MICROSCAN DRIED GRAM-POSITIVE MIC/COMBO PANELS NEW ANTIMICROBIAL TEST CEFOXITIN SCREEN
Applicant
Siemens Healthcare Diagnostics
Product Code
LRG · Microbiology
Decision Date
Sep 8, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Indications for Use
The MicroScan® Dried Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive cocci. After inoculation, panels are incubated for 16 - 24 hours at 35℃ +/- 1℃ in a non-CO2 incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert. The MicroScan Cefoxitin Screen is intended to determine the susceptibility of staphylococci to the penicillinase stable beta-lactams. This particular submission is for the addition of the antimicrobial test the Cefoxitin Screen, at a concentration of 4 mcg/ml, to the test panel. The gram-positive organisms which may be used for Cefoxitin Screen susceptibility testing in this panel are: Staphylococcus aureus Staphylococcus lugdunensis
Device Story
MicroScan® Dried Gram-Positive MIC/Combo Panels utilize dehydrated Mueller Hinton Broth and nutrients in microdilution wells. Inoculum prepared via turbidity or Prompt® method; panels rehydrated and incubated at 35°C for 16-20 hours. Device detects qualitative antimicrobial susceptibility; Cefoxitin Screen (CfxS) well (4 µg/mL) used alongside oxacillin MIC to assess resistance. Read manually or automatically via autoScan® 4 or WalkAway® instruments. Output provides MIC and categorical interpretation (SIR). Used in clinical microbiology laboratories to guide antibiotic selection for S. aureus and S. lugdunensis infections.
Clinical Evidence
Clinical evaluation compared Cefoxitin Screen performance against CLSI cefoxitin disk diffusion and mecA PCR using fresh/stock clinical isolates and challenge strains. Results: 97.8% categorical agreement vs. disk diffusion; 98.7% vs. mecA PCR. Sensitivity/Specificity vs. disk diffusion: 98.6%/96.9%. Sensitivity/Specificity vs. mecA PCR: 99.7%/97.7%. Reproducibility testing confirmed performance across different inoculum methods (Turbidity, Prompt) and instruments (autoSCAN-4, WalkAway).
Technological Characteristics
Miniaturized broth microdilution panels; dehydrated antimicrobial agents in wells. Rehydrated with water. Incubation 16-24 hours at 35°C. Compatible with visual reading or automated instrumentation (autoSCAN-4, WalkAway).
Indications for Use
Indicated for the addition of Cefoxitin Screen at 4 µg/mL to gram positive susceptibility panels for testing Staphylococcus aureus and Staphylococcus lugdunensis only.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Predicate Devices
MicroScan Dried Gram-Positive and Gram-Negative MIC/Combo Panels (K862140)
Related Devices
K020160 — DRIED GRAM-POSITIVE MIC/COMBO PANELS WITH AMPICILLIN · Dade Behring, Inc. · Mar 14, 2002
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k081013
B. Purpose for Submission:
Addition of Cefoxitin screen to the MicroScan® Dried Gram-Positive MIC/Combo Panels
C. Measurand:
Cefoxitin 4 µg/mL
D. Type of Test:
Qualitative growth based detection algorithm
E. Applicant:
Siemens Healthcare Diagnostics, Inc.
F. Proprietary and Established Names:
MicroScan® Dried Gram-Positive MIC/Combo Panels
G. Regulatory Information:
1. Regulation section:
866.1640 Antimicrobial Susceptibility Test (AST) Powder
2. Classification:
Class II
3. Product code:
LRG-Instrument for Auto Reader & Interpretation of Overnight Antimicrobial Susceptibility Systems
JWY - Manual Antimicrobial Susceptibility Test Systems
LTT – Panels, Test, Susceptibility, Antimicrobial
LTW – Susceptibility Test Cards, Antimicrobial
4. Panel:
83 Microbiology
H. Intended Use:
1. Intended use(s):
{1}
For use with MicroScan® Dried Gram Positive MIC/Combo, Dried Gram Positive Breakpoint Combo and Dried Gram Positive ID Type 2 panels. MicroScan® Positive panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram positive cocci and Listeria monocytogenes. Refer to Limitation of Procedure Section for use with fastidious streptococci.
The MicroScan® Dried Gram-Positive MIC/Combo Panel is used to determine qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative gram-positive cocci.
