K Number
K111133

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-06-24

(63 days)

Product Code
Regulation Number
880.6250
Age Range
18 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Non-Sterile, powder-free latex examination gloves with protein claim (50 micrograms or less).

AI/ML Overview

This looks like a 510(k) premarket notification for non-sterile, powder-free latex examination gloves, which typically doesn't involve complex algorithms or AI. The acceptance criteria and "study" described here refer to conformance with established performance standards for medical gloves, rather than a clinical trial or AI model validation.

Therefore, many of the typical AI/ML study components (like sample sizes for test/training sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, etc.) are not applicable to this type of device submission.

Here's a breakdown of the available information relating to acceptance criteria and device performance:

1. Table of Acceptance Criteria and Reported Device Performance

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCE
Dimensions: overall length; width, palm and finger thicknessASTM D 3578-2005 (reapproved 2010)Meets
Tensile strength: before and after agingASTM D 3578-2005 (reapproved 2010)Meets
Ultimate elongation: before and after agingASTM D 3578-2005 (reapproved 2010)Meets
Freedom from holes: pinholes AQL 2.5ASTM D 3578-2005 (reapproved 2010)Meets
Powder Free ResidueASTM D 3578-2005 (reapproved 2010)Meets
Protein LevelASTM D 3578-2005 (reapproved 2010)Meets
Biocompatability
Primary Skin Irritation in RabbitsISO 10993-10/2002Passes
Guinea Pig SensitizationISO 10993-10/2002Passes

2. Sample size used for the test set and the data provenance

  • Sample Size: Not explicitly stated in terms of a numerical count of individual gloves tested. The submission refers to meeting standards like "AQL 2.5" for pinholes, which implies a sampling plan defined by ASTM D3578. However, the specific number of gloves sampled for each test is not provided in this summary.
  • Data Provenance: The testing was performed by "Siam Sempermed Laboratories in Thailand" for physical properties and "Consumer Product Testing" in Fairfield, New Jersey, USA for biocompatibility. This suggests a combination of in-house (Thailand) and third-party (USA) testing. The data is prospective in the sense that the manufacturer tested their actual product to demonstrate compliance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This is not applicable. The "ground truth" for glove performance is defined by the objective, quantifiable metrics within the referenced ASTM and ISO consensus standards (e.g., specific dimensions, tensile strength, protein limits, pinhole AQL). There are no "experts" establishing ground truth in the way a clinical study would for medical imaging.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. Adjudication methods are relevant for subjective interpretations, whereas these tests involve objective measurements against predefined thresholds in standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a physical medical glove, not an AI or imaging device that uses "human readers" or "AI assistance."

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not applicable. This is not an algorithmic or AI device. The performance tests are for the physical properties of the glove itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" is established by consensus standards (ASTM D3578-05, ISO 10993-10/2002) which specify measurable physical, chemical, and biological properties that the gloves must meet for their intended use. These standards are developed through a consensus process involving experts in the field.

8. The sample size for the training set

  • Not applicable. There is no "training set" as this is not an AI/ML device.

9. How the ground truth for the training set was established

  • Not applicable.

{0}------------------------------------------------

510 (k) Summary

As Required by 21 section 807.92 ( c)

KIII133

    1. Submitter Name: SSC Surat Thani
      JUN 2 4 2011
  • 189 , Moo 7, Phaiwat , Kanjanadit, 2. Address: Suratthani, 84160, THAILAND

  • (+66) 77 277888 3. Phone:

  • (+66) 77 277889 4. Fax:

  1. Contract Person: Mrs. Sureerat Choosri (Quality manager)

  2. Date summary prepared: 1 Feb , 2011

Official Correspondent: Sempermed USA Inc. 7.

  • 13900 49th Street North 8. Address: Clearwater, USA , FL 33762
  • 727 787 7250 9. Phone:
  • 727 787 7558 10. Fax:
  • Mr. William E. Harris 11. Contact person:
  1. Device Trade or Proprietary Name: Non-sterile, powder-free latex examination gloves with protein claim (50 micrograms or less).

  2. Device Common or usual name: Examination glove

  3. Device Classification Name: Glove , Patient Examination , Latex

  4. Description of the Device:

  5. Description of the Device:
    Non-Sterile, powder-free latex examination gloves with protein claim (50 micrograms or less). 16. Indications for use of the device: Based on 21 CFR Part 880.6250: "Patient examination glove." A patient examination glove is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

  6. Summary of The Technological Characteristics of The devices :(According Guidance for Industry and FDA Staff - Medical Glove Guidance Manual(January 22, 2008))

Non-Sterile, powder-free latex examination gloves with protein claim (50 micrograms or less) are summarized with the following technological characteristics:

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCE
Dimensions: overall length; width, palmand finger thicknessASTM D 3578-2005(reapproved 2010)Meets
Tensile strength: before and after agingASTM D 3578-2005(reapproved 2010)Meets
Ultimate elongation: before and afteragingASTM D 3578-2005(reapproved 2010)Meets
Freedom from holes: pinholes AQL 2.5ASTM D 3578-2005(reapproved 2010)Meets
Powder Free ResidueASTM D 3578-2005(reapproved 2010)Meets
Protein LevelASTM D 3578-2005(reapproved 2010)Meets

{1}------------------------------------------------

K 111133

BiocompatabilityPrimary Skin Irritation inRabbitsGuinea Pig SensitizationPasses
Passes

18. Substantial Equivalents Based on Assessment of Non-Clinical Performance Data

Stantial Equivalents Dased of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above.

