Intended Use

The Focus Technologies West Nile Virus IgM Capture ELISA is intended for qualitatively detecting IgM antibodies to West Nile virus in human serum. In conjunction with the Focus Technologies West Nile Virus ELISA IgG, the test is indicated for testing persons having symptoms of meningioencephalitis, as an aid in the presumptive laboratory diagnosis of West Nile virus infection. Positive results must be confirmed by neutralization test, or by using the current CDC guidelines for diagnosing West Nile encephalitis. This test is not intended for self-testing, and this test is not FDA cleared nor approved for testing blood or plasma donors. Assay performance characteristics have not been established for automated instruments.

Device Description

Indirect Enzyme-linked immunosorbent assay for qualitatively detecting human serum IgM antibodies to West Nile virus.

AI/ML Overview

Here's a summary of the acceptance criteria and the studies performed for the Focus Technologies West Nile Virus IgM Capture ELISA, based on the provided text:

Acceptance Criteria and Device Performance

The acceptance criteria for this device are not explicitly stated in a quantitative table format (e.g., "sensitivity must be >X%"). Instead, the performance characteristics are presented as observed percentages and 95% confidence intervals from various studies, which are then compared to relevant reference methods. The implicit acceptance criteria are that the device performs comparably or acceptably against established reference methods like the CDC IgM ELISA and Plaque Reduction Neutralization Test (PRNT) for diagnosing West Nile virus infection.

Table of Acceptance Criteria (Implicit) and Reported Device Performance:

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
Clinical Sensitivity (Confirmed WNV)High positive detection rate compared to reference methods.90.9% (40/44) when compared to CDC WNV IgM ELISA positive and WNV PRNT positive in encephalitis/meningitis patients.
Positive Agreement with Presumptive CDC WNV IgM ELISAHigh positive agreement.100% (2/2) for presumed positive WNV encephalitis patients (from encephalitis/meningitis study).100% (1/1) for presumptive positive CDC WNV IgM ELISA samples (from non-flavivirus test samples).
Negative Agreement with Presumptive CDC WNV IgM ELISA (Encephalitis/Meningitis)High negative agreement.98.8% (251/254) for presumed negative patients (from encephalitis/meningitis study).
Serological Sensitivity (WNV PRNT Positives)High positive detection rate in PRNT-confirmed samples.100% (75/75) for WNV PRNT positive samples.100% (1/1) for WNV PRNT confirmed sample (from suspected encephalitis/meningitis study).
Negative Agreement with Presumptive WNV IFAHigh negative agreement.96.1% (99/103) for WNV IgM IFA negative samples.
Negative Agreement with Presumptive CDC WNV IgM ELISA (Suspected Encephalitis/Meningitis)High negative agreement for presumptive negatives.98.0% (48/49) for WNV presumptive negative samples (from suspected encephalitis/meningitis study).
Negative Agreement with Presumptive CDC WNV IgM ELISA (Non-flavivirus samples)High negative agreement in general population (non-flavivirus).99.4% (468/471) for CDC ELISA IgM negative samples (from non-flavivirus test samples).
Cross-reactivityLow rates of false positives with other pathogens.Varies: Dengue (40%), St. Louis encephalitis (53.8%), Herpes simplex (10%), Cytomegalovirus (7.1%), Borrelia burgdorferi (15%), Rheumatoid factor (25%), Anti-nuclear antibodies (5%), Eastern Equine Encephalitis (0%), Epstein-Barr virus (0%). Note: High cross-reactivity with some flaviviruses (Dengue, SLE) observed.
Freeze-Thaw StabilityNo change in interpretation after multiple freeze-thaw cycles.No changes in interpretation across 8 sera (5 positive, 3 negative) after up to 5 freeze-thaw cycles.
ReproducibilityLow variability across lots, assays, and laboratories.Inter- & Intra-assay %CV: 0.3% - 20.0%Inter-lot %CV: 0.4% - 3.6%Inter-lab %CV: 2.3% - 13.2%

