K Number
K122386

Validate with FDA (Live)

Date Cleared
2013-02-08

(186 days)

Product Code
Regulation Number
886.4390
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VICTUS Femtosecond Laser Platform is indicated for use in:

  • the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea
  • for anterior capsulotomy during cataract surgery
  • the creation of penetrating arcuate cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea.
Device Description

The VICTUS Femtosecond Laser Platform (hercafter referred to as the VICTUS Platform) is a precision ophthalmic surgical laser indicated for use in patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea.
The VICTUS Platform is an ophthalmic surgical femtosecond laser designed for cutting a penetrating arcuate cut of pre-selected depth, radius, opening angle, and position angle. The system works by first being programmed with the depth, radius, opening angle and position angle at which the penetrating arcuate cut is desired. The surgeon then fixates the eye with a sterile disposable PMMA contact lens (hercafter referred to as the Patient Interface) that is connected to the laser via a vacuum tube. Depths of -120µm, radius of 1500 - 4750 um, opening angles of 1-360°, and position angles of 0 - 359° are available.
Laser pulses are delivered through the sterile Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eye relative to the delivery of the laser beam. Surgical effects are produced by scanning thousands of individual pulses, producing continuous incisions. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the VICTUS Femtosecond Laser Platform, based on the provided text:


1. Acceptance Criteria and Reported Device Performance

The provided text does not explicitly state quantitative acceptance criteria in a table format for the arcuate cuts/incisions. However, the performance data section implies the following outcomes were evaluated and met a criteria of being "highly reproducible and accurate." The postmarket surveillance review indicates "comparability to the standard manual procedure" for clinical endpoints.

Acceptance Criteria (Implied/Evaluated)Reported Device Performance (VICTUS Femtosecond Laser Platform)
Bench/Pre-Clinical Testing:
Reproducibility of rim depthHighly reproducible
Accuracy of rim depthAccurate
Reproducibility of diameterHighly reproducible
Accuracy of diameterAccurate
Reproducibility of open angleHighly reproducible
Accuracy of open angleAccurate
Reproducibility of position angleHighly reproducible
Accuracy of position angleAccurate
Reproducibility of centrationHighly reproducible
Accuracy of centrationAccurate
Visual comparison to predicate devicesQualitatively comparable (based on scanning electron microscopy)
Postmarket Surveillance Review (Clinical):
Post-surgical loss of lines of BCDVAComparable to standard manual procedure
Post-surgical overall change in refractionComparable to standard manual procedure
Postoperative induced cylinderComparable to standard manual procedure
Adverse events (rate and type)Comparable to standard manual procedure
Serious adverse events (rate and type)Comparable to standard manual procedure

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 49 eyes from 34 subjects.
  • Data Provenance: Retrospective, single site in Europe.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The text states that "all surgeries were performed by the same surgeon" at the single clinical location. This surgeon's performance was compared between manual and laser-assisted procedures. While this surgeon's outcomes define the "ground truth" for the comparison within the study, the text does not specify the number of experts or their qualifications that established an independent "ground truth" for evaluating the quality of the surgical outcomes themselves. The comparison is against the surgeon's own manual technique.


4. Adjudication Method for the Test Set

The provided text does not specify an adjudication method (e.g., 2+1, 3+1). The "ground truth" appears to be the clinical outcomes as assessed by the operating surgeon and the postmarket surveillance review.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly described. The study compared laser-assisted procedures to manual procedures performed by the same surgeon at a single site. The effect size of how much human readers improve with AI vs. without AI assistance is not applicable in this context as the device is a surgical laser, not an AI diagnostic tool.


6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) was done

Yes, a standalone performance evaluation was conducted in the Non-Clinical Performance Data section. This involved bench testing using various materials (porcine eyes, agarose gel, PETG, and PMMA) and scanning electron microscopy to assess the laser's cutting accuracy and reproducibility. This evaluate the device's inherent mechanical and cutting capabilities independent of human intervention in the cutting process itself.


7. The Type of Ground Truth Used

  • For Non-Clinical Performance Data: The ground truth was established through measurements (rim depth, diameter, open angle, position angle, centration) and visual assessment (scanning electron microscopy) of the cuts produced in various test materials (porcine eyes, agarose gel, PETG, PMMA).
  • For Postmarket Surveillance Clinical Review: The ground truth was based on clinical outcomes data (post-surgical loss of best corrected distance visual acuity (BCDVA), post-surgical overall change in refraction, postoperative induced cylinder, adverse events, and serious adverse events) observed in patients after both laser-assisted and manual arcuate incisions, as assessed by the operating surgeon.

