K Number
K121633

Validate with FDA (Live)

Date Cleared
2012-09-18

(106 days)

Product Code
Regulation Number
866.3480
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Veritor™ System for Rapid Detection of Respiratory Syncytial Virus (RSV) is a chromatographic immunoassay with an instrumented read for the direct and qualitative detection of RSV fusion protein from nasopharyngeal washes/aspirates and nasopharyngeal swabs in transport media from patients suspected of having a viral respiratory infection. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in infants and pediatric patients under the age of 20 years. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an alternative method, such as a FDA-cleared molecular assay. The test is intended for professional and laboratory use. It is to be used in conjunction with the BD Veritor ™ System Reader.

Device Description

The BD RSV test is a chromatographic assay to qualitatively detect RSV fusion protein in samples processed from respiratory specimens. The processed specimen is added to the test device where RSV viral antigens bind to anti-RSV antibodies conjugated to detector particles on the RSV test strip. The antigen-conjugate complex migrates across the test strip to the reaction area and is captured by an antibody line on the membrane. Results are interpreted by the BD Veritor™ System Reader, a portable electronic device which uses a reflectance-based measurement method to evaluate the line signal intensities on the assay test strip, and applies specific algorithms to determine the presence or absence of any target analyte(s). A liquid crystal display (LCD) on the instrument communicates the results to the operator.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the BD Veritor™ System for Rapid Detection of RSV, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" as a separate table. However, based on the clinical study results and comparisons, the implicit criteria are high sensitivity and specificity relative to viral cell culture.

MetricAcceptance Criteria (Implicit from Context - comparable to predicate/clinical needs)Reported Device Performance (BD Veritor™ System for RSV)
Clinical Performance (NPS)High sensitivity and specificity acceptable for an aid in diagnosisSensitivity: 88.4% (95% CI: 82.8%, 92.4%)Specificity: 98.3% (95% CI: 96.8%, 99.1%)
Clinical Performance (NPWA)High sensitivity and specificity acceptable for an aid in diagnosisSensitivity: 91.6% (95% CI: 86.3%, 94.9%)Specificity: 94.5% (95% CI: 91.2%, 96.7%)
Analytical Sensitivity (LOD) - Viral Strain VR-26 (Long Subgroup A)Low concentration producing ≥95% positivity$1.43 \times 10^5$ TCID50/mL (95.0% positive)
Analytical Sensitivity (LOD) - Viral Strain VR-955 (9320 subgroup B)Low concentration producing ≥95% positivity$3.98 \times 10^4$ TCID50/mL (95.0% positive)
Analytical Sensitivity (LOD) - Viral Strain VR-1540 (A-2)Low concentration producing ≥95% positivity$1.94 \times 10^3$ TCID50/mL (98.3% positive)
Analytical Sensitivity (LOD) - Viral Strain VR-1580 (Washington subgroup B)Low concentration producing ≥95% positivity$1.08 \times 10^4$ TCID50/mL (96.7% positive)
Analytical Sensitivity (LOD) - Viral Strain VR-1400 (Wild Type subgroup B)Low concentration producing ≥95% positivity$2.96 \times 10^3$ TCID50/mL (95.0% positive)
Analytical Specificity (Cross-Reactivity)No cross-reactivity with common respiratory pathogens and floraNone showed cross-reactivity with tested microorganisms
Interfering SubstancesNo interference with tested substancesNo interference noted for any of the tested substances
Media CompatibilityNo interference or compatibility issues with common transport mediaNo interference or compatibility issues seen with tested media
ReproducibilityConsistent results across sites and operatorsHigh consistency for moderate and negative samples, acceptable for low positive samples (e.g., Moderate positive RSV: 100% (90/90) total positive)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 1146 prospectively collected specimens.
    • 440 Nasopharyngeal Wash / Aspirates (NPWA)
    • 706 Nasopharyngeal Swabs (NPS)
  • Data Provenance: From five U.S. trial sites during the 2011-2012 respiratory season. The data is prospective, as indicated by "prospectively collected specimens."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not specify the number or qualifications of experts used to establish the ground truth. It states that the ground truth was "an FDA cleared D3 Duet™ DFA on R-Mix cell culture," which is a laboratory method, not directly human experts.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method for the test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, an MRMC comparative effectiveness study involving human readers with and without AI assistance (or in this case, instrumented read) was not explicitly described. This is a point-of-care diagnostic device with an instrumented reader, not an AI for image analysis where human readers' performance might be augmented. The device itself performs the interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Yes, this study represents a standalone evaluation of the device. The BD Veritor™ System for Rapid Detection of RSV test is an "instrumented read" system where the results are "interpreted by the BD Veritor™ System Reader... [which] applies specific algorithms to determine the presence or absence of any target analyte(s)." The clinical study compares this instrument's performance directly against the reference method (viral cell culture), indicating a standalone assessment.