2. Indication(s) for use:
This submission is indicated for the addition of Cefoxitin Screen at a cefoxitin concentration of 4 µg/mL to the gram positive susceptibility panel for testing Staphylococcus aureus, and Staphylococcus lugdunensis only.
3. Special conditions for use statement(s):
For prescription use only
The Prompt® method of inoculation is an alternate method of inoculation preparation that is supported in the methodology along with the turbidity method. The stationary and log inoculum methods should not be used with this antibiotic.
4. Special instrument requirements:
These panels can be read at ≥ 16 hours of incubation either manually, automatically on the autoScan® 4, or with the WalkAway® instrument systems.
I. Device Description:
The MicroScan® Dried Gram-Positive MIC/Combo Panel contains microdilutions of each antimicrobial agent in various concentrations with Mueller Hinton Broth and various nutrients which are dehydrated and dried in panels. Each panel contains two control wells: a no-growth control well (contains water only/no nutrients or broth), and a growth control well (contains test medium without antibiotic). The panel is rehydrated and inoculated at the same time with 0.1 ml of suspension prepared by the turbidity method (inoculum prepared in water, then 0.1 ml transferred to 25 ml of inoculum water containing pluronic-D/F-a wetting solution) for a final inoculum concentration of 3-7 X 10⁵. The Prompt® method of inoculation is also recommended as an alternate means of preparing the inoculum. The panels are incubated at 35°C in a non-CO₂ for 16-20 hours and read by visual observation of growth. Panels may also be read automatically with the WalkAway® or the AutoScan®4.
J. Substantial Equivalence Information:
{2}
1. Predicate device name(s): MicroScan Dried Gram-Positive and Gram-Negative MIC/Combo Panels
2. Predicate 510(k) number(s): K862140
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | | |
| Intended Use | See above | Same |
| Inoculum preparation | Inoculum prepared from isolated colonies using either the Turbidity method or Prompt® system | Same |
| Technology | Growth based after 16 hours incubation | Same |
| Results | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | Same |
| Instrument | autoScan® -4 or WalkAway® | Same |
| | | |
| Differences | | |
| Item | Device | Predicate |
| Components | Dried Cefoxitin Screen Well (CfxS) 4 ug/mL | Different concentrations depending on the antibiotic |
| Test organism | Gram-Positive organisms | Varies according to the antibiotic |
| Limitations | The performance for Cefoxitin Screen Well has not been established with Stationary and Log Inoculum methods. Inoculum should be prepared with turbidity or Prompt® method. | None |
# K. Standard/Guidance Document Referenced (if applicable):
Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA"; CLSI M7 (M100-S18) "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically;
{3}
Approved Standard".
## L. Test Principle:
The Cefoxitin Screen uses the 18 hour result from a well containing cefoxitin at 4 µg/mL and growth media, labeled CfxS, and the oxacillin MIC at 18 hours. Isolates that are CfxS positive (MIC>4) are considered to be resistant to oxacillin. If the result is negative (MIC≤4), the oxacillin MIC is used to make a final assessment of the susceptibility to penicilline stable beta-lactams.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Reproducibility was demonstrated using 10 isolates tested at 3 sites on 3 separate days in triplicate. Acceptable reproducibility was demonstrated with only category agreement (Negative, Positive) since that is all that is detected. The study included the testing of the following inoculum and reading variables; turbidity inoculum method and Prompt® method of inoculation with reading performed manually using a touchScan® SR, autoScan 4® or the WalkAway® instrument.
#### b. Linearity/assay reportable range:
Not Applicable
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Quality Control was performed daily with the turbidity method, and Prompt® method with the following results and expected ranges as stated. The values repeat the number of times a result was obtained at each concentration.
| Organism | Conc. In ug/mL | Reference result Turbidity inoc. | Turbidity inoculation with Read methods | | | Prompt® inoculation with Read | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Manual | Walk-Away® | Auto-Scan® | Manual | Walk-Away® | Auto-Scan® |
| | | | | | | | | |
| S. aureus ATCC 29213
Exp. Range: <=4 µg/mL | 4 | 112 | 109 | 69 | 71 | 122 | 117 | 124 |
| | 6 | | | | | 1 | 1 | 1 |
| | 8 | | | | | 1 | 1 | 1 |
| | 16 | | | | | | | |
| | >16 | | | | | | 1 | 1 |
| | | | | | | | | |
| | | | | | | | | |
| S. aureus ATCC 43300
Exp. Range: >4 µg/mL | 8 | | | 1 | 1 | 1 | 1 | 1 |
| | 16 | 96 | 105 | 67 | 69 | 76 | 94 | 102 |
| | >16 | 15 | 5 | 1 | | 45 | 23 | 21 |
| | | | | | | | | |
| | | | | | | | | |
{4}
Quality control results demonstrated the ability of all variables of the procedure (reading and inoculation) to produce acceptable results most of the time. Although the modes are the same, the Prompt® inoculation method appears to have more results at MIC of >4 with the S. aureus ATCC 29213. The Prompt® inoculation method with S. aureus ATCC 43300 also had more results at the >16 when compared to the Turbidity method.
