LVIS NeuroMatch

K222450 · LVIS Corporation · OMB · Jun 9, 2023 · Neurology

Device Facts

Record IDK222450
Device NameLVIS NeuroMatch
ApplicantLVIS Corporation
Product CodeOMB · Neurology
Decision DateJun 9, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1400
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

1. LVIS NeuroMatch Software is intended for the review, monitoring and analysis of electroencephalogram (EEG) recordings made by EEG devices using scalp electrodes and to aid neurologists in the assessment of EEG. The device is intended to be used by qualified medical practitioners who will exercise professional judgement in using the information. 2. LVIS NeuroMatch includes the calculation and display of a set of quantitative measures intended to monitor and analyze EEG waveforms. These include Artifact Strength, Asymmetry Spectrogram, Autocorrelation Spectrogram, and Fast Fourier Transform (FFT) Spectrogram. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveforms. 3. LVIS NeuroMatch displays physiological signals such as electrocardiogram (ECG/EKC) if it is provided in the EEG recording. 4. The aEEG functionality included in LVIS NeuroMatch is intended to monitor the state of the brain. 5. LVIS NeuroMatch Artifact Reduction (AR) is intended to reduce muscle and eye movements in EEG signals from the International Standard 10-20 placement. AR does not remove the entifact signal and is not effective for other types of artifacts. AR may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified medical practitioner trained in recognizing artifacts, and any interpretation or diagnosis must be made with reference to the original waveforms. 6. This device does not provide any diagnostic conclusion about the patient's condition to the user.

Device Story

LVIS NeuroMatch is software for review, monitoring, and analysis of EEG recordings; inputs include scalp EEG signals and optional ECG/EKG data. Device performs quantitative analysis, including Artifact Strength, Asymmetry Spectrogram, Autocorrelation Spectrogram, and FFT Spectrogram; features Artifact Reduction (AR) to mitigate muscle/eye movement interference in 10-20 placement signals. Used in clinical settings by neurologists/qualified practitioners; output displayed for clinician review. AR modifies waveforms; clinicians must verify against original data. Device does not provide diagnostic conclusions; aids clinical decision-making by providing quantitative measures and signal processing to support brain state monitoring.

Indications for Use

Indicated for review, monitoring, and analysis of EEG recordings in patients requiring neurological assessment. Intended for use by qualified medical practitioners. Not for diagnostic conclusions.

Regulatory Classification

Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 9, 2023 LVIS Corporation % Zane Liu Senior Manager, Regulatory Affairs Veranex. Inc. 224 Airport Parkway, Ste. 250 San Jose, CA 95110 Re: K222450 Trade/Device Name: LVIS NeuroMatch® Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLT, OMA Dated: May 11, 2023 Received: May 12, 2023 Dear Zane Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Patrick Antkowiak -S Patrick Antkowiak, Ph.D. Acting Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K222450 Device Name LVIS NeuroMatch® #### Indications for Use (Describe) 1. LVIS NeuroMatch Software is intended for the review, monitoring and analysis of electroencephalogram (EEG) recordings made by EEG devices using scalp electrodes and to aid neurologists in the assessment of EEG. The device is intended to be used by qualified medical practitioners who will exercise professional judgement in using the information. 2. LVIS NeuroMatch includes the calculation and display of a set of quantitative measures intended to monitor and analyze EEG waveforms. These include Artifact Strength, Asymmetry Spectrogram, Autocorrelation Spectrogram, and Fast Fourier Transform (FFT) Spectrogram. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveforms. 3. LVIS NeuroMatch displays physiological signals such as electrocardiogram (ECG/EKC) if it is provided in the EEG recording. 4. The aEEG functionality included in LVIS NeuroMatch is intended to monitor the state of the brain. 5. LVIS NeuroMatch Artifact Reduction (AR) is intended to reduce muscle and eye movements in EEG signals from the International Standard 10-20 placement. AR does not remove the entifact signal and is not effective for other types of artifacts. AR may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified medical practitioner trained in recognizing artifacts, and any interpretation or diagnosis must be made with reference to the original waveforms. 6. This device does not provide any diagnostic conclusion about the patient's condition to the user. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|--------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 9, 2023 LVIS Corporation % Zane Liu Senior Manager, Regulatory Affairs Veranex. Inc. 224 Airport Parkway, Ste. 250 San Jose, CA 95110 Re: K222450 Trade/Device Name: LVIS NeuroMatch® Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLT, OMA Dated: May 11, 2023 Received: May 12, 2023 Dear Zane Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {4}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Patrick Antkowiak -S Patrick Antkowiak, Ph.D. Acting Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...