Study Details:

  1. Sample sizes used for the test set and the data provenance:

    • Study Site 1 (Encephalitis/Meningitis Patients):
      • Test Set Size: 300 patients.
      • Data Provenance: Sera sequentially submitted to a state department of health laboratory in the northeastern U.S.; archived and masked. Likely retrospective, given the archival nature.
    • Study Site 2 (WNV PRNT Positives):
      • Test Set Size: 75 samples.
      • Data Provenance: Samples prescreened positive (by Focus) with a WNV native antigen ELISA and confirmed WNV positive by PRNT. From a clinical laboratory in the mid-western U.S.; archived. Likely retrospective.
    • Study Site 3 (West Nile IFA Negatives):
      • Test Set Size: 103 retrospective samples.
      • Data Provenance: From a clinical laboratory in the southwestern U.S.; retrospective.
    • Study Site 4 (Suspected Encephalitis/Meningitis Patients):
      • Test Set Size: 50 samples.
      • Data Provenance: Archived and masked sera provided by a U.S. federal government laboratory. Likely retrospective.
    • Study Site 4 (Non-Flavivirus Test Samples):
      • Test Set Size: 476 samples.
      • Data Provenance: Prospectively collected from North America during August 2003, submitted to a clinical laboratory in Southern California for non-flavivirus tests. Prospective.
    • Cross-reactivity Study:
      • Test Set Size: 75 samples (various pathogens).
      • Data Provenance: From Study Site 4 (Focus) and Study Site 1 (DOH for SLE positives). Sera were retrospective and masked.
    • Specificity of IgM Capture Wells Study:
      • Test Set Size: 15 sera.
      • Data Provenance: Not explicitly stated, but likely in-house from Focus (Study Site 4) using known WNV IgM and IgG positive samples.
    • Sera Freeze-Thaw Study:
      • Test Set Size: 8 sera (5 positive, 3 negative).
      • Data Provenance: Not explicitly stated, likely in-house from Focus (Study Site 4).
    • Reproducibility Studies:
      • Inter-lot and Inter/Intra-assay: 5-7 samples in duplicates/triplicates across different lots/days.
      • Inter-laboratory: Samples run in triplicate at 3 different labs on 3 different days.
      • Data Provenance: Focus (Study Site 4), a state department of health laboratory in the northeastern U.S. (Study Site 1), and a clinical laboratory in the mid-western U.S. (Study Site 2).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not specify the number of experts or their qualifications. The ground truth was established by reference laboratory assays (e.g., CDC IgM ELISA, WNV PRNT, WNV IgM IFA), which were performed by trained personnel in state department of health or clinical laboratories. The expertise lies in the established protocols and interpretation guidelines of these reference methods rather than individual expert adjudication of each case.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • There was no explicit "adjudication method" in the sense of multiple human readers/experts evaluating the cases. The ground truth was determined by comparing the device's results against established reference laboratory tests (CDC IgM ELISA, WNV PRNT, WNV IgM IFA). The results of these reference tests served as the primary basis for classifying samples as positive or negative. Certain equivocal results from the Focus assay were calculated as positive or negative for performance metric purposes, indicating a predefined rule for handling these, but not human adjudication.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is an in vitro diagnostic (IVD) ELISA kit, not an AI-powered image analysis tool or a system designed to assist human readers in interpreting complex images. Its output is a quantitative optical density that is used to derive a qualitative (positive, negative, equivocal) result, which is then interpreted by laboratory personnel. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, this study represents a standalone performance evaluation of the Focus Technologies West Nile Virus IgM Capture ELISA. The device is a laboratory assay (ELISA kit) designed to provide a result directly from a patient sample. Its performance is evaluated intrinsically through its chemical and immunological reactions, and the optical density reading is interpreted by a predefined algorithm/cutoff. Human involvement is in performing the assay and reading the result, but the "performance" itself is that of the assay kit. This is the typical way IVD assays are evaluated.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The ground truth was primarily established using established reference laboratory tests:
      • Plaque Reduction Neutralization Test (PRNT): Considered the gold standard for confirming WNV infection.
      • CDC West Nile Virus IgM ELISA: A widely accepted and validated method for detecting WNV IgM antibodies.
      • West Nile IFA (Immunofluorescence Assay): Another method used for WNV IgM antibody detection.
      • Clinical Criteria: For the encephalitis/meningitis patient cohort, a combination of clinical symptoms (fever, altered mental status, CSF pleocytosis) in conjunction with laboratory results defined the diagnosis.
  7. The sample size for the training set:

    • The document does not explicitly describe a "training set" for the device in the context of machine learning. For an ELISA kit, development typically involves internal optimization and validation using various samples, but these are not defined as a distinct "training set" like in AI/ML. The performance data presented are from validation/test sets.
  8. How the ground truth for the training set was established:

    • Since a formal "training set" as understood in AI/ML is not described, the method for establishing its ground truth is not applicable or detailed in this document. The development and internal validation of such an assay would involve using well-characterized samples (often confirmed by gold-standard methods like PRNT or a previously validated ELISA) to optimize reagent concentrations, incubation times, and cutoff values, but this process isn't reported as a "training set" with established ground truth in the same way as for AI models.

{0}------------------------------------------------

OCT 2 2 2003

Image /page/0/Picture/1 description: The image shows the logo for Focus Technologies. The logo consists of the word "FOCUS" in a bold, sans-serif font, with two crescent moon shapes forming the letter "O". Below the word "FOCUS" is the word "technologies" in a smaller, sans-serif font. The logo is black and white.

Page 1 of 6

ApplicantFocus Technologies, Inc.10703 Progress WayCypress, California 90630USA
Establishment RegistrationNo.2023365
Contact PersonMichael J. Wagner, Esq.tel (714) 220-1900fax (714) 995-6921mwagner@focustechnologies.com
Summary DateOctober 17, 2003
Proprietary NameWest Nile Virus IgM Capture ELISA
Generic NameWest Nile Virus IgM Capture ELISA
ClassificationWest Nile Virus Serological Reagents21 CFR §866.3940Class II
Predicate DeviceFocus Technologies Arbovirus IFA IgM (K913618)Focus Technologies HSV-2 ELISA (K993724)CDC West Nile Virus IgM Capture ELISAWest Nile Virus Plaque Reduction Neutralization Test

Device Description

Indirect Enzyme-linked immunosorbent assay for qualitatively detecting human serum IgM antibodies to West Nile virus.

Intended Use

The Focus Technologies West Nile Virus IgM Capture ELISA is intended for qualitatively detecting IgM antibodies to West Nile virus in human serum. In conjunction with the Focus Technologies West Nile Virus ELISA IgG, the test is indicated for testing persons having symptoms of meningioencephalitis, as an aid in the presumptive laboratory diagnosis of West Nile virus infection. Positive results must be confirmed by neutralization test, or by using the current CDC guidelines for diagnosing West Nile encephalitis. This test is not intended for self-testing, and this test is not FDA cleared nor approved for testing blood or plasma donors. Assay performance characteristics have not been established for automated instruments.

Test Principle

In the Focus Technologies West Nile Virus IgM Capture ELISA, the polystyrene microwells are coated with anti-human antibody specific for IgM (u-chain). Diluted serum samples and controls are incubated in the wells, and IgM present in the sample binds to the anti-human antibody (IgM specific reactants are removed by washing. Recombinant WNV antigen is then added to the wells and incubated; and, if anti-WNV IgM is present in the sample, the WNV antigen binds to the anti-WNV in the well. Unbound WNV antigen is then removed by washing the well. Mouse anti-flavivirus conjugated with horseradish peroxidase (HRPO) is then added to the wells and incubated; and, if WNV antigen has been retained in the well by the anti-flavivirus in the sample, the mouse anti-flavivirus: HRPO binds to the WNV antigen in the wells. Excess conjugate is removed by washing. Enzyme substrate and chromogen are added, and the color is allowed to develop. After adding the Stop Reagent, the resultant color change is quantified by a spectrophotometric reading of optical density (OD) that is directly proportional to the amount of antigen-specific IgM present in the sample. Sample optical density readings are compared with reference cut-off OD readings to determine results.