8. The Sample Size for the Training Set

The provided text does not mention a training set or details about its sample size. This device is a surgical laser, and the studies described are performance evaluations, not the development or training of an AI algorithm.


9. How the Ground Truth for the Training Set Was Established

As no training set is mentioned (since this is not an AI/algorithm-driven device in the context of typical AI device submissions), this question is not applicable.

{0}------------------------------------------------

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Applicant:Technolas Perfect Vision GmbHMesserschmittstr. 1.+380992 Munich, GermanyTel: 49 174 3494062
Contact:Betty M. Johnson、Sr. Manager, Regulatory AffairsTechnolas Perfect Visionc/o PO Box 17190Anaheim, CA 92817Ph: (714) 273-8254Fax: (714) 855-2794Email : b.johnson@technolaspv.com
Trade Name:Common Name:Classification Name:Device Classification:Product CodeVICTUS Femtosecond Laser PlatformOphthalmic LaserLaser, OphthalmicClass IIOOE (Ophthalmic Femtosecond Laser)HQF (Laser, Ophthalmic)

Substantially Equivalent To:

510(k) NumberProduct Trade NameCurrent Manufacturer
K120426VICTUS Femtosecond LaserPlatformTechnolas Perfect Vision
K110427Femtec Laser System for AnteriorCapsulotomyTechnolas Perfect VisionGmbH
K113151iFS Laser SystemAbbott Medical Optics
K101626LenSx LaserLenSx, an Alcon company

{1}------------------------------------------------

Description of the Device Subject to Premarket Notification:

The VICTUS Femtosecond Laser Platform (hercafter referred to as the VICTUS Platform) is a precision ophthalmic surgical laser indicated for use in patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea.

The VICTUS Platform is an ophthalmic surgical femtosecond laser designed for cutting a penetrating arcuate cut of pre-selected depth, radius, opening angle, and position angle. The system works by first being programmed with the depth, radius, opening angle and position angle at which the penetrating arcuate cut is desired. The surgeon then fixates the eye with a sterile disposable PMMA contact lens (hercafter referred to as the Patient Interface) that is connected to the laser via a vacuum tube. Depths of -120µm, radius of 1500 - 4750 um, opening angles of 1-360°, and position angles of 0 - 359° are available.

Laser pulses are delivered through the sterile Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eye relative to the delivery of the laser beam. Surgical effects are produced by scanning thousands of individual pulses, producing continuous incisions. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location.

Indications for Use:

The VICTUS Platform is indicated for use in:

  • the creation of a corneal flap in patients undergoing LASIK surgery or other treatment . requiring initial lamellar resection of the cornea
  • for anterior capsulotomy during cataract surgery .
  • the creation of penetrating arcuate cuts / incisions in the cornea in patients undergoing ● cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea.

Technical Characteristics Comparison:

The design principle of the VICTUS Platform is fundamentally the same as the predicate devices. The VICTUS Platform mode of operation is the same as the previously cleared VICTUS Femtosecond Laser Platform cleared in K120426, iFS Laser System cleared in K113151, and the LenSx Laser System cleared in K101626, all of which deliver femtosecond pulses to produce a pattern of photodisruption to create cuts / separation in ophthalmic tissue. The VICTUS Platform delivers femtosecond pulses to produce a pattern of photodisruption for creation of penetrating arcuate cuts / incisions in the cornea, as do the predicate devices.

The means of fixation of the patient contact portion of the VICTUS Platform is substantially equivalent to that present in the VICTUS Femtosecond Laser Platform cleared in K120426, the Femtec Laser System for Anterior Capsulotomy cleared in K110427, iFS Laser System cleared in K113151, and the LenSx Laser cleared in K101626. All of these predicate systems use suction vacuum to affix a suction ring to the corneal surface prior to use.

{2}------------------------------------------------

The VICTUS Platform mode of operation and the technology used to create the cutting action are similar or identical to these previously mentioned devices, and therefore substantially equivalent to these legally marketed predicate devices.

Performance Data:

The VICTUS Femtosecond Laser Platform has undergone testing and is in compliance with applicable safety standards as listed in the following table.

