K220739 · Zhejiang Gongdong Medical Technology Co., Ltd. · JSM · Nov 17, 2022 · Microbiology
Device Facts
Record ID
K220739
Device Name
Disposable Urine Collection Tube
Applicant
Zhejiang Gongdong Medical Technology Co., Ltd.
Product Code
JSM · Microbiology
Decision Date
Nov 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2390
Device Class
Class 1
Indications for Use
Disposable Urine Collection Tube is intended for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory. In the laboratory, urine specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts.
Device Story
Disposable Urine Collection Tube system facilitates safe collection and transport of urine specimens. Device consists of PET tube, butyl-rubber safety plug, and PE cap containing boric acid preservative. Vacuum-based mechanism enables automated filling of urine sample via puncture needle. Used by healthcare professionals in clinical settings to transport specimens to laboratories. Boric acid additive prevents overgrowth of uropathogenic bacteria and yeasts during transport. Healthcare providers process samples using standard clinical laboratory procedures for cultivation. System maintains specimen viability for up to 48 hours at 2-4°C and 22-25°C, supporting accurate diagnostic testing and clinical decision-making.
Clinical Evidence
Bench testing only. No clinical data. Microbial recovery studies evaluated viability of C. albicans, E. coli, E. faecalis, P. aeruginosa, P. mirabilis, and S. saprophyticus at 0, 24, and 48 hours at 2-4°C and 22-25°C. All results showed <1 log10 change. Fill volume impact study (90% and 110% fill) confirmed <1 log10 change for E. coli, P. aeruginosa, and P. mirabilis after 48 hours. Shelf-life validated for 18 months via real-time aging. Sterilization validated to SAL 10^-6 per ISO 11137-1/2.
Indicated for the collection, transport, and preservation of urine specimens from the collection site to the testing laboratory for the cultivation of uropathogenic bacteria and yeasts. Intended for use by healthcare professionals.
Regulatory Classification
Identification
A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.
Predicate Devices
UriSwab™ - Urine Collection, Transport and Preservation System (K180052)
Related Devices
K163436 — Greiner Vacuette Urine Count and Culture, Mannitol tube · Greiner Bio-One Na, Inc. · Mar 6, 2017
K024240 — BD VACUTAINER PLUS PLASTIC URINE C&S PRESERVAYIVE TUBES AND KITS · Bd · Mar 26, 2003
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K220739
B Applicant
Zhejiang Gongdong Medical Technology Co., Ltd.
C Proprietary and Established Names
Disposable Urine Collection Tube
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JSM | Class I, reserved | 21 CFR 866.2390 - Transport Culture Medium | MI - Microbiology |
| LIO | Class I, reserved | 21 CFR 866.2900 - Microbiological specimen collection and transport device | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain a substantial equivalence determination for the Disposable Urine Collection Tubes for the collection, transport, and preservation of urine specimen from the collection site to the testing laboratory for the cultivation of uropathogenic bacteria and yeasts.
B Measurand:
Not applicable
C Type of Test:
Non-propagating Transport Device
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K220739 - Page 2 of 8
# III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
Disposable Urine Collection Tube is intended for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory. In the laboratory, urine specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
None
# IV Device/System Characteristics:
A Device Description:
The Disposable Urine Collection Tubes for the collection, transport and preservation system consist of a safety rubber plug, a plastic cap and boric acid as an additive inside the tube. The tube is made of Polyethylene terephthalate (PET) and the safety plug is made of butyl-rubber while the disposable cap is made of Polyethylene (PE). The disposable tube device sustains the viability of organisms during transport for up to 48 hrs. at 2-4°C and 22-25°C. There are two main tube types (round and conical bottom), and four configurations by volume of urine for the round bottom tubes (4 ml, 7 ml, 8 ml, 10 ml). The conical bottom tubes only have one configuration by volume (8 ml). Each volume is available in yellow and white caps. The different device configurations and information on the amount of preservative contained in each tube is provided in table 1.
