URINE COLLECTION, PRESERVATION AND TRANSPORT SYSTEM
K013819 · Sierra Diagnostic, LLC · JTW · Feb 28, 2002 · Microbiology
Device Facts
Record ID
K013819
Device Name
URINE COLLECTION, PRESERVATION AND TRANSPORT SYSTEM
Applicant
Sierra Diagnostic, LLC
Product Code
JTW · Microbiology
Decision Date
Feb 28, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2900
Device Class
Class 1
Indications for Use
The Sierra Diagnostics L.L.C. Urine Collection, Preservation and Transport System is intended for use in the collection, preservation, and transportation of urine specimens at temperatures not exceeding 60°C for testing with the Abbott LCx® Neisseria qonorrhoeae and Chlamydia trachomatis assays.
Device Story
Device consists of urine collection cup containing nucleic acid chemical preservative; includes inert indicator beads to verify preservative presence. Used for collection, preservation, and transport of urine specimens for subsequent laboratory testing with Abbott LCx® assays. Unlike predicate devices requiring refrigeration, this system allows specimen storage for up to 6 days at temperatures up to 60°C. Healthcare providers use the system to maintain sample integrity for gonococcal and chlamydial nucleic acid detection. Preserved samples are transported to laboratories for analysis; results assist clinicians in diagnosing infections in symptomatic or asymptomatic patients.
Clinical Evidence
Multi-site clinical study confirmed device effectively preserved gonococcal and chlamydial nucleic acid targets in urine from symptomatic and asymptomatic males and females. Bench testing established preservative effectiveness: 100% correlation between refrigerated samples (24 hours) and samples preserved with the device (144 hours at 60°C). Sensitivity studies using 10 gonococcal serovars confirmed preservation down to the LCx® limit of detection with a 1:10 to 1:15 preservative-to-urine ratio.
Technological Characteristics
Urine collection cup containing chemical nucleic acid preservative and inert indicator beads. Preserves nucleic acid targets at temperatures up to 60°C. Preservative-to-urine ratio 1:10 to 1:15. Non-electronic, mechanical/chemical device.
Indications for Use
Indicated for collection, preservation, and transport of urine specimens from symptomatic and asymptomatic males and females for testing with Abbott LCx® Neisseria gonorrhoeae and Chlamydia trachomatis assays at temperatures up to 60°C.
Regulatory Classification
Identification
A microbiological specimen collection and transport device is a specimen collecting chamber intended for medical purposes to preserve the viability or integrity of microorganisms in specimens during storage of specimens after their collection and during their transport from the collecting area to the laboratory. The device may be labeled or otherwise represented as sterile. The device aids in the diagnosis of disease caused by pathogenic microorganisms.
Predicate Devices
Standard urine collection cup referenced in Abbott LCx® Neisseria gonorrhoeae (K935833) and Chlamydia trachomatis (K934622) package inserts
Related Devices
K163436 — Greiner Vacuette Urine Count and Culture, Mannitol tube · Greiner Bio-One Na, Inc. · Mar 6, 2017
K220739 — Disposable Urine Collection Tube · Zhejiang Gongdong Medical Technology Co., Ltd. · Nov 17, 2022
K070062 — COPROSET SALMONELLA AND SHIGELLA, URISET PRESERVATIVE TUBES AND SWAB SET GENERAL USE · Diesse Diagnostica Senese S.P.A. · Jul 30, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
FEB 2 8 2002
# 510(k) Summary
K013819
### I. General Information on Submitter
Name: Address:
Telephone:
Contact Person:
Date Prepared:
Fax:
Sierra Diagnostics, L.L.C. 21109 Longeway #C Sonora, CA 95370 (209) 536-0886 (209) 536-0853 Tony Baker October __ , 2001
#### II. General Information on Device
| Name: | Sierra Diagnostics L.L.C. Urine Collection,<br>Preservation and Transport System |
|----------------------|----------------------------------------------------------------------------------|
| Classification Name: | Accessory to Neisseria spp. and Chlamydia<br>serological reagents |
### III. Predicate Device
The standard urine collection cup used to collect specimens for testing with the Abbott LCx® Neisseria qonorrhoeae and Chlamydia trachomatis assays and referenced in the package inserts for the LCx® devices (See 510/k) Nos. K935833 (Neisseria gonorrhoeae) and K934622 (Chlamydia trachomatis)).
### Description of Device IV.
The device is comprised of a urine collection cup containing of a nucleic acid chemical preservative. The device allows urine specimens for LCx® gonococcal or chlamydial testing to be preserved for up to 6 days at temperatures not to exceed 60°C. Inert indicator beads are included in the urine cup as an indicator that a preservative is present in the sample.
#### V. Intended Use
The Sierra Diagnostics L.L.C. Urine Collection, Preservation and Transport System is intended for use in the collection, preservation, and transportation of urine specimens at temperatures not exceeding 60°C for testing with the Abbott LCx® Neisseria qonorrhoeae and Chlamydia trachomatis assays.
{1}------------------------------------------------
## Technological Characteristics of Device Compared to Predicate Device VI.
The Sierra Urine Collection, Preservation, and Transport System and the predicate device share the same technological characteristics with the exception of the method of preservation. The predicate device employs a temperature preservation method while the Sierra device uses chemical preservation.
### Summary of Performance Data VII.
The effectiveness of the Sierra Urine Collection, Preservation, and Transport System was established by the comparative testing of fresh and preserved urine spiked with gonococcal and chlamydial DNA. LCx® testing of samples that were preserved through refrigeration for 24 hours were compared with results for specimens preserved with the Sierra device and tested after being held for 144 hrs. at 60°C. There was a 100% correlation between the refrigerated and preserved samples.
Effectiveness was further established by a multi-site clinical study. The results of this study demonstrated that the device effectively preserved gonococcal and chlamydial nucleic acid targets in urine specimens from symptomatic and asymptomatic males and females.
The effective preservative concentration range and effect on LCx® sensitivity was established by a study using urine specimens spiked with less than 10 cfu of 10 different gonococcal serovars. Results from this test proved that Sierra's device effectively preserved nucleic acid targets down to the LCx® level of detection with a preservative to urine ratio ranging from 1:10 to 1:15.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# FEB 2 8 2002
Sierra Diagnostics, L.L.C. c/o Donald R. Stone, Esq. Kirkpatrick and Lockhart, LLP 1800 Massachusetts Avenue, NW Suite 200 Washington, DC 20036-1221
k013819 Re:
Ro13017
Trade/Device Name: Sierra Diagnostics L.L.C. Urine Collection, Preservation and Transport System Regulation Number: 21 CFR 866.2900 Regulation Name: Microbiological Specimen Collection and Transport System Regulatory Class: Class I Product Code: JTW Dated: February 11, 2002 Received: February 12, 2002
Dear Mr. Stone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can he found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, (201) 59 react the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small mornation on Jour responsiver Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 510(k) Number:
Sierra Diagnostics L.L.C. Urine Collection, Preservation and Transport Device Name: System
## Indications for Use:
The Sierra Diagnostics L.L.C. Urine Collection, Preservation and Transport System is indicated for 1 use in the collection, preservation, and transportation of urine specimens at temperatures not ase in the concession, proses with the Abbott LCx® Neisseria gonorrhoeae and Chlamydia trachomatis assays.
# PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IN NEEDED
` Concurrence of CDRH, Office of Device Evaluation (ODE)
**Prescription Use** √
(Per 21 CFR 801.109)
(Division Sio ver-The Counter Use Inical Laboratory L Division of C
510(k) Number_KO13
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.