K052519 · Spectral Diagnostics, Inc. · NOP · Nov 30, 2006 · Microbiology
Device Facts
Record ID
K052519
Device Name
SPECTRAL WEST NILE VIRUS IGM STATUS TEST
Applicant
Spectral Diagnostics, Inc.
Product Code
NOP · Microbiology
Decision Date
Nov 30, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3940
Device Class
Class 2
Indications for Use
The Spectral West Nile virus IgM STATus™ test is a rapid immunochromatographic lateral flow assay that utilizes recombinant West Nile virus (WNV) antigen (E glycoprotein) for the qualitative detection of IgM antibodies to WNV in human serum or plasma (sodium heparin or sodium citrate). This test is for the presumptive laboratory diagnosis of West Nile virus infection in patients having signs and symptoms of meningoencephalitis. Positive results must be confirmed by the PRNT (Plaque Neutralization Reduction Test), or by using the current CDC guidelines for diagnosis of this disease. The Spectral test is not intended for point of care testing, home use or in screening blood donor samples.
Device Story
The Spectral WNV IgM STATus™ test is a rapid, lateral flow immunochromatographic assay for qualitative detection of WNV IgM antibodies in human serum or plasma. The device uses recombinant WNV E protein antigen to capture IgM antibodies from the patient sample. A tertiary complex is formed with gold-labeled monoclonal murine anti-flavivirus E glycoprotein antibody, which migrates to a test line containing goat anti-human IgM. A control line confirms test validity. The test is performed in a laboratory setting by trained personnel. Results are visually interpreted within 15 minutes as a pinkish-purple band. Positive results require confirmation via PRNT or CDC-recommended methods. The device provides a faster alternative to traditional multi-step ELISA methods, aiding in the presumptive diagnosis of WNV infection in symptomatic patients.
Clinical Evidence
Performance evaluated across multiple sites using clinical serum specimens. Sensitivity: 95% (n=40) and 100% (n=24) in PRNT-confirmed acute cases; 80% (n=25) in late-stage cases. Negative agreement: 98.8% (n=346) in non-flavivirus populations. Cross-reactivity tested against various pathogens (e.g., Dengue, HSV, HIV, SLE); some cross-reactivity observed with ANA and RF. Precision/reproducibility confirmed across 3 sites/operators using 15-member panel.
Technological Characteristics
Solid-phase immunochromatographic lateral flow assay. Recombinant WNV E glycoprotein antigen. Gold-labeled monoclonal murine anti-flavivirus conjugate. Goat anti-human IgM capture antibody. Manual visual readout. Compatible with serum, sodium heparin plasma, and sodium citrate plasma. No electronic components or software.
Indications for Use
Indicated for presumptive laboratory diagnosis of West Nile virus infection in patients presenting with signs and symptoms of meningoencephalitis. Prescription use only.
Regulatory Classification
Identification
West Nile virus serological reagents are devices that consist of antigens and antisera for the detection of anti-West Nile virus IgM antibodies, in human serum, from individuals who have signs and symptoms consistent with viral meningitis/encephalitis. The detection aids in the clinical laboratory diagnosis of viral meningitis/encephalitis caused by West Nile virus.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Serological Reagents for the Laboratory Diagnosis of West Nile Virus.” See § 866.1(e) for the availability of this guidance document.
K041817 — WEST NILE DETECT IGM ELISA · Inbios Intl., Inc. · Nov 19, 2004
K041231 — WEST NILE VIRUS IGM CAPTURE ELISA, MODEL E-WNV02M · Panbio Limited · Aug 10, 2004
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K052519
B. Purpose for Submission:
This is an original application for a new rapid immunochromatographic assay for the qualitative determination of West Nile Virus IgM in human serum or plasma in patients having signs and symptoms of meningoencephalitis.
C. Measurand:
West Nile Virus IgM
D. Type of Test:
Rapid Immunochromatographic Assay
E. Applicant:
Spectral Diagnostics, Inc.
