K Number
K202641

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-12-14

(94 days)

Product Code
Regulation Number
866.2950
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DNA/RNA Shield™ collection tube is intended for the stabilization and inactivation of upper and lower respiratory human specimens suspected of containing SARS-CoV-2. These devices can be used for collection transport and storage of specimens at ambient temperatures (20-25°C). Specimens collected and stored in a DNA/RNA Shield™ collection tube are suitable for use with legally marketed molecular diagnostic devices.

Device Description

The DNA/RNA Shield™ collection tube and reagent consist of a tube pre-filled with DNA/RNA Shield™ transport media. DNA/RNA Shield™ is a transport media that ensures stability of SARS-CoV-2 RNA during sample transport/storage at ambient temperatures and is intended to inactivate SARS-CoV-2, effectively lyses cells from collected upper and lower respiratory biological specimens. The DNA/RNA Shield™ transport media may be kitted with a swab, sputum collection kit or as a tube alone.

AI/ML Overview

The provided text describes the DNA/RNA Shield Collection Tube's performance data, which supports its substantial equivalence to a predicate device. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance Criteria (from text)Reported Device Performance (from text)
Inactivation- Inactivation of SARS-CoV-2 virus efficiently as other legally marketed devices. - Equivalent performance to PrimeStore MTM for inactivation.> 2.0 log reduction in viral titer after 30 min exposure.
Limit of Detection (LoD)- Determine SARS-CoV-2 LoD in combination with a legally marketed rRT-PCR Kit.250 GEC/mL (15 GEC/reaction) for sputum and oral swab.
Specimen Stability- SARS-CoV-2 RNA from sputum and oral swab preserved and stabilized. - Acceptance criteria of +/- 3.0 Ct after 4 weeks storage at 20-25°C. - Equivalent performance to the legally marketed predicate device.RNA stabilized in DNA/RNA Shield met the acceptance criteria of +/- 3.0 Ct after 4 weeks storage at 20-25°C.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size for the test set in each study. However, it indicates:

  • Inactivation Study: High concentrations of SARS-CoV-2 virus (9x10^6 PFU) were inoculated.
  • Limit of Detection Study: "Samples" were extracted and amplified.
  • Specimen Stability Study: "Contrived specimens with sample matrices and SARS-CoV-2 RNA spiked into DNA/RNA Shield" were used.

The data provenance is retrospective in the sense that the studies were conducted to gather performance data on the device. The specific country of origin of the data is not mentioned, but the submitting company is Zymo Research in Irvine, California, USA, suggesting the studies were likely conducted in the US.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The studies described are laboratory-based performance evaluations (inactivation, LoD, stability) and do not involve human interpretation or expert consensus for establishing ground truth in the typical clinical sense.

4. Adjudication Method for the Test Set

This information is not applicable as the studies are laboratory-based performance evaluations, not involving human interpretation or adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

This information is not applicable. The device is a "DNA/RNA Shield Collection Tube," a sample collection and stabilization device, not an AI-powered diagnostic or imaging tool that would typically involve human readers or AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This information is not applicable. The device is a sample collection and stabilization tube, not an algorithm, so the concept of "standalone performance" in this context is irrelevant. The performance evaluated is the physical and chemical properties of the shield media itself.

7. The Type of Ground Truth Used

  • Inactivation Study: The ground truth for inactivation was established by measuring the viability of the virus (cytopathic effect, CPE) after exposure to the DNA/RNA Shield medium. This is a direct biological measure.
  • Limit of Detection Study: The ground truth for LoD was established by spiking known concentrations of SARS-CoV-2 RNA (measured in GEC/mL) into the medium and performing rRT-PCR. This is an analytical measurement based on a known standard.
  • Specimen Stability Study: The ground truth for stability was established by spiking SARS-CoV-2 RNA into contrived specimens and then measuring the Ct values from rRT-PCR over time, comparing them to a baseline. This also relies on analytical measurements against a known standard.

In all cases, the ground truth is based on pre-defined analytical standards and direct biological or molecular measurements, not expert consensus, pathology, or outcomes data in a clinical setting.