Table 1. Disposable Tubes Description and Packaging configurations
| Ref No. | Draw volume (mL) | Tube size (Diameter × Length) in mm | Additive Boric acid in mg | Material | Type of tube | Cap color |
| --- | --- | --- | --- | --- | --- | --- |
| GD040UB | 4 | 13×75 | 40 | PET | Round bottom | yellow |
| GD040UBW | 4 | 13×75 | 40 | PET | Round bottom | white |
| GD070UB | 7 | 13×100 | 70 | PET | Round bottom | yellow |
| GD070UBW | 7 | 13×100 | 70 | PET | Round bottom | white |
| GD080UB | 8 | 16×100 | 80 | PET | Round bottom | yellow |
{2}
K220739 - Page 3 of 8
| GD080UBW | 8 | 16×100 | 80 | PET | Round bottom | white |
| --- | --- | --- | --- | --- | --- | --- |
| GD100UB | 10 | 16×100 | 100 | PET | Round bottom | yellow |
| GD100UBW | 10 | 16×100 | 100 | PET | Round bottom | white |
| GD080UAC | 8 | 16×100 | 80 | PET | Conical bottom | yellow |
| GD080UACW | 8 | 16×100 | 80 | PET | Conical bottom | white |
## B Principle of Operation:
The Disposable Urine Collection Tube system is used to safely collect and transport urine from collection sites to the testing laboratories. It is intended to be used by health care professionals. The vacuum in the collection tube allows autofill the urine sample through the puncture needle and the boric acid in the tubes prevents overgrowth of uropathogenic bacteria and yeasts.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
UriSwab-Urine Collection, Transport and Preservation System
### B Predicate 510(k) Number(s):
K180052
### C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device: K220739 | Predicate: K180052 |
| --- | --- | --- |
| Device Trade Name | Disposable Urine Collection Tube | UriSwab-Urine Collection, Transport and Preservation System |
| Device Product Code and Classification | JSM, LIO, Class I | Same |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Disposable Urine Collection Tube is intended for the collection, transport, and preservation of urine specimens from the collection site to the testing laboratory. In the laboratory, urine | Copan UriSwab - Urine Collection, Transport and Preservation System is intended for the collection, transport, and preservation of urine specimens from the collection site to the testing laboratory. In |
{3}
| | specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts. | the laboratory, UriSwab specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts. |
| --- | --- | --- |
| Specimen Stability | 2-4°C and 22-25°C; 48 hrs. | Same |
| Device Storage Temperature | 22-25°C | Same |
| Sterilization | Radiation | Same |
| Sterile | Yes; 10-6SAL | Same |
| General Device Characteristic Differences | | |
| Tube material | Polyethylene Terephthalate (PET) | Plastic |
| Tube type | Round and Conical bottom | Conical bottom |
| Tube size | 13 mm × 75 mm; 13 mm × 100 mm; 16 mm × 100 mm | 12 mm × 80 mm; 16 mm × 100 mm |
| Tube volume | 4 ml, 7 ml, 8 ml, 10 ml | 1.5 ml, 3.2 ml |
| Cap | Safety cap, yellow and white. | Screw cap, yellow |
| Additive | Boric acid | Boric acid and Sodium formate |
| Collection device | PET tube with additive | Plastic tube, stick with sponge |
| Tube filling | Predefined vacuum | Sponge applicator |
VI Standards/Guidance Documents Referenced:
Quality Control of Microbiological Transport Systems, Clinical and Laboratory Standards Institute (CLSI) guideline M40-A2, 2015.
Sterilization of health care products - Radiation - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices, 1st edition, 11137-1, International Organization for Standardization (ISO), 2006.
Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose, 3rd edition, 11137-2, International Organization for Standardization (ISO), 2013.
VII Performance Characteristics (if/when applicable):
K220739 - Page 4 of 8
{4}
K220739 - Page 5 of 8
# A Analytical Performance:
1. Precision/Reproducibility:
Not Applicable
2. Linearity:
Not Applicable
3. Analytical Specificity/Interference:
Not Applicable
4. Assay Reportable Range:
Not Applicable
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
## Stability Verification Report:
The shelf life for the Disposable Urine Collection Tube was determined to be 18 months from the date of manufacture when stored in well-ventilated clean room with the environment of relative humidity not more than 80% and at temperature 22 – 25°C. The shelf life of the Disposable Urine Collection Tube was evaluated using real-time aging performance test at time points T = 0, 6, 12 and 18 months. At each time point, the following performance tests were conducted as described below:
- Appearance test was carried out by inspecting visually for clear label, transparent PET tube and any sign of deformation or breakage.
- Vacuum test was performed by measuring volume drawn into the Disposable Urine Collection Tube between 90 to 110% of the nominal liquid capacity allowed.
- Physical and mechanical performance tests included leak test, strength test, and a 50 KPa pressure test.
- Mechanical stress test was conducted by checking the tubes could withstand centrifugation.
- The strength test, and urine collection volume capacity test were carried out as part of the evaluation.
- Sterility test and the standard of boric acid content used were also assessed.
- The package integrity was evaluated by visually checking for any changes or damage to the package. Six lots of varying production were used for the studies and results showed that all samples passed the real time aging test for 18 months at 22-25°C.
All the results were acceptable and support the claim that the Disposable Urine Collection Tube is physically or visually stable for 18 months.
{5}
# Sterilization:
The Disposable Urine Collection Tubes were sterilized by electron beam radiation with an average dose of $20\mathrm{kGy}$ to achieve a sterility assurance level (SAL) of $10^{-6}$ with a dosimeter attached to monitor dosage through the sterilization process. Sterilization was validated following ISO 11137-1: 2006 and ISO 11137-2: 2013 Sterilization of Health Care Products Radiation guidelines. Sterility test was carried out according to ISO11137-2:2009 Medical Devices Radiation Sterilization-Microbiological Methods-Part 2. In this sterility test, 100 samples after irradiated at the verification dose were inoculated with $200\mathrm{ml}$ soybean casein liquid medium and incubated at $30\pm 2^{\circ}\mathrm{C}$ for 14 days. The results were observed with no growth. Sterilization studies of the tube were acceptable.