7. The Type of Ground Truth Used

The primary ground truth used was viral cell culture (specifically, an "FDA cleared D3 Duet™ DFA on R-Mix cell culture"). For some discrepant results (BD Veritor RSV Positive, Viral Cell Culture negative specimens), an "FDA cleared Prodesse Pro Flu+ molecular assay" was used as a secondary confirmation method.

8. The Sample Size for the Training Set

The document does not specify a separate training set size. The device uses "pre-set thresholds" and "specific algorithms" but the development and training details for these algorithms are not provided in this summary. The clinical study described is for validation/testing, not training.

9. How the Ground Truth for the Training Set Was Established

Since a dedicated training set is not explicitly mentioned, the method for establishing ground truth for any internal algorithm development (if applicable) is not detailed in this 510(k) summary. The provided performance data pertains to the validation of the final device.

{0}------------------------------------------------

K121633

BD Veritor™ System for Rapid Detection of RSV Clinical Laboratory Product

510(k) SUMMARY

SEP 1 8 2012

SUBMITTED BY:BECTON, DICKINSON AND COMPANY10865 Road to the Cure, Suite 200San Diego, CA 92121Phone: 858-795-7890Fax: 858-812-8505
CONTACT NAME:Gregory Payne

DATE PREPARED: September 12, 2012

DEVICE TRADE NAME: BD Veritor™ System for Rapid Detection of RSV

DEVICE COMMON NAME: Antigens Cf (including Cf Controls) Respiratory Syncytial Virus

DEVICE CLASSIFICATION: 21 CFR § 866.3480

PREDICATE DEVICES: Quidel QuickVue RSV 10 test

INTENDED USE:

The BD Veritor™ System for Rapid Detection of Respiratory Syncytial Virus (RSV) is a chromatographic immunoassay with an instrumented read for the direct and qualitative detection of RSV fusion protein from nasopharyngeal washes/aspirates and nasopharyngeal swabs in transport media from patients suspected of having a viral respiratory infection. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in infants and pediatric patients under the age of 20 years. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an alternative method, such as a FDA-cleared molecular assay. The test is intended for professional and laboratory use. It is to be used in conjunction with the BD Veritor TM System Reader.

DEVICE DESCRIPTION:

The BD RSV test is a chromatographic assay to qualitatively detect RSV fusion protein in samples processed from respiratory specimens. The processed specimen is added to the test device where RSV viral antigens bind to anti-RSV antibodies conjugated to detector particles on the RSV test strip. The antigen-conjugate complex migrates across the test strip to the reaction area and is captured by an antibody line on the membrane. Results are interpreted by the BD Veritor™ System Reader, a portable electronic device which uses a reflectance-based measurement method to evaluate the line signal intensities on the assay test strip, and applies specific algorithms to determine the presence or absence of any target analyte(s). A liquid crystal display (LCD) on the instrument communicates the results to the operator.

BD Diagnostic Systems Becton, Dickinson and Company

{1}------------------------------------------------

DEVICE COMPARISON:

