TC Gloves

K221185 · Thanh Cong Pharmaceutical and Trading Company Limited · LYY · Jun 22, 2022 · General Hospital

Device Facts

Record IDK221185
Device NameTC Gloves
ApplicantThanh Cong Pharmaceutical and Trading Company Limited
Product CodeLYY · General Hospital
Decision DateJun 22, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

TC Gloves Powder-Free Latex Examination Gloves is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Story

Disposable, non-sterile, ambidextrous, powder-free natural rubber latex examination gloves; finger-textured surface. Used by healthcare examiners to provide a physical barrier against contamination between patient and examiner. Device is worn on the hand; no electronic or mechanical components. Performance verified against ASTM D3578-19 standards for dimensions, physical properties (tensile strength, elongation), and barrier integrity (AQL 2.5). Biocompatibility confirmed via ISO 10993-1:2018 testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity).

Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM D3578-19 (dimensions, physical properties), ASTM D5151-19 (barrier integrity/AQL 2.5), ASTM D6124-06 (powder residue), and ASTM D5712-15 (extractable protein). Biocompatibility evaluated per ISO 10993-1:2018, confirming non-cytotoxic, non-sensitizing, non-irritating, non-systemically toxic, and non-pyrogenic properties.

Technological Characteristics

Material: Natural rubber latex. Form factor: Ambidextrous, finger-textured, powder-free. Dimensions: Length min 230mm, palm/finger thickness min 0.08mm. Standards: ASTM D3578-19 (rubber examination gloves), ASTM D5151-19 (holes), ASTM D6124-06 (powder residue), ASTM D5712-15 (protein). Biocompatibility: ISO 10993-1:2018. Sterility: Non-sterile.