StandardTitle
EN ISO 60601-1Medical electrical equipment - Part I: General requirementsfor safety
EN ISO 60601-1-2Medical electrical equipment - Part 1: General requirementsfor safety; 2. Collateral standard: electromagneticcompatibility; requirements and tests
EN ISO 60601-1-4Medical electrical equipment - Part 1-4: General requirementsfor safety; Collateral standard: Programmable electricalmedical systems
EN ISO 60601-2-22Medical electrical equipment – Part 2: Particular requirementsfor the safety of diagnostic and therapeutic laser equipment

The VICTUS Femtosecond Laser Platform has been found to perform equivalently to the predicate devices in patients undergoing ophthalmic surgery or other treatment requiring penetrating arcuate cuts / incisions in the cornea. The VICTUS Femtosecond Laser Platform and the predicate devices therefore have similar performance profiles.

Non-Clinical Performance Data

A variety of test procedures were conducted to demonstrate the performance of the proposed VICTUS Platform in support of this premarket submission. The collected data were evaluated by comparing the mean values to the specified acceptance criteria and their 95% confidence intervals. Four different materials were used for the bench performance testing; porcine eyes, agarose gel, polyethylene terephthalate (PETG), and polymethyl methacrylate (PMMA). Scanning electronic microscopy has also been utilized to visually assess the results of the arcuate cuts / incisions procedures with the VICTUS Platform. The resulting micrographs are compared qualitatively with images obtained from the predicate devices.

The testing showed that laser-assisted arcuate cuts / incisions performed with the VICTUS Femtosecond Laser Platform resulted in highly reproducible and accurate rim depth, diameter, open angle, position angle, and centration.

{3}------------------------------------------------

K122386

Postmarket Surveillance Clinical Review

A retrospective postmarket surveillance review was performed at a single site in Europe to compare manual arcuate incisions using a diamond blade to laser-assisted arcuate incisions using the VICTUS™ Platform. This review was conducted at a single clinical location, and all surgeries were performed by the same surgeon.

A total of 49 eyes in 34 subjects were included in the analysis. The mean age of the study population was 57.4 ± 11.7 years (range: 18 to 84 years). Twenty four (24) eyes received laserassisted arcuate cuts, of which 11 received arcuate cuts alone and 13 received both arcuate cuts and cataract surgery. Twenty five (25) eyes received manual arcuate cuts, of which four received arcuate cuts alone and 21 received both arcuate cuts and cataract surgery. One adverse event was reported in the laser-assisted group, which was considered by the surgeon to not be related to the laser-assisted procedure.

The laser-assisted arcuate cuts procedure was shown to be comparable to the standard manual procedure when assessed for common clinical endpoints such as post-surgical loss of lines of best corrected distance visual acuity (BCDVA), post-surgical overall change in refraction, postoperative induced cylinder, adverse events, and serious adverse events.

The postmarket surveillance review data confirm that laser-assisted arcuate cuts are comparable to the standard manual procedure.

Basis for Determination of Substantial Equivalence:

The technological characteristics of the VICTUS Femtosecond Laser Platform are substantially equivalent to the technological characteristics of the VICTUS Femtosecond Laser Platform (K120426), the Femtec Laser System for Anterior Capsulotomy (K110427), iFS Laser System (K113151), the LenSx Laser (K101626).

The proposed additional indication for use of arcuate cuts / incisions for the VICTUS Femtosecond Laser Platform is very similar to the indications for use cleared for the iFS Femtosecond Laser in K113151 and for the LenSx Laser in K101426; both include arcuate corneal incisions.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.

February 8, 2013

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Technolas Perfect Vision GmbH % Ms. Betty M. Johnson Senior Manager, Regulatory Affairs, Americas PO Box 17190 Anaheim Hills, CA 92817

Re: K122386

Trade/Device Name: VICTUS Femtosecond Laser Platform Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Femtosecond Laser Regulatory Class: Class II Product Code: OOE, HQF Dated: February 1, 2013 Received: February 4, 2013

Dear Ms. Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{5}------------------------------------------------

Page 2 - Ms. Betty M. Johnson

CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Kesia YAlexander -S.

Malvina B. Eydelman, M.D. Director Division of Opthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known): K122386

Device Name(s): VICTUS Femtosecond Laser Platform

Indications for Use:

The VICTUS Femtosecond Laser Platform is indicated for use in:

  • the creation of a corneal flap in patients undergoing LASIK surgery or other treatment ● requiring initial lamellar resection of the cornea
  • . for anterior capsulotomy during cataract surgery
  • . the creation of penetrating arcuate cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODF)

(Division Sign-Off) (Division of Ophthalmic and Ear, Nose and Throat Devices 510(k) Number

Page of

§ 886.4390 Ophthalmic laser.

(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.