Puritan PurSafe Plus Collection and Transport System

K242820 · Puritan Medical Products, LLC · QBD · May 27, 2025 · Microbiology

Device Facts

Record IDK242820
Device NamePuritan PurSafe Plus Collection and Transport System
ApplicantPuritan Medical Products, LLC
Product CodeQBD · Microbiology
Decision DateMay 27, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.2950
Device ClassClass 2

Indications for Use

Puritan PurSafe Plus Collection and Transport System is an enclosed system intended for the collection, inactivation, and preservation of human upper respiratory specimens suspected of containing SARS-CoV-2. Puritan PurSafe Plus Collection and Transport System can be used for collection, transport, and storage of specimens at 2-4°C and 25-30°C. Specimens collected and stored in the PurSafe Plus Collection and Transport System are suitable for use with legally marketed molecular diagnostic devices.

Device Story

System consists of sterile polyester flock swab and polypropylene vial containing 1mL PurSafe Plus buffer; used for collection of human upper respiratory specimens. Buffer inactivates SARS-CoV-2 and preserves viral RNA for molecular diagnostic testing. Used in clinical settings; operated by healthcare professionals. Specimens stored at 2-4°C or 25-30°C. Output is stabilized specimen in transport medium; processed by downstream molecular diagnostic instruments (e.g., Cepheid GeneXpert). Benefits include safe handling of potentially infectious samples and reliable preservation of viral RNA for diagnostic accuracy.

Clinical Evidence

Bench testing only. LOD studies (N=24) confirmed detection of SARS-CoV-2 at 64 copies/mL using Cepheid GeneXpert, equivalent to predicate. Stability studies (N=72) confirmed RNA preservation for 28 days at 4°C and 30°C with no significant Ct value shifts. Inactivation studies confirmed SARS-CoV-2 inactivation after 1 minute of exposure to buffer.

Technological Characteristics

System includes sterile polyester flock swab and polypropylene vial with 1 mL MK buffer (Guanidine thiocyanate, HEPES, Tris Base, Triton X-100, EDTA). pH 8.3 ± 0.2. Conductivity 50-100 mS/cm. Shelf-life 24 months at 2-4°C or 30-35°C. Single-use, non-sterile buffer, sterile swab.

Indications for Use

Indicated for the collection, inactivation, and preservation of human upper respiratory specimens suspected of containing SARS-CoV-2 for use with molecular diagnostic devices.

Regulatory Classification

Identification

A microbial nucleic acid storage and stabilization device is a device that consists of a container and reagents intended to stabilize microbial nucleic acids in human specimens for subsequent isolation and purification of nucleic acids for further molecular testing. The device is not intended for preserving morphology or viability of microorganisms.

Special Controls

The special controls for this device are: (1) The intended use for the 21 CFR 809.10 labeling must include a detailed description of microorganisms and types of human specimens intended to be preserved. (2) The 21 CFR 809.10(b) labeling must include: (i) A detailed device description, including all device components. (ii) Performance characteristics from applicable analytical studies, including but not limited to, nucleic acid stability and microorganism inactivation. (iii) A limiting statement that erroneous results may occur when the transport device is not compatible with molecular testing. (iv) A limiting statement that the device has only been validated to preserve the representative microorganisms used in the analytical studies. (3) Design verification and validation must include the following: (i) Overall device design including all device components and all control elements incorporated into the analytical validation procedures. (ii) Thorough description of the microorganisms and methodology used in the validation of the device including, but not limited to, extraction platforms and assays used for the detection of preserved nucleic acids. (iii) The limit of detection (LoD) of the molecular test used to establish microorganism nucleic acid stability.