The BD Veritor™ System for Rapid Detection of RSV was compared to the Quidel .
QuickVue RSV 10 test (K101918)
ProductFeatureBD Veritor™ System for RSVQuidel QuickView RSV 10 test (K101918)
Intended UseThe BD Veritor™ System for Rapid Detection ofRespiratory Syncytial Virus (RSV) is achromatographic immunoassay with aninstrumented read for the direct and qualitativedetection of RSV fusion protein fromnasopharyngeal washes/aspirates andnasopharyngeal swabs in transport media frompatients suspected of having a viral respiratoryinfection. This test is intended for in vitrodiagnostic use to aid in the diagnosis of RSVinfections in infants and pediatric patients underthe age of 20 years. Negative results do notpreclude RSV infection and should not be usedas the sole basis for treatment or for othermanagement decisions. A negative test ispresumptive. It is recommended that negativetest results be confirmed by viral cell culture oran alternative method, such as a FDA-clearedmolecular assay. The test is intended forprofessional and laboratory use. It is to be usedin conjunction with the BD Veritor ™ SystemReader.The QuickVue RSV 10 test is animmunoassay that allows for the rapid,qualitative detection of respiratorysyncytial virus (RSV) antigen directly fromnasopharyngeal swab 'and nasopharyngealaspirate/wash specimens for symptomaticpediatric patients (less than six years old). Thetest is intended for use as an aid in the rapiddiagnosis of acute RSV infection. Negativeresults do not preclude RSV infection andshould not be used as the sole basis fortreatment or for other management decisions.A negative test is presumptive. It isrecommended that negative test results beconfirmed by cell culture. The test is intendedfor professional and laboratory use.
SpecimenTypesNasopharyngeal swab in transport media,nasopharyngeal wash/aspirateNasopharyngeal swab, nasopharyngealwash/aspirate
AssayTechnologyImmunochromatographicImmunochromatographic
DetectionFormatAn opto-electronic reader determines the lineintensity at each of the spatially-defined test andcontrol line positions, interprets the results usingthe scoring algorithm, and reports a positive,negative, or invalid result on the LCD screenbased on pre-set thresholds.Visual determination of presence or absenceof pink-to-red Test Line and the appearance ofa blue Procedural Control Line on the test stripindicate the presence of RSV antigen.
QualitativeYesYes
Total AssayTimeApproximately 10 minutes10 minutes
Control format• Kit RSV positive and RSV negative dry swabprocedural control• Internal positive control• Internal negative control• Kit RSV positive control swab• Kit RSV negative control swab• Internal control lines

.

.

{2}------------------------------------------------

SUMMARY OF PERFORMANCE DATA:

Analytical Sensitivity

The limit of detection (LOD) for the BD Veritor System for Rapid Detection of RSV test was established for the following RSV strains The LOD for each strain represents the lowest concentration producing a positivity rate of ≥95% based on testing 60 or more replicates.

Viral StrainCalculated LOD(TCID50/mL)No. Positive / Total% Positive
VR-26 (Long Subgroup A)$1.43X10^5$57/6095.0
VR-955 (9320 subgroup B)$3.98X10^4$57/6095.0
VR-1540 (A-2)$1.94X10^3$59/6098.3
VR-1580 (Washington subgroup B)$1.08X10^4$58/6096.7
VR-1400 (Wild Type subgroup B)$2.96X10^3$76/8095.0

TCID50/mL = 50% Tissue Culture Infectious Dose

Analytical Specificity (Cross Reactivity)

The BD Veritor System for Rapid Detection of RSV test was evaluated with bacteria and veast at a target concentration of approximately 10° CFU/mL (CFU – Colony Forming Units) with the exception of Fusobacterium nucleatum which was tested at 1.5 X 10°. The viruses were evaluated at concentrations of 103 TCIDsomiL or greater. Of the microorganisms tested, none showed cross-reactivity in the RSV test.

Bacteriodes fragilis Bordetella pertussis Candida albicans Chlamydia pneumoniae Corynebacterium diphtherium Escherichia coli Fusobacterium nucleatum

Haemophilus influenzae Haemophilus parainfluenzae Kingella kingae Klebsiella pneumoniae Lactobacillus sp. Legionella sp. Moraxella catarrhalis Mycobacterium tuberculosis Mycoplasma pneumoniae Neisseria gonorrhoeae Neisseria meningitidis Neisseria mucosa Neisseria sp.(Neisseria perflaus) Neisseria subflava Peptostreptococcus anaerobius Porphyromonas asaccharolyticus Prevotella oralis Propionibacterium acnes Proteus mirabilis Pseudomonas aeruginosa

BD Diagnostic Systems Becton, Dickinson and Company

Adenovirus, type 1 Adenovirus, type 7 Cytomegalovirus Enterovirus HSV Type 1 Human Coronavirus OC43 Human metapneumovirus (HMPV-27 A2) Human Parainfluenza Measles virus Mumps virus Rhinovirus

{3}------------------------------------------------

BD Veritor™ System for Rapid Detection of RSV Clinical Laboratory Product

Serratia marcescens Staphylococcus aureus Staphylococcus epidermidis Streptococcus mutans Streptococcus pneumoniae Streptococcus pyogenes Streptococcus sp. Group C Streptococcus sp. Group G Streptococcus salivarius Veillonella parvula

Interfering Substances

Various substances were evaluated with the BD Veritor System for Rapid Detection of RSV test. These substances included whole blood (2%) and various medications. No interference was noted with this assay for any of the substances at the concentrations tested.

