MICROSCAN MICROSTREP PLUS PANEL CHLORAMPHENICOL (1 TO 32 MCG/ML)
K062293 · Dade Behring, Inc. · LRG · Aug 24, 2006 · Microbiology
Device Facts
Record ID
K062293
Device Name
MICROSCAN MICROSTREP PLUS PANEL CHLORAMPHENICOL (1 TO 32 MCG/ML)
Applicant
Dade Behring, Inc.
Product Code
LRG · Microbiology
Decision Date
Aug 24, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Indications for Use
To determine bacterial antimicrobial agent susceptibility. The MicroScan MICroSTREP plus ® Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae . After inoculation, panels are incubated for 20 - 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read visually according to the Package Insert. Additionally, the panels may be incubated in and read by a MicroScan ® WalkAway instrument. This particular submission is for the addition of instrument read capability of the antimicrobial Chloramphenicol, at concentrations of 1 to 32 mcg/ml on the MicroScan MICroSTREP plus ® Panel. The organisms which may be used for Chloramphenicol susceptibility testing on this panel are: Susceptible streptococci, including Streptococcus pneumoniae
Device Story
MicroScan MICroSTREP plus Panel is a miniaturized broth dilution susceptibility test. Input: bacterial colonies grown on solid media. Process: panels rehydrated with Mueller-Hinton broth, lysed horse blood, and HEPES buffer; inoculated with standardized organism suspension; incubated 20-24 hours at 35°C. Output: Minimum Inhibitory Concentration (MIC) determined by visual observation or automated reading via MicroScan WalkAway instrument. Used in clinical laboratories by technicians. Instrument reading provides quantitative/qualitative susceptibility data to clinicians to guide antibiotic therapy for streptococcal infections.
Clinical Evidence
Bench testing only. Performance evaluated using stock and CDC Challenge strains. Compared instrument read results against expected results derived from CLSI frozen Reference Panels. Demonstrated 100% Essential Agreement for Chloramphenicol. Reproducibility and precision testing with the WalkAway instrument were acceptable.
Technological Characteristics
Miniaturized broth microdilution panel. Reagents: Mueller-Hinton broth, 2-5% lysed horse blood, 50 mM HEPES buffer. Energy source: non-CO2 incubator and automated WalkAway instrument. Connectivity: WalkAway instrument integration. Sterilization: Not specified.
Indications for Use
Indicated for determining antimicrobial susceptibility of aerobic streptococci, including Streptococcus pneumoniae, to Chloramphenicol (1-32 mcg/ml).
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
K063099 — MICROSCAN MICROSTREP PLUS PANEL CEFEPIME (O.015 - 8 MCG/ML) · Dade Behring, Inc. · Nov 1, 2006
K020939 — MICROSCAN MICROSTREP PLUS PANEL-TETRACYCLINE · Dade Microscan, Inc. · May 17, 2002
K020686 — MICROSCAN MICROSTREP PLUS PANEL NEW ANTIMICROBIAL - TRIMETHOPRIM/SULFAMETHOXAZOLE · Dade Microscan, Inc. · Apr 30, 2002
K063064 — MICROSCAN MICROSTREP PLUS PANEL CEFTRIAXONE (0.015-8MCG/ML) · Dade Behring, Inc. · Oct 26, 2006
K020820 — MICROSCAN MICROSTEP PLUS PANEL NEW ANTIMICROBIAL - CEFUROXIME · Dade Behring, Inc. · May 9, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
K062293
AUG 2 4 2006
## 510(k) Summary Information:
| Device Manufacturer: | Dade Behring Inc. |
|----------------------|-------------------------------------------------------------------------------------------------------------|
| Contact name: | Maureen Mende, Regulatory Affairs Group Manager |
| Phone/Fax: | 916-374-3174/916-374-3144 |
| Date prepared: | July 25, 2006 |
| Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panels |
| Trade Name: | MicroScan MICroSTREP plus® Panel |
| Intended Use: | To determine bacterial susceptibility to Chloramphenicol |
| Indication for Use: | For determining antimicrobic susceptibility with aerobic<br>streptococci including Streptococcus pneumoniae |
| Predicate device: | MicroScan® MICroSTREP plus® Panel |
## 510(k) Summary:
The MicroScan MICroSTREP plus® Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae. After inoculation, panels are incubated for 20 - 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read according to the Package Insert.
