BD VERITOR SYSTEM FOR THE RAPID DETECTION OF RSV

K133140 · Becton, Dickinson & CO · GQG · Nov 15, 2013 · Microbiology

Device Facts

Record IDK133140
Device NameBD VERITOR SYSTEM FOR THE RAPID DETECTION OF RSV
ApplicantBecton, Dickinson & CO
Product CodeGQG · Microbiology
Decision DateNov 15, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.3480
Device ClassClass 1
AttributesPediatric

Indications for Use

The BD Veritor™ System for Rapid Detection of Respiratory Syncytial Virus (RSV) is a chromatographic immunoassay with an instrumented read for the direct and qualitative detection of RSV antigen from nasopharyngeal washes/aspirates and nasopharyngeal swabs in transport media from patients suspected of having a viral respiratory infection. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in neonatal and pediatric patients under the age of 20. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an alternative method, such as a FDA-cleared molecular assay. The test is intended for professional and laboratory use. It is to be used in conjunction with the BD Veritor™ System Reader.

Device Story

Device is a chromatographic immunoassay for qualitative detection of RSV antigen. Input: nasopharyngeal aspirate, wash, or swab in transport media. Process: specimen processed in RV Reagent C tube; 3 drops dispensed into RSV test device and 3 drops into Flu A+B test device; devices inserted into BD Veritor System Reader for instrumented analysis. Output: qualitative RSV and Flu A+B results displayed on reader. Used in professional/laboratory settings. Enables simultaneous testing for RSV and Flu A+B from single specimen; aids clinical diagnosis of viral respiratory infections.

Clinical Evidence

No clinical data provided in this submission; substantial equivalence is based on the identical nature of the reagent and assay technology to the predicate device.

Technological Characteristics

Chromatographic immunoassay; utilizes proprietary enhanced colloidal-gold particles. Includes unitized reagent tubes with mucolytic RV Reagent C. Five-zone test strip design (positive/negative controls, test lines, background). Instrumented read via BD Veritor System Reader. Professional use.

Indications for Use

Indicated for neonatal and pediatric patients <20 years old suspected of viral respiratory infection to aid in RSV diagnosis. Professional/laboratory use only.