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand to prevent cross-contamination between patient and examiner. Suitable for over-the-counter use.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it. June 22, 2022 Thanh Cong Pharmaceutical and Trading Company Limited % Manoj Zacharias US Agent Liberty Management Group Limited 75 Executive Drive, Suite 114 Aurora, Illinois 60504 Re: K221185 Trade/Device Name: TC Gloves Powder Free Latex Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LYY Dated: March 29, 2022 Received: April 25, 2022 Dear Manoj Zacharias: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K221185 Device Name TC Gloves Powder Free Latex Examination Gloves Indications for Use (Describe) TC Gloves Powder Free Latex Examination Gloves is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for TC Pharma. The logo features three green leaves sprouting upwards, symbolizing growth and nature. The text "TC Pharma" is written in green, with "TC" stacked above "Pharma", and the entire logo is encircled by two curved green lines, creating a sense of completeness and harmony. # 510(k) SUMMARY K221185 [AS REQUIRED BY 21CFR807.92] ### I. SUBMITTER DETAILS | 510(k) Owner's Name | THANH CONG PHARMACEUTICAL AND TRADING COMPANY<br>LIMITED | |--------------------------|---------------------------------------------------------------------------------------| | Address | KM 6+ 200 National Road No.38, Hap Linh, Bac Ninh City, Bac<br>Ninh province, Vietnam | | Contact person | Ms. Nguyễn Thị Tú Anh | | Contact Designation | Director of Production | | Contact Phone Number | 098 171 2992 | | Contact Email | tuanhnquyentcpharma@gmail.com | | Date of Summary Prepared | 07-04-2022 | ### II. DEVICE DETAILS | Brand Name | TC Gloves | |----------------------------|----------------------------------------| | Device Common Name | Powder-Free Latex Examination Gloves | | Device Classification name | Non-powdered patient examination glove | | Regulation Number | 21 CFR 880.6250 | | Class | I | | Product Code | LYY | ## III. PREDICATE DEVICE DETAILS | Predicate Device Name | Palm Care Latex Examination Powder Free Gloves | |-----------------------|------------------------------------------------| | 510(k) Number | K202377 | | Regulation Number | 21 CFR 880.6250 | | Class | I | | Product Code | LYY | # Manufacturer: THANH CONG PHARMACEUTICAL AND TRADING CO., LTD Km 6 + 200 National Road No.38, Hap Linh, Bac Ninh City, Bac Ninh province, Vietnam Tel: (+84) 222 3720031, (+84) 24 38563948 Email: support@tcpharma.vn * Website: http://www.tcpharma.vn {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for TC Pharma. The logo features two green leaves above the text "TC Pharma" in green font. The logo is surrounded by a gray, curved shape. ## IV. DEVICE DESCRIPTION TC Gloves Powder Free Latex Examination Gloves is a Class I device bearing the product code LYY (21CFR 880.6250). They meet all the current specifications listed under the ASTM D3578 - 19, Standard Specification for Rubber Examination Gloves. They are made from natural rubber latex. These gloves are white in color having Finger Texture / Ambidextrous and are powder free. The product is non-sterile. #### V. INDICATION FOR USE TC Gloves Powder-Free Latex Examination Gloves is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. # VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE | SI.<br>No | Features compared | Proposed Device | Predicate Device | Result | | | |-----------|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-------------|--| | | General Information | | | | | | | 1. | 510(k) Number | K221185 | K202377 | - | | | | 2. | Manufacturer | Thanh Cong Pharmaceutical And<br>Trading Company Limited | Hi-Care Thai Gloves Co. Ltd. | - | | | | 3. | Classification | I | I | Same | | | | 4. | Regulation number | 21 CFR 880.6250 | 21 CFR 880.6250 | Same | | | | 5. | Product Code | LYY | LYY | Same | | | | 6. | Indication For Use | TC Gloves Powder-Free Latex<br>Examination<br>Gloves are<br>disposable devices intended for<br>medical purpose that are worn on<br>the examiner's hand to prevent<br>contamination between patient<br>and examiner. | Palm Care Latex Examination<br>Powder Free Gloves are<br>disposable devices intended for<br>medical purpose that are worn on<br>the examiner's hand to prevent<br>contamination between patient<br>and examiner. | Same | | | | 7. | Material | Latex | Latex | Same | | | | 8. | Color | White | Natural (No color is added) | Different-1 | | | | 9. | Texture | Finger Texture | Finger texture | Same | | | | 10. | Ambidextrous | Yes | Yes | Same | | | | 11. | Size | S, M, L, XL | XS, S, M, L | Similar-1 | | | | SI.