K Number
K020939

Validate with FDA (Live)

Date Cleared
2002-05-17

(56 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To determine bacterial susceptibility to Tetracycline
For determining antimicrobic susceptibility with aerobic streptococci, including Streptococcus pneumoniae

Device Description

The MicroScan® MICroSTREP plus™ Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae. After inoculation, panels are incubated for 20 - 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read visually according to the Package Insert.
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test. Various antimicrobial agents are diluted in water, buffer or minute concentrations of broth to concentrations bridging the range of clinical interest. Panels are rehydrated with 115 ul Mueller-Hinton broth supplemented with 2-5% lysed horse blood (LHB) and buffered with 50 mM HEPES, after inoculation of the broth with a standardized suspension of the organism in saline. After incubation in a non-CO2 incubator for 20-24 hours, the minimum inhibitory concentration (MIC) for the test organism is manually read by observing the lowest antimicrobial concentration showing inhibition of growth.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the MicroScan® MICroSTREP plus™ Panel with Tetracycline, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Acceptable performance for streptococcal isolates when compared with an NCCLS frozen Reference Panel, as defined in the FDA DRAFT document "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", dated March 8, 2000.Overall Essential Agreement of 92.9% for Tetracycline when compared with the frozen Reference panel.
Acceptable reproducibility and precision with Tetracycline.Reproducibility testing demonstrated acceptable reproducibility and precision with Tetracycline.
Acceptable Quality Control results for Tetracycline.Quality Control testing demonstrated acceptable results for Tetracycline.

Study Details

  1. Sample size used for the test set and the data provenance:

    • The document mentions "fresh and stock Efficacy isolates and stock Challenge strains" for the external evaluation. However, the specific number of isolates used in the test set is not provided in the document.
    • Data Provenance: The document does not explicitly state the country of origin. It indicates it was an "external evaluation," suggesting it was not conducted internally by the manufacturer. It's a prospective evaluation comparing the new panel against a reference standard.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. The ground truth was established by an "NCCLS frozen Reference Panel," which implicitly suggests expert-defined reference standards, but without specifying the number or qualifications of experts involved in its establishment.
  3. Adjudication method for the test set:

    • This information is not provided in the document. The comparison was made against a reference panel, but the method of discrepancy resolution (adjudication) if any, is not described.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is a quantitative and/or qualitative antimicrobial agent susceptibility panel, not an AI-assisted diagnostic tool for human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This device is a diagnostic panel that requires visual interpretation. The document states, "panels are incubated for 20 - 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read visually according to the Package Insert," and "the minimum inhibitory concentration (MIC) for the test organism is manually read by observing the lowest antimicrobial concentration showing inhibition of growth." Therefore, it is not a standalone algorithm-only device nor is it explicitly described as having an 'algorithm only' performance study. The core of its performance relies on the visual manual reading against the reference.
  6. The type of ground truth used:

    • The ground truth was established by an NCCLS frozen Reference Panel. This type of ground truth represents a highly standardized and accepted method for determining antimicrobial susceptibility, implicitly involving expert-defined and validated reference standards for MIC determination.
  7. The sample size for the training set:

    • The document does not mention a training set. This type of device (a diagnostic panel for antimicrobial susceptibility) is typically validated against a reference method rather than developed using a machine learning training set.
  8. How the ground truth for the training set was established:

    • As no training set is mentioned, this information is not applicable.

{0}------------------------------------------------

162020939

MAY 1 7 2002

510(k) Summary Information:

Device Manufacturer:Dade MicroScan Inc.
Contact name:Cynthia Van Duker, Regulatory Affairs Manager
Fax:916-374-3144
Date prepared:March 20, 2002
Product Name:Microdilution Minimum Inhibitory Concentration (MIC) Panels
Trade Name:MicroScan® MICroSTREP plus™ Panel
Intended Use:To determine bacterial susceptibility to Tetracycline
Indication for UseFor determining antimicrobic susceptibility with aerobic streptococci, including
Streptococcus pneumoniae
Predicate device:MicroScan® Streptococcus MIC Panel (K963641).

510(k) Summary:

The MicroScan® MICroSTREP plus™ Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae. After inoculation, panels are incubated for 20 - 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read visually according to the Package Insert.

The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test. Various antimicrobial agents are diluted in water, buffer or minute concentrations of broth to concentrations bridging the range of clinical interest. Panels are rehydrated with 115 ul Mueller-Hinton broth supplemented with 2-5% lysed horse blood (LHB) and buffered with 50 mM HEPES, after inoculation of the broth with a standardized suspension of the organism in saline. After incubation in a non-CO2 incubator for 20-24 hours, the minimum inhibitory concentration (MIC) for the test organism is manually read by observing the lowest antimicrobial concentration showing inhibition of growth.

The proposed MicroScan® MICroSTREP plus™ Panel demonstrated substantially equivalent performance with streptococcal isolates when compared with an NCCLS frozen Reference Panel, as defined in the FDA DRAFT document "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", dated March 8, 2000.

The Premarket Notification (510[k]) presents data in support of the new MICroSTREP plus™ Panel with Tetracycline.

The external evaluation was conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed MICroSTREP plus™ Panel by comparing its performance with an NCCLS frozen Reference panel. The MICroSTREP plus™ Panel demonstrated acceptable performance with an overall Essential Agreement of 92.9% for Tetracycline when compared with the frozen Reference panel.

Reproducibility testing demonstrated acceptable reproducibility and precision with Tetracycline.

Quality Control testing demonstrated acceptable results for Tetracycline.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the caduceus. The logo is black and white.

MAY 1 7 2002

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Cynthia Van Duker Regulatory Affairs Manager Dade Behring Inc. 1584 Enterprise Boulevard West Sacramento, CA 95691

Re: K020939

Trade/Device Name: MicroScan® MICroSTREP plus™ Panels with Tetracyclin 0.06-16ug/ml Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: Class II Product Code: LTT, JWY Dated: March 20, 2002 Received: March 22, 2002

Dear Ms. Van Duker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and v additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indication for Use Statement

510(k) No .:

Device Name:

Intended Use

Indications for Use:

K020939 (To be assigned by FDA)

MicroScan® MICroSTREP plus™ Panel

To determine bacterial antimicrobial agent susceptibility

The MicroScan® MICroSTREP plus™ Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae. After inoculation, panels are incubated for 20 - 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read visually according to the Package Insert.

This particular submission is for the addition of the antimicrobial Tetracycline at concentrations of 0.06 to 16 mcg/ml to the test panel

The organisms which may be used for Tetracycline susceptibility testing in this panel are:

Streptococcus spp. Streptococcus pneumoniae

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) NumberK020939
Prescription UseOver-The-Counter Use
(Per 21 CFR 801.109)OR

(Optional Format 1-2-96)

8

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).