ADVIA Centaur CMV IgG and ADVIA Centaur CMV IgG Quality Control
Applicant
Siemens Healthcare Diagnostics, Inc.
Product Code
LFZ · Microbiology
Decision Date
Jul 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3175
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The ADVIA Centaur® CMV IgG (CMV IgG) assay is for in vitro diagnostic use in the qualitative detection of IgG antibodies to cytomegalovirus (CMV) in human pediatric and adult serum and plasma (dipotassium EDTA, lithium heparin) using the ADVIA Centaur XP system. The assay is used to determine CMV IgG serological status and as an aid in the diagnosis of CMV infection in individuals for whom a CMV IgG test was ordered, including pregnant women. The ADVIA Centaur CMV IgG assay is not intended for blood and tissue donor screening. The ADVIA Centaur® CMV IgG Quality Control material is for in vitro diagnostic use for monitoring the performance of the ADVIA Centaur CMV IgG (CMV IgG) assay on ADVIA Centaur systems.
Device Story
Fully automated, 2-step sandwich immunoassay using indirect chemiluminometric technology on ADVIA Centaur XP system. Input: human serum or plasma samples. Process: sample diluted; incubated with biotinylated CMV viral lysate antigens on streptavidin-coated magnetic particles; CMV-specific antibodies captured; washed; Lite reagent (acridinium-ester-labeled anti-human IgG mouse monoclonal antibody) added; complex washed; chemiluminescent reaction initiated by acid/base reagents. Output: relative light units (RLUs) proportional to bound anti-CMV IgG; results reported as reactive or nonreactive based on Index value (<1.00 nonreactive, ≥1.00 reactive). Used in clinical laboratories by trained personnel. Benefits: aids in diagnosis of CMV infection and determination of serological status.
Clinical Evidence
Clinical performance evaluated via method comparison study with 1,842 samples (1,699 prospective, 143 retrospective) across diverse populations (general, pregnant, pediatric, HIV-positive, transplant patients). Compared to a predicate CMV IgG assay, overall positive agreement was 99.6% (1037/1041) and negative agreement was 96.8% (637/658). CDC panel of 80 characterized samples showed 100% agreement. Precision/reproducibility studies (CLSI EP05-A3) showed within-lab CVs 3.0-3.9% and reproducibility CVs 4.6-7.0%. Interference and cross-reactivity testing showed no significant impact on clinical interpretation.
Indicated for qualitative detection of IgG antibodies to cytomegalovirus (CMV) in human pediatric and adult serum and plasma. Used to determine CMV IgG serological status and aid in diagnosis of CMV infection in patients, including pregnant women. Not for blood or tissue donor screening.
Regulatory Classification
Identification
Cytomegalovirus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to cytomegalovirus in serum. The identification aids in the diagnosis of diseases caused by cytomegaloviruses (principally cytomegalic inclusion disease) and provides epidemiological information on these diseases. Cytomegalic inclusion disease is a generalized infection of infants and is caused by intrauterine or early postnatal infection with the virus. The disease may cause severe congenital abnormalities, such as microcephaly (abnormal smallness of the head), motor disability, and mental retardation. Cytomegalovirus infection has also been associated with acquired hemolytic anemia, acute and chronic hepatitis, and an infectious mononucleosis-like syndrome.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
July 30, 2018
Siemens Healthcare Diagnostics, Inc. Kira Gordon Sr. Regulatory Affairs Specialist 511 Benedict Ave Tarrytown, New York 10591
#### Re: K181213
Trade/Device Name: ADVIA Centaur CMV IgG ADVIA Centaur CMV IgG Quality Control Regulation Number: 21 CFR 866.3175 Regulation Name: Cytomegalovirus serological reagents Regulatory Class: Class II Product Code: LFZ, OCH Dated: May 1, 2018 Received: May 7, 2018
#### Dear Kira Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR
{1}------------------------------------------------
Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Steven R. Gitterman -S for
Uwe Scherf, Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181213
Device Name ADVIA Centaur® CMV IgG ADVIA Centaur® CMV IgG Quality Control
Indications for Use (Describe)
#### ADVIA Centaur CMV lgG:
The ADVIA Centaur® CMV IgG (CMV IgG) assay is for in vitro diagnostic use in the qualitative detection of IgG antibodies to cytomegalovirus (CMV) in human pediatric and adult serum and plasma (dipotassium EDTA, lithium heparin) using the ADVIA Centaur XP system. The assay is used to determine CMV IgG serological status and as an aid in the diagnosis of CMV infection in individuals for whom a CMV IgG test was ordered, including pregnant women. The ADVIA Centaur CMV IgG assay is not intended for blood and tissue donor screening.
