K Number
K020160

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-03-14

(56 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MicroScan® Dried Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative Gram-Positive cocci. After inoculation, panels are incubated for 16 - 20 hours at 35°C +/- 1°C in a non-CO2 incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert.

This particular submission is for a revised QC range of ≥ 0.5 mcg/ml for S. aureus ATCC #29213 with Ampicillin.

The Gram-Positive organisms which may be used for Ampicillin susceptibility testing in this panel are:

Staphylococcus aureus (beta-lactamase and non-beta-lactamase producing) Staphylococcus epidermidis (beta-lactamase and non-beta-lactamase producing) Staphyloccus saprophyticus (beta-lactamase and non-beta-lactamase producing) Streptococcus faecalis (Enterococcus) Streptococcus pyogenes

The MicroScan® Dried Gram-Positive MIC/Combo Panels with Ampicillin are not intended for use with Streptococcus pneumoniae and viridans streptococci.

Device Description

MicroScan® Dried Gram-Positive MIC/Combo Panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative Gram-Positive cocci.

The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in broth and dehydrated. Various antimicrobial agents are diluted in broth to concentrations bridging the range of clinical interest. Panels are rehydrated with water, after inoculation with a standardized suspension of the organism. After incubation in a non-CO, incubator for 16-20 hours, the minimum inhibitory concentration (MIC) for the test organism is read by determining the lowest antimicrobial concentration showing inhibition of growth.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the MicroScan® Dried Gram-Positive MIC/Combo Panel, based on the provided documents:

The 510(k) submission (K020140, later referenced as K020160) is specifically for a revised Quality Control (QC) range for Ampicillin with S. aureus ATCC 29213 on an already marketed device, the MicroScan® Dried Gram-Positive MIC/Combo Panel. Therefore, the information primarily focuses on validating this specific QC range change rather than the entire device's performance.

1. A table of acceptance criteria and the reported device performance

The document states that the proposed MicroScan® Dried Gram-Positive MIC/Combo Panel demonstrated "substantially equivalent performance when compared with an NCCLS frozen Reference Panel, as defined in the FDA DRAFT document 'Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices', dated March 8, 2000."

While specific numerical acceptance criteria for the revised QC range are not explicitly detailed in the provided text, the standard for antimicrobial susceptibility devices typically involves:

  • Essential Agreement (EA): Agreement within +/- 1 two-fold dilution between the test device and the reference method.
  • Categorical Agreement (CA): Agreement in the interpretive category (Susceptible, Intermediate, Resistant) between the test device and the reference method.
  • Discrepancy Rates: Acceptable rates for Very Major Errors (VME), Major Errors (ME), and Minor Errors (mE). VME occurs when a resistant isolate is reported as susceptible. ME occurs when a susceptible isolate is reported as resistant. mE occurs when an intermediate isolate is reported as susceptible or resistant, or vice versa.
  • Quality Control (QC) Range Agreement: The QC results must fall within the established acceptable range.

Given this submission is for a revised QC range, the acceptance criteria would primarily revolve around validating the new range for Ampicillin with S. aureus ATCC 29213. This typically means that a high percentage of QC tests performed with the new range must fall within the specified bounds, demonstrating adequate reproducibility and accuracy for internal quality control.

Acceptance Criteria (Inferred from regulatory guidance for AST devices)Reported Device Performance (Implied for the revised QC range)
For QC Range Validation:
QC results for S. aureus ATCC 29213 with Ampicillin fall within the new range of ≥ 0.5 mcg/ml consistently."Substantially equivalent performance" to the NCCLS frozen Reference Panel, supporting the revised QC range. (Specific numerical data are not provided in the summary.)
(General performance for entire panel, though not the primary focus of this submission)"Substantially equivalent performance" to the NCCLS frozen Reference Panel for the broader panel.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: The document does not specify the exact sample size for the test set used to validate the revised QC range. It only broadly refers to "data in support of a revised Quality Control range." For AST devices, test sets typically involve a significant number of clinical isolates and challenge organisms. However, for a QC range revision, the sample size would focus on repeated testing of the S. aureus ATCC 29213 strain.
  • Data Provenance: Not specified in the provided text. It's likely a controlled laboratory study, but the country of origin or whether it was retrospective/prospective is not mentioned.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided. For antimicrobial susceptibility testing, the "ground truth" (or reference method) is typically established by well-defined reference broth microdilution methods, often performed by trained microbiologists following standardized protocols, not individual "experts" in the same sense as image interpretation. The document mentions comparison to "an NCCLS frozen Reference Panel," which implicitly establishes the reference method.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable and not provided. Adjudication methods like 2+1 are used for resolving disagreements among human readers in studies, particularly in image interpretation, where subjective decisions are made. For a quantitative test like MIC determination, the NCCLS (now CLSI) reference method provides the objective "ground truth" against which the device is measured.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is a diagnostic an in vitro diagnostic (IVD) device for antimicrobial susceptibility testing, not an AI-assisted diagnostic tool that aids human readers in interpretation. Therefore, an MRMC study comparing human reader performance with and without AI assistance is not relevant here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device is an automated or semi-automated system for determining Minimum Inhibitory Concentrations (MICs). While it can be read visually, the "MicroScan instrumentation" implies an automated reading component. The "substantially equivalent performance" claim would inherently reflect the standalone performance of the device (or its automated reading capabilities if used) compared to the reference standard. The study validates the device's ability to accurately determine MICs, which is its standalone function.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth used is an NCCLS frozen Reference Panel. This implies that the reference method is based on the National Committee for Clinical Laboratory Standards (NCCLS, now Clinical and Laboratory Standards Institute - CLSI) standard broth microdilution method, which is the recognized gold standard for antimicrobial susceptibility testing. The frozen reference panels are carefully prepared and verified to ensure accurate MICs.