8. The Sample Size for the Training Set

This information is not applicable. The device is a physical product (collection tube and media), not a machine learning model, so there is no concept of a "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 14, 2020

Zymo Research Julie Ogi Regulatory Officer 17062 Murphy Ave. Irvine, California 92614

Re: K202641

Trade/Device Name: DNA/RNA Shield Collection Tube Regulation Number: 21 CFR 866.2950 Regulation Name: Microbial Nucleic Acid Storage and Stabilization Device Regulatory Class: Class II Product Code: QBD Dated: September 8, 2020 Received: September 11, 2020

Dear Julie Ogi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kristian Roth, Ph.D. Branch Chief Bacterial Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K202641

Device Name

DNA/RNA Shield Collection Tube

Indications for Use (Describe)

The DNA/RNA Shield™ collection tube is intended for the stabilization and inactivation of upper and lower respiratory human specimens suspected of containing SARS-CoV-2. These devices can be used for collection transport and storage of specimens at ambient temperatures (20-25°C). Specimens collected in a DNA/RNA Shield™ collection tube are suitable for use with legally marketed molecular diagnostic devices.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Zymo Research. The logo consists of three green oval shapes arranged in a triangular pattern above the company name. Below the company name is the tagline, "The Beauty of Science is to Make Things Simple."

ApplicantZymo Research17062 Murphy Ave.Irvine, CA 92614USA
Establishment Registration No.Pending / User Fees Paid
Contact PersonJulie OgiRegulatory OfficerPhone (949) 679-1190Fax (949) 266-9452e-mail jogi@zymoresearch.com
Summary DateSeptember 8, 2020
Proprietary NameDNA/RNA Shield™ collection tube
Common nameDNA/RNA Shield
DeviceTransport device for the stabilization of microbialnucleic acids
Product Codes/Regulation NumbersQBD - Transport device for the stabilization ofmicrobial nucleic acids / 21 CRF 866.2950
ClassificationClass II
Review PanelMicrobiology 83
Predicate DevicesPrimeStore MTM (DEN170029)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Zymo Research. The logo consists of three green oval shapes arranged in a triangular pattern. Each oval shape contains two smaller circles inside. Below the shapes, the words "ZYMO RESEARCH" are written in a bold, sans-serif font. Underneath the company name is the text "The Beauty of Science is to Make Things Simple".

Intended Use

The DNA/RNA Shield™ collection tube is intended for the stabilization and inactivation of upper and lower respiratory human specimens suspected of containing SARS-CoV-2. These devices can be used for collection transport and storage of specimens at ambient temperatures (20-25°C). Specimens collected and stored in a DNA/RNA Shield™ collection tube are suitable for use with legally marketed molecular diagnostic devices.

Device Description

The DNA/RNA Shield™ collection tube and reagent consist of a tube pre-filled with DNA/RNA Shield™ transport media. DNA/RNA Shield™ is a transport media that ensures stability of SARS-CoV-2 RNA during sample transport/storage at ambient temperatures and is intended to inactivate SARS-CoV-2, effectively lyses cells from collected upper and lower respiratory biological specimens. The DNA/RNA Shield™ transport media may be kitted with a swab, sputum collection kit or as a tube alone.

Comparison to Predicate Device

Similarities
ItemPredicate Device:PrimeStore MTM (DEN170029)Proposed Device:DNA/RNA Shield™ collectiontube and reagent
Intended UsePrimeStore MTM is a deviceintended for the stabilization,transportation and inactivation ofinfectious unprocessed nasalwashes suspected of containingInfluenza A (Flu A) virus RNA orunprocessed sputum samplessuspected of containingMycobacteriumtuberculosis (MTB)DNA from human samples.Transport device can transport andstore specimens between 36-77 °F(2-25 °C).The DNA/RNA Shield™ collectiontube is intended for the stabilizationand inactivation of upper and lowerrespiratory human specimenssuspected of containing SARS-CoV-2. These devices can be usedfor collection transport and storageof specimens at ambienttemperatures (20-25°C). Specimenscollected and stored in aDNA/RNA Shield™ collection tubeare suitable for use with legallymarketed molecular diagnosticdevices.
Special conditions foruse1. For in vitro diagnostic use only2. For prescription use only1. For in vitro diagnostic use only2. For prescription use only
DNA/RNA stabilizationat ambienttemperature28 days: MTB8 days: Flu A≤28 days sputum and OP
Sample sourceHuman respiratoryHuman respiratory

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Zymo Research. The logo consists of three green shapes resembling cells arranged in a triangular pattern above the company name. Below the company name, the tagline "The Beauty of Science is to Make Things Simple" is displayed.

