QBC ACCUTUBE(423406)/QBC AUTOREAD(424571)

K960827 · Becton Dickinson Diagnostic Instrument Systems · GKZ · May 17, 1996 · Hematology

Device Facts

Record IDK960827
Device NameQBC ACCUTUBE(423406)/QBC AUTOREAD(424571)
ApplicantBecton Dickinson Diagnostic Instrument Systems
Product CodeGKZ · Hematology
Decision DateMay 17, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

The QBC® AUTOREAD™ System provides a diagnostic hematology profile on venous or capillary blood: hematocrit, hemoglobin, MCHC, platelet count, white blood cell count, granulocyte count (% and number), and lymphocyte/monocyte count (% and number).

Device Story

System performs automated hematology analysis using QBC AccuTube disposables. Input: venous or capillary blood sample collected in AccuTube; sample centrifuged. Analyzer processes spun tube to measure hematocrit, hemoglobin, MCHC, platelet count, WBC count, granulocyte count, and lymphocyte/monocyte count. Used in clinical settings; operated by healthcare personnel. Output: diagnostic hematology report provided via printer/display. System flags out-of-range values. Benefits: provides rapid, automated blood profile analysis for clinical decision-making.

Clinical Evidence

Bench testing only. Study compared modified AccuTube/AUTOREAD system against original AccuTube/AUTOREAD and microhematocrit/impedance methods using 160 fresh EDTA whole blood samples. Hematocrit correlation (r=0.993 vs microhematocrit; r=0.999 vs predicate). Hemoglobin correlation (r=0.995 vs impedance; r=0.999 vs predicate). Other parameters (platelets, WBC, granulocytes, lymph/mono) showed high correlation (r=0.967-0.989) against predicate.

Technological Characteristics

System includes analyzer, centrifuge, printer, and QBC AccuTube disposables. AccuTubes contain reagent coatings. Sensing principle involves optical analysis of centrifuged blood columns. Connectivity includes barcode identification. Software uses calculation algorithms for hematology parameters.