Inoculum density control: A turbidity meter was used for the turbidity inoculation method. The Prompt® method of inoculation had colony counts (CC) performed periodically throughout the study to determine the average inoculum density since there is no visual check of the inoculum using this device. The Prompt® method of inoculation had an average of 1.05 x 10⁶ CFU/mL for S. aureus ATCC 43300 with a range of 1.9 x 10⁵ to 2.38 x 10⁶ CFU/mL and an average of 1.82 x 10⁶ CFU/mL for S. aureus ATCC 29213 with a range of 2.7 x 10⁵ to 7.9 x 10⁶ CFU/mL. The CFU study demonstrated that the S. aureus has a higher concentration of organism that reproduces a more resistant result which would explain the more resistant results with the QC study. The inoculum of the Prompt® method generally provides a higher number of CFU with more variability than a method using a turbidity meter.
d. Detection limit: Not Applicable
e. Analytical specificity: Not Applicable
f. Assay cut-off: Not Applicable
2. Comparison studies:
a. Method comparison with predicate device: Clinical testing was performed at three sites using fresh isolates supplemented with stock isolates of gram positive cocci. A comparison of the MicroScan® Dried Gram-Positive test panel results was made to the reference method as recommended in the CLSI standard M7-A7 with the following deviations from that recommendation: Pluronic-F is used as the inoculum in the frozen reference panels. This is composed of water which contains a very small amount of Pluronic to provide a smoother draw of liquid into the inoculator. Testing of the reference method and the MicroScan panels was performed at the same time. A challenge set was also tested at one site and compared to the reference broth dilution result mode that was determined by previous testing of each isolate multiple times in the recommended reference panel. All isolates tested grew in the MicroScan panels.
{5}
Summary Table
| | total | CA N | %CA | # Resistant | # Susceptible | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Efficacy | 306 | 305 | 99.6 | 150 | 156 | 1 | 0 |
| Challenge | 75 | 75 | 100 | 34 | 41 | 0 | 0 |
| Combined | 381 | 380 | 99.7 | 184 | 197 | 1 | 0 |
CA-Category Agreement
maj-major discrepancies
vmj-very major discrepancies
CA is when the interpretation of the reference method agrees exactly with the interpretation of the MicroScan® results.
The challenge set of organisms was also tested using the Prompt® method and turbidity method of inoculation with all reading methods. This included 75 challenge isolates that were tested at two sites. The inoculum was prepared by the turbidity or Prompt® method and incubated in the WalkAway® instrument. All panels had additional readings performed after the WalkAway® reading was completed using the autoScan®-4 and then manually on the touchSCAN®-SR. The table below demonstrates the numbers that were in exact agreement with the reference method results.
| Inoculation method | Read method | No. Tested | CA N | % CA | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- |
| Turbidity | Manual | 75 | 75 | 100 | 0 | 0 |
| Turbidity | WalkAway ® | 75 | 75 | 100 | 0 | 0 |
| Turbidity | autoScan® 4 | 75 | 75 | 100 | 0 | 0 |
| Prompt® | Manual | 75 | 75 | 100 | 0 | 0 |
| Prompt® | WalkAway ® | 75 | 75 | 100 | 0 | 0 |
| Prompt® | autoScan® 4 | 75 | 75 | 100 | 0 | 0 |
b. Matrix comparison: Not Applicable
3. Clinical studies:
a. Clinical Sensitivity: Not Applicable
b. Clinical specificity: Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable
4. Clinical cut-off:
{6}
Not Applicable
5. Expected values/Reference range:
S. aureus and S. lugdunensis ≤ 2 (S), ≥ 4 R
**N. Proposed Labeling:**
The expected value range, interpretive criteria and QC are included in the package insert. The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.
**O. Conclusion:**
The information submitted in this premarket notification is complete and supports a substantial equivalence decision.
7
{7}
8
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.