Differences
ItemPredicate Device:PrimeStore MTM(DEN170029)Proposed Device:DNA/RNA Shield™ collection tubeand reagent
RNA stabilizationatrefrigeration≤8 days flu≤36 days MTBno refrigeration claims
AnalyteDNA, RNARNA
Sample collectionNasal wash suspected ofcontaining Influenza A virus.Sputum samples suspected ofcontaining MTB.Lower and Upper RespiratorySpecimens for SARS-CoV-2.

Performance data:

Inactivation

An inactivation study was conducted to verify that DNA/RNA Shield inactivates SARS-CoV-2 virus as efficiently as other legally marketed devices.

High concentrations of SARS-CoV-2 virus (9x106 PFU) were inoculated into DNA/RNA Shield. The viability of the virus was measured after 30 min in DNA/RNA Shield medium by inoculating aliquots onto veroE6 cells, incubating for two days and measuring the cytopathic effect (CPE).

The results for the inactivation study indicated a CPE of < 3.0 log reduction because of the DNA/RNA Shield media alone cause cell toxicity. The media and virus confirmed > 2.0 log reduction in viral titer after a 30 min exposure and demonstrates equivalent performance of DNA/RNA Shield with PrimeStore MTM™ for inactivation.

Limit of detection

An analytical sensitivity study was conducted to determine the SARS-CoV-2 Limit of Detection (LoD) obtained by DNA/RNA Shield in combination with the Quick SARS-CoV-2 rRT-PCR Kit. The RNA was spiked into DNA/RNA Shield medium then samples were extracted using the Quick-DNA/RNA Viral MagBead kit performed on the KingFisher™ Flex Purification System and then amplified using the Authorized Quick SARS-CoV-2 rRT-PCR Kit. The DNA/RNA Shield reached an LoD of 250 GEC/mL (15 GEC/reaction) for sputum and oral swab.

Specimen Stability

A stability study was designed to demonstrate that SARS-CoV-2 RNA from sputum and oral swab was preserved and stabilized in DNA/RNA Shield media. The stability studies used contrived specimens with sample matrices and SARS-CoV-2 RNA spiked into DNA/RNA Shield. The specimens were then amplified using the Quick SARS-CoV-2 rRT-PCR Kit. The results of SARS-CoV-2 RNA stability study confirmed that RNA stabilized in DNA/RNA Shield met the acceptance criteria of +/- 3.0 Ct after 4 weeks storage at 20 - 25 °C. This study demonstrates equivalent performance of DNA/RNA Shield with the legally marketed predicate device.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Zymo Research. The logo consists of three green oval shapes arranged in a triangular pattern. Below the shapes, the words "ZYMO RESEARCH" are written in bold, black letters. Underneath the company name is the tagline "The Beauty of Science is to Make Things Simple."

Based on the above, Zymo Research believes that the DNA/RNA Shield collection tube is substantially equivalent to the legally marketed predicate device for the inactivation, LoD and stabilization of specimens containing SARS-CoV-2 for collection, stabilization and recovery. No new issues of safety or effectiveness were found for the DNA/RNA Shield collection tube.

§ 866.2950 Microbial nucleic acid storage and stabilization device.

(a)
Identification. A microbial nucleic acid storage and stabilization device is a device that consists of a container and reagents intended to stabilize microbial nucleic acids in human specimens for subsequent isolation and purification of nucleic acids for further molecular testing. The device is not intended for preserving morphology or viability of microorganisms.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of microorganisms and types of human specimens intended to be preserved.
(2) The labeling required under § 809.10(b) of this chapter must include the following:
(i) A detailed device description, including all device components;
(ii) Performance characteristics from applicable analytical studies, including nucleic acid stability and microorganism inactivation;
(iii) A limiting statement that erroneous results may occur when the transport device is not compatible with molecular testing; and
(iv) A limiting statement that the device has only been validated to preserve the representative microorganisms used in the analytical studies.
(3) Design verification and validation must include the following:
(i) Overall device design, including all device components and all control elements incorporated into the analytical validation procedures;
(ii) Thorough description of the microorganisms and methodology used in the validation of the device including, extraction platforms and assays used for the detection of preserved nucleic acids; and
(iii) The limit of detection (LoD) of the molecular test used to establish microorganism nucleic acid stability.