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test. Various antimicrobial agents are diluted in water, buffer or minute concentrations of broth to concentrations bridging the range of clinical interest. Panels are rehydrated with 115 ul Mueller-Hinton broth supplemented with 2-5% lysed horse blood (LHB) and buffered with 50 mM HEPES, after inoculation of the broth with a standardized suspension of the organism in saline. After incubation in a non-CO2 incubator for 20-24 hours, the minimum inhibitory concentration (MIC) for the test organism is manually read by observing the lowest antimicrobial concentration showing inhibition of growth. Additionally, the panels may be incubated in and read by a MicroScan® WalkAway instrument.
The proposed instrument read method for the MicroScan MICroSTREP plus® Panel demonstrated substantially equivalent performance with streptococcal isolates when compared with an expected result generated on a CLSI frozen Reference Panel, as defined in the FDA document "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", dated February 5, 2003.
This Premarket Notification (510[k]) presents data in support of reading the MICroSTREP plus® Panel with Chloramphenicol on the MicroScan® WalkAway instrument.
The external evaluation was conducted with stock and CDC Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed instrument read method for the MICroSTREP plus® Panel by comparing its performance with Expected Results determined before the evaluation. The MICroSTREP plus® Panel demonstrated acceptable
{1}------------------------------------------------
performance with an overall Essential Agreement of 100% for Chloramphenicol instrument read results compared with the Expected Result.
Instrument reproducibility testing demonstrated acceptable reproducibility and precision with Chloramphenicol and the WalkAway® instrument.
Quality Control testing demonstrated acceptable results for Chloramphenicol.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Maureen Mende Regulatory Affairs Group Manager Dade Behring, Inc. 2040 Enterprise Boulevard West Sacramento, CA 95691-9972
AUG 2 4 2006
Re: k062293
> Trade/Device Name: MicroScan MICroSTREP plus® Panel Chloramphenicol (1 to 32 mcg/ml) Regulation Number: 21 CFR § 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: II Product Code: LRG, LTT Dated: July 25, 2006 Received: August 7, 2006
Dear Ms. Mende:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours,
Sally attorn
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indication for Use Statement
| 510(k) No.: | K062293<br>(To be assigned by FDA) |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | MicroScan MICroSTREP plus <sup>®</sup> Panel<br>Chloramphenicol (1 to 32 mcg/ml) |
| Intended Use | To determine bacterial antimicrobial agent susceptibility |
| Indications for Use: | The MicroScan MICroSTREP plus <sup>®</sup> Panel is used to determine<br>quantitative and/or qualitative antimicrobial agent susceptibility of<br>colonies grown on solid media of aerobic streptococci, including<br><i>Streptococcus pneumoniae</i> . After inoculation, panels are incubated<br>for 20 - 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and<br>read visually according to the Package Insert. Additionally, the<br>panels may be incubated in and read by a MicroScan <sup>®</sup> WalkAway<br>instrument. |
| | This particular submission is for the addition of instrument read<br>capability of the antimicrobial Chloramphenicol, at concentrations<br>of 1 to 32 mcg/ml on the MicroScan MICroSTREP plus <sup>®</sup> Panel. |
| | The organisms which may be used for Chloramphenicol<br>susceptibility testing on this panel are: |
| | Susceptible streptococci, including <i>Streptococcus pneumoniae</i> |
Prescription Use ___ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
__ﺍ__Page 1 of__1__
Division Sign-Off
reddu
Office of In Viter Diagnostic Device Evaluation and Safety
్రార 1 -----
SINIKI KO6 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.