MODIFIED PEDIKAIR

K954320 · Kinetic Concepts, Inc. · FNM · Jul 18, 1997 · General Hospital

Device Facts

Record IDK954320
Device NameMODIFIED PEDIKAIR
ApplicantKinetic Concepts, Inc.
Product CodeFNM · General Hospital
Decision DateJul 18, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5550
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

Modified PediKair is a critical care bed with an air mattress for supporting bedridden pediatric patients in a manner that helps prevent and treat complications of immobility such as skin breakdown and decubitus ulcers.

Device Story

Modified PediKair is a critical care bed system for pediatric patients. It features a low-pressure air mattress with transverse inflatable cushions for pulsation therapy, an air bladder for vibration/percussion therapy, and lateral bladders for automated patient rotation (up to 40°). Inflatable side cushions and collapsible rails ensure patient safety during rotation. Operated in clinical settings, the device aims to reduce pressure below capillary levels (25-35 mmHg) to prevent decubitus ulcers and mitigate pulmonary complications. It allows immediate access for CPR without removing the patient.

Clinical Evidence

No clinical data provided; substantial equivalence is based on design and technological similarities to the predicate device.

Technological Characteristics

Low-pressure air mattress system; transverse inflatable cushions for pulsation; air bladder for vibration/percussion; lateral bladders for rotation; collapsible side rails; removable head/foot rails. Operates via pneumatic inflation/deflation cycles to manage pressure and patient positioning.

Indications for Use

Indicated for bedridden pediatric patients requiring support to prevent or treat complications of immobility, including skin breakdown, decubitus ulcers, and pulmonary complications.

Regulatory Classification

Identification

An alternating pressure air flotation mattress is a device intended for medical purposes that consists of a mattress with multiple air cells that can be filled and emptied in an alternating pattern by an associated control unit to provide regular, frequent, and automatic changes in the distribution of body pressure. The device is used to prevent and treat decubitus ulcers (bed sores).

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} CONFIDENTIAL K 954320 JUL 18 1997 # Summary ## Introduction Modified PediKair is a critical care bed with an air mattress for supporting bedridden pediatric patients in a manner that helps prevent and treat complications of immobility such as skin breakdown and decubitus ulcers. The air mattress consists of a series of transverse, inflatable cushions that can be alternately inflated and deflated to provide pulsation therapy (as on the predicate device), an air bladder beneath the patient's chest region to provide vibration/percussion therapy, and left and right turning bladders to rotate the patient side to side. Inflatable support cushions on each side of the patient's head and body keep the patient positioned properly during rotation. PediKair includes the collapsible side rails and removable head and foot rails of the predicate device for maximum and immediate access to the patient—CPR can be performed without the hindrance of rails and without removing the patient from the bed. ## Summary As with the predicate device, the re-designed PediKair mattress is also a low-pressure air mattress. Low-pressure support mattresses are effective in preventing or reducing the incidence of decubitus ulcers in immobile patients by lowering the pressure on any portion of the body to below capillary pressure (25-35 mmHg). Modified PediKair also provides kinetic therapy, which is effective in the care management and prevention of decubitus ulcers and in the treatment or mitigation of pulmonary complications caused by prolonged immobility. A continuously rotating surface, similar in concept and design to that of the KCI Pediatric RotoRest P-30 product (510(k) No. K822402), automatically turns the patient 40° (maximum) on one side of the body, to lying on his/her back, to 40° (maximum) on the opposite side of the body. The side support cushions, which automatically inflate when Modified PediKair is in use, and the bed's side rails keep the patient from falling out of the bed while in a rotated position, or while being rotated. 510(k) Premarket Notification • Modified PediKair Summary & Certification • 2-1 {1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. William H. Quirk Director, Regulatory Affairs Kinetic Concepts, Incorporated P.O. Box 659508 San Antonio, Texas 78265 Re: K954320 Trade Name: Modified Pedikair Regulatory Class: II Product Code: FNM Dated: June 12, 1997 Received: June 13, 1997 JUL 18 1997 Dear Mr. Quirk: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2} Page 2 - Mr. Quirk This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), promotion, or advertising, please contact the Office of Compliance, Promotion and Advertising Policy Staff (HFZ-302) at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597. Sincerely yours, Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...