PWB · Fetal Head Elevator

Obstetrics/Gynecology · 21 CFR 884.4350 · Class 2

Overview

Product CodePWB
Device NameFetal Head Elevator
Regulation21 CFR 884.4350
Device ClassClass 2
Review PanelObstetrics/Gynecology

Identification

A fetal head elevator is a prescription device consisting of a mechanism that elevates the fetal head to facilitate delivery during a Caesarean section.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the fetal head elevator is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible. (2) Performance data must demonstrate the sterility of patient-contacting components of the device. (3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life. (4) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Reliability testing of device deployment and retrieval under relevant use conditions must be conducted. (ii) Testing of the maximum force applied to the fetal head in an anatomic model must be conducted. (iii) Testing of uniform application of the elevator mechanism on the fetal head must be conducted. (5) Labeling must include the following: (i) Contraindication for use in the presence of active genital infection; (ii) Specific instructions regarding the proper placement and use of the device; and (iii) A shelf life.

Recent Cleared Devices (2 of 2)

RecordDevice NameApplicantDecision DateDecision
K243799Fetal PillowCooperSurgical, Inc.Aug 29, 2025SESE
DEN150053FETAL PILLOWSafe Obstetric Systems, Ltd.Jul 27, 2017DENG

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...