Regulatory Classification

Identification

Respiratory syncytial virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to respiratory syncytial virus in serum. Additionally, some of these reagents consist of respiratory syncytial virus antisera conjugated with a fluorescent dye (immunofluorescent reagents) and used to identify respiratory syncytial viruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of respiratory syncytial virus infections and provides epidemiological information on diseases caused by these viruses. Respiratory syncytial viruses cause a number of respiratory tract infections, including the common cold, pharyngitis, and infantile bronchopneumonia.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification Decision Summary To: Becton, Dickinson and Company RE: K133140 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class I device requiring 510(k). The following items are present and acceptable 1. The name and 510(k) number of the SUBMITTER'S previously cleared device: Trade Name: BD Veritor™ System RSV assay 510(k) number: k121633 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling. 3. A description of the device MODIFICATION(S). The modification presented in this 510(k) consisted of adding a procedure for using the obtained clinical specimen to run the BD Veritor™ System RSV assay in addition to the Flu A+B assay, both assays will be performed separately, but using a single nasopharyngeal aspirate/wash or nasopharyngeal swab in transport media. The following text and accompanying diagrams were added to the package insert: ALTERNATIVE TEST PROCEDURE USE OF BD VERITOR SYSTEM FOR RAPID DETECTION OF BD VERITOR SYSTEM FOR RAPID DETECTION OF RSV (CAT# 256042) and INFLUENZA A+B (CAT# 256041) FROM A SINGLE NP WASH, ASPIRATE OR SWAB SPECIMEN IN TRANSPORT MEDIA USE THIS PROCEDURE IF BOTH RSV AND INFLUENZA ARE TO BE TESTED FROM A SINGLE PATIENT IMPORTANT NOTE: THE SAMPLE MAY BE TESTED UP TO 15 MINUTES AFTER PROCESSING. NOTES: Reagents, specimens and devices must be at room temperature (15–30°C) for testing. Thoroughly mix all specimens prior to removal of an aliquot for processing. Do not centrifuge specimens. Note: The BD Veritor System for Rapid Detection of Flu A+B (CAT #256041) is required for this procedure. 1. For each patient specimen and control swab, remove one RV Reagent C tube/tip and one BD Veritor System RSV device and one BD Veritor System Flu A+B from its foil pouch immediately before testing. ![img-0.jpeg](img-0.jpeg) 2. Label the BD Veritor System devices and one RV Reagent C tube for each specimen and control to be tested. 3. Place the labeled RV Reagent C tube(s) in the designated area of the tube rack. 4. Vortex or thoroughly mix specimen. Do not centrifuge. 5. 2. Remove and discard the cap from the RV Reagent C tube corresponding to the sample to be tested. 6. Using the transfer pipette, transfer 300 μL of specimen into the RV Reagent C tube. Discard pipette after use. ![img-1.jpeg](img-1.jpeg) {1} 2 7. Press the attached tip firmly onto the RV Reagent C tube containing the processed specimen or control (threading/twisting not required). NOTE: Do not use tips from any other product, including other products from BD or other manufacturers. 8. Vortex or mix thoroughly. 9. Invert the RV Reagent C tube and hold the tube vertically (approximately one inch above the BD Veritor System RSV device sample well). Holding the tube at the ridged area, squeeze gently allowing three (3) drops of the processed sample to be dispensed into the sample well of the appropriately labeled BD Veritor System RSV device. NOTE: Squeezing the tube too close to the tip may cause leakage. 10. Continue to test for Flu A+B. Invert the RV Reagent C tube and hold the tube vertically (approximately one inch above the BD Veritor System Flu A+B device sample well). Holding the tube at the ridged area, squeeze gently allowing three (3) drops of the processed sample to be dispensed into the sample well of the appropriately labeled BD Veritor System Flu A+B device. NOTE: Squeezing the tube too close to the tip may cause leakage. 11. After adding the sample, allow the tests to run for 10 minutes before inserting into the reader. 12. When the test is ready, insert the BD Veritor System RSV device into the BD Veritor System Reader. (The BD Veritor System Reader should be powered-on prior to use and will indicate when it is ready for insertion of the BD Veritor System device.) 13. Follow the reader on-screen prompts to complete the procedure and obtain the test result 13. After the RSV result is obtained and recorded, insert the BD Veritor System Flu A+B device into the BD Veritor System Reader. 14. Follow the reader on-screen prompts to complete the procedure and obtain the test result ![img-2.jpeg](img-2.jpeg) ![img-3.jpeg](img-3.jpeg) 4. The FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. 5. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, and physical characteristics: Similarities | Item | Proposed Device | Proposed Device | | --- | --- | --- | {2} 3 | Features | BD Veritor™ System Flu A+B assay | BD Veritor™ System Flu A+B assay | | --- | --- | --- | | Intended Use | The BD Veritor™ System for Rapid Detection of Respiratory Syncytial Virus (RSV) is a chromatographic immunoassay with an instrumented read for the direct and qualitative detection of RSV antigen from nasopharyngeal washes/aspirates and nasopharyngeal swabs in transport media from patients suspected of having a viral respiratory infection. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in infants and pediatric patients under the age of 20 years. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an alternative method, such as a FDA-cleared molecular assay. The test is intended for professional and laboratory use. It is to be used in conjunction with the BD Veritor™ System Reader. | Same. | | Read Results | BD Veritor™ System Reader | BD Veritor™ System Reader | | Specimen Types | nasopharyngeal swab in transport media, nasal aspirate, and nasal wash | nasopharyngeal swab in transport media, nasal aspirate, and nasal wash | | Read Result Time | 10 minutes | 10 minutes | | External Controls | Test kit contains Positive and Negative Control swabs | Test kit contains Positive and Negative Control swabs | ## Differences The package insert has been updated to include an additional procedure for using the obtained clinical specimen to run the BD Veritor™ System Flu A+B assay in addition to the RSV assay. Both assays will be performed separately, but using a single nasopharyngeal aspirate/wash or nasopharyngeal swab in transport media. ## 6. Design Control Activities Summary: a) Analytical Reactivity Testing was conducted as described in section 3, Device Modifications. b) Declaration of Conformity A "Declaration of Conformity" statement was submitted for the manufacturing facility and validation activities and signed by the Director of Quality Assurance and the Senior Director of Technical Operations respectively. The statements indicate that; {3} 4 1. The manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 2. The validation activities, as required by the risk analysis, for the modification were performed by the designated individuals and the results demonstrated that the predetermined acceptance criteria were met. In conclusion, based on the results of the analytical reactivity testing the modified labeling is truthful and accurate. The changes do not affect the performance of the test and it is therefore substantially equivalent to the current cleared test. 7. A Truthful and Accurate Statement, a 510(k) Summary, and the Indications for Use Enclosure. The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. On this basis, I recommend the device be determined substantially equivalent to the previously cleared device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...