The VICTUS Femtosecond Laser Platform is indicated for use for: the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea. for anterior capsulotomy during cataract surgery. the creation of cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring cuts / incisions in the cornea. laser-assisted lens fragmentation of nuclear cataracts during cataract surgery, not for fragmentation of posterior subcapsular (PSC) and cortical cataracts
Device Story
VICTUS Femtosecond Laser Platform is a precision ophthalmic surgical laser for anterior segment procedures. System uses Yb:KYW femtosecond laser to create Laser-Induced Optical Breakdown (LIOB) in corneal tissue. Physician operates system via touchscreen, user controls, and footswitch pedals while patient is supine. Integrated Fourier-based OCT provides real-time visualization and positioning overlay on video microscope feed. Sterile, single-use Patient Interface (PI) kit—comprising a contact lens and suction clip—immobilizes the eye, provides a reference surface for depth control, and couples the laser beam. System includes assistant workstation for data management, chiller, controller, and treatment illumination. Output consists of scanned laser patterns for flaps, incisions, capsulotomy, and lens fragmentation. Device assists surgeons in performing precise ophthalmic surgeries, potentially improving procedural outcomes through automated depth control and visualization.
Clinical Evidence
No clinical data was provided. Substantial equivalence was established through bench testing comparing the modified Patient Interface (PI) kit to the predicate. Bench testing demonstrated equivalent cut quality, pattern fidelity, and docking performance (suction integrity/IOP safety) for all cleared indications.
Technological Characteristics
Yb:KYW femtosecond laser source; LIOB mechanism. Patient Interface (PI) kit uses injection-molded COP Zeonex contact lens (10.5 mm radius, 1.53 refractive index) and suction clip. Integrated Fourier-based OCT for visualization. Complies with EN ISO 60601-1, 60601-1-2, and 60601-2-22. Sterilization via gamma irradiation.
Indications for Use
Indicated for patients undergoing LASIK, cataract surgery, or other ophthalmic procedures requiring corneal incisions, corneal flaps, anterior capsulotomy, or nuclear cataract lens fragmentation. Contraindicated for fragmentation of posterior subcapsular (PSC) and cortical cataracts.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below.
September 1, 2020
Technolas Perfect Vision GmbH % Ken Nehmer Director, Regulatory Affairs 351 Buena Vista Ave. E, Unit 501E San Francisco, CA 94117
Re: K200724
Trade/Device Name: VICTUS Femtosecond Laser Platform Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic laser Regulatory Class: Class II Product Code: OOE Dated: March 9, 2020 Received: March 19, 2020
Dear Ken Nehmer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for LT Charles Chiang Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K200724
Device Name VICTUS Femtosecond Laser Platform
Indications for Use (Describe) The VICTUS Femtosecond Laser Platform is indicated for use for:
· the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea.
• for anterior capsulotomy during cataract surgery.
· the creation of cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring cuts / incisions in the cornea.
· laser-assisted lens fragmentation of nuclear cataracts during cataract surgery, not for fragmentation of posterior subcapsular (PSC) and cortical cataracts
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| I. | Submitter | Technolas Perfect Vision GmbH<br>Messerschmittstr. 1 +3<br>80992 Munich, Germany<br>Tel: 49 174 3494062 |
|----|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact | Ken Nehmer<br>Director, Regulatory Affairs<br>Technolas Perfect Vision GmbH<br>351 Buena Vista Ave. E<br>Unit 501E<br>San Francisco, CA 94117<br>Ph: 415-297-0408<br>Email: ken.nehmer@bausch.com |
| | Date Prepared | September 1, 2020 |
#### II. Device
| Name of Device | VICTUS Femtosecond Laser Platform |
|-------------------------|-------------------------------------------------------------------------------------------------------------------|
| Common Name | Ophthalmic Laser |
| Classification Name | Laser, Ophthalmic |
| Regulatory Class | Class II |
| Product Code | OOE (Ophthalmic Femtosecond Laser) |
| Regulation Number | 21 CFR 886.4390 |
| III. Predicate Device | VICTUS Femtosecond Laser Platform (K171014)<br>This predicate has not been subject to a design-related<br>recall. |
## IV. Device Description
The VICTUS Femtosecond Laser is a precision ophthalmic surgical laser indicated for use in patients undergoing ophthalmic surgery in the anterior segment of the eye. The VICTUS Femtosecond Laser system produces scanned patterns designed to create flaps, produce corneal incisions, arcuate incisions, capsulotomy cuts, and crystalline lens fragmentation patterns for use in cataract surgery. The patient is treated while supine on a patient bed that the physician can position electromechanically via simple controls. The physician controls the
{4}------------------------------------------------
VICTUS laser system itself through a touchscreen display and other user controls. Laser emission and of PI suction vacuum are actuated by the physician using separate footswitch pedals.
