Quantra QPlus System

K201513 · Hemosonics, LLC · QFR · Aug 21, 2020 · Hematology

Device Facts

Record IDK201513
Device NameQuantra QPlus System
ApplicantHemosonics, LLC
Product CodeQFR · Hematology
Decision DateAug 21, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 864.5430
Device ClassClass 2

Indications for Use

The Quantra® QPlus® System is composed of the Quantra Hemostasis Analyzer, QPlus Cartridge, and Quantra Quality Controls Level 1 and 2. The Quantra QPlus System is intended for in vitro diagnostic use. The Quantra Hemostasis Analyzer uses Sonic Estimation of Elasticity via Resonance (SEER) Sonorheometry, an ultrasound-based technology, to measure the shear modulus of whole blood during coagulation. The QPlus Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation state of a 3.2% citrated venous whole blood sample. The QPlus Cartridge includes tests to assess coagulation characteristics via the intrinsic pathway, via the extrinsic pathway, and includes tests with a heparin neutralizer. The system is intended to be used by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood by means of the following functional parameters: Clot Time (CT), Clot Time with Heparinase (CTH), Clot Stiffness (CS), Fibrinogen Contribution to Clot Stiffness (FCS), Platelet Contribution to Clot Stiffness (PCS) and Clot Time Ratio (CTR). The Quantra QPlus System is indicated for the evaluation of blood coagulation in perioperative patients age 18 years and older to assess possible hypocoagulable and hypercoagulable conditions in cardiovascular or major orthopedic surgeries before, during, and following the procedure. Results obtained with the Quantra QPlus System should not be the sole basis for patient diagnosis.

Device Story

Quantra QPlus System; modification adds Quantra Desktop Remote Viewer (QDRV) accessory. System analyzes whole blood viscoelastic properties; provides coagulation status. QDRV allows remote viewing of results from locations separate from analyzer. Used in clinical settings; operated by healthcare professionals. Output facilitates clinical decision-making regarding hemostasis management.

Clinical Evidence

No clinical data; substantial equivalence based on design control activities, risk analysis, and verification/validation of software modification.

Technological Characteristics

Viscoelastic coagulation analyzer; software-based remote viewing accessory (QDRV). Fundamental scientific technology unchanged from predicate. Connectivity: networked remote viewing capability.

Indications for Use

Indicated for use as a coagulation system for the measurement of whole blood viscoelastic properties.

Regulatory Classification

Identification

A coagulation system for the measurement of whole blood viscoelastic properties in perioperative patients is an in vitro diagnostic device used to evaluate blood coagulation, fibrinolysis, or both, in perioperative patients, as an aid in the assessment of coagulopathies when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.

Special Controls

The special controls for this device are:

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include detailed documentation of, and results from, the following: (i) A study assessing precision using protocols determined to be acceptable by FDA, to cover the measurement range for each reported parameter (test output). Testing must include native specimens with coagulation profiles representative of the intended use population. In order to cover the measuring range, testing may include a limited number of contrived specimens, not to exceed 10 to 20 percent, or as otherwise deemed appropriate by FDA. The contrived specimens must be prepared to resemble clinical specimens. This testing must evaluate repeatability and reproducibility and provide assessments of within-run, within-day, between-run, between-day, between-reagent lot, between-instrument, between-site, and between-operator precision, as applicable to the system; (ii) Studies that demonstrate the performance of each parameter (test output) throughout the claimed measurement range, to include linearity studies or dose-response studies, as applicable to the parameter (test output); (iii) Potential interferent study that includes evaluation of hemolyzed and lipemic samples as potential interferents; exogenous and endogenous interferents associated with each patient population intended for use with the device, and which might be expected to affect assay performance, must be evaluated; and potential interferents that are specific for, or related to, the technology or methodology of the device. Evaluation of all potential interferents must be performed using a protocol determined to be acceptable to the FDA ( *e.g.,* an FDA-recognized standard) and include both normal and abnormal specimens covering coagulation profiles representative of the intended use population;(iv) A study that evaluates specimen stability under the intended conditions for specimen collection, handling, and storage, using samples that cover the coagulation profiles representative of the intended use population, and using protocols determined to be acceptable by FDA; (v) A multisite clinical study, determined to be acceptable by FDA, demonstrating performance, relative to clinically relevant and clinically validated laboratory test(s) for each parameter (test output). Further, the study must meet all of the following criteria: (A) The study must be performed in the intended use population and include representation from all patient populations for whom the device is intended to be used. Potential endogenous and exogenous interferents for each target patient population must be evaluated or known prior to the study; (B) The study must be conducted at a minimum of three external sites representative of the intended use setting by the intended operators; (C) Test samples must be collected at time intervals relevant to the device's use in the intended use population; (D) Clinical specimens, which cover coagulation profiles representative of the intended use population, must be evaluated at each of the three clinical sites in the study; (E) Analysis of the concordance of clinical interpretation of patient coagulation status made from individual test parameter (test output) results as compared to clinical interpretation of coagulation status from a clinically relevant laboratory test or tests ( *e.g.,* a comparative viscoelastic device or standard laboratory tests) must be conducted; and(F) Expected (reference) values for each parameter (test output) must be demonstrated by testing a statistically appropriate number of samples from apparently healthy normal individuals; (vi) For a device with a user interface that has information that needs to be interpreted by the user in correctly using the device to achieve the intended test results or a device that does not provide a final output that is a comprehensive interpretation of all parameter (test output) results, a study evaluating the ability of device users to correctly interpret results; (vii) For any device indicated to guide blood product use, a clinical outcome study determined to be acceptable by FDA that specifically validates the device's indicated use in guiding blood product use; and (viii) For any device indicated to guide use of medication, a clinical outcome study determined to be acceptable by FDA that specifically validates the device's indicated use in guiding use of medication. (2) The labeling required under § 809.10(b) of this chapter must include the following: (i) A summary of results from the study required by paragraph (b)(1)(i) of this section, including repeatability, reproducibility, and assessments of within-run, within-day, between-run, between-day, between-reagent lot, between-instrument, between-site, and between-operator precision, as applicable to the system. (ii) The claimed measurement range of each parameter (test output), as supported by demonstrated performance of the parameter (test output) throughout the claimed measurement range, including studies required by paragraphs (b)(1)(i) through (iii) and (v) of this section, and, if applicable, paragraphs (b)(1)(vii) and (viii) of this section. (iii) Identification of known interferents, including all endogenous, exogenous, technology-specific, and patient population-specific interferents, specific to each parameter (test output). The information must include the concentration(s) or level(s) at which interference was found to occur and the concentration range or levels at which interference was not found to occur. (iv) Information regarding the multisite clinical study required by paragraph (b)(1)(v) of this section, including: (A) Each patient population evaluated; (B) Each intended use setting and the operators; (C) A summary of the results, including the concordance analysis to clinically relevant laboratory test(s); and (D) Demonstrated expected (reference) values for each parameter (test output). (3) The labeling required under § 809.10 of this chapter must include the following: (i) A limiting statement that the result(s) from the device is(are) not intended to be used as the sole basis for a patient diagnosis. (ii) Unless appropriate clinical outcome studies are done in accordance with paragraph (b)(1)(vii) of this section that specifically validate an indication for the device's use in guiding blood product use, a limiting statement that the device has not been evaluated to guide blood product use. (iii) Unless appropriate clinical outcome studies are done in accordance with paragraph (b)(1)(viii) of this section that specifically validate an indication for the device's use in guiding use of medication, a limiting statement that the device has not been evaluated to guide use of medication.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K201513 B Applicant HemoSonics, LLC C Proprietary and Established Names Quantra QPlus System D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QFR | Class II | 21 CFR 864.5430 - Coagulation system for the measurement of whole blood viscoelastic properties | HE - Hematology | ## II Review Summary: This 510(k) submission contains information/data on modifications made to the submitter's own CLASS II device requiring 510(k). The following items are present and acceptable. 1. The name and submission number of the SUBMITTER'S previously authorized device (DEN180017). 2. Submitter's statement that the INDICATIONS FOR USE/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} 3. A description of the device **MODIFICATION(S)**, including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the **FUNDAMENTAL SCIENTIFIC TECHNOLOGY** of the modified device has not changed. This change was for implementing a software modification to the Quantra Hemostasis Analyzer that allows an optional accessory, Quantra Desktop Remote Viewer (QDRV), for viewing results from a location separate from the analyzer. 4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously authorized (or their preamendment) device. K201513 - Page 2 of 2
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...