SubstanceConcentrationTested
Whole Blood2%
4-Acetamidophenol10 mg/mL
Acetylsalicylic acid20 mg/mL
Chlorpheniramine maleate5 mg/mL
Dextromethorphan10 mg/mL
Diphenhydramine HCI5 mg/mL
Guaiacol Glyceryl Ether20 mg/mL
Ibuprofen10 mg/mL
Loratidine100 ng/mL
Menthol Throat Lozenges10 mg/mL
Ayr Saline Nasal Gel10 mg/mL
Oxymetazoline0.05 mg/mL
Phenylephrine1 mg/mL
Pseudoephedrine HCI20 mg/mL
Three OTC mouthwashes5 %
Four OTC nasal sprays10 %
Four OTC throat drops25 %
Homeopathic AllergyMedicine10 mg/mL
Albuterol0.083 mg/mL
Synagis4 ug/mL
Amantadine Hydrochloride500 ng/mL
Beclomethasone500 ng/mL
Budesonide500 ng/mL
Dexamethasone10 mg/mL
Fexofenadine500 ng/mL
FluMist1%
Flunisolide500 ng/mL
Fluticasone500 ng/mL
Mometasone500 ng/mL
Mupirocin500 ng/mL
Oseltamivir500 ng/mL
Purified Mucin Protein1 mg/mL

BD Diagnostic Systems Becton, Dickinson and Company

{4}------------------------------------------------

BD Veritor TM System for Rapid Detection of RSV Clinical Laboratory Product

Ribavirin500 ng/mL
Rimantadine500 ng/mL
Tobramycin500 ng/mL
Triamcinolone500 ng/mL
Zanamivir1 mg/mL

Media Compatibility

Different types of transport media commonly used for the preservation and transport of respiratory specimens were evaluated for compatibility with the BD Veritor™ System for Rapid Detection of RSV test. The effects of frozen storage of transport media samples on the stability of the antigen were also evaluated in this study by storing them for 24 hours at -20 ± 5°C. The media tested were: Amies. Bartel ViraTrans. BD Universal. Earle's Minimal Essential, Hank's Balanced Salts, M4, M4-RT, M5, M6, Normal Saline, and Phosphate Buffered Saline. No interference or compatibility issues were seen.

CLINICAL STUDIES

Performance characteristics for the BD Veritor System for Rapid Detection of RSV test were established in multi-center clinical studies conducted at five U.S. trial sites during the 2011-2012 respiratory season. A total of 1174 prospectively collected specimens received in the laboratory with an order for respiratory virus testing were enrolled in the study, of which. 26 were noncompliant with the study protocol and one was noncompliant on the viral cell culture reference testing level. Removal of these specimens yields a total of 1147 specimens. One additional specimen had a final undetermined viral cell culture reference result which could not be verified. Removal of this specimen results in a total of 1146 specimens. A total of 1146 were evaluated using the BD Veritor System for Rapid Detection of RSV test and viral cell culture. The prospective specimens consisted of 440 Nasopharyngeal Wash /Aspirates (NPWA) and 706 nasopharyngeal swabs (NPS) in transport media from symptomatic patients. 44.3% of the samples were from females and 55.7% from males. 80% of patients were 2 years and under.

The performance of the BD Veritor System for Rapid Detection of RSV test was compared to an FDA cleared D3 Duet™ DFA on R-Mix cell culture and is presented in the following tables

{5}------------------------------------------------

Summary of the performance of the BD Veritor System for Rapid Detection of RSV Test compared to viral cell culture by specimen type, all sites.