F. Proprietary and Established Names:
Spectral West Nile Virus IgM STATus™ Test
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3940
2. Classification:
Class II
3. Product code:
NOP
4. Panel:
{1}
Microbiology (83)
## H. Intended Use:
1. Intended use(s):
The Spectral West Nile virus IgM STATus™ test is a rapid immunochromatographic lateral flow assay that utilizes recombinant West Nile virus (WNV) antigen (E glycoprotein) for the qualitative detection of IgM antibodies to WNV in human serum or plasma (sodium heparin or sodium citrate). This test is for the presumptive laboratory diagnosis of West Nile virus infection in patients having signs and symptoms of meningoencephalitis. Positive results must be confirmed by the PRNT (Plaque Neutralization Reduction Test), or by using the current CDC guidelines for diagnosis of this disease.
2. Indication(s) for use:
See Intended Use above
3. Special conditions for use statement(s):
This device is for prescription use only
4. Special instrument requirements:
Not Applicable
## I. Device Description:
The Spectral WNV IgM STATusTM test employs solid-phase immunochromatographic assay technology to qualitatively detect the presence of WNV IgM antibodies in serum or plasma. See Test Principle below for more details.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Focus West Nile Virus IgM Capture ELISA
2. Predicate 510(k) number(s):
K031952
3. Comparison with predicate:
{2}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Antigen | Recombinant West Nile Virus antigen (E protein) | Recombinant West Nile Virus antigen (E protein) |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay format | Rapid immunochromatographic | ELISA |
| Specimen types | Serum and plasma (sodium heparin or sodium citrate) | Serum only |
| Readout | Visible by operator | Optical Density by plate reader |
| Conjugate | Gold-labeled mouse anti-flavivirus | Mouse anti-flavivirus HRP |
# K. Standard/Guidance Document Referenced (if applicable):
Not Applicable
# L. Test Principle:
The Spectral WNV IgM STATus™ test employs solid-phase immunochromatographic assay technology to qualitatively detect the presence of WNV IgM antibodies in serum or plasma. When the specimen to be tested is dispensed into the sample well of the Spectral device, anti-WNV IgM in the sample will bind to the recombinant WNV antigen (envelop glycoprotein (E) of West Nile virus, NY99 strain) to form a tertiary complex with gold-labeled monoclonal murine antibody against flavivirus family glycoprotein E. This tertiary complex will migrate through reaction strip and be captured by goat anti-human IgM antibodies at the Test area. Excess, unreacted gold complex detector is captured by immobilized anti-mouse IgG antibodies at the Control area. Assay results are determined by visual assessment of the presence or absence of control and test bands.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Reproducibility studies were performed using 15 patient serum samples at 3 study sites over a three day period. Each day a new lot was used and a different operator was employed. The 15 member reproducibility panel consisted of 6 clinical specimens with a mean index value $25\%$ above the cutoff, 6 clinical specimens with a mean index value $15\%$ below the cutoff,
{3}
and 3 clinical specimens with a mean index value 6 times the cutoff level of a commercially available WNV IgM Capture ELISA device (comparator device). All sites and operators produced the expected result for all panel members on every day of testing.
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not Applicable
d. Detection limit:
Not Applicable
e. Analytical specificity:
## Interfering Substances
The following constituents of human blood were tested to ensure that they did not interfere with the performance of the Spectral device. The concentration of the serum proteins or lipids shown in Table below were added to positive and negative serum specimens and tested with the Spectral device. No interference up to the concentrations listed below was observed.
