The TEG 6s System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a blood sample. The TEG 6s System records the kinetic changes in a venous sample of 3.2% citrated whole blood as the sample clots, and retracts in real time. The system output consists of a table of numerical values for parameters R, K, Angle, MA, and FLEV. The TEG 6s System provides specific blood modifiers, in the form of reagents dried-in-place within TEG 6s cartridges. Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests. The indication for use for the TEG 6s System is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure.
Device Story
TEG Manager 2.0.0; optional software accessory for TEG 6s Hemostasis System. Inputs: clinical data from TEG 6s analyzers; patient info from Hospital Information System (HIS). Function: remote viewing of patient tracings/test results; administration of connected TEG 6s devices. Output: visual display of patient data. Used in clinical environments by healthcare providers. Users cannot manipulate or input clinical data. Benefits: centralized monitoring and management of hemostasis testing across multiple analyzers.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by design control activities, risk management documentation, and verification/validation testing of the software modification.
Technological Characteristics
Software application (TEG Manager 2.0.0). Interfaces with TEG 6s analyzers and HIS. Cybersecurity controls included in risk analysis. Operates as a remote viewing and administration tool for connected hardware.
Indications for Use
Indicated for remote viewing of patient tracing and test results from TEG 6s analyzers and administration of connected TEG 6s devices in clinical settings.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
K243858 — TEG 6s Hemostasis System Citrated: K, KH, RT, FF Assay Cartridge · Haemonetics Corporation · Jan 15, 2025
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification OIR Decision Summary
To: THE FILE
RE: DOCUMENT NUMBER K160502
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable:
1. The name and 510(k) number of the SUBMITTER'S previously cleared device: TEG® 6s Hemostasis System; K140893 and K150041.
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials.
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for the TEG manager 2.0.0, an optional accