K Number
K163436

Validate with FDA (Live)

Date Cleared
2017-03-06

(89 days)

Product Code
Regulation Number
866.2390
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Greiner Vacuette Urine Count and Culture, Mannitol tube is a urine stabilization device intended for collection, transport and storage of urine for bacterial and yeast culture. Urine samples collected in the Greiner Vacuette Urine Count and Culture, Mannitol tube can be stored at 20-25°C for up to 48 h prior to culture. This device is intended for professional use only.

Device Description

Vacuette Urine CCM tubes are tubes made of PET with a pre-defined vacuum to exact draw volumes. They are fitted with yellow Vacuette Safety Caps. The tube interior is sterile. The evacuated tube contains a fine particle granulate stabilizer to preserve the urine sample at 20-25°C for up to 48 hours. The stabilizer helps to preserve and protect the level of bacteria at collection and maintain typically formed elements in urine sediment.

AI/ML Overview

The provided document describes the Greiner Vacuette Urine Count and Culture, Mannitol tube, a device intended for the collection, transport, and storage of urine for bacterial and yeast culture. The performance data section outlines the study conducted to demonstrate the device's suitability.

  1. A table of acceptance criteria and the reported device performance
Acceptance CriteriaReported Device Performance
Recovered organism count within +/- 1 log from the original spiked concentration.Acceptable performance was demonstrated for all organisms tested.
  1. Sample size used for the test set and the data provenance

    • Sample size: All analyses were performed in triplicates (for each organism, at room temperature and refrigerated temperature). The organisms tested were 10 potential pathogens (Escherichia coli, Enterococcus faecalis, Proteus mirabilis, Pseudomonas aeruginosa, Staphylococcus saprophyticus, Enterobacter cloacae, Klebsiella pneumoniae, Streptococcus agalactiae, Candida albicans, Candida glabrata). Therefore, the total number of samples processed for recovery testing was 10 organisms * 3 replicates * 2 temperature conditions = 60 samples.
    • Data provenance: Not explicitly stated, but the study used ATCC cultures (American Type Culture Collection), which are standardized and universally recognized reference strains. The urine used was "sterile filtered urine." The study was conducted in accordance with CLSI M40-A2, a globally recognized standard.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the study did not involve human interpretation or expert medical assessment for establishing ground truth. The study focused on the performance of the device in maintaining organism viability, which is a quantitative laboratory measurement.

  3. Adjudication method for the test set

    This is not applicable, as the study does not involve expert review or interpretation of results that would require an adjudication method. The ground truth (original spiked concentration) is a predefined quantitative value.

  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a urine stabilization tube, not an AI-assisted diagnostic tool for human readers.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a physical product (a urine stabilization tube), not an algorithm or AI. The study evaluated the standalone performance of the device in maintaining microbial concentrations.

  6. The type of ground truth used

    The ground truth was the original spiked concentration of ATCC cultures in sterile filtered urine. This is a form of controlled experimental data.

  7. The sample size for the training set

    This is not applicable. The device is a physical product, not a machine learning model that requires a training set.

  8. How the ground truth for the training set was established

    This is not applicable, as there is no training set for this type of device and study.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads in profile, representing the department's mission to protect the health of all Americans. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

March 6, 2017

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

GREINER BIO-ONE NA INC. MANFRED ABEL QUALITY SYSTEM AND REGULATORY AFFAIRS 4238 CAPITAL DRIVE MONROE NC 28110

Re: K163436

Trade/Device Name: Greiner Vacuette Urine Count and Culture, Mannitol Tube Regulation Number: 21 CFR 866.2390 Regulation Name: Transport culture medium Regulatory Class: I Product Code: JSM, LIO Dated: December 5, 2016 Received: December 7, 2016

Dear Mr. Abel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Steven R. Gitterman -S

for Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163436

Device Name

Greiner Vacuette Urine Count and Culture, Mannitol tube

Indications for Use (Describe)

The Greiner Vacuette Urine Count and Culture, Mannitol tube is a urine stabilization device intended for collection, transport and storage of urine for bacterial and yeast culture. Urine samples collected in the Greiner Vacuette Urine Count and Culture, Mannitol tube can be stored at 20-25°C for up to 48 h prior to culture. This device is intended for professional use only.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5. Premarket notification 510(k) Summary

1. SUBMITTERApplicant Name:Greiner Bio-One GmbH.Bad Haller Strasse 324550 Kremsmuenster,Austria
Contact person:Manfred Abel, M.S., MBAGreiner Bio-One NA Inc.4238 Capital DriveMonroe, NC 28110704 763 0855Manfred.Abel@gbo.com
Establishment registration number:8020040
Date prepared:Nov 23, 2016

2. DEVICE - CLASSIFICATION

Trade Name:Greiner Vacuette Urine Count and Culture, Mannitol tube
Common name:Greiner Vacuette CCM tube
Device:Culture Media, Non-Propagating Transport
Regulatory Description:Transport culture medium
Regulation Medical Specialty:Microbiology
Product Code:JSM
Regulation No:866.2390
Class:1
Review Panel:Microbiology

3. PREDICATE DEVICE - CLASSIFICATION

Device name:BD Vacutainer Plus Plastic Urine C&S Tube
510(k) number:K024240
Device:Culture Media, Non-Propagating Transport
Regulatory Description:Transport culture medium
Regulation Medical Specialty:Microbiology
Product Code:JSM
Regulation No:866.2390
Class:1
Review Panel:Microbiology

4. DEVICE DESCRIPTION

Vacuette Urine CCM tubes are tubes made of PET with a pre-defined vacuum to exact draw volumes. They are fitted with yellow Vacuette Safety Caps. The tube interior is sterile.

{4}------------------------------------------------

The evacuated tube contains a fine particle granulate stabilizer to preserve the urine sample at 20-25°C for up to 48 hours. The stabilizer helps to preserve and protect the level of bacteria at collection and maintain typically formed elements in urine sediment.

5. INDICATION FOR USE

The Greiner Vacuette Urine Count and Culture, Mannitol tube is a urine stabilization device intended for collection, transport and storage of urine for bacterial and yeast culture. Urine samples collected in the Greiner Vacuette Urine Count and Culture, Mannitol tube can be stored at 20-25°C for up to 48 h prior to culture. This device is intended for professional use only.

6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Both the subject device and the predicate have the same fundamental scientific technology and technological characteristics. Both devices are comprised of clear plastic tubes.

Greiner Vacuette Urine CCM tube(Count & Culture Mannitol)BD Vacutainer Plus Plastic UrineC&S Tube(Culture and Sensitivity)
FDA statusUnder reviewK024240
ClassificationClass 1Class 1
Regulation866.2390866.2390
Product CodeJSMJSM
Intended UseThe Vacuette Urine CCM tube is a urinestabilization device intended forcollection, transport and storage of urinefor bacterial and yeast culture. Urinesamples collected in the Vacuette® UrineCCM tube can be stored at 20-25°C forup to 48 h prior to culture. This device isintended for professional use only.The BD Vacutainer™ PLUS PlasticUrine C&S Preservative Tubes andKits are intended for the collectionand transport of urine samples forculture and sensitivity testing ofbacteria.The BD Vacutainer Plus PlasticUrine C&S Preservation Tube isfilled with lyophilized urinemaintenance formula and evacuatedto draw approximately 4 - 10 ml(depending on the tube size) ofurine. The lyophilized urinemaintenance formula can maintainthe bacterial population in the urinespecimen for a period of 48 hours atroom temperature levels comparableto those urine specimens withoutadditive, held under refrigeration forthe same period of time
ManufacturerGreiner Bio-One (GBO)Becton, Dickinson and company(BD)
Tubeclear plasticclear plastic and glass
Tube volume13x75 mm - 4ml16x100 mm - 5ml16x100 mm - 10ml13x75 mm - 4ml16x75 mm - 5ml16x100 mm - 10ml
CapSafety cap, yellowConventional stopper, gray
Vacuumpredefined Vacuumpredefined Vacuum
Additiveboric acid, sodium formate, sodiumborate, mannitolboric acid, sodium formate, sodiumborate, mannitol

{5}------------------------------------------------

Storage20 – 25 °Cnormal room condition
Storage stability atroom temperature48 hours48 hours
SterileYesyes

The intended use of the Greiner Vacuette Urine CCM tube is similar to the predicate device. Both devices are intended for the collection and storage of urine samples for culture at room temperature for up to 48 hours. Based on the device features, principles of operation, and technological characteristics we believe Greiner CCM tube is substantially equivalent to the predicate device.

7. PERFORMANCE DATA

Testing was performed for the suitability of Greiner CCM tube as sampling and transport system for microbiological specimen using procedures outlined in the Clinical Laboratory Standard Institute (CLSI) M40-A2 Quality Control of Microbiological Transport Systems. Viability and recovery testing was performed with recommendation from M40-A2 and FDA. ATCC cultures from 10 potential pathogens of Escherichia coli, Enterococcus faecalis, Proteus mirabilis, Pseudomonas aeruginosa, Staphylococcus saprophyticus, Enterobacter cloacae, Klebsiella pneumoniae, Streptococcus agalactiae, Candida albicans. Candida glabrata were used to spike sterile filtered urine to 10° and 10° cfu/ml. All analyses were performed in triplicates at room temperature and refrigerated temperature.

8. DISCUSSION / CONCLUSION

The GBO CCM tube fulfills the requirements set in the CLSI standard M40-A2, "Quality Control of Microbiological Transport Systems" demonstrating that the recovered organism are within +/- 1 log from the original spiked concentration is considered acceptable performance. The data provided shows that acceptable performance was demonstrated for all organisms tested.

§ 866.2390 Transport culture medium.

(a)
Identification. A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.(b)
Classification. Class I (general controls).