The mode of operation for the VICTUS Platform is using Yb:KYW Femtosecond laser to produce Laser-Induced Optical Breakdown (LIOB) of the corneal tissue. Scanned patterns of laser pulses from the VICTUS are delivered through a sterile, disposable Patient Interface (PI). The PI consists of two components: (i) a contacting lens, and (ii) a suction clip. Together, the suction clip and contacting lens connect to make a single assembly during a VICTUS procedure. The surface of the contacting lens that contacts the eye is spherically curved to match the curvature of the human eye. The suction clip assembly gently immobilizes the eye with respect to the VICTUS laser beam and optical visualization systems. The PI contacting lens and suction clip together allow for proper optical coupling of the laser and optical paths into the patient's eye and create a precise opto-mechanical reference surface for depth control of the scanning laser beam.
| Component | Description |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Assistant workstation | Allows to perform the system tests, to enter and manage the patient data, to select and program procedures, to apply vacuum for patient interface and suction clip, to adjust the procedure according to the patient's eye and to export treatment records. |
| Chiller | Helps to ensure that the main components are kept at a steady temperature. |
| Controller | Checks the connectivity among all electronic components. |
| GUI PC | Interface between the user and the laser system. |
| Laser source | Generates the laser beam. |
| Optical unit | Controls the complete laser beam path. |
| Surgeon control screen | Shows a live camera image, a live OCT image, and all relevant parameters for the selected treatment. |
| Treatment illumination | Contains the laser ring light module. It illuminates the treatment area. |
| Video microscope | Used as part of treatment planning and shows the top view of the eye to be treated. |
| OCT | Fourier based integrated optical coherence tomography unit which is used to visualize and/or position the treatment as an overlay on the video image. The OCT is not used for diagnostic purposes. |
The VICTUS Femtosecond Laser Platform main unit is composed of the following:
{5}------------------------------------------------
| Additional feature | Description |
|-----------------------------------------------------|-------------------------------------------------------------------------|
| Surgical microscope (optional feature) | Allows for centering the suction clip |
| Surgical microscope illumination (optional feature) | Provides a uniform and adjustable illumination |
| Patient Bed | Used to position a supine patient for treatment by the VICTUS Platform. |
The following components are additional features to the main unit:
The following components are accessories to the main unit:
| Accessory | Description |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient Interface Kit (PIK) | The patient interface kit is sterile, single use kit<br>composed of a contact lens component and a suction<br>clip component. The contact lens and suction clip<br>assembly create a reference surface for depth control<br>and fix the eye relative to the delivery of the laser beam. |
#### V. Indications for Use
The VICTUS Femtosecond Laser Platform is indicated for use for:
- · the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea.
- · for anterior capsulotomy during cataract surgery.
- · the creation of cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring cuts / incisions in the cornea.
- · laser-assisted lens fragmentation of nuclear cataracts during cataract surgery, not for fragmentation of posterior subcapsular (PSC) and cortical cataracts
The indications for use remain unchanged from those previously cleared via K171014.
# VI. Comparison of Technological Characteristics with the Predicate
Modifications to the VICTUS Femtosecond Laser described in this premarket notification are limited to the contacting lens component of the Patient Interface accessory and to the PI mating hardware (i.e. the Spacer Cone) that is permanently installed on the VICTUS laser. No other modifications are proposed to the cleared device.
A comparison of the technological characteristics for both the current and modified Patient Interface Kit components are provided in the following table.