*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of microorganisms and types of human specimens intended to be preserved. (2) The labeling required under § 809.10(b) of this chapter must include the following: (i) A detailed device description, including all device components; (ii) Performance characteristics from applicable analytical studies, including nucleic acid stability and microorganism inactivation; (iii) A limiting statement that erroneous results may occur when the transport device is not compatible with molecular testing; and (iv) A limiting statement that the device has only been validated to preserve the representative microorganisms used in the analytical studies. (3) Design verification and validation must include the following: (i) Overall device design, including all device components and all control elements incorporated into the analytical validation procedures; (ii) Thorough description of the microorganisms and methodology used in the validation of the device including, extraction platforms and assays used for the detection of preserved nucleic acids; and (iii) The limit of detection (LoD) of the molecular test used to establish microorganism nucleic acid stability.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K242820 B Applicant Puritan Medical Products LLC C Proprietary and Established Names Puritan PurSafe Plus Collection and Transport System D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QBD | Class II | 21 CFR 866.2950 - Microbial Nucleic Acid Storage And Stabilization Device | MI - Microbiology | ## II Submission/Device Overview: A Purpose for Submission: To obtain a substantial equivalence determination for the Puritan PurSafe Plus Collection and Transport System for use in collection, inactivation, and preservation of upper respiratory specimens suspected of containing SARS-CoV-2. B Measurand: Storage and stability of nucleic acids from SARS-CoV-2. C Type of Test: Microbial nucleic acid storage and stabilization device ## III Intended Use/Indications for Use: Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} K242820 - Page 2 of 9 A Intended Use(s): See Indications for Use below. B Indication(s) for Use: Puritan PurSafe Plus Collection and Transport System is an enclosed system intended for the collection, inactivation, and preservation of human upper respiratory specimens suspected of containing SARS-CoV-2. Puritan PurSafe Plus Collection and Transport System can be used for collection, transport, and storage of specimens at 2-4°C and 25-30°C. Specimens collected and stored in the PurSafe Plus Collection and Transport System are suitable for use with legally marketed molecular diagnostic devices. C Special Conditions for Use Statement(s): Rx - For Prescription Use Only D Special Instrument Requirements: None IV Device/System Characteristics: A Device Description: The Puritan PurSafe Plus Collection and Transport System is comprised of a sterile polyester flock swab applicator with a standard tip for collecting human upper respiratory specimens and a polypropylene vial containing 1 mL PurSafe Plus MK buffer (media) intended for inactivation and preservation of viral nucleic acid present in specimens suspected of containing SARS-CoV-2. Each Puritan PurSafe Plus Collection and Transport System peel pouch is provided with the following materials: - One pre-labeled screw-cap polypropylene vial containing 1 mL of MK buffer solution preservative (non-sterile; for single use only). - One scored sterile HydraFlock flocked swab with a standard tip. The Puritan PurSafe Plus Collection and Transport System maintains the integrity of SARS-CoV-2 viral nucleic acid when properly stored. Prior to use, the kit reagents should be stored at 2-4°C or 30-35°C for up to 24 months. After collection, the PurSafe Plus MK buffer tube containing the specimen can be stored for up to 28 days at 2-4°C or 25-30°C, for specimen storage and transportation to the laboratory, without a significant loss of RNA integrity. B Principle of Operation: The product is intended for use in collection, inactivation, transport, and preservation of SARS-CoV-2 viral nucleic acid from human upper respiratory specimens using the sterile HydraFlock flocked swab applicator with a standard tip included with the kit. After sample collection, the swab is transferred into the pre-labeled screw-cap polypropylene vial containing 1 mL of MK buffer solution preservative. Users then break the