#### ADVIA Centaur® CMV IgG Quality Control:
The ADVIA Centaur® CMV IgG) Quality Control material is for in vitro diagnostic use for monitoring the performance of the ADVIA Centaur CMV IgG (CMV IgG) assay on ADVIA Centaur systems.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### 1. Applicant:
| Contact: | Kira Gordon, PhD<br>Sr. Regulatory Affairs Specialist |
|----------|--------------------------------------------------------------------------------|
| Address: | Siemens Healthcare Diagnostics Inc.<br>511 Benedict Ave<br>Tarrytown, NY 10591 |
| Phone: | (914) 524-2996 |
| FAX: | (914) 524-3579 |
| Email: | kira.gordon@siemens-healthineers.com |
- 2. Date: May 01, 2018
### 3. Proprietary and Established Names:
ADVIA® Centaur CMV IgG ADVIA® Centaur CMV IgG Quality Control
#### 4. Regulatory Information:
| Classification Names: | Cytomegalovirus serological reagents |
|-----------------------|--------------------------------------|
| Regulation Number: | 21 CFR § 866.3175 |
| Classification: | Class II |
| Product Code: | LFZ |
| Panel: | Microbiology (83) |
### 4. Predicate Device:
Device Name: bioMerieux VIDAS CMV IgG (CMVG) Assay 510(k) Number: k920661 Manufacturer: bioMerieux
### 5. Indications for Use:
ADVIA Centaur CMV lgG:
The ADVIA Centaur® CMV IgG Assay is for in vitro diagnostic use in the qualitative detection of IgG antibodies to cytomegalovirus (CMV) in human pediatric and adult serum and plasma (dipotassium EDTA, lithium heparin) using the ADVIA Centaur XP system. The assay is used to determine CMV IgG serological status and as an aid in the diagnosis of CMV infection in individuals for whom a CMV IgG test was ordered, including pregnant women.
The ADVIA Centaur CMV IgG assay is not intended for blood and tissue donor screening.
ADVIA Centaur® CMV IgG Quality Control: The ADVIA Centaur® CMV IgG (CMV IgG) Quality Control material is for in vitro
{4}------------------------------------------------
diagnostic use for monitoring the performance of the ADVIA Centaur CMV IgG (CMV IgG) assay on ADVIA Centaur systems.
### 9. Device Description:
CMV IgG Assay Kit (100-Tests) consists of 1 ReadyPack containing ADVIA Centaur CMV IgG Lite Reagent and Solid Phase reagent, and 1 ReadyPack ancillary reagent pack that contains ADVIA Centaur CMV IgG Diluent, and a set of Calibrators (1 vial each of Low and High, with fill volume of 2 mL each). The reagents and calibrators are packaged together in the ADVIA Centaur CMV IgG Assay kit, while the associated ADVIA Centaur CMV IgG Quality Control is packaged separately.
# 10. Test Principle
The ADVIA Centaur CMV IgG assay is a fully automated, 2-step sandwich immunoassay using indirect chemiluminometric technology. The patient specimen is diluted with ADVIA Centaur CMV IgG Diluent and incubated with the Solid Phase reagent. The Solid Phase reagent contains a heterogeneous mixture of biotinylated CMV viral lysate antigens, preformed to streptavidin-coated magnetic particles. The antigen-coated particles subsequently capture CMV-specific antibodies in the specimen. The antibody-antigen complex is washed and Lite reagent is added. The Lite reagent consists of an acridinium-ester (AE)-labeled anti-human IgG mouse monoclonal antibody. The entire complex is washed and the signal is generated in the presence of Lite Reagent bound to the Solid Phase via the CMV IgG-CMV antigen complex.
The system automatically performs the following steps:
- Dispenses 20 uL of sample into a cuvette. -
- Dispenses 195 µL of ADVIA Centaur CMV IgG Diluent into the cuvette with the sample. -
- Removes 100 uL of the diluted sample from the cuvette and dispenses it into a second cuvette.
- -Dispenses 200 uL of Solid Phase and incubates the mixture for 18.0 minutes at 37°C.
- -Separates the Solid Phase from the mixture and aspirates the unbound reagent.
- -Washes the cuvette with ADVIA Centaur Wash 1.
- -Re-suspends the washed particles in 250 uL of ADVIA Centaur Wash 1.
- Dispenses 100 uL Lite Reagent and incubates the mixture for 18.0 minutes at 37°C. -
- -Separates the Solid Phase from the mixture and aspirates the unbound reagent.
- Washes the cuvette with ADVIA Centaur Wash 1. -
- Dispenses 300 uL of ADVIA Centaur Acid Reagent and 300 uL of ADVIA Centaur Base -Reagent to initiate the chemiluminescent reaction.
- Reports results according to the selected option, as described in the system operating -Instructions.
A direct relationship exists between the amount of bound anti-CMV IgG present in the patient specimen and the amount of relative light units (RLUs) detected by the system. A result of reactive or nonreactive is determined according to the Index established with the calibrators:
- . Samples with an Index Value of < 1.00 are considered nonreactive for CMV
{5}------------------------------------------------
IgG antibodies.
- . Samples with an Index Value of ≥ 1.00 are considered reactive for CMV IgG antibodies.
#### 11. Substantial Equivalence Information:
Device Name: bioMerieux VIDAS CMV IgG (CMVG) Assay 510(k) Number: k920661
Comparison with Predicate:
| Item | New Device: | Predicate Device: |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The ADVIA Centaur® CMV IgG Assay is<br>for in vitro diagnostic use in the qualitative<br>detection of IgG antibodies to<br>cytomegalovirus (CMV) in human pediatric<br>and adult serum and plasma (dipotassium<br>EDTA, lithium heparin) using the ADVIA<br>Centaur XP system. The assay is used to<br>determine CMV IgG serological status and<br>as an aid in the diagnosis of CMV infection<br>in individuals for whom a CMV IgG test<br>was ordered, including pregnant women.<br>The ADVIA Centaur CMV IgG assay is not<br>intended for blood and tissue donor<br>screening. | The VIDAS CMV IgG (CMVG)<br>Assay is intended for use on the<br>instruments of the VIDAS family<br>(Vitek ImmunoDiagnostic Assay<br>System) as a semi-quantitative<br>automated enzyme-linked<br>fluorescent immunoassay (ELFA).<br>It is intended for use in<br>determination of CMV<br>immunological experience from a<br>single serum sample, or as an aid in<br>the diagnosis of current CMV<br>infection through evaluation of<br>paired sera for a significant increase<br>in CMV-specific IgG. It is not<br>intended for use in testing<br>(screening) blood or plasma donors. |
| Instrument | ADVIA Centaur | VIDAS/mini-VIDAS |
| Measurement | Qualitative | Semi-Quantitative |
| Technology | Chemiluminescense | enzyme-linked fluorescent<br>immunoassay |
| Assay Protocol | sandwich immunoassay | sandwich immunoassay |
| Sample type | Serum, Plasma | Serum |
| Cut-Offs | < 1.0 (Index) - nonreactive<br>≥ 1.0 (Index) - reactive | < 4 AU/mL - negative<br>> 4 to < 6 AU/mL - equivocal<br>≥ 6 AU/mL - positive |
| Sample Volume | 20 µL | 100 µL |
| Calibrators matrix | Human plasma | Human serum |
| Calibrator fill volume | Liquid, 2 mL | Liquid, 2 mL |
| Controls | Negative and Positive | Negative and Positive |
### Non-Clinical Studies
#### a. Assay Cut-off
A comparison study between ADVIA Centaur CMV IgG assay and a comparator CMV IgG assay was performed to determine the placement of the cutoff value at a 1.00 Index by testing 389 remnant clinical samples. Of these, 196 samples were negative, 186 samples were positive, and the rest were equivocal by the comparator device. The cutoff value was set at ≥ 95% positive agreement and ≥ 95% negative agreement to the comparator device. Assigned values for
{6}------------------------------------------------
calibrators and controls are based on the cutoff Index determination.
### b. Precision
Precision was evaluated according to CLSI document EP05-A3. Repeatability and Within-Lab imprecision were evaluated by testing 5 serum-based samples (serum sample pools), and 2 plasma-based samples (controls. negative and positive). The samples were assayed in duplicate over the course of 20 days, 2 runs per day, for a total of 40 runs and 80 replicates. The results
{7}------------------------------------------------
| are shown below | | | | | | | | | | |
|----------------------|----|------|---------------|------|-------------|------|-------------|------|------------|------|
| Sample<br>Type | N | Mean | Repeatability | | Between-Run | | Between-Day | | Within-Lab | |
| | | | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| Control 1 (negative) | 80 | 0.08 | 0.00 | NAa | 0.00 | 0.0 | 0.00 | 0.0 | 0.01 | NA |
| Control 2 (positive) | 80 | 4.54 | 0.09 | 2.0 | 0.12 | 2.6 | 0.09 | 2.0 | 0.18 | 3.9 |
| Serum Pool 1 | 80 | 0.07 | 0.00 | NA | 0.00 | 0.6 | 0.00 | NA | 0.01 | NA |
| Serum Pool 2 | 80 | 0.90 | 0.02 | 2.1 | 0.01 | 1.3 | 0.02 | 2.3 | 0.03 | 3.4 |
| Serum Pool 3 | 80 | 1.36 | 0.03 | 2.5 | 0.02 | 1.1 | 0.02 | 2.1 | 0.05 | 3.5 |
| Serum Pool 4 | 80 | 2.70 | 0.06 | 2.2 | 0.07 | 2.5 | 0.07 | 1.7 | 0.10 | 3.7 |
| Serum Pool 5 | 80 | 9.68 | 0.20 | 2.1 | 0.12 | 1.3 | 0.12 | 1.8 | 0.29 | 3.0 |
a NA = not applicable
#### c. Reproducibility
The reproducibility study was conducted at three external sites using 2 reagent lots. The protocol was run over 5 days, 2 runs per day, and 3 replicates per run for the sample pools, and 6 replicates per run for the negative and positive control materials. Reproducibility data was pooled for each reagent lot across three sites. Data is presented for 1 representative reagent lot.
| Sample<br>Type | N | Mean | Repeatability | | Between-Run | | Between-Day | | Between-Site | | Reproducibility | |
|-------------------------|-----|-------|---------------|------|-------------|------|-------------|------|--------------|------|-----------------|------|
| | | | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| Control 1<br>(negative) | 180 | 0.10 | 0.01 | NAa | 0.00 | NA | 0.00 | NA | 0.02 | NA | 0.02 | NA |
| Control 2<br>(positive) | 180 | 4.55 | 0.12 | 2.7 | 0.07 | 1.5 | 0.02 | 0.3 | 0.15 | 3.4 | 0.21 | 4.6 |
| Serum<br>Pool 1 | 90 | 0.10 | 0.01 | NA | 0.00 | NA | 0.00 | NA | 0.01 | NA | 0.01 | NA |
| Serum<br>Pool 2 | 89b | 0.80 | 0.01 | 1.8 | 0.02 | 1.9 | 0.01 | 1.1 | 0.05 | 6.1 | 0.05 | 6.7 |
| Serum<br>Pool 3 | 90 | 1.23 | 0.02 | 1.9 | 0.02 | 2.0 | 0.00 | 0.0 | 0.07 | 5.4 | 0.07 | 6.0 |
| Serum<br>Pool 4 | 90 | 2.98 | 0.06 | 1.9 | 0.06 | 1.9 | 0.05 | 1.7 | 0.13 | 4.2 | 0.16 | 5.3 |
| Serum<br>Pool 5 | 90 | 25.25 | 0.45 | 1.8 | 0.40 | 1.6 | 0.26 | 1.0 | 1.64 | 6.5 | 1.77 | 7.0 |
a NA = not applicable.
b One run had 2 replicates instead of 3.
{8}------------------------------------------------
#### d. Interference
Interference by endogenous substances in the ADVIA Centaur CMV IgG assay was evaluated at four CMV IgG levels (low negative, high negative, low positive, and high positive). Interfering substances at the levels indicated were tested as described in CLSI Document EP07-A2. There was no change in clinical interpretation throughout the assay range at the levels indicated.
| Interferent | Concentration |
|------------------|-------------------------------------------|
| Hemolyzed | 500 mg/dL of hemoglobin |
| Icteric | 20 mg/dL of conjugated bilirubin |
| Icteric | 20 mg/dL of unconjugated bilirubin |
| Lipemic | 3000 mg/dL of intralipids (Triglycerides) |
| Total Protein* | 9g/dL of protein |
| Immunoglobulins* | 3g/dL of immunoglobulin |
| Biotin | 4500 ng/mL of biotin |
| Cholesterol | 400 mg/dL of cholesterol |
* Total protein and immunoglobulins were tested using clinical samples with high serum protein and samples from multiple myeloma patients respectively.
#### e. Cross-reactivity
The ADVIA Centaur CMV IgG assay was evaluated for potential cross-reactivity in specimens with other viral and microbial antibodies and other disease states. The CMV IgG status of each
{9}------------------------------------------------
sample was compared using the comparator assay.
| Clinical Category | Number Tested | ADVIA Centaur CMV IgG Reactive | Comparator CMV IgG Reactive |
|-----------------------------------------------|---------------|--------------------------------|-----------------------------|
| Chlamydia IgG | 11 | 4 | 4 |
| CMV IgM | 10 | 6 | 6 |
| Epstein-Barr virus (EBV) IgG | 13 | 0 | 0 |
| Epstein-Barr virus (EBV) IgM | 10 | 4 | 4 |
| Graves' disease | 3* | 0 | 0 |
| Hepatitis A infection (HAV) IgG | 10 | 3 | 3 |
| Hepatitis B Core (HBc) IgG | 10 | 3 | 4 |
| Hepatitis C infection (HCV) IgG | 10 | 1 | 1 |
| Herpes simplex virus 1 (HSV1) IgG | 13 | 3 | 3 |
| Herpes simplex virus 2 (HSV2) IgG | 12 | 0 | 0 |
| Human anti-mouse antibody (HAMA) | 14 | 7 | 6 |
| Human chorionic gonadotropin (hCG) | 11 | 0 | 0 |
| Human herpes virus (HHV6) IgG | 11 | 0 | 0 |
| Human immunodeficiency virus (HIV) Antibodies | 16 | 9 | 9 |
| Influenza Antibodies | 11 | 7 | 7 |
| Measles IgG | 11 | 0 | 0 |
| Multiparity | 20 | 18 | 18 |
| Multiple myeloma | 23 | 11 | 11 |
| Parvovirus B19 IgG | 11 | 3 | 3 |
| Rheumatoid factor (RF) | 10 | 1 | 1 |
| Rubella IgG | 13 | 0 | 0 |
| Sjogren's Syndrome | 4* | 0 | 0 |
| Systemic lupus erythematosus (SLE)* | 3* | 0 | 0 |
| Syphilis IgG | 11 | 4 | 4 |
| Toxoplasma IgG | 10 | 4 | 4 |
| Varicella zoster virus (VZV) IgG | 16 | 0 | 0 |
| Total | 297 | 88 | 88 |
*Results may not be conclusive due to low number of samples tested.
# f. Matrix Comparison
The ADVIA Centaur CMV IgG assay was evaluated using different specimen matrices.
{10}------------------------------------------------
ADVIA Centaur CMV IgG results that ranged from 0.74-28.55 Index were analyzed using orthogonal regression. The following results were obtained:
| Serum (x) vs. | N | Mean<br>(Index) | Slope | Intercept<br>(Index) | Correlation Coefficient (r) |
|----------------------------|----|-----------------|-------|----------------------|-----------------------------|
| Dipotassium EDTA plasma | 38 | 10.76 | 1.01 | -0.08 | 0.99 |
| Lithium heparin plasma (y) | 38 | 10.76 | 1.01 | -0.07 | 0.96 |
# 12. Clinical Studies
- a. Method Comparison with predicate device
Percent agreement was determined by comparing the performance of the ADVIA Centaur CMV IgG assay to a comparator CMV IgG assay. A total of 1842 samples that were sent for CMV IgG testing were analyzed, including:
- · 1699 prospectively collected specimens
- 684 general population subjects sent for CMV IgG testing o
- 348 pregnant subjects o
- o 229 pediatric subjects (2-21 years old)
- 44 HIV-positive subjects o
- o 394 transplant-patient subjects
- · 143 retrospective HIV-positive specimens
#### Prospective Study
A total of 1699 clinical routine specimens (from people aged 6 months-91 years, both male and female) were obtained from 6 collection sites in the United States. The specimens were from subjects sent for CMV IgG testing, pediatric subjects (aged 2-21 years), pregnant women. HIVpositive subjects, and transplant patients. The following results were obtained:
#### Prospective Study - Combined Population
| | Comparator-CMV IgG Assay | | | |
|-----------------------------|--------------------------|-----------|----------|-------|
| ADVIA Centaur CMV IgG Assay | Positive | Equivocal | Negative | Total |
| Reactive | 1037 | 11 | 10 | 1058 |
| Nonreactive | 1 | 3 | 637 | 641 |
| Total | 1038 | 14 | 647 | 1699 |
| ADVIA Centaur CMV IgG Assay | Agreement (%) | 95% Confidence Interval |
|-----------------------------|-------------------|-------------------------|
| Positive agreement | 99.6% (1037/1041) | 99.0%-99.9% |
| Negative agreement | 96.8% (637/658) | 95.2%-98.0% |
Note: Fourteen samples that tested equivocal on the comparative assay were further tested on 2 other CMV IgG comparator assays. Of these 14 samples, 10 remained equivocal, 3 agreed and 1 did not agree with the ADVIA Centaur CMV IgG assay when compared to the two out of three consensus result.
The results obtained with the different subgroup prospective populations are presented in the
{11}------------------------------------------------
sections that follow.
#### Subjects Sent for CMV IgG Testing
A total of 684 general population prospective serum samples sent for CMV IgG testing were analyzed. The following results were obtained:
| ADVIA Centaur CMV IgG Assay | Comparator CMV IgG Assay | | | |
|-----------------------------|--------------------------|-----------|----------|-------|
| | Positive | Equivocal | Negative | Total |
| Reactive | 375 | 6 | 7 | 388 |
| Nonreactive | 1 | 1 | 294 | 296 |
| Total | 376 | 7 | 301 | 684 |
#### Subjects Sent for CMV IgG Testing
| ADVIA Centaur CMV IgG Assay | Agreement (%) | 95% Confidence Interval |
|-----------------------------|-----------------|-------------------------|
| Positive agreement | 99.5% (375/377) | 98.1%-99.9% |
| Negative agreement | 95.8% (294/307) | 92.9%-97.7% |
Note: Seven samples that tested equivocal on the comparator assay were further tested on 2 other CMV IgG assays. Of these 7 samples, 5 remained equivocal and 2 agreed with the ADVIA Centaur CMV IgG assay when compared to the two out of three consensus result.
#### Pregnant Women Subgroup
Serum samples sent for CMV IgG testing from 348 pregnant women were prospectively collected from 4 sites and tested. The following results were obtained:
| | Comparator CMV IgG Assay | | | |
|-----------------------------|--------------------------|-----------|----------|-------|
| ADVIA Centaur CMV IgG Assay | Positive | Equivocal | Negative | Total |
| Reactive | 287 | 0 | 1 | 288 |
| Nonreactive | 0 | 0 | 60 | 60 |
| Total | 287 | 0 | 61 | 348 |
#### Pregnant Women Subgroup
| ADVIA Centaur CMV IgG Assay | Agreement (%) | 95% Confidence Interval |
|-----------------------------|------------------|-------------------------|
| Positive agreement | 100.0% (287/287) | 98.7%-100.0% |
| Negative agreement | 98.4% (60/61) | 91.2%-99.9% |
### Pediatric Subgroup
A total of 229 pediatric prospective serum samples were tested at 3 sites. Specimens were obtained from 4 sites, from males who were not pregnant. The subjects' ages
{12}------------------------------------------------
ranged from 2-21 years. The following results were obtained:
| Pediatric Subgroup | | | | |
|-----------------------------|--------------------------|-----------|----------|-------|
| | Comparator CMV IgG Assay | | | |
| ADVIA Centaur CMV IgG Assay | Positive | Equivocal | Negative | Total |
| Reactive | 80 | 1 | 1 | 82 |
| Nonreactive | 0 | 1 | 146 | 147 |
| Total | 80 | 2 | 147 | 229 |
| ADVIA Centaur CMV IgG Assay | Agreement (%) | 95% Confidence Interval |
|-----------------------------|-----------------|-------------------------|
| Positive agreement | 98.8% (80/81) | 93.3%-99.9% |
| Negative agreement | 98.6% (146/148) | 95.2%-99.8% |
Note: Two samples that tested equivocal on the comparator assay were further tested on 2 other CMV IgG assays and the two out of three consensus result remained equivocal.
### Human Immunodeficiency Virus (HIV) Patient Subgroup
Serum specimens sent for CMV IgG testing were prospectively collected at 5 sites, from 44 HIV-positive patients and tested. The following results were obtained:
#### HIV-Positive Patient Subgroup
| | Comparator CMV IgG Assay | | | |
|-----------------------------|--------------------------|-----------|----------|-------|
| ADVIA Centaur CMV IgG Assay | Positive | Equivocal | Negative | Total |
| Reactive | 43 | 0 | 0 | 43 |
| Nonreactive | 0 | 0 | 1 | 1 |
| Total | 43 | 0 | 1 | 44 |
| ADVIA Centaur CMV IgG Assay | Agreement (%) | 95% Confidence Interval |
|-----------------------------|----------------|-------------------------|
| Positive agreement | 100.0% (43/43) | 91.8%–100.0% |
| Negative agreement | 100.0% (1/1) | 2.5%–100.0% |
#### Transplant Patient Subgroup
Serum samples from 238 preoperative transplant patients from 3 sites were tested. The
{13}------------------------------------------------
following results were obtained:
| | | Transplant Patient Subgroup - Preoperative |
|--|--|--------------------------------------------|
|--|--|--------------------------------------------|
| ADVIA Centaur CMV IgG Assay | Comparator CMV IgG Assay | | | |
|-----------------------------|--------------------------|-----------|----------|-------|
| | Positive | Equivocal | Negative | Total |
| Reactive | 151 | 0 | 0 | 151 |
| Nonreactive | 0 | 0 | 87 | 87 |
| Total | 151 | 0 | 87 | 238 |
| ADVIA Centaur CMV IgG Assay | Agreement (%) | 95% Confidence Interval |
|-----------------------------|------------------|-------------------------|
| Positive agreement | 100.0% (151/151) | 97.6%–100.0% |
| Negative agreement | 100.0% (87/87) | 95.8%–100.0% |
Serum samples from 156 transplant patients were collected from 3 sites that perform kidney, pancreas, bone and marrow stem cell, liver, face and limb, heart, and lung transplants. After testing the following results were obtained
#### Transplant Patient Subgroup - Postoperative
| | Comparator CMV IgG Assay | | | |
|-----------------------------|--------------------------|-----------|----------|-------|
| ADVIA Centaur CMV IgG Assay | Positive | Equivocal | Negative | Total |
| Reactive | 101 | 4 | 1 | 106 |
| Nonreactive | 0 | 1 | 49 | 50 |
| Total | 101 | 5 | 50 | 156 |
| ADVIA Centaur CMV IgG Assay | Agreement (%) | 95% Confidence Interval |
|-----------------------------|-----------------|-------------------------|
| Positive agreement | 99.0% (101/102) | 94.7%-99.9% |
| Negative agreement | 90.7% (49/54) | 79.7%-96.9% |
Note: Five samples that tested equivocal by the comparator assay were further tested on 2 other CMV IgG assays. Of these 5 samples, 3 remained equivocal, 1 agreed and 1 did not agree with the ADVIA Centaur CMV IgG assay when compared to the two of three consensus result.
### Retrospective Study - HIV-Positive Patient Subgroup
A total of 143 retrospective remnant samples from the HIV-positive patient subgroup collected at 1
{14}------------------------------------------------
site were tested. The following results were obtained:
| Comparator CMV IgG Assay | | | | |
|-----------------------------|----------|-----------|----------|-------|
| ADVIA Centaur CMV IgG Assay | Positive | Equivocal | Negative | Total |
| Reactive | 135 | 0 | 0 | 135 |
| Nonreactive | 0 | 0 | 8 | 8 |
| Total | 135 | 0 | 8 | 143 |
#### Retrospective HIV-Positive Patient Subgroup
| ADVIA Centaur CMV IgG Assay | Agreement (%) | 95% Confidence Interval |
|-----------------------------|------------------|-------------------------|
| Positive agreement | 100.0% (135/135) | 97.3%-100.0% |
| Negative agreement | 100.0% (8/8) | 63.1%-100.0% |
### g. Centers for Disease Control (CDC) Panel:
A panel of 80 previously characterized serum samples was obtained from the CDC and evaluated with the ADVIA Centaur CMV IgG assay to determine the performance of the assay. There was 100% agreement with the serological status provided by the CDC.
| | Expected CDC Panel Results | | |
|-----------------------------|----------------------------|----------|-------|
| ADVIA Centaur CMV IgG Assay | Positive | Negative | Total |
| Reactive | 39 | 0 | 39 |
| Nonreactive | 0 | 41 | 41 |
| Total | 39 | 41 | 80 |
Note: The results are presented as a means to convey further information on the performance of this assay with a characterized serum panel from the CDC. This does not imply an endorsement of the assay by the CDC.
#### 13. Conclusions
From the above comparative testing results of the ADVIA Centaur CMV IgG assay we conclude that it is substantially equivalent to the bioMerieux VIDAS CMV IgG Assay.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.