Potential Interferents and Concentrations
| Potential Interfering material | Concentration |
| --- | --- |
| Human serum albumin | 16 g/dL |
| Bilirubin (unconj.) | 60 mg/dL |
| Hemoglobin | 25 g/dL |
| Triglycerides | 1.3 g/dL |
| Human IgM | 300 mg/dL |
## IgM Specificity:
The IgM specificity of the Spectral WNV IgM test was assessed in 14 paired positive samples by treating with 5mM dithiothreitol (DTT) solution for one hr at room temperature (22-24°C). All 14 treated samples produced negative results following the treatment with DTT and their respective untreated paired samples remained positive, indicating the presence of IgM class antibodies in the positive samples (10).
## Cross Reactivity:
Cross-reactivity studies with sera that were sero-positive to other potentially cross-reactive pathogens were conducted at a Public Health Laboratory in the North Eastern USA (site 1). Three out of seven Dengue post infection samples were tested at Public Health Laboratory at Mid-West Canada (site 3); 29 out of 33 RF positive sera and 23 out
{4}
of 28 ANA positive sera were tested in-house (site 2). Cross reactivity with Eastern Equine Encephalitis specimens was studied at a CDC laboratory at Mid-West USA (site 4). Results of these studies are tabulated below.
Results: Cross Reactivity Results
| Potentially Cross-Reactive Agents | Spectral WNV IgM STATus™ test Results | | | | |
| --- | --- | --- | --- | --- | --- |
| | Site | Number of Samples | Negative | Positive | % Cross-Reactive Samples |
| Herpes Simplex Virus 1 and/or 2 | 1 | 10 | 9 | 1 | 10% |
| Cytomegalovirus Polyvalent Test, all reactive | 1 | 5 | 5 | 0 | 0% |
| Syphilis (trep Reactive) | 1 | 10 | 10 | 0 | 0% |
| Epstein Barr Virus, IgG and IgM reactive | 1 | 5 | 5 | 0 | 0% |
| ANA (anti-nuclear antibody) | 1,2 | 28 | 17 | 11 | 39% |
| Rheumatoid Factor | 1,2 | 33 | 23 | 10 | 30% |
| HIV(human immunodeficiency virus) | 1 | 5 | 4 | 1 | 20% |
| Japanese Encephalitis Virus (vaccine recipients) IgM/IgG reactive,(CDC) | 1 | 5 | 5 | 0 | 0% |
| Saint Louis Encephalitis Virus. SLE IgM (4/5) and IgG (5/5) reactive, CDC | 1 | 5 | 5 | 0 | 0% |
| Dengue Virus (5 recent and 2 post infection)- DEN IgM & IgG reactive, CDC | 1,3 | 10 | 9 | 1 | 10% |
| Hepatitis B Virus (HbsAg) | 1 | 5 | 5 | 0 | 0% |
| Hepatitis C Virus (HCAb) | 1 | 5 | 5 | 0 | 0% |
| California Encephalitis PRNT reactive as JC or LAX | 1 | 6 | 6 | 0 | 0% |
| Legionella, Polyvalent Test, all reactive | 1 | 5 | 5 | 0 | 0% |
| Yellow Fever (vaccine recipients) | 1 | 5 | 4 | 1 | 20% |
| E. coli infection (culture confirmed) | 2 | 4 | 4 | 0 | 0% |
| Eastern Equine Encephalitis | 4 | 2 | 2 | 0 | 0% |
{5}
f. Assay cut-off: Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
# Study sites 1-3: Negative agreement of Spectral device with comparator in non-flavivirus IgM samples:
Studies were performed at three sites on clinical serum specimens collected prospectively from patients with a variety of non-flavivirus ailments (rash, febrile, bronchitis, diarrhea, drug, neuropathy, acute coronary syndrome etc) and from endemic normal populations either at emergency departments of hospitals or at in-house collection. Site 1 represents an endemic region at the South-Western State of America; sites 2 and 3 represent endemic regions of South Eastern Provinces of Canada. These specimens were tested at individual sites concomitantly with the Spectral and a commercially available comparator device. The results are tabulated as "Pos" (positive), or "Neg" (negative). Indeterminate results produced by the comparator device are listed as equivocal ("Eqv")
Study Site 1: Negative agreement of Spectral device with comparator in non-flavivirus IgM samples $(n = 160)$ .