swab shaft at the score point leaving the swab tip in the transport tube and screw the lid of the tube tightly. {2} Puritan MK buffer solution contains the following components: Guanidine thiocyanate, HEPES, Tris Base, Triton X-100, EDTA, and molecular grade water. Puritan MK buffer solution preserves viral nucleic acid present in human upper respiratory specimen for up to 28 days when stored at 2-4°C and 25-30°C. The preserved and stabilized SARS-CoV-2 RNA maintains its integrity for downstream molecular based detection. V Substantial Equivalence Information: A Predicate Device Name(s): DNA/RNA Shield Collection Tube B Predicate 510(k) Number(s): K202641 C Comparison with Predicate(s): | Device & Predicate Device(s): | Subject Device: K242820 | Predicate: K202641 | | --- | --- | --- | | Device Trade Name | Puritan PurSafe Plus Collection and Transport System | DNA/RNA Shield | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | Puritan PurSafe Plus Collection and Transport System is an enclosed system intended for the collection, inactivation, and preservation of human upper respiratory specimens suspected of containing SARS-CoV-2. Puritan PurSafe Plus Collection and Transport System can be used for collection, transport, and storage of specimens at 2-4°C and 25-30°C. Specimens collected and stored in the PurSafe Plus Collection and Transport System are suitable for use with legally marketed molecular diagnostic devices. | The DNA/RNA Shield collection tube is intended for the stabilization and inactivation of upper and lower respiratory human specimens suspected of containing SARS-CoV-2. These devices can be used for collection transport and storage of specimens at ambient temperatures (20-25°C). Specimens collected and stored in a DNA/RNA Shield collection tube are suitable for use with legally marketed molecular diagnostic devices. | K242820 - Page 3 of 9 {3} | Single-use Device | Yes | Same | | --- | --- | --- | | Product Configuration | Medium in vial and cap System including swab. | Same | | Swab Tip | Flocked Swab | Same | | Analyte | RNA | Same | | Inactivation | SARS-CoV-2 virus | Same | | General Device Characteristic Differences | | | | pH | 8.3 ± 0.2 | 5.0-7.0 | | Specimen type | Upper respiratory specimens | Lower, upper respiratory and saliva specimens | | Specimen stability | Up to 28 days at 2-4°C and 25-30°C. | Up to 28 days at 20-25°C. | | Shelf-life stability | 24 months when refrigerated (2-4°C) or stored at room temperature (30-35°C) | 24 months at 20-25°C (ambient temperature) | VI Standards/Guidance Documents Referenced: Special controls that are applicable to regulation 21 CFR 866.2950. VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Not applicable (N/A) 2. Linearity: N/A 3. Analytical Specificity/Interference: N/A 4. Assay Reportable Range: N/A K242820 - Page 4 of 9 {4} 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): # Shelf Life A shelf-life stability study was conducted using three lots of Puritan PurSafe Plus Collection and Transport System. Device lots were stored at $2 - 4^{\circ}\mathrm{C}$ (refrigerated) and $30 - 35^{\circ}\mathrm{C}$ (RT). The shelf-life study evaluated multiple time points at each storage temperature condition. To evaluate device shelf-life stability during the study, various parameters like pH measurements, conductivity, and physical characteristics (namely, media appearance, media level, box/case appearance, package seal, print legibility and fading of ink/color) were evaluated. Acceptance criteria included the following parameters used for evaluation, pH range of $8.3 \pm 0.2$ , conductivity range $50 - 100 \mathrm{~mS/cm}$ , media level within $5\%$ tolerance of fill volume target, no discoloration or turbidity in media appearance, legible printing with no ink fading, and a sealed package. For each parameter evaluated in this study, the acceptance criteria were met indicating that Puritan PurSafe Plus Collection and Transport System remains stable when stored at $2 - 4^{\circ}\mathrm{C}$ or $30-35^{\circ}\mathrm{C}$ for up to 24 months from the date of manufacturing. # 6. Detection Limit: Limit of detection (LoD) studies were carried out to assess any impact or interference of the Puritan PurSafe Plus Collection and Transport System on the established performance of an FDA-cleared downstream in vitro diagnostic assay. For this study, the Cepheid Xpert Xpress Cov-2/Flu/RSV plus assay with an established limit