{1}------------------------------------------------

K031952 510(k) Summary of Safety and Effectiveness West Nile Virus IgM Capture ELISA Catalog No. EL0300M Prepared October 17, 2003 Page 2 of 6

Expected Values

The prevalence of West Nile antibodies varies depending age, geographic location, testing method used, and other factors. A community based serosurvey for West Nile infection conducted in New York in 2000 found that 0.2% (5/2433) of persons tested overall had antibodies indicating recent West Nile infection, and that 1.1% (2/176) of persons reporting a recent headache and fever had antiboding a recent West Nile infection. Two serosurveys conducted in New York City (NYC) in 1999 and 2000 showed that approximately 1 in 150 infections (<1%) resulted in meningits or encephalitis. The NYC results are consistent with a 1996 Romanian serosurvey indicating that 1:140 to 1:320 infections resulted in meningitis or encephalitis.

Prevalence in Samples Submitted for Non-Flavivirus Testing (n=476)

Focus assessed reactivity with 476 samples prospectively collected from North America during August 2003. The samples had been submitted to a clinical laboratory located in Southern California for non-flavivirus tests (e.g., tests for other infectious diseases). The samples consisted of 64.1% males, and 1.5% from persons of unspecified gender.

AgeNegEqvPos% Positive95%CI
0 to 924000.0% (0/24)0.0-14.2%
10 to 1928013.5% (1/29)0.1-17.8%
20 to 2970000.0% (0/70)0.0-5.1%
30 to 3982000.0% (0/82)0.0-4.4%
40 to 4977011.3% (1/78)0.0-6.9%
50 to 5948123.9% (2/51)0.5-13.5%
60 to 6938012.6% (1/39)0.1-13.5%
70 to 7934000.0% (0/34)0.0-10.3%
80+17100.0% (0/18)0.0-18.5%
Unknown50100.0% (0/51)0.0-7.0%
Overall468351.1% (5/476)0.3-2.4%

IgM Prevalence with Samples Submitted for Non-Flavivirus Testing (n=476)

tib
for

{2}------------------------------------------------

K031952 510(k) Summary of Safety and Effectiveness West Nile Virus IgM Capture ELISA Catalog No. EL0300M Prepared October 17, 2003 Page 3 of 6

Performance Characteristics

Study Site 1: Focus Reactivity with Reactivity with Encephalitis/Meningitis Patients (n = 300)

A state department of health laboratory located in the northeastern U.S. assessed the device's reactivity from encephalitis/meningitis patients (n = 300). Patients were suspected of having either viral meningitis. Viral encephalitis criteria included: 1) fever; 2) altered mental status and/or other evidence of cortical involvement; and 3) CSF pleocytosis with predominant lymphocytes and/or elevated protein and a negative gram stain and culture. Viral meningitis criteria included: 1) fever, 2) headache, stiff neck and/or other meningeal signs; and 3) CSF pleocytosis with predominant lymphocytes and/or elevated protein and a negative gram stain and culture). The sera were sequentially submitted to the laboratory, archived, and masked. The reference methods were the CDC IgM ELISAs, and a plaque reduction neutralization test (PRNT) for West Nile virus.

Of 300 encephalitis/meningitis patients, 44 were classified as confirmed positive West Nile encephalitis patients (encephalitis/meningitis symptoms, CDC IgM ELISA positive and WNV PRNT positive) and 256 had presumptive assay results (CDC WNV IgM ELISA). The Focus IgM assay was positive with 90.9% (40/44) of the confirmed positive WNV encephalitis patients (including 2 Focus equivocals calculated as negatives). Of the 256 patients with presumptive assay results, 254 were classified as presumed negative patients (CDC WNV IgM ELISA negative),), and 2 were classified as presumed positive West Nile encephalitis patients (CDC WNV IgM ELISA positive). The Focus IgM assay was positive with 100% (2/2) of the presumed positive WNV encephalitis patients. The Focus IgM assay was negative with 98.8% (251/254) of the presumed negative patients (including 2 Focus equivocals calculated as positives).

Specimens Characterized by Reference AssaysFocus WNV IgM ELISA Results
NegEqvPosTotal%95% CI
Clinical sensitivity (encephalitis/meningitis symptoms, CDCWNV IgM ELISA positive and WNV PRNT positive)22404490.9% (40/44)78.3-97.5%
Positive agreement with presumptive CDC WNV IgM ELISA0022100% (2/2)15.8-100%
Negative agreement with presumptive CDC WNV IgM ELISA2512125498.8% (251/254)96.6-99.8%

Study Site 1: Focus Reactivity with Encephalitis/Meningitis Patients (n=300)

Study Site 2: Focus Reactivity with WNV PRNT Positives (n = 75)

A clinical laboratory located in the mid-western U.S. assessed the device's reactivity with 75 samples that were prescreened positive (by Focus) with a West Nile virus native antigen ELISA, and confirmed West Nile positive by plaque reduction neutralization test (PRNT). The sera were sequentially submitted to the laboratory, archived, The Focus IgM ELISA was positive with 100% (75/75) of the WNV PRNT positive samples.

Study Site 2: Focus Reactivity with WNV PRNT Positives (n = 75)

Specimens Characterized by Reference AssaysFocus WNV IgM ELISA Results
NegEqvPosTotal%95% CI
Serological sensitivity (CDC WNV IgM ELISA positive andWNV PRNT positive)007575100% (75/75)95.2-100%

{3}------------------------------------------------

K031952

Image /page/3/Picture/1 description: The image shows the logo for Focus Technologies. The logo is in black and white. The word "FOCUS" is in large, bold letters, with a crescent moon shape replacing the letter "O". Below the word "FOCUS" is the word "technologies" in smaller letters.

Page 4 of 6

Performance Characteristics (continued)

Study Site 3: Focus Reactivity with West Nile IFA Negatives (n=103)

A clinical laboratory located in the southwestern U.S. assessed reactivity with 103 retrospective samples that were West Nile IFA negative. The Focus IgM ELISA was negative with 96.1% (99/103) of WNV IgM IFA negative samples (including one equivocal calculated as positive).

Study Site 3: Focus Reactivity with West Nile IFA Negatives (n=103)

Specimens Characterized by Reference AssaysFocus WNV IgM ELISA Results
NegEqvPosTotal%95% CI
Negative agreement with presumptive WNV IFA991310396.1% (99/103)90.3-98.9%

Study Site 4: Focus Reactivity with Suspected Encephalitis/Meningitis Patients (n= 50)

Focus assessed the device's reactivity with 50 samples from patients suspected of encephalitis/meningitis. A U.S. federal government laboratory provided the archived and masked sera. One sample was confirmed positive by WNV PRNT, and the other 49 were presumptively negative (CDC ELISA) for arboviruses present in North America (LAC, EEE, SLE and WNV). The Focus IgM ELISA was negative with 98.0% (48/49) of the WNV presumptive negative samples, and positive with the one WNV PRNT confirmed sample.

Study Site 4: Focus Reactivity with Suspected Encephalitis/Meningitis Patients (n= 50)

Specimens Characterized by Reference AssaysFocus WNV IgM ELISA Results
NegEqvPosTotal%95% CI
Serological sensitivity (CDC WNV IgM ELISA positive andWNV PRNT positive)0011100% (1/1)NA
Negative agreement with presumptive CDC WNV IgM ELISA48014998.0% (48/49)89.1-99.9%

Study Site 4: Focus Reactivity with Non-Flavivirus Test Samples (n = 476)

Focus assessed the device's reactivity with 476 samples prospectively collected from North America during August 2003. The samples had been submitted to a clinical laboratory located in Southern California for non-flavivirus tests (e.g., tests for other infectious diseases). Postive samples were tested with a CDC WNV IgM ELISA. The Focus West Nile IgM Capture ELISA was negative with 99.4% (468/471) of the CDC ELISA IgM negative samples (including 3 Focus equivocals calculated as positive with 100% (1/1) of the CDC ELISA IgM positive samples. Four CDC ELISA IgM indeterminant samples were excluded from the calculations.

Study Site 4: Focus Reactivity with Non-Flavivirus Test Samples (n = 476)*

Specimens Characterized by Reference AssaysFocus WNV IgM ELISA Results
NegEqvPosTotal%95% CI
Positive agreement with presumptive CDC WNV IgM ELISA0011100% (1/1)NA
Negative agreement with presumptive CDC WNV IgM ELISA4683047199.4% (468/471)98.1-99.9%
  • Excludes four samples that were indeterminant with the CDC IgM ELISA.

{4}------------------------------------------------

Performance Characteristics (continued)

Focus Cross-reactivity

Focus (Study Site 4) and a state department of health laboratory located in the northeastern U.S. (DOH) (Study Site 1) assessed the device's cross-reactivity with sera that were sero-positive to other potentially cross-reactive pathogens (n = 75). The DOH tested the SLE positives, and Focus tested the other retrospective and masked. The results of the studies are summarized in the table below.

PopulationSiteFocus WNV IgM ELISA Results
NegEqvPosTotal% Positive95% CI
Dengue virus (secondary infections)414151540.0% (6/15)16.3-67.7%
St. Louis encephalitis virus16071353.8% (7/13)25.1-80.8%
Eastern Equine Encephalitis virus420020.0% (0/2)0.0-84.2%
Herpes simplex virus418112010.0% (2/20)1.2-31.7%
Epstein-Barr virus41900190.0% (0/19)0.0-17.6%
Cytomegalovirus41301147.1% (1/14)0.2-33.9%
Borrelia burgdorferi40032015.0% (3/20)3.2-37.9%
Rheumatoid factor40142025.0% (5/20)3.7-49.1%
Anti-nuclear antibodies4001205.0% (1/20)0.1-24.9%

Specificity of the Focus IgM Capture Wells

Focus (Study Site 4) assessed specificity of the IgM Capture Wells by selecting fifteen different sera that were positive for both WNV IgM and IgG, and treating the sera in four different ways:

  • No Treatment: The sera were not treated with DTT nor 2-ME, and the sera were diluted in the kit Sample Diluent .. . 1) (no goat anti-human IgG);
    • 2) Goat anti-human IgG (GtalgG): The sera were treated with diluent containing Goat anti-human-IgG;
    • Dithiothreitol (DTT): The sera were treated with 5 uL of 50 mM DTT and the sera were diluted in the kit Sample 3) Diluent (no goat anti-human IgG);
      1. 2-Mercaptoethanol (2-ME): The sera were treated with 5 uL of 1.43 M (10% v/v) 2-mercaptoethanol, and the sera were diluted in the kit Sample Diluent (no goat anti-human IgG).

All treatment groups were tested with IgM ELISA, and the first two groups were tested with the Focus WNV IgG ELISA. The "No Treatment" groups showed that all 15 samples are clearly IgM and IgG positive, with indices ranging from 2.75 to 4.99. Treating with DTT caused 100% (15/15) of the samples to show at least a 50% decrease in reactivity, with two samples remaining positive, three samples becoming equivocal, and ten samples becoming negative. Treating with mercaptoethanol caused 100% (15/15) of the samples to become IgM negative. Treatment with goat antihuman IgG precipitating reagent caused 100% (14/14) of the samples to become IgG negative, while 100% (15/15) of the samples remained IgM positive.

{5}------------------------------------------------

Performance Characteristics (continued)

Sera Freeze-Thaw Study

Focus (Study Site 4) assessed the impact on the WNV IgM assay's reactivity by selecting 8 sera (5 positive and 3 negative), subjecting them to up to 5 repeated freeze-thaw cycles, and testing them in parallel with all not been frozen. There were no changes in interpretation in any of the sera. Positive samples trended slightly towards increasing indices, while negative sera did not appear to change.

Focus Reproducibility

Reproducibility studies included Inter-lot Reproducibility, Inter/Intra-assay Reproducibility, and Inter-laboratory Reproducibility. In each study, two sets of samples were masked duplicates. Focus (Study Site 4) assessed the device's Inter-lot Reproducibility by testing five samples on three separate lots. For one lot, the samples were run in triplicate, and run in duplicate with the other two lots. Each of the three lots had a different lot of Antigen and Capture Wells. Focus (Study Site 4) assessed the device's Inter/Intra-assay Reproducibility by testing seven samples in triplicate, once a day, for three days, for a total of 63 data points. A state department of health laboratory located in the northeastern U.S. (Study Site 1), and a clinical laboratory located in the mid-western U.S. (Study Site 2), Focus (Study Site 4), assessed the device's Inter-laboratory Reproducibility. Each of the three laboratories in triplicate on three different days.

Focus Reproducibility
SampleInter- & Intra-assayInter-lotInter-Lab
Index MeanIntra-assay%CVInter-assay%CVIndex MeanIndex %CVIndex MeanIndex %CV
M2*0.212.910.30.221.20.239.7
M6*0.233.420.00.230.40.2413.2
M50.691.65.70.700.70.716.4
M1*1.431.52.91.412.61.454.0
M7*1.531.84.01.542.11.4912.8
M32.372.71.72.333.62.232.5
M42.991.90.32.981.92.782.3
  • There were two sets of masked pairs (same sample, different labeled identity): M2 & M6 were the second masked pair.

{6}------------------------------------------------

Public Health Service

Image /page/6/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines representing the staff and two snakes intertwined around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

OCT 2 2 2003

Michael J. Wagner, Esq. Senior Regulatory Affairs Specialist Focus Technologies, Inc. 10703 Progress Way Cypress, CA 90630

Re: K031952

Trade/Device Name: West Nile Virus IgM Capture ELISA Regulation Number: 21 CFR 866.3940 Regulation Name: West Nile Virus Serological Reagents Regulatory Class: Class II Product Code: NOP Dated: October 17, 2003 Received: October 20, 2003

Dear Mr. Wagner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{7}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Page 1 of 1

510(k) Number (if known): K031952

West Nile Virus IgM Capture ELISA Device Name:

The Focus Technologies West Nile Virus IgM Capture ELISA is Indications for Use: intended for qualitatively detecting IgM antibodies to West Nile virus in human serum. In conjunction with the Focus Technologies West Nile Virus ELISA IgG, the test is indicated for testing persons having symptoms of meningioencephalitis, as an aid in the presumptive laboratory diagnosis of West Nile virus infection. Positive results must be confirmed by neutralization test, or by using the current CDC guidelines for diagnosing West Nile encephalitis. This test is not intended for self-testing, and this test is not FDA cleared nor approved for testing blood or plasma donors. Assay performance characteristics have not been established for automated instruments.

(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Saqattar
Division Sign-Off 10/21/03

Office of In Vitro Diagnostic Device Evaluation and Safety

(Optional Format 3-10-98)

510(k)K031952

/

Abersons

OTC

§ 866.3940 West Nile virus serological reagents.

(a)
Identification. West Nile virus serological reagents are devices that consist of antigens and antisera for the detection of anti-West Nile virus IgM antibodies, in human serum, from individuals who have signs and symptoms consistent with viral meningitis/encephalitis. The detection aids in the clinical laboratory diagnosis of viral meningitis/encephalitis caused by West Nile virus.(b)
Classification. Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Serological Reagents for the Laboratory Diagnosis of West Nile Virus.” See § 866.1(e) for the availability of this guidance document.