{6}------------------------------------------------
| CLEARED DEVICE<br>K171014 | | PROPOSED<br>DEVICE |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| Trade Name –<br>Laser System | VICTUS Femtosecond Laser Platform | Same |
| Indication for Use | The VICTUS Femtosecond Laser Platform is<br>indicated for use for:<br>the creation of a corneal flap in patients<br>undergoing LASIK surgery or other<br>treatment requiring initial lamellar resection<br>of the cornea.<br>for anterior capsulotomy during cataract<br>surgery.<br>the creation of cuts / incisions in the cornea<br>in patients undergoing cataract surgery or<br>other ophthalmic treatment requiring cuts /<br>incisions in the cornea.<br>laser-assisted lens fragmentation of nuclear<br>cataracts during cataract surgery, not for<br>fragmentation of posterior subcapsular<br>(PSC) and cortical cataracts | Same |
| Trade Name –<br>Patient Interface Kit | VERAFIT Patient Interface Kit<br>(SKU: 90000200TPV) | Patient Interface (PI) Kit<br>(SKU: 90000145) |
| Function | Sterile, single use accessory designed to<br>(i) provide reference surface for laser depth<br>control and (ii) affix eye for delivery of laser<br>beam | Same |
| Usage | Sterile, single use, disposable | Same |
| Accessory Design | Two components:<br>contact lens<br>suction assembly: spring clip with suction<br>ring, tubing, vacuum reservoir w/ console<br>connector | Same |
| Eye Fixation Method | Suction-fixated, curved contact | Same |
| Applanation Force | Shear Force: ≤ 2 N<br>Vertical Force:<br>Cataract indications ("soft dock"): 0.2 – 1.0 N<br>Flaps/incisions ("hard dock"): 2.0 – 5.0 N | Same |
| Sterilization Method | Gamma irradiation | Same |
| Suction Clip Component<br>Design | Limbal suction ring with spring clip | Same |
| Suction Method | Integrated suction ring with vacuum tubing<br>and connector | Same |
| | CLEARED DEVICE<br>K171014 | PROPOSED<br>DEVICE |
| | Contacting Lens Component | |
| Manufacturing method | Machined<br>polymeric contacting lens | Injection molded<br>polymeric contacting lens |
| Contact Lens Material | PMMA | COP Zeonex |
| Distal (eye-contacting)<br>radius of curvature | 9.5 mm | 10.5 mm |
| Proximal<br>radius of curvature | 17 mm | Same |
| Index of refraction | 1.49 | 1.53 |
| Maximum treatment<br>diameter | 12.0 mm | 11.7 mm |
| Outside diameter of<br>contacting surface | 15.6 mm | same |
## TECHNOLOGICAL COMPARISON OF THE CLEARED AND PROPOSED PATIENT INTERFACE KIT COMPONENTS
{7}------------------------------------------------
## VII. Performance Data
## Biocompatibility testing
Biocompatibility assessment for the modified Patient Interface contacting lens component of the PI Kit has been performed in accordance with the requirements of ISO 10993-1 to evaluate the impact of the patient contacting material modification. Test results satisfied the acceptance criteria as defined by the associated ISO standards.
## Limulus Amebocyte Lysate (LAL) testing
The VICTUS Patient Interface (PI) Kit includes a patient interface contacting lens and the suction clip ring which contact the corneal surface only (non-intraocular). Limulus Amebocyte Lysate (LAL) Endotoxin testing had been performed with results of < 1.0 EU/sample. Results are below the defined endotoxin limit for a medical device (non-intraocular) of 20 EU/medical device.
## Electrical safety and electromagnetic compatibility (EMC)
No new electrical safety or EMC testing was required to support this premarket notification. The VICTUS Femtosecond Laser Platform has previously undergone testing and complies with the applicable following safety standards:
| STANDARD | TITLE |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| EN ISO 60601-1: 2005 + A1 | Medical electrical equipment – Part I: General requirements for safety |
| EN ISO 60601-1-2: 2007 | Medical electrical equipment – Part 1: General requirements for safety; 2. Collateral standard: electromagnetic compatibility; requirements and tests |
| EN ISO 60601-2-22: 2007 | Medical electrical equipment – Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipment |
{8}------------------------------------------------
The VICTUS Femtosecond Laser Platform has been found to perform equivalently to the predicate device. The VICTUS Femtosecond Laser Platform and the predicate device therefore have a similar performance profile.
# Software Verification and Validation Testing
No new software testing was performed to support this premarket notification
## Mechanical and acoustic testing
There was no mechanical or acoustical testing performed to support substantial equivalence of this premarket notification.
# Animal Study
There was no animal study performed to support substantial equivalence of this premarket notification.
# Non-Clinical Performance Data
Nonclinical performance testing was conducted to establish substantial equivalence between the Patient Interface (PI) Kit (SKU: 90000145) with the modified Patient Interface contacting lens and the PI Kit (SKU: 90000200TPV) cleared under K171014.
All bench testing results demonstrated equivalent performance between the modified and the current 510(k)-cleared PI Kits. Cuts made using the modified PI Kit were shown to be equivalent in terms of both cut pattern fidelity and cut quality to those produced with the current 510(k)cleared PI for all cleared cataract and corneal indications for use. The docking behavior in terms of IOP safety and limbal suction integrity are equivalent between the current 510(k)-cleared and the modified (proposed) PI Kit models.
## Clinical Studies
There was no clinical study performed to support substantial equivalence of this premarket notification.
## VIII. Conclusion
The minor modifications proposed in this 510(k) Premarket Notification did not impact the product design requirements or the conformance to applicable standards specific to this device. Any differences between the predicate and proposed devices do not affect the substantial equivalence of the device as demonstrated by the performance testing.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.