Indications for Use

Indicated for patients requiring a diagnostic hematology profile from venous or capillary blood samples. No specific age or gender contraindications stated.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} MAY 17 1996 K960827 Becton Dickinson Primary Care Diagnostics 7 Loveton Circle Sparks, Maryland 21152-0370 Telephone (410) 316-4000 # BECTON DICKINSON 510(K) SUMMARY SUBMITTED BY: BECTON DICKINSON PRIMARY CARE DIAGNOSTICS BECTON DICKINSON AND COMPANY 7 LOVETON CIRCLE SPARKS, MD 21152 CONTACT: Rocky Ganske, Director of Quality Management and Regulatory Compliance TELEPHONE: (410) 316-4713 PREPARED: February 21, 1996 TRADE NAME: QBC® AUTOREAD™ System COMMON NAME: Automated Differential Cell Counter CLASSIFICATION NAME: Automated Differential Cell Counter PREDICATE DEVICE: QBC® AUTOREAD™ System (K910834) QBC® AccuTube for QBC® AUTOREAD™ (K953340) INTENDED USE: The QBC® AUTOREAD™ System provides a diagnostic hematology profile on venous or capillary blood: hematocrit, hemoglobin, MCHC, platelet count, white blood cell count, granulocyte count (% and number), and lymphocyte/monocyte count (% and number). DEVICE DESCRIPTION: The QBC® AUTOREAD™ System includes an analyzer, power pack, printer, centrifuge, and test accessories. The system uses specially designed tubes known as the QBC® AccuTube. AccuTubes are filled with blood and centrifuged. The spun tube is placed in the QBC® AUTOREAD™ analyzer which provides a report. DEVICE TECHNOLOGICAL CHARACTERISTICS: Tables 1 summarizes the similarities to and differences from currently marketed and/or cleared devices. {1} TABLE 1: SUMMARY OF SIMILARITIES TO AND DIFFERENCES FROM CURRENTLY MARKETED AND/OR CLEARED DEVICES | PRODUCT ATTRIBUTES | PRODUCT NAME | | | | --- | --- | --- | --- | | | QBC AUTOREAD System K910834 | QBC AccuTube for AUTOREAD System K953340 | QBC AUTOREAD System (as configured and presented in this submission) | | SYSTEM HARDWARE | QBC AUTOREAD | QBC AUTOREAD | QBC AUTOREAD | | TUBE TYPES | 2 tube types: venous and capillary | 1 tube type: QBC AccuTube [or] 2 tube types: venous and capillary | 1 tube type: QBC AccuTube (modified) [or] 2 tube types: venous and capillary | | TUBE FILLING MECHANISM FOR VENOUS BLOOD SAMPLES | QBC Pipetter | QBC Pipetter | QBC Pipetter [or] AccuTube filled via capillary action | | TUBE FILLING MECHANISM FOR CAPILLARY (FINGER STICK) BLOOD SAMPLES | Capillary action | Capillary action | Capillary action | | SYSTEM HEMATOLOGY PARAMETERS | White Blood Cell Granulocyte Lymph/Mono Platelet Hematocrit Hemoglobin MCHC | White Blood Cell Granulocyte Lymph/Mono Platelet Hematocrit Hemoglobin MCHC | White Blood Cell Granulocyte Lymph/Mono Platelet Hematocrit Hemoglobin MCHC | | FIBRINOGEN TESTING CAPABILITIES | Available | Available | Available | | BAR CODE/POSITIVE PATIENT IDENTIFICATION | Not available | Not available | Available for QBC AccuTube | | HEMATOLOGY DIAGNOSTICS REPORTS | Available | Available | Available | {2} TABLE 1 (Continued): SUMMARY OF SIMILARITIES TO AND DIFFERENCES FROM CURRENTLY MARKETED AND/OR CLEARED DEVICES | PRODUCT ATTRIBUTES | PRODUCT NAME | | | | --- | --- | --- | --- | | | QBC AUTOREAD System K910834 | QBC AccuTube for AUTOREAD System K953340 | QBC AUTOREAD System (as configured and presented in this submission) | | OUT OF RANGE VALUES | Values outside the validated operating range are shown blinking on the AUTOREAD display; values are flagged as out of range on the sample printout | Values outside the validated operating range are shown blinking on the AUTOREAD display; values are flagged as out of range on the sample printout | Values outside the validated operating range are shown blinking on the AUTOREAD display; values are flagged as out of range on the sample printout | | MODE FEATURES/SELECTION FOR CONTROL, PROFICIENCY TEST, CHECK ROD, AND PATIENT SPECIMEN TESTING | Modes manually selected by operator 6 modes | Modes manually selected by operator 5 modes | Modes manually selected by operator 5 modes | {3} # SUMMARY OF DEVICE TESTING: ## ACCURACY STUDIES Accuracy data for the QBC AUTOREAD System is derived from K953340; performance of the QBC AUTOREAD System with the modified QBC AccuTube disposable is equivalent to that of the QBC AUTOREAD System with the original QBC AccuTube disposable. The hardware utilized by both systems to read the AccuTube is identical. The disposable is identical with the exception of movement of the reagent coating to facilitate filling of the AccuTube via capillary action and incorporation of a fluorescent stripe on the outside of the tube for positive tube type identification. Calculation algorithms are identical with the exception of adjustments made to the hematocrit and hemoglobin algorithms to compensate for movement of the reagent coating within the AccuTube. A study was performed to demonstrate equivalency between the relocated reagent coating AccuTube/AUTOREAD System and the original AccuTube/AUTOREAD System in terms of hematocrit and hemoglobin values. The study was conducted at two hospital sites with a total of 160 fresh EDTA whole blood samples. Each sample was prepared in duplicate QBC AccuTubes with relocated reagent coating, duplicate QBC AccuTubes with original reagent coating location and duplicate microhematocrit tubes. A scatter plot comparing the relocated QBC AccuTube read with the modified QBC AUTOREAD algorithm and the spun microhematocrit method was performed. Regression statistics are summarized below: | HEMATOCRIT STUDY | | | | --- | --- | --- | | CORRELATION COEFFICIENT | SLOPE | INTERCEPT | | 0.993 | 0.991 | 0.270 | An additional scatter plot comparing the relocated QBC AccuTube read with the modified QBC AUTOREAD algorithm and the original QBC AccuTube read with the original QBC AUTOREAD algorithm was performed. Regression statistics are summarized below: | HEMATOCRIT STUDY | | | | --- | --- | --- | | CORRELATION COEFFICIENT | SLOPE | INTERCEPT | | 0.999 | 0.975 | 1.015 | This study demonstrates that QBC AccuTubes with relocated reagent coating read with a QBC AUTOREAD System provide equivalent results when compared to the spun microhematocrit method and to the original QBC AccuTube read with the original QBC AUTOREAD algorithm. {4} Impedance values for hemoglobin were collected from the hospitals and compared to QBC AccuTubes and non-relocated QBC AccuTubes. A scatter plot comparing the relocated QBC AccuTube read with the modified QBC AUTOREAD System data and the impedance system data was performed. Regression statistics are summarized below: | HEMOGLOBIN STUDY | | | | --- | --- | --- | | CORRELATION COEFFICIENT | SLOPE | INTERCEPT | | 0.995 | 0.969 | 0.410 | An additional scatter plot comparing the relocated QBC AccuTube read with the modified QBC AUTOREAD algorithm and the original QBC AccuTube read with the original QBC AUTOREAD algorithm was performed. Regression statistics are summarized below: | HEMOGLOBIN STUDY | | | | --- | --- | --- | | CORRELATION COEFFICIENT | SLOPE | INTERCEPT | | 0.999 | 0.981 | 0.239 | These studies demonstrate that QBC AccuTubes with relocated reagent coating read with a QBC AUTOREAD System provide equivalent results when compared to the impedance cell counter method and to the original QBC AccuTube read with the original QBC AUTOREAD algorithm. Scatter plots were prepared for all other hematological parameters comparing the relocated QBC AccuTube read with the modified QBC AUTOREAD algorithm and the original QBC AccuTube read with the original QBC AUTOREAD algorithm. Regression statistics are summarized below: | PARAMETER | CORRELATION COEFFICIENT | SLOPE | INTERCEPT | | --- | --- | --- | --- | | PLATELETS | 0.989 | 0.932 | 11.633 | | WHITE BLOOD CELLS | 0.984 | 1.031 | -0.116 | | GRANULOCYTES (#) | 0.985 | 1.045 | 0.002 | | LYMPHOCYTES/MONOCYTES (#) | 0.967 | 0.921 | 0.080 | {5} # PRECISION STUDY Accuracy data for the QBC AUTOREAD System is derived from K953340; performance of the QBC AUTOREAD System with the modified QBC AccuTube disposable is equivalent to that of the QBC AUTOREAD System with the original QBC AccuTube disposable. # CONCLUSION: Evaluation of the QBC® AUTOREAD™ System has demonstrated its equivalency to the predicate device in terms of accuracy and precision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...