Clinical Performance- Veritor RSV to ViralCell Culture, By Specimen Type
Culture
SpecimenTypeVeritorRSVPN
NPSP1539*162
NPSN20524544
173533706
Reference Method: Culture
Sensitivity: 88.4% (95% CI: 82.8%, 92.4%)
Specificity: 98.3% (95% CI: 96.8%, 99.1%)
NPWAP15215**167
NPWAN14259273
166274440

Reference Method: Culture

Sensitivity: 91.6% (95% C1: 86.3%, 94.9%)

Specificity: 94.5% (95% C1: 91.2%, 96.7%)

*of the 9 BD Veritor RSV Positive, Viral Cell Culture negative specimens, 6 were positive by FDA cleared Prodesse Pro Flu+ molecular assay

** of the 15 BD Veritor RSV Positive, Viral Cell Culture negative specimens, 8 were positive by FDA cleared Prodesse Pro Flu+ molecular assay

Reproducibility

The reproducibility of the BD Veritor System for Rapid Detection of RSV test was evaluated at three clinical laboratory sites. The reproducibility panel was composed of 12 simulated RSV samples. These included moderate positive samples, low positive samples (near the assay limit of detection), and high negative samples (i.e., containing very low concentrations of virus) and negative samples. The panel was tested by two operators at each site for five consecutive days. The results are summarized below.

BD Veritor™ RSV Reproducibility (% RSV positive results)
SampleS1S3S5Total
High negativeRSV0% (0/30)(0%,11.3%)3.3% (1/30)(0.6%,16.7%)3.3% (1/30)(0.6%,16.7%)2.2% (2/90)(0.6%,7.7%)
Low positiveRSV93.3% (28/30)(78.7%,98.2%)76.7% (23/30)(59.1%,88.2%)93.3% (28/30)(78.7%,98.2%)87.8% (79/90)(79.4%,93%)
Moderatepositive RSV100% (30/30)(88.6%,100%)100% (30/30)(88.6%,100%)100% (30/30)(88.6%,100%)100% (90/90)(95.9%,100%)
Negative0% (0/30)(0%,11.3%)0% (0/30)(0%,11.3%)0% (0/30)(0%,11.3%)0% (0/90)(0%,4.1%)

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure with stylized wings.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

2012

Becton, Dickinson and Company
c/o Gregory P. Payne, RAC
Director, Quality Systems and Regulatory Affairs
10865 Road to the Cure, Suite 200
San Diego, CA 92121

Re: K121633

Trade/Device Name: BD Veritor™ System for Rapid Detection of RSV Regulation Number: 21 CFR §866.3480 Regulation Name: Respiratory syncytial virus serological reagents Regulatory Class: Class I · Product Code: GQG Dated: August 20, 2012 Received: August 22, 2012

Dear Mr. Payne:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, additions of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r rease be actived a determination that your device complies with other requirements of the Act that I Dr Hao mass and regulations administered by other Federal agencies. You must of any I occur statules and securements, including, but not limited to: registration and listing (21

{7}------------------------------------------------

Page 2 - Gregory P. Payne, RAC

CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nredicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Vayain

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

BD Veritor " System for Rapid Detection of RSV Clinical Laboratory Product

510(k) Number:

Device Name: BD Veritor™ System for Rapid Detection of RSV

Indications for Use:

The BD Veritor™ System for Rapid Detection of Respiratory Syncytial Virus (RSV) is a chromatographic immunoassay with an instrumented read for the direct and qualitative detection of RSV fusion protein from nasopharyngeal washes/aspirates and nasopharyngeal swabs in transport media from patients suspected of having a viral respiratory infection. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in infants and pediatric patients under the age of 20 years. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an alternative method, such as a FDAcleared molecular assay. The test is intended for professional and laboratory use. It is to be used in conjunction with the BD Veritor TM System Reader.

Prescription Use ー V -(Part 21 CFR 801 Subpart D) AND/OR

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices Evaluation and Safety (OIVD)

Tamara Feldbly
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K121633

§ 866.3480 Respiratory syncytial virus serological reagents.

(a)
Identification. Respiratory syncytial virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to respiratory syncytial virus in serum. Additionally, some of these reagents consist of respiratory syncytial virus antisera conjugated with a fluorescent dye (immunofluorescent reagents) and used to identify respiratory syncytial viruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of respiratory syncytial virus infections and provides epidemiological information on diseases caused by these viruses. Respiratory syncytial viruses cause a number of respiratory tract infections, including the common cold, pharyngitis, and infantile bronchopneumonia.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.