iSWAB-Respiratory Tract Sample Collection Media-Extraction Less (iSWAB-RC-EL)

K221802 · Mawi Dna Technologies · QBD · Apr 3, 2023 · Microbiology

Device Facts

Record IDK221802
Device NameiSWAB-Respiratory Tract Sample Collection Media-Extraction Less (iSWAB-RC-EL)
ApplicantMawi Dna Technologies
Product CodeQBD · Microbiology
Decision DateApr 3, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.2950
Device ClassClass 2

Indications for Use

The iSWAB-Respiratory Tract Sample Collection Media-Extraction Less collection device is intended for the stabilization and inactivation of upper respiratory and saliva human specimens suspected of containing SARS-CoV-2. This device can be used for the collection, transport, and storage of specimens at ambient temperature. Specimens collected in the iSWAB-Respiratory Tract Sample Collection Media-Extraction Less collection device are suitable for use with legally marketed molecular diagnostic tests.

Device Story

Collection tube pre-filled with 800 µL stabilizing buffer; includes proprietary insert to optimize specimen release from swabs; allows swab-free transport. Used for upper respiratory and saliva specimens suspected of SARS-CoV-2. Eliminates RNA isolation step in diagnostic workflows. Used in clinical settings; operated by healthcare professionals. Stabilizes specimens at ambient temperature (15-30°C) for up to 33 days. Output is a stabilized specimen suitable for direct use with molecular diagnostic tests; benefits patient by simplifying workflow and reducing processing time.

Clinical Evidence

No clinical data. Bench testing only. Performance supported by inactivation studies, analytical sensitivity (LoD) studies using BGI Real-Time Fluorescent RT-PCR Kit, and stability studies demonstrating SARS-CoV-2 preservation at ambient temperature for 28-33 days. Compatibility verified with multiple molecular assays including CDC 2019-nCoV RT-PCR Panel, BGI, Bio-Rad, and RT-LAMP assays.

Technological Characteristics

Non-sterile 12x50 mm screwcap vial with proprietary insert. Contains 800 µL stabilizing buffer. Passive device; no electrical, software, or patient-contact components. Shelf-life 15 months.

Indications for Use

Indicated for stabilization and inactivation of upper respiratory and saliva human specimens suspected of containing SARS-CoV-2 for use with molecular diagnostic tests.

Regulatory Classification

Identification

A microbial nucleic acid storage and stabilization device is a device that consists of a container and reagents intended to stabilize microbial nucleic acids in human specimens for subsequent isolation and purification of nucleic acids for further molecular testing. The device is not intended for preserving morphology or viability of microorganisms.

Special Controls

The special controls for this device are: (1) The intended use for the 21 CFR 809.10 labeling must include a detailed description of microorganisms and types of human specimens intended to be preserved. (2) The 21 CFR 809.10(b) labeling must include: (i) A detailed device description, including all device components. (ii) Performance characteristics from applicable analytical studies, including but not limited to, nucleic acid stability and microorganism inactivation. (iii) A limiting statement that erroneous results may occur when the transport device is not compatible with molecular testing. (iv) A limiting statement that the device has only been validated to preserve the representative microorganisms used in the analytical studies. (3) Design verification and validation must include the following: (i) Overall device design including all device components and all control elements incorporated into the analytical validation procedures. (ii) Thorough description of the microorganisms and methodology used in the validation of the device including, but not limited to, extraction platforms and assays used for the detection of preserved nucleic acids. (iii) The limit of detection (LoD) of the molecular test used to establish microorganism nucleic acid stability.

*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of microorganisms and types of human specimens intended to be preserved. (2) The labeling required under § 809.10(b) of this chapter must include the following: (i) A detailed device description, including all device components; (ii) Performance characteristics from applicable analytical studies, including nucleic acid stability and microorganism inactivation; (iii) A limiting statement that erroneous results may occur when the transport device is not compatible with molecular testing; and (iv) A limiting statement that the device has only been validated to preserve the representative microorganisms used in the analytical studies. (3) Design verification and validation must include the following: (i) Overall device design, including all device components and all control elements incorporated into the analytical validation procedures; (ii) Thorough description of the microorganisms and methodology used in the validation of the device including, extraction platforms and assays used for the detection of preserved nucleic acids; and (iii) The limit of detection (LoD) of the molecular test used to establish microorganism nucleic acid stability.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K221802 - Page 1 of 10 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K221802 B Applicant Mawi DNA Technologies C Proprietary and Established Names iSWAB-Respiratory Tract Sample Collection Media-Extraction Less (iSWAB-RC-EL) D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QBD | Class II | 21 CFR 866.2950 - Microbial Nucleic Acid Storage And Stabilization Device | MI - Microbiology | ## II Submission/Device Overview: A Purpose for Submission: The purpose of this submission is to obtain 510(k) substantial equivalence for the iSWAB-Respiratory Tract Sample Collection Media-Extraction Less (iSWAB-RC-EL) for the collection, transport and storage of unprocessed samples containing SARS-CoV-2. B Measurand: Nucleic acids from SARS-CoV-2 C Type of Test: Microbial nucleic acid storage and stabilization device ## III Intended Use/Indications for Use: {1} K221802 - Page 2 of 10 A Intended Use(s): See Indications for Use below. B Indication(s) for Use: The iSWAB-Respiratory Tract Sample Collection Media-Extraction Less collection device is intended for the stabilization and inactivation of upper respiratory and saliva human specimens suspected of containing SARS-CoV-2. This device can be used for the collection, transport, and storage of specimens at ambient temperature. Specimens collected in the iSWAB-Respiratory Tract Sample Collection Media-Extraction Less collection device are suitable for use with legally marketed molecular diagnostic tests. C Special Conditions for Use Statement(s): Rx - For Prescription Use Only D Special Instrument Requirements: N/A IV Device/System Characteristics: A Device Description: The iSWAB-Respiratory Tract Sample Collection Media-Extraction Less (iSWAB-RC-EL) collection device consists of a collection tube that is pre-filled with 800 µL of the iSWAB-RC-EL non-toxic, stabilizing buffer and fitted with a proprietary insert. The insert is designed to optimize the release of specimens collected with swabs into the stabilizing buffer allowing for swab-free transport of specimens. The iSWAB-RC-EL collection device is designed to eliminate the RNA extraction step from diagnostic workflows. The media is designed to skip the solid phase extraction step of a traditional two step lysis and extraction of nucleic acids from specimens prior to amplification using traditional PCR techniques. B Principle of Operation: The iSWAB-RC-EL stabilizing buffer is intended to inactivate SARS CoV-2 by disrupting/lysing the capsid lipid membranes, denature proteins, inactivate enzymes, and stabilize SARS CoV-2 RNA. The iSWAB-RC-EL collection device is designed for storage of specimens between 15-30 °C for 28 days in nasal samples and 33 days in saliva samples. The stabilizing buffer contains the following components: - Nontoxic inactivation agents - Salts - pH buffer - Water {2} V Substantial Equivalence Information: A Predicate Device Name(s): DNA/RNA Shield Collection Tube B Predicate 510(k) Number(s): K202641 C Comparison with Predicate(s): | Device & Predicate Device(s): | Device: K221802 | Predicate: K202641 | | --- | --- | --- | | Device Trade Name | iSWAB-RC-EL | DNA/RNA Shield | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | The iSWAB-Respiratory Tract Sample Collection Media-Extraction Less (iSWAB-RC-EL) collection device is intended for the stabilization and inactivation of upper respiratory and saliva human specimens suspected of containing SARS-CoV-2. This device can be used for the collection, transport, and storage of specimens at ambient temperature. Specimens collected in the iSWAB-Respiratory Tract Sample Collection Media-Extraction Less collection device are suitable for use with legally marketed molecular diagnostic tests. | The DNA/RNA Shield collection tube is intended for the stabilization and inactivation of upper and lower respiratory human specimens suspected of containing SARS-CoV-2. These devices can be used for collection transport and storage of specimens at ambient temperatures (20-25°C). Specimens collected and stored in a DNA/RNA Shield collection tube are suitable for use with legally marketed molecular diagnostic devices. | | Viral Inactivation | SARS-CoV-2 inactivation | Same | | Analyte | RNA from SARS-CoV-2 | Same | | General Device Characteristic Differences | | | | Specimen Storage | SARS-CoV-2 RNA: 28 days with nasal specimens and 33 days with saliva specimens at | SARS-CoV-2 RNA: 28 days at 20-25°C | K221802 - Page 3 of 10 {3} | | 15-30°C | | | --- | --- | --- | | Specimen type | Nasal and Saliva Specimens for SARS-CoV-2 | Lower, Upper Respiratory and Saliva Specimens for SARS-CoV-2 | | Sample processing type | RNA extraction optional | Requires RNA extraction before use | | Shelf Life | 15 months | 24 months | VI Standards/Guidance Documents Referenced: Special controls that are applicable to regulation 21 CFR 866.2950. VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility N/A 2. Linearity: N/A 3. Analytical Specificity/Interference: N/A 4. Assay Reportable Range: N/A 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): Shelf Life The shelf life for the iSWAB-RC-EL is 15 months after the date of manufacture. The stability of the iSWAB-RC-EL was performed using real time stability on a total of three lots. Stability studies assessed media appearance, pH, density, conductivity, and bacterial and fungal growth. 6. Detection Limit: Nasal samples: The limit of detection (LoD) of SARS-CoV-2 for iSWAB-RC-EL was established using an EUA authorized assay to determine the lowest concentration of virus that can be repeatedly K221802 - Page 4 of 10 {4} recovered from the sample tube with an accuracy greater than $95\%$ . Negative clinical nasal matrix was collected and pooled. 20 individual replicates of matrix were spiked with $25~\mathrm{cp} / \mu \mathrm{L}$ of ATCC VR-1986 and then added to the media. Samples were assessed using the BGI's "Real-Time Fluorescent RT-PCR Kit for Detecting SARS-CoV-2" assay. $10~\mu \mathrm{L}$ of sample in media was directly mixed with $20~\mu \mathrm{L}$ of RT-PCR master mix targeting the ORF1ab region of SARS-CoV-2 genome (FAM channel) and the human housekeeping gene $\beta$ -Actin (ACTB, VIC/HEX channel) as a sample control. At $25~\mathrm{cp} / \mu \mathrm{L}$ , 19/20 samples were positive indicating that the LoD for nasal clinical matrix is $25~\mathrm{cp} / \mu \mathrm{L}$ , The Ct values for each replicate and target are listed Table 1 below. Table 1. LOD Studies in Nasal Matrix (25 cp/μL) | Replicate # | ORF1ab (average Ct) | SD | ACTB (average Ct) | SD | | --- | --- | --- | --- | --- | | 1 | 35.00 | 0.12 | 27.99 | 0.02 | | 2 | 34.27 | 0.26 | 27.93 | 0.06 | | 3 | 35.02 | 0.46 | 27.57 | 0.20 | | 4 | 35.07 | 0.45 | 27.54 | 0.01 | | 5 | 35.36 | 0.43 | 28.08 | 0.01 | | 6 | 34.96 | 0.37 | 27.91 | 0.17 | | 7 | 34.93 | 0.41 | 27.71 | 0.10 | | 8 | 35.16 | 0.62 | 27.88 | 0.12 | | 9 | 34.91 | 0.64 | 28.16 | 0.03 | | 10 | 34.91 | 0.13 | 28.20 | 0.02 | | 11 | 34.96 | 0.64 | 28.02 | 0.16 | | 12 | 35.31 | 0.41 | 28.12 | 0.11 | | 13 | 35.19 | 0.13 | 28.03 | 0.02 | | 14 | 35.20 | 0.14 | 26.75 | 0.91 | | 15 | 35.78 | 0.52 | 27.82 | 0.03 | | 16 | 35.23 | 0.17 | 27.98 | 0.24 | | 17 | 35.27 | 0.30 | 28.20 | 0.02 | | 18 | 36.65 | 1.06 | 29.21 | 0.19 | | 19 | 35.55 | 0.41 | 27.17 | 1.06 | | 20 | 34.99 | 0.60 | 28.16 | 0.09 | # Saliva samples: The limit of detection (LoD) of SARS-CoV-2 for iSWAB-RC-EL was established using an EUA authorized assay to determine the lowest concentration of virus that can be repeatedly recovered from the sample tube with an accuracy greater than $95\%$ . Negative clinical saliva matrix was collected and pooled. 20 individual replicates of matrix were spiked with $30~\mathrm{cp} / \mu \mathrm{L}$ of ATCC VR-1986. Samples were assessed using the BGI's "Real-Time Fluorescent RT-PCR Kit for Detecting SARS-CoV-2" assay. $10~\mu \mathrm{L}$ of sample in media was directly mixed with $20~\mu \mathrm{L}$ of RT-PCR master mix targeting the ORF1ab region of SARS-CoV-2 genome (FAM channel) and the human housekeeping gene $\beta$ -Actin (ACTB, VIC/HEX channel) as a sample control. At $30~\mathrm{cp} / \mu \mathrm{L}$ , 20/20 samples were positive, however at $25~\mathrm{cp} / \mu \mathrm{L}$ only 13/20 samples K221802 - Page 5 of 10 {5} were positive, indicating that the LoD for nasal clinical matrix is $30~\mathrm{cp} / \mu \mathrm{L}$ The Ct values for each replicate and target are listed in Table 2 below. Table 2. LOD studies in Saliva Matrix (30 cp/μL) | Replicate # | ORF1ab (average Ct) | SD | ACTB (average Ct) | SD | | --- | --- | --- | --- | --- | | 1 | 35.11 | 0.09 | 32.30 | 0.21 | | 2 | 35.55 | 0.46 | 31.24 | 0.72 | | 3 | 35.85 | 0.60 | 32.08 | 0.22 | | 4 | 36.36 | 0.29 | 31.05 | 0.40 | | 5 | 35.80 | 0.12 | 31.29 | 0.60 | | 6 | 35.65 | 0.83 | 32.32 | 0.19 | | 7 | 36.23 | 0.30 | 32.39 | 0.55 | | 8 | 35.51 | 0.13 | 32.13 | 0.91 | | 9 | 35.44 | 0.35 | 32.76 | 0.50 | | 10 | 35.26 | 0.64 | 31.21 | 0.55 | | 11 | 34.62 | 0.07 | 31.79 | 0.11 | | 12 | 35.96 | 0.65 | 32.14 | 0.50 | | 13 | 35.67 | 0.77 | 32.19 | 0.49 | | 14 | 35.99 | 1.11 | 32.57 | 0.82 | | 15 | 35.20 | 0.23 | 32.73 | 0.05 | | 16 | 36.05 | 0.40 | 32.76 | 0.14 | | 17 | 35.02 | 0.24 | 32.50 | 0.43 | | 18 | 34.89 | 0.71 | 31.72 | 0.58 | | 19 | 35.08 | 0.05 | 32.14 | 0.81 | | 20 | 34.59 | 0.69 | 32.51 | 0.54 | # 7. Specimen Stability # Nasal Matrix The specimen stability of SARS-CoV-2 virus was determined in iSWAB-RC-EL media by spiking $110~\mathrm{cp} / \mu \mathrm{L}$ of ATCC VR-1986 in iSWAB-RC-EL media with negative clinical nasal matrix. $10~\mu \mathrm{L}$ of sample was mixed directly without extraction with $20~\mu \mathrm{L}$ of master mix from BGI Real Time Fluorescent RT-PCR Kit for Detecting SARS-CoV-2. A pre-defined acceptance criteria of $\pm 3$ Ct from time zero was used to establish stability after storage at K221802 - Page 6 of 10 {6} room temperature, $20 - 25^{\circ}\mathrm{C}$ . The study results support that the transport media stabilizes SARS-CoV-2 RNA up to 28 days at $20 - 25^{\circ}\mathrm{C}$ , as summarized in Table 3 below. Table 3. Specimen Stability Results in Nasal Matrix | | Orf1ab average Ct | Delta Ct from baseline | ACTB average Ct | Delta Ct from baseline | | --- | --- | --- | --- | --- | | Day 0 | 33.74 | 0.00 | 30.79 | 0.00 | | Day1 | 33.22 | -0.52 | 31.84 | 1.05 | | Day 4 | 32.51 | -1.23 | 28.62 | -2.17 | | Day 7 | 32.77 | -0.97 | 29.64 | -1.15 | | Day 14 | 32.75 | -0.99 | 28.59 | -2.20 | | Day 21 | 33.20 | -0.54 | 29.42 | -1.37 | | Day 28 | 34.83 | 1.09 | 28.59 | -2.20 | # Saliva Matrix The specimen stability of SARS-CoV-2 virus was determined in iSWAB-RC-EL media by spiking $110~\mathrm{cp} / \mu \mathrm{L}$ of ATCC VR-1986 in iSWAB-RC-EL media with negative clinical saliva matrix. $10~\mu \mathrm{L}$ of sample was mixed directly without an extraction step with $20~\mu \mathrm{L}$ of master mix from BGI Real Time Fluorescent RT-PCR Kit for Detecting SARS-CoV-2. A predefined acceptance criteria of $\pm 3$ Ct from time zero was used to establish stability after storage at room temperature, $20 - 25^{\circ}\mathrm{C}$ . The study results support that the transport media stabilizes SARS-CoV-2 RNA up to 33 days at $20 - 25^{\circ}\mathrm{C}$ , as summarized in Table 4 below. Table 4. Specimen Stability Results in Saliva Matrix | | Orf1ab average Ct | Delta Ct from baseline | ACTB average Ct | Delta Ct from baseline | | --- | --- | --- | --- | --- | | Day 0 | 33.62 | 0.00 | 31.34 | 0.00 | | Day1 | 35.00 | 1.38 | 29.00 | -2.33 | | Day 8 | 35.56 | 1.94 | 29.02 | -2.31 | | Day 19 | 32.65 | -0.97 | 29.89 | -1.45 | | Day 26 | 34.25 | 0.63 | 30.80 | -0.54 | | Day 33 | 34.90 | 1.29 | 31.48 | 0.14 | # 8. Inactivation: # Media cytopathic effect Cell lines were spiked with iSWAB-RC-EL media without dilution and no cytopathic effect was observed over 2 hours. This supported the addition of media without dilution K221802 - Page 7 of 10 {7} to cell culture monolayers for less than 2 hours for the viral inactivation phase of the