settings for sternotomy rewiring and long bone fracture repair. The device provides mechanical stabilization of bone segments. It is not intended for use with other bone fixation implants like plates. The device is provided sterile via Ethylene Oxide.

Clinical Evidence

Bench testing only. No clinical data included. Mechanical performance (chemical composition, creep, tension static, and fatigue tests) was evaluated using methodologies derived from literature (Hagerich et al., 2022) and compared to the predicate device.

Technological Characteristics

Materials: UHMWPE suture tapes, AISI300 steel needles (ASTM F899), ABS loader. Principle: Mechanical bone fixation/cerclage. Sterilization: Ethylene Oxide. No software or electronic components.

Indications for Use

Indicated for orthopedic surgery patients requiring bone fixation cerclage, specifically for sternotomy (rewiring osteomized sternums) and repair of long bone fractures resulting from trauma or reconstruction.

Regulatory Classification

Identification

A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION August 14, 2025 GM Dos Reis Industria e Comercio Ltda. Guilherme Esteves Pontes Regulatory Affairs Analyst Avenida Pierre Simon de Laplace, 600 Campinas, SP 13069320 Brazil Re: K243905 Trade/Device Name: Stitch - Cerclage Suture System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: JDQ Dated: July 16, 2025 Received: July 16, 2025 Dear Guilherme Esteves Pontes: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1} K243905 - Guilherme Esteves Pontes Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K243905 - Guilherme Esteves Pontes Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # CHRISTOPHER FERREIRA -S Christopher Ferreira, M.S. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K243905 Device Name Stitch Cerclage - Suture Tapes Indications for Use (Describe) Stitch Cerclage - Suture Tapes are indicated in orthopedic surgeries as bone fixation cerclage, specifically: - Sternotomy indications including the "rewiring" of osteomized sternums; - Repair of long bone fractures due to trauma or reconstruction. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} # GMReis Qualidade para Vida K243905 - Stitch Cerclage - Suture Tapes # Section 5 - 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92. # I. Submitter: GM Dos Reis Industria e Comercio Ltda Avenida Pierre Simon de La Place 600 Campinas, São Paulo, Brazil 13069-320 Guilherme Esteves Pontes, Senior Regulatory Affairs Analyst Telephone: +55 (19) 3765-9900 Email: guilherme.qualidade@gmreis.com.br Date prepared: December 16, 2024 # II. Device Name: Trade Name: Stitch Cerclage - Suture Tapes Common Name: Cerclage, Fixation Classification Name: Bone fixation cerclage Device Class: II Product Codes: JDQ Regulation Number: 21 CFR 888.3010 # III. Predicate Devices: Legally marketed device to which we are claiming "Substantial Equivalence" are the following: Arthrex FiberTape and TigerTape Cerclage Sutures - Arthrex (K221485) (Predicate Device) Suture Anchors - HTA Headless Titanium Anchor and ZIP Anchors - GMReis (K223114) (Reference Device) EXPERT - Joint Fixation System (K200332) (Reference Device) # IV. Device Description: Stitch Cerclage - Suture Tapes are indicated in orthopedic surgeries for internal fixation of discontinued bone structures. The product are not to be used in conjunction with other bone fixation implants, like plates. The surgical sutures and tapes are made of UHMWPE (ultra-high molecular weight polyethylene), the needles are made in AISI300 steel series, according to the standard ASTM F899 and cerclage is assembled on an GM dos Reis Industria e Comercio Ltda. Pierre Simon de Laplace Ave., 600, Block 3F9677 Techno Park, Campinas, SP, Brazil, Zip Code 13069320 Phone: +551937659900, Email: gmreis@gmreis.com.br Website: www.gmreis.com.br {5} # GMReis Qualidade para Vida K243905 - Stitch Cerclage - Suture Tapes ABS loader, additional instruments may be used as well. Those materials are identical to those used in other cleared GMReis devices like the reference device cleared under K223114. It is provided in sterile condition sterilized by Ethylene Oxide. ## V. Statement of Indications for Use of the Device: Stitch Cerclage - Suture Tapes are indicated in orthopedic surgeries as bone fixation cerclage, specifically: - Sternotomy indications including the "rewiring" of osteomized sternums; - Repair of long bone fractures due to trauma or reconstruction. ## VI. Comparison of Technological Characteristics with The Predicate Device: The subject device is substantially equivalent in indications and design principles to the following predicate device: K221485 - Arthrex FiberTape and TigerTape Cerclage Sutures - Arthrex Inc. The subject and predicate device have equivalent intended use and equivalent technological characteristics. Both devices are manufactured from identical materials and share equivalent design characteristics as well as physical dimensions. Any difference in technological characteristics do not raise new issues of safety or efficacy. The performance of the subject device was demonstrated through mechanical testing according to literature and predicate comparison. No clinical data were included in this submission. The subject device is provided sterile by Ethylene Oxide method. ## VII. Performance Data: The chemical composition, creep test, tension static and fatigue tests conducted with the products were based on the article "Cerclage Performance Analysis – a Biomechanical Comparison of Different Techniques and Materials", L. M. Hagerich, et al, Musculoskeletal Disorders (2022) and compared to the predicate device. The methodology of this literature was used because there is no international standard in the ISO – International Standardization Organization nor in the ASTM – American Standardization for Testing and Materials for testing these types of products. We therefore consider the subject device equivalent to its predicate device in the mechanical tests performed. ## VIII. Conclusions: As was established in this submission, the subject Stitch Cerclage - Suture Tapes are substantially equivalent to the predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics, intended use, indications for use, material composition, anatomical region, multiple sizes, and basic design features GM dos Reis Industria e Comercio Ltda. Pierre Simon de Laplace Ave., 600, Block 3F9677 Techno Park, Campinas, SP, Brazil, Zip Code 13069320 Phone: +551937659900, Email: gmreis@gmreis.com.br Website: www.gmreis.com.br {6} GMReis Qualidade para Vida K243905 - Stitch Cerclage - Suture Tapes compared to its predicate devices. Any differences between the subject and the predicate devices are considered minor and do not raise different questions of safety or effectiveness. GM dos Reis Industria e Comercio Ltda. Pierre Simon de Laplace Ave., 600, Block 3F9677 Techno Park, Campinas, SP, Brazil, Zip Code 13069320 Phone: +551937659900, Email: gmreis@gmreis.com.br Website: www.gmreis.com.br
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...