<br>No | Features compared | Proposed Device | Predicate Device | Result | | | | 12. | OTC Use | Yes | Yes | Same | | | | 13. | Reusability | Single use | Single use | Same | | | | 14. | Sterility | Non- sterile | Non- sterile | Same | | | | 15. | Dimensions | Length Min 230 mm<br>Width Min 95+/-10 mm (for<br>medium size) | Length Min 230 mm<br>Width Min 95+/-10 mm (for medium size) | Same | | | | 16. | Thickness | Palm min 0.08 mm<br>Finger min 0.08 mm | Palm min 0.08 mm<br>Finger min 0.08 mm | Same | | | | 17. | Physical Properties | Before Aging<br>Tensile Strength min 18 Mpa<br>Ultimate Elongation Min 650%<br>After Aging<br>Tensile Strength min 14 Mpa<br>Ultimate Elongation Min 500% | Before Aging<br>Tensile Strength min 18 Mpa<br>Ultimate Elongation Min 650%<br>After Aging<br>Tensile Strength min 14 Mpa<br>Ultimate Elongation Min 500% | Same | | | | 18. | Detection of Holes | Passes AQL 2.5 | Passes AQL 1.5 | Different-2 | | | | 19. | Powder Free Residue | ≤2 mg/glove | ≤2 mg/glove | same | | | | 20. | Extractable Protein<br>Testing | 200 μg/ dm² max | 200 μg/ dm² max | same | | | | 21. | Biocompatibility Study | Invitro Cytotoxicity | Under the conditions of the study, cytotoxic to L-929 cells. Additional Testing was performed to determine if this was a systemic toxicity concern. | Under the conditions of the study, non-cytotoxic. | Different-3 | | | | | Skin Sensitization | Under the conditions of the study not a sensitizer | Under the conditions of the study not a sensitizer | same | | | | | Skin Irritation | Under the condition of study not an irritant | Under the condition of study not an irritant | same | | | | | Acute Systemic<br>Toxicity | Under the condition of study, the device extracts do not pose a systemic toxicity. | Under the condition of study, the device extracts do not pose a systemic toxicity. | same | | | | | Material Mediated<br>Pyrogenicity | Under the conditions of the study, the device did not demonstrate a material mediated Pyrogenicity response. | Under the conditions of the study, the device did not demonstrate a material mediated Pyrogenicity response. | same | | ## Table 1: General Comparison {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for TC Pharma. The logo features two green leaves above the text "TC Pharma" in green font. A grey swoosh design is behind the leaves and text. There are no significant differences between the two products and are identical in terms of intended use, materials, design, manufacturing methods. Both devices met the performance standards. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for TC Pharma. The logo features two green leaves above the text "TC Pharma" in green font. The leaves are positioned in the center of the logo, and the text is placed below the leaves. A gray curved line surrounds the leaves and text, creating a circular shape. ## VII. PERFORMANCE DATA #### Non- Clinical Data A. ## Performance Tests TC Gloves Powder-Free Latex Examination Gloves is subjected to the following performance tests according to the requirements of Guidance for Industry and FDA Staff - Medical Glove Guidance Manual and found to be safe and efficient with respect to its intended use: - Dimension . - . Physical property - . Barrier property tests - A Detection of Holes in Medical Gloves - . Powder Free Residue - Extractable protein testing ● ### Table 2: Performance Testing Summary | SI<br>No. | Tests | Proposed Device actual Data | | | Acceptance Criteria | | | Result | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|---------------------|-----------------|-----------------------------|-----------------------------|-----------------|--------| | 1. | Dimension<br>Length, Width and<br>Thickness<br>ASTM D3578 - 19,<br>Standard Specification<br>for Rubber Examination<br>Gloves | Size | Length | Width | Size | Length | Width | Pass | | | | S | 240.9mm | 85mm | S | 230mm min | 80mm ±10 | | | | | M | 240.6mm | 95.8mm | M | | 95mm ±10 | | | | | L | 240.7mm | 104.5mm | L | | 111mm ±10 | | | | | XL | 240.3mm | 113.5mm | XL | | 120mm ±10 | | | Thickness | | | Thickness | | | | | | | Size | Palm | Finger | Size | Palm | Finger | | | | | S | 0.13mm | 0.15mm | S | 0.08 mm min | 0.08 mm<br>min | | | | | M | 0.14mm | 0.15mm | M | | | | | | | L | 0.14mm | 0.15mm | L | | | | | | | XL | 0.13mm | 0.15mm | XL | | | | | | | 2. | | Tensile strength | | | Tensile strength | | | Pass | | | Physical property<br>Tensile strength<br>and Ultimate<br>Elongation<br>ASTM D3578 - 19,<br>Standard Specification<br>for Rubber Examination<br>Gloves | Size | Before<br>ageing | After<br>ageing | Size | Before<br>ageing | After<br>ageing | | | | S | 19.9 Mpa | 19.4 Mpa | S | 18 Mpa Min<br>for all sizes | 14 Mpa Min<br>for all sizes | | | | | M | 19.8 Mpa | 19.4 Mpa | M | | | | | | | L | 20.0 Mpa | 19.3 Mpa | L | | | | | | XL | 20.0 Mpa | 19.2 Mpa | XL | | | | | | | Ultimate elongation | | | Ultimate elongation | | | | | | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for TC Pharma. The logo features two green leaves above the text "TC Pharma" in green font. There are two gray curved lines that surround the leaves and text. | SI<br>No. | Tests | Proposed Device actual Data | | | Acceptance Criteria | | | Result | |-----------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|--------------------------------|-----------------|---------------------|--------------------------------|---------------------------|--------| | | | Size | Before<br>ageing | After<br>ageing | Size | Before<br>ageing | After<br>ageing | | | | | S | 805.5% | 790.5% | S | | | | | | | M | 803.1% | 794.9% | M | 650% Min for<br>all sizes | 500% Min for<br>all sizes | | | | | L | 804.6% | 791% | L | | | | | | | XL | 800.9% | 794.7% | XL | | | | | 3. | Detection of Holes<br>in Medical Gloves<br>ASTM D5151-19,<br>Standard Test Method<br>for Detection of Holes<br>in Medical Gloves | Size | AQL 2.5 | | Size | AQL 2.5 | | Pass | | | | S | | | S | | | | | | | M | | | M | | | | | | | L | | | L | | | | | | XL | XL | | | | | | | | 4. | Powder Free<br>Residue<br>ASTM D6124-06<br>(Reapproved 2017)<br>Standard Test Method<br>for Residual Powder on<br>Medical Gloves | Size | Residual powder content | | Size | Residual powder content | | | | | | S | 0.31 mg | | S | ≤2 mg/Glove Max | | | | | | M | 0.42 mg | | M | | | Pass | | | | L | 0.57 mg | | L | | | | | | XL | 0.40 mg | | XL | | | | | | 5. | Extractable<br>Protein Testing<br>ASTM D5712-15<br>Standard Test Method<br>for Analysis of Aqueous<br>Extractable Protein in<br>Latex, Natural Rubber,<br>and Elastomeric<br>Products Using the<br>Modified Lowry Method | Size | Extractable protein<br>content | | Size | Extractable protein<br>content | | | | | | S | 47.96 (µg/dm²) | | S | | | | | | | M | 50.0 (µg/dm²) | | M | | | | | | | L | 34.82 (µg/dm²) | | L | 200 µg/ dm² max | | Pass | | | | XL | 49.27 (µg/dm²) | | XL | | | | #### B. Biocompatibility The materials used in the TC Gloves Powder-Free Latex Examination Gloves are biocompatible based on the biocompatibility tests mentioned in the Guidance for Industry and FDA Staff - Medical Glove Guidance Manual: - Invitro Cytotoxicity - . Skin Sensitization - . Skin Irritation - Acute Systemic Toxicity - . Material-Mediated Pyrogenicity These tests are performed according to ISO 10993-1:2018, Biological Evaluation of Medical Devices - Part 1, Evaluation and Testing within a Risk Management Process. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for TC Pharma. The logo features two green leaves above the text "TC Pharma". The leaves are arranged in a V-shape, with one leaf slightly overlapping the other. The text "TC Pharma" is in a bold, sans-serif font. The logo is surrounded by two curved, gray lines that form a partial oval shape. | SI.<br>No | Test Performed | Proposed Device | Acceptance Criteria | Result | |-----------|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------| | 1. | Invitro Cytotoxicity | Under the conditions of the study,<br>cytotoxic to L-929 cells. Additional<br>Testing was performed to<br>determine if this was a systemic<br>toxicity concern | Under the conditions of the<br>study, non-cytotoxic. | - | | 2. | Skin Sensitization | Under the conditions of the study<br>not a sensitizer | Under the conditions of the study<br>not a sensitizer | Pass | | 3. | Skin Irritation | Under the condition of study not<br>an irritant | Under the condition of study not<br>an irritant | Pass | | 4. | Acute Systemic Toxicity | Under the condition of study, the<br>device extracts do not pose a<br>systemic toxicity. | Under the condition of study, the<br>device extracts do not pose a<br>systemic toxicity. | Pass | | 5. | Material-Mediated<br>Pyrogenicity | Under the conditions of the study,<br>the device did not demonstrate a<br>material mediated Pyrogenicity<br>response. | Under the conditions of the<br>study, the device did not<br>demonstrate a material<br>mediated Pyrogenicity<br>response. | Pass | ## Table 3: Biocompatibility Test Summary ### C. Clinical Test Data Clinical study was not conducted as clinical data is not needed for TC Gloves Powder-Free Latex Examination Gloves. ## VIII. CONCLUSION The conclusion drawn from the non-clinical tests demonstrate that the subject device, TC Gloves Powder-Free Latex Examination Gloves are as safe, as effective and perform as well as or better than legally marketed predicated device in K202377.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...