inactivation study. # Inactivation time The viral inactivation phase of the study was conducted to assess the time the virus needs to be exposed to the media to render the virus inactive. The iSWAB-RC-EL was spiked with known concentrations of SARS-CoV-2 virus and negative nasal matrix (Table 5) or saliva matrix (Table 6), and incubated at room temperature for 30 minutes, 2 hours and 6 hours. At each time point $100~\mu \mathrm{L}$ of the iSWAB-RC-EL spiked with respective matrix and SARS-CoV-2 virus was inoculated onto MDCK II cell culture. The cultures were incubated for up to 48 hours post infection and observed by inverted light microscopy for cytopathic effect (CPE). The Reed and Muench method was used to determine the $50\%$ tissue culture infectious dose per mL (TCID50/mL). In cases where sample showed no sign of detectable virus, statistical analysis based on the Poisson distribution was carried out to ascertain the theoretical maximum potential titer. Cells spiked with virus without incubation in iSWAB-RC-EL media showed CPE within 10 minutes. Table 5. Inactivation Study Results in Nasal Clinical Matrix | Volume (mL) | Replicate | Input Viral Load (Log10TCID50) | Output Viral Load (Log10TCID50) | Reduction (Log10TCID50) | | --- | --- | --- | --- | --- | | 30 minutes | 1 | 5.93 ± 0.28 | 1.48 ± 0.18 | 4.45 ± 0.34 | | | 2 | | 1.35 ± 0.16 | 4.58 ± 0.33 | | | 3 | | 1.35 ± 0.16 | 4.58 ± 0.33 | | | Average | | 1.39 ± 0.17 | 4.54 ± 0.33 | | 2 hours | 1 | 5.81 ± 0.27 | 0.60 ± 0.19 | 5.21 ± 0.33 | | | 2 | | 0.48 ± 0.18 | 5.33 ± 0.32 | | | 3 | | 0.23 ± 0.12 | 5.58 ± 0.29 | | | Average | | 0.44 ± 0.16 | 5.37 ± 0.31 | | 6 hours | 1 | 5.81 ± 0.30 | 0.06 ± 0.02 | 5.75 ± 0.35 | | | 2 | | 0.05 ± 0.03 | 5.76 ± 0.35 | | | 3 | | 0.05 ± 0.02 | 5.76 ± 0.32 | | | Average | | 0.18 ± 0.02 | 5.76 ± 0.34 | Table 6. Inactivation Study Results in Saliva Clinical Matrix | Volume (mL) | Replicate | Input Viral Load (Log10TCID50) | Output Viral Load (Log10TCID50) | Reduction (Log10TCID50) | | --- | --- | --- | --- | --- | | 30 minutes | 1 | 5.93 ± 0.28 | 1.35 ± 0.16 | 4.58 ± 0.33 | | | 2 | | 1.35 ± 0.16 | 4.58 ± 0.33 | | | 3 | | 1.60 ± 0.19 | 4.33 ± 0.34 | | | Average | | 1.43 ± 0.17 | 4.50 ± 0.33 | K221802 - Page 8 of 10 {8} | 2 hours | 1 | 5.81 ± 0.27 | 0.10 ± 0.17 | 5.71 ± 0.32 | | --- | --- | --- | --- | --- | | | 2 | | 0.23 ± 0.12 | 5.58 ± 0.29 | | | 3 | | 0.35 ± 0.16 | 5.46 ± 0.31 | | | Average | | 0.23 ± 0.15 | 5.58 ± 0.31 | | 6 hours | 1 | 5.81 ± 0.30 | 0.05 ± 0.01 | 5.80 ± 0.34 | | | 2 | | 0.07 ± 0.03 | 5.80 ± 0.35 | | | 3 | | 0.04 ± 0.03 | 5.81 ± 0.35 | | | Average | | 0.05 ± 0.02 | 5.80 ± 0.35 | The results support that the iSWAB-RC-EL media requires a minimum of 6 hours to inactivate SARS-CoV-2 at room temperature as shown by a >5 log reduction in TCID50/mL values when used with nasal or saliva matrix. 9. Assay Cut-Off: N/A B Comparison Studies: 1. Method Comparison with Predicate Device: N/A 2. Matrix Comparison: N/A C Clinical Studies: 1. Clinical Sensitivity: N/A 2. Clinical Specificity: N/A 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): N/A D Clinical Cut-Off: N/A K221802 - Page 9 of 10 {9} E Expected Values/Reference Range: N/A VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K221802 - Page 10 of 10
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...