# 6. Detection Limit:
# Microbial Recovery Studies:
Microbial recovery (viability) studies were conducted to support the claim of the Disposable Urine Collection Tube for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory. Sample panels were prepared from the listed bacteria and yeast (below) that were commonly associated with urinary tract infections. Cultures from $1.5 \times 10^{8}$ (equivalent 0.5 McFarland standard) were diluted into the pooled human negative clinical urine matrix to give a final concentration of $1.5 \times 10^{2} \mathrm{CFU} / \mathrm{ml}$ . The appropriate volume of urine samples spiked with each organism was introduced into the Disposable Urine Collection Tubes. Three lots of the Disposable Urine Collection Tubes with three samples per lot were evaluated for the microbial recovery (viability) assay. The inoculated sample tubes were initially tested to establish a time point zero then stored at $2-4^{\circ} \mathrm{C}$ and $22-25^{\circ} \mathrm{C}$ for 24 hrs., and 48 hrs. After each storage time and temperature, the samples were plated to grow colonies which were counted to determine CFU/ml for each time points. According to CLSI document M40-A2, the acceptance criteria for microbial recovery method was set to be within $1 \log_{10}$ of the initial microorganism count. Results of recovery studies are presented in table 2.
Table 2. Performance of Disposable Urine Collection Tube in Microbial Recovery Studies
| Strain tested | Storage temperature | Average CFU/ml recovered (N=9) at time points | | | Change in Log10(0 - 48 hrs) |
| --- | --- | --- | --- | --- | --- |
| | | 0 hr | 24 hrs | 48 hrs | |
| Candida albicans(ATCC 24433) | 2-4°C | 140 | 107 | 85 | -0.22 |
| | 22-25°C | 140 | 74 | 69 | -0.31 |
| Escherichia coli(ATCC 25922) | 2-4°C | 137 | 104 | 92 | -0.17 |
| | 22-25°C | 137 | 85 | 42 | -0.51 |
| Enterococcus faecalis(ATCC 29212) | 2-4°C | 145 | 102 | 91 | -0.20 |
| | 22-25°C | 145 | 80 | 54 | -0.43 |
| Pseudomonas aeruginosa(ATCC 27853) | 2-4°C | 143 | 103 | 69 | -0.32 |
| | 22-25°C | 143 | 106 | 72 | -0.30 |
| Proteus mirabilis(ATCC 7002) | 2-4°C | 140 | 111 | 72 | -0.29 |
| | 22-25°C | 140 | 105 | 71 | -0.29 |
| Staphylococcus saprophyticus | 2-4°C | 139 | 109 | 69 | -0.30 |
K220739 - Page 6 of 8
{6}
The microbial recovery studies demonstrated acceptable results.
## Fill Volume Impact Recovery Study:
A fill volume impact recovery study was performed to demonstrate that underfilling of Disposable Urine Collection Tube containing preservative does not negatively impact the viability of urine microorganism. Sample panels were prepared by listed microorganisms (below) diluted from $1.5 \times 10^{8}$ CFU/mL (equivalent to a $0.5$ McFarland standard) into the pooled human urine matrix to give a final concentration of $1.5 \times 10^{2}$ CFU/ml. Three production lots were tested in three replicates with each urine samples spiked with $E.$ coli, $P.$ aeruginosa or $P.$ mirabilis at the sample fill volume of $90\%$ and $110\%$. The inoculated sample tubes were initially tested to establish a time point zero then stored at $22-25^{\circ}\mathrm{C}$ for 48 hrs. After the storage time, the samples were plated to grow colonies which were counted to determine CFU/ml for each time points. The acceptance criteria for the study is less than 1 $\log_{10}$ change when the fill volume is above or below the line. The results were shown in table 3.
Table 3. Performance of Disposable Urine Collection Tube in fill volume impact recovery study.
| Strain tested | Fill volume | Average CFU/ml recovered (N=9) at time points | | Change in Log_{10} (0 - 48 hrs) |
| --- | --- | --- | --- | --- |
| | | 0 hr | 48 hrs | |
| E. coli (ATCC 25922) | 90% | 125 | 48 | -0.42 |
| | 110% | 127 | 46 | -0.44 |
| P. aeruginosa (ATCC 27853) | 90% | 143 | 49 | -0.47 |
| | 110% | 142 | 50 | -0.45 |
| P. mirabilis (ATCC 7002) | 90% | 140 | 40 | -0.54 |
| | 110% | 139 | 39 | -0.55 |
Fill volume recovery study results were acceptable.
## 7. Assay Cut-Off:
Not applicable.
K220739 - Page 7 of 8
{7}
B Comparison Studies:
1. Method Comparison with Predicate Device:
Method comparison is not applicable for a bacterial transport medium. The device itself does not provide a result that can be used in making a clinical decision. Bench testing studies were conducted to determine the ability of the Disposable Urine Collection Tube for urine collection, transport & preservation to maintain viability of microorganisms commonly found as the etiological agents of urinary tract infections. The analytical studies showed recovery of representative bacteria are within the acceptance criteria recommended in CLSI-M40-A2.
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
Not applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K220739 - Page 8 of 8
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.