19. Conclusion

It can be concluded that Non-Sterile, powder-free latex examination gloves with protein casting If can be concinded that Noll-Sterlic, powere not according to the glove performance standards referenced in section (50 micrograms of less) will perform according to the grows personemaly, this device is substantially I / above and meet ASTM standards, and ITA requirements. Sonse two predicate device Sizm equivalent to currently dialketed devices. I ms conver free with protein claim (0) micrograms or less).
Sempermed Latex Patient Examination Glove , Powder free with protein c Sempermed Lates I anem Exammation Giover expressed in the tabulated comparison as below.

Technical comparison of specific elements is attached in the main submission.

FDA file reference number510k number : K100907
Attachments inside notification submission fileREFER TO APPENDIX 1
TECHNOLOGICAL CHARACTERISTICSComparison result
Indications for useIdentical
Target populationIdentical
DesignIdentical
MaterialsIdentical
PerformanceIdentical
SterilityNot applicable
BiocompatibilityIdentical
Mechanical safetyIdentical
Chemical safetyIdentical
Anatomical sitesIdentical
Human factorsIdentical
Energy used and/or deliveredIdentical (Not applicable)
Compatibility with environment and other devicesIdentical
Where usedIdentical
Standards metIdentical
Electrical safetyIdentical (Not applicable)
Thermal safetyIdentical (Not applicable)
Radiation safetyIdentical (Not applicable)

{2}------------------------------------------------

K 111133 SSC Surat Thani
DECLARATION OF CONFORMITY
This is to declare and to confirm that SSC Surat Thani conforms in its manufacturing of Non-Sterile, powder-free latex examination gloves with protein claim (50 micrograms or less) to thefollowing recognized standards, latest released revisions.
Standards met and standards tested againstInapplicablerequirementsor deviationsIdentification of anyway(s) in which thestandard wasadapted for theapplication of thedevice, i.e.,identification of analternative series oftests that wereperformedSpecification ofany deviationsfrom eachapplicablestandardSpecification of the differencesthat may exist between thetested device and the device tobe marketed and justificationof the test resultsName and address of any test laboratory orcertification body involved in determining theconformance of the device with the standardand reference to any accreditation of thoseorganization
Gloves:
ASTM D 3578-05 (Reapproved 2010)
Dimensions: overall length, width, palm and fingerthickness: ASTM D3578-05(Reapproved 2010)Tested according to ASTM D3767-03(reapproved2008)Not applicableSiam Sempermed Laboratories in Thailand
Tensile strength & Ultimate elongation: before andafter aging: ASTM D3578-05 (Reapproved 2010)Tested according to ASTM D 412-2006ae 2 andASTM D 573-04 (Reapproved 2010)Not applicableSiam Sempermed Laboratories in Thailand
Donning support: ASTM D 3578-05 (Reapproved2010) : Powder Free residue: Tested according toASTM D6124-06Not applicableSiam Sempermed Laboratories in Thailand
Water soluble Proteins: ASTM D 3578-05(Reapproved 2010)Tested according to ASTM D5712-2010Not applicableSiam Sempermed Laboratories in Thailand
Freedom from holes: ASTM D 3578-05(Reapproved 2010)Test according to ASTM D5151-06Not applicableSiam Sempermed Laboratories in Thailand
Sampling inspection: ISO 2859-1:1999/Cor1:2001
Packages storage: ISO 2230-2002Not applicable
Supervision of product and design: QSR, MDDdirectives 93/42Not applicable
Biocompatibility:ISO 10993-10/2002Consumer Product Testing ,70 New DutchLane , Fairfield, New Jersey , 07004-2514Tel : 973 808 7111 Fax :973 808 7234
Name, Signature and Position

Swank

  1. B

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. William E. Harris President & CEO SSC Surat Thani 13900 49th Street North Clearwater, Florida 33762

JUN 2 4 2011

Re: K111133

Trade/Device Name: Non-Sterile Powder-Free Latex Examination Gloves with Protein Claim (50 Micrograms or Less) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: June 9, 2011 Received: June 9, 2011

Dear Mr. Harris:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.\

Sincerely vours,

Anthony D. mahr

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Non-Sterile, Powder-Free Latex Examination Glove with Protein Claim (50 micrograms or less).

Indications For Use: A patient examination glove is a disposable device intended for A patient oxamiles that is worn on the examiner's hand or finger medical purposes that on between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

CZ

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Elaine S. Mashall for Elizabeth Clavenie willians
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
Page 1 of 1

510(k) Number:K111133
-------------------------

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.