of detection of 64 copies/mL for nasal swab specimens was used on the Cepheid GeneXpert IV system. A preliminary LoD study was performed with clinically negative nasal matrix spiked with heat-inactivated SARS-CoV-2 (BEI Resources; ATCC # VR-1986HK, Lot # 70037781; initial concentration of $4.2 \times 10^{5}$ genome copies/ $\mu$ L) at different viral RNA concentrations of 256, 128, 64, 32, 16 and 0 genome copies/ $\mu$ L, covering the pre-established LoD value of the Cepheid Xpert Xpress Cov-2/Flu/RSV plus assay (see Table 1 below). Per the assay protocol, $300 \mu$ L of the specimen+buffer mix was loaded onto the Cepheid Xpert Xpress CoV-2/Flu/RSV plus cartridge. All replicates with SARS-CoV-2 RNA concentrations of 64 genome copies/ $\mu$ L and higher resulted in positive detection in the preliminary LoD study. These results are consistent with the established limit of detection for the Cepheid Xpert Xpress Cov-2/Flu/RSV plus assay. Table 1. Preliminary LoD study results | ID | Lot | Genome copies/mL | SARS-CoV-2 test result | Cepheid Probe Check | | --- | --- | --- | --- | --- | | P1 | Puritan Plus | 256 | Positive | Pass | | P2 | Puritan Plus | 256 | Positive | Pass | | P3 | Puritan Plus | 128 | Positive | Pass | | P4 | Puritan Plus | 128 | Positive | Pass | | P5 | Puritan Plus | 64 | Positive | Pass | | P6 | Puritan Plus | 64 | Positive | Pass | | P7 | Puritan Plus | 32 | Negative | Pass | | P8 | Puritan Plus | 32 | Positive | Pass | | P9 | Puritan Plus | 16 | Negative | Pass | K242820 - Page 5 of 9 {5} To confirm the LoD, 20 additional sample replicates were evaluated at the lowest concentration at which all replicates were positive in the preliminary LoD (i.e., 64 genome copies/mL). Similar to the preliminary LoD study, samples for the confirmatory LoD study were prepared by spiking clinically negative nasal matrix with heat-inactivated SARS-CoV-2 and using three Puritan PurSafe Plus Collection and Transport System lots. The study was conducted over six days (2-8 samples/day). All samples indicated positive detection of SARS-CoV-2 viral RNA. All samples differed by less than two Ct values. The Ct values and results of the confirmatory LoD study for samples using nasal matrix in Puritan PurSafe Plus Collection and Transport System can be seen in Table 2 below. These results demonstrate that the Puritan PurSafe Plus Collection and Transport System does not introduce bias into the downstream assay results and are acceptable. Table 2. Confirmatory LoD study results | Rep | Assay Day | SARS-CoV-2 test results | Cepheid Probe Check | Ct Values | | --- | --- | --- | --- | --- | | 1 | 1 | Positive | Pass | 36.6 | | 2 | 1 | Positive | Pass | 36.7 | | 3 | 1 | Positive | Pass | 36.9 | | 4 | 1 | Positive | Pass | 37.6 | | 5 | 1 | Positive | Pass | 37.6 | | 6 | 1 | Positive | Pass | 37.1 | | 7 | 2 | Positive | Pass | 36.4 | | 8 | 2 | Positive | Pass | 37.0 | | 9 | 2 | Positive | Pass | 36.5 | | 10 | 2 | Positive | Pass | 36.0 | | 11 | 3 | Positive | Pass | 37.1 | | 12 | 3 | Positive | Pass | 36.9 | | 13 | 3 | Positive | Pass | 36.6 | | 14 | 3 | Positive | Pass | 36.6 | | 15 | 3 | Positive | Pass | 36.6 | | 16 | 3 | Positive | Pass | 37.2 | | 17 | 3 | Positive | Pass | 36.3 | | 18 | 3 | Positive | Pass | 36.3 | | 19 | 6 | Positive | Pass | 36.1 | | 20 | 6 | Positive | Pass | 36.9 | | | Min | 36.0 | | | | | | | Max | 37.6 | # 7. Specimen Stability: A specimen stability study was conducted to evaluate the preservation of SARS-CoV-2 viral RNA in clinically negative nasal matrix stored in multiple lots of PurSafe Plus Collection and Transport System. Samples prepared with SARS-CoV-2 concentrations corresponding to 2X LoD (i.e., 100 genome copies/mL) were used for the stability studies. Storage conditions (2- $4^{\circ}\mathrm{C}$ and $30 - 35^{\circ}\mathrm{C}$ ) were evaluated across multiple timepoints including at 0-, 1-, 7-, and 28-days. K242820 - Page 6 of 9 {6} Following the assay protocol, $300~\mu \mathrm{L}$ of the specimen+buffer mix was evaluated using the Cepheid Xpert Xpress CoV-2/Flu/RSV plus assay. Two replicates were assayed at each time point and storage temperature $(N = 72)$ . Results are summarized in Table 3 and 4 below. Table 3. Specimen stability results: | | | Lot 1 (240312) New | | Lot 2 (230207) Mid | | Lot 3 (221109) Exp | | | --- | --- | --- | --- | --- | --- | --- | --- | | Temperature | Time (days) | Assay results | Ct Values | Assay results | Ct Values | Assay results | Ct Values | | | 0 | Positive | 36.3 | Positive | 37.1 | Positive | 35.8 | | | 0 | Positive | 36.1 | Positive | 36.2 | Positive | 35.4 | | 2-4°C | 1 | Positive | 35.6 | Positive | 36.6 | Positive | 37.1 | | | 1 | Positive | 35.3 | Positive | 35.3 | Positive | 37.5 | | | 7 | Positive | 36.4 | Positive | 36.6 | Positive | 36.7 | | | 7 | Positive | 35.9 | Positive | 36.3 | Positive | 36.4 | | | 28 | Positive | 36.2 | Positive | 36.2 | Positive | 36.9 | | | 28 | Positive | 36.1 | Positive | 36.5 | Positive | 36.6 | | 25-30°C | 1 | Positive | 33.9 | Positive | 36.6 | Positive | 37.4 | | | 1 | Positive | 37 | Positive | 36 | Positive | 36.8 | | | 7 | Positive | 35.9 | Positive | 38.3 | Positive | 37.1 | | | 7 | Positive | 38.4 | Positive | 39 | Positive | 36.2 | | | 28 | Positive | 37.9 | Positive | 39.9 | Positive | 38.3 | | | 28 | Positive | 37.5 | Positive | 37.7 | Positive | 36.8 | Table 4. Specimen stability- difference from baseline: | | | Lot 1 (240312) New | | Lot 2 (230207) Mid | | Lot 3 (221109) Exp | | | --- | --- | --- | --- | --- | --- | --- | --- | | Temperature | Time (days) | Mean Ct Values | Difference from baseline (Time 0) | Mean Ct Values | Difference from baseline (Time 0) | Mean Ct Values | Difference from baseline (Time 0) | | | 0 | 36.2 | -- | 36.6 | -- | 35.6 | -- | | 2-4°C | 1 | 35.5 | 0.75 | 36.0 | 0.65 | 37.3 | 1.7 | | | 7 | 36.2 | 0.05 | 36.5 | 0.15 | 36.6 | 0.95 | | | 28 | 36.2 | 0.05 | 36.4 | 0.25 | 36.8 | 1.15 | | 25-30°C | 1 | 35.5 | 0.75 | 36.3 | 0.3 | 37.1 | 1.5 | | | 7 | 37.2 | 0.95 | 38.7 | 2.05 | 36.7 | 1.05 | | | 28 | 37.7 | 1.5 | 38.8 | 2.2 | 37.6 | 1.95 | All samples yielded positive detection at 100 genome copies/mL at storage temperatures of $2 - 4^{\circ}\mathrm{C}$ and $25 - 30^{\circ}\mathrm{C}$ and each time point evaluated in the study. All internal Cepheid probe checks passed. Mean Ct values for all samples across all time points and temperatures had less than 3 Ct difference from the baseline values (Time = 0). There were no significant differences among lots and with time and temperature. These results support the sample handling claims in the device labeling. # 8. Inactivation: Uninfected Vero E6 cells were exposed to the PurSafe Plus Buffer to observe possible cytotoxicity. Serial dilutions of each buffer (3 PurSafe Plus Lots) were carried out across a 96-well plate ranging from 1:10 dilution to 1:2160 dilution. Some cytotoxicity was observed with the PurSafe Plus buffer at a dilution of 1:10, but the 1:60 dilution had healthy cells with K242820 - Page 7 of 9 {7} no cytotoxic effects. Therefore, for the subsequent inactivation studies, a 1:60 dilution of the Puritan MK buffer solution was used. Subsequently, inactivation of SARS CoV-2 was evaluated in a cell culture-based assay using multiple lots of Puritan PurSafe Plus. The culture media from cells infected with SARS-CoV-2 virus (Isolate: USA-WA1/2020; Titer = 1.02 x 10⁸ TCID₅₀/mL) was mixed at a ratio of 1:10 with the Puritan MK buffer and incubated for different time periods. Samples were collected at 0-, 1-, 5- and 30-minutes time points and mixed with 2% media to prepare serial dilutions starting at 1:60 for infectivity testing in 96-well plates. Four replicates were evaluated for each condition. Vero E6 cells were added to the wells to evaluate viral infectivity. Cytopathic effects (CPE) were visually observed over seven days to determine the outcome of the study. For the cell-only controls no cytopathic effect was observed. The positive control SARS-CoV-2 culture fluid and virus treated with PBS resulted in positive CPE over the course of the assay. No visible CPE was observed over the course of the assay in wells with virus mixed with Puritan MK buffer for any exposure time evaluated in the study. SARS-CoV-2 was inactivated after exposure to each lot of PurSafe Plus MK Buffer at all timepoints evaluated in the study. Therefore, the PurSafe Plus MK Buffer was effective at inactivating the SARS-CoV-2 virus. 9. Assay Cut-Off: N/A B Comparison Studies: 1. Method Comparison with Predicate Device: N/A 2. Matrix Comparison: N/A C Clinical Studies: 1. Clinical Sensitivity: N/A 2. Clinical Specificity: N/A 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): K242820 - Page 8 of 9 {8} N/A D Clinical Cut-Off: N/A E Expected Values/Reference Range: N/A VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K242820 - Page 9 of 9
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...