8. The sample size for the training set

The document does not provide information about a "training set." For this type of IVD device, the development process would involve internal validation and optimization, but the 510(k) summary focuses on the performance of the finalized device (the "test set" in broader terms) against a recognized reference method. It's less about training a machine learning model and more about validating the analytical performance of a chemical/biological assay.

9. How the ground truth for the training set was established

As there's no explicit mention of a "training set" in the context of machine learning, this question is not directly applicable. For the overall development of the MicroScan panels, the "ground truth" for calibrating and developing the various antimicrobial concentrations and interpretive criteria would have been established through extensive studies using standard reference methods (like NCCLS/CLSI broth microdilution) over many years, against a wide range of bacterial isolates.

{0}------------------------------------------------

MAR 1 4 2002

K020140

510(k) Summary

510(k) Submission Information:

Device Manufacturer:Dade MicroScan Inc.
Contact name:Cynthia Van Duker, Regulatory Affairs Manager
Fax:916-374-3144
Date prepared:January 16, 2002
Product Name:Microdilution Minimum Inhibitory Concentration (MIC) Panel
Trade Name:MicroScan® Dried Gram-Positive MIC/Combo Panels
Intended Use:To determine antimicrobial agent susceptibility
510(k) Notification:New QC range - Ampicillin
Predicate device:MicroScan Dried Gram Positive MIC/Combo Panels

510(k) Summary:

MicroScan® Dried Gram-Positive MIC/Combo Panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative Gram-Positive cocci.

The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in broth and dehydrated. Various antimicrobial agents are diluted in broth to concentrations bridging the range of clinical interest. Panels are rehydrated with water, after inoculation with a standardized suspension of the organism. After incubation in a non-CO, incubator for 16-20 hours, the minimum inhibitory concentration (MIC) for the test organism is read by determining the lowest antimicrobial concentration showing inhibition of growth.

The proposed MicroScan® Dried Gram-Positive MIC/Combo Panel demonstrated substantially equivalent performance when compared with an NCCLS frozen Reference Panel, as defined in the FDA DRAFT document "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", dated March 8, 2000. The Premarket Notification (510[k]) presents data in support of a revised Quality Control range for Ampicillin with S. aureus ATCC 29213 on the MicroScan® Dried Gram-Positive MIC/Combo Panel.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The text is written in all capital letters.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAR 1 4 2002

Ms. Cynthia Van Duker Regulatory Affairs Manager Dade Behring Inc. 1584 Enterprise Boulevard West Sacramento, CA 95691

K020160 Re:

Trade/Device Name: MicroScan® Dried Gram-Positive MIC/Combo Panels with Ampicillin (0.008-128 mcg/ml) Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: Class II Product Code: LTT, JWY Dated: January 16, 2002 Received: January 17, 2002

Dear Ms. Van Duker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

INDICATIONS FOR USE STATEMENT

Page 1 of 1 .

510(k) Number (if known): K O2O l 00

Device Name: MicroScan® Dried Gram-Positive MIC/Combo Panels with Ampicillin (0.008 - 128 mcg/ml)

Indications For Use:

The MicroScan® Dried Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative Gram-Positive cocci. After inoculation, panels are incubated for 16 - 20 hours at 35°C +/- 1°C in a non-CO2 incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert.

This particular submission is for a revised QC range of ≥ 0.5 mcg/ml for S. aureus ATCC #29213 with Ampicillin.

The Gram-Positive organisms which may be used for Ampicillin susceptibility testing in this panel are:

Staphylococcus aureus (beta-lactamase and non-beta-lactamase producing) Staphylococcus epidermidis (beta-lactamase and non-beta-lactamase producing) Staphyloccus saprophyticus (beta-lactamase and non-beta-lactamase producing) Streptococcus faecalis (Enterococcus) Streptococcus pyogenes

The MicroScan® Dried Gram-Positive MIC/Combo Panels with Ampicillin are not intended for use with Streptococcus pneumoniae and viridans streptococci.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

OR

510(k) Number k02 0160

Prescription Use J (Per 21 CFR 801.109)

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).