| Spectral Device | | Comparator device | | | Total |
| --- | --- | --- | --- | --- | --- |
| | | Pos | Eqv | Neg | |
| | Pos | 0 | 0 | 1** | 1 |
| | Neg | 0 | 2* | 157 | 159* |
| | Total | 0 | 2 | 158 | 160 |
Negative agreement with the comparator device $= 157 / 158^{*}99.4\%$ 96.5-99.9% (95% CI)
*Note: Two specimens that produced equivocal results with the comparator device were determined to have interference from crossreacting antibodies.
**One Spectral false positive result was observed and determined to be due to interference from crossreacting antibodies
Study Site 2: Negative agreement of Spectral device with comparator in non-flavivirus IgM samples $(n = 114)$ .
| Spectral Device | | Comparator device | | | Total |
| --- | --- | --- | --- | --- | --- |
| | | Pos | Eqv | Neg | |
| | Pos | 0 | 0 | 2* | 2 |
| | Neg | 0 | - | 112 | 112 |
{6}
7
| | Total | 0 | 0 | 114 | 114 |
| --- | --- | --- | --- | --- | --- |
Negative agreement with the comparator device = 112/114 98.2% 93.8-99.8% (95% CI)
*Two Spectral false positive results were observed and determined to be due to interference from crossreacting antibodies
Study Site 3: Negative agreement of Spectral device with comparator in non-flavivirus IgM samples (n=72).
| Spectral Device | | Comparator device | | | Total |
| --- | --- | --- | --- | --- | --- |
| | | Pos | Eqv | Neg | |
| | Pos | 1* | 0 | 1 | 2** |
| | Neg | 0 | 0 | 70 | 70 |
| | Total | 1 | 0 | 71 | 72 |
Negative agreement with the comparator device = 70/71 98.6% 92.4-99.9% (95% CI)
*False positive due to cross reacting antibodies, as the sample remained positive with both devices when antigen was removed.
**Two Spectral false positive results were observed and determined to be due to interference from crossreacting antibodies.
All Sites (1-3): Negative agreement of Spectral device with comparator in non-flavivirus IgM samples (n= 346)
| Spectral Device | | Comparator device | | | Total |
| --- | --- | --- | --- | --- | --- |
| | | Pos | Eqv | Neg | |
| | Pos | 1* | 0 | 4 | 5*** |
| | Neg | 0 | 2** | 339 | 341* |
| | Total | 1 | 2 | 343 | 346 |
Negative agreement with the comparator device = 339/343 98.8% 97.1-99.7% (95% CI)
*False positive due to cross reacting antibodies (e.g. RF & heterophiles), as the sample remained positive with both devices when antigen was removed.
**Two specimens that produced equivocal results with the comparator device were determined to have interference from crossreacting antibodies.
***Five Spectral false positive results were observed and determined to be due to interference from crossreacting antibodies
Study Site 4: A provincial health laboratory located at South Eastern Province of Canada assessed the performance of the Spectral device with randomized retrospective patient serum specimens (n=90) that were sent to the laboratory with suspected WNV infection based on clinical signs and symptoms. This batch of samples constituted 40 WNV PRNT confirmed and 50 CDC WNV IgM ELISA negative specimens. All specimens were randomized and tested blinded with the Spectral device.
Results Study Site 4: Spectral WNV IgM STATus™ test reactivity with WNV PRNT Positive/CDC WNV IgM ELISA positive and CDC WNV IgM ELISA negative specimens
{7}
8
| Specimen Characterized by Reference Assays | Spectral WNV IgM STATus™ Results | | | | |
| --- | --- | --- | --- | --- | --- |
| | Pos | Neg | Total | Pos or Neg Agreement | 95% CI |
| Serological sensitivity (CDC WNV IgM and IgG ELISA Positive and WNV PRNT Positive), n=40 | 38 | 2 | 40 | 95% | 83.1%-99.4% |
| Negative Agreement with presumptive CDC WNV IgM ELISA Negative, n=50 | 2 | 48 | 50 | 96% | 86.3%-99.5% |
## Study Site 5:
Spectral device reactivity was assessed at a public health laboratory located in a prairie Province at Central Canada using a banked panel of clinical serum specimens that were drawn from patients (n=79) with clinical symptomology consistent with the West Nile virus infection. These samples also included symptomatic acutely infected patients (PRNT confirmed, n=24) exhibiting either febrile illness or neuro-invasive disease such as meningitis or encephalitis and, others specimens from asymptomatic patients (n=25) that were sent to the laboratory to detect previous West Nile virus exposures. All these samples were then randomized, blinded and tested with the Spectral device.
Results Study Site 5: Spectral WNV IgM STATus™ test reactivity with WNV PRNT Positive/CDC WNV IgM Elisa positive and CDC WNV IgM Elisa negative specimens
| Specimen Characterized by Reference Assays | Spectral WNV IgM STATus™ Results | | | | |
| --- | --- | --- | --- | --- | --- |
| | Pos | Neg | Total | Pos or Neg Agreement | 95% CI |
| Serological sensitivity for acute PRNT Pos, CDC WNV IgM & IgG Elisa Pos (n=24) | 24 | 0 | 24 | 100% | 88.3%-100% |
| Serological sensitivity for late PRNT Pos, CDC WNV IgM & IgG Elisa Pos (n=25) | 20 | 5 | 25 | 80% | 59.3%-93.2% |
| Negative agreement with presumptive CDC WNV IgM Elisa Neg (n=30) | 1 | 29 | 30 | 97% | 82.8%-99.9% |
## b. Matrix comparison:
In order to show that the Spectral device produces similar performance for all indicated matrices, 63 sodium citrate plasma and serum paired samples and 59 sodium heparin plasma and serum paired samples (obtained by spiking with the known WNV positive clinical specimens) were tested concomitantly with the Spectral device. Comparable results were obtained for sodium citrate plasma vs. serum and sodium heparin plasma vs. serum.
3. Clinical studies:
Not Applicable
4. Clinical cut-off:
Not Applicable
{8}
9
5. Expected values/Reference range:
Serum specimens from a total of 346 presumably healthy individuals (n=67) or non-WNV ailments (n=279) including 61 febrile patients were tested with the Spectral WNV IgM STATus™ test (henceforth referred to as the Spectral device) in a general population from endemic regions of USA and Canada. The samples were about equally distributed between male (47.1%) and female (52.9%) populations. Positive results were observed and determined to be positive due to heterophile antibodies interferences.
Prevalence of WNV IgM and Efficacy of Spectral Device in Endemic Population
| Age (Years) | Negative | Positive | % Positive | 95% CI |
| --- | --- | --- | --- | --- |
| 0 to 9 | 0 | 0 | 0.0% (0/0) | NA |
| 10 to 19 | 8 | 0 | 0.0% (0/8) | 0.00%, 31.23% |
| 20 to 29 | 39 | 1 | 2.5% (1/40) | 0.06%, 13.16% |
| 30 to 39 | 56 | 1 | 1.7% (1/57) | 0.04%, 9.39% |
| 40 to 49 | 53 | 1 | 1.8% (1/54) | 0.05%, 9.89% |
| 50 to 59 | 56 | 0 | 0.0% (0/55) | 0.00%, 5.30% |
| 60 to 69 | 49 | 1 | 2.0% (1/50) | 0.05%, 10.65% |
| 70 to 79 | 39 | 1 | 2.5% (1/40) | 0.06%, 13.16% |
| 80+ | 41 | 0 | 0.0% (0/41) | 0.00%, 7.05% |
| Overall | 341 | 5 | 1.4% (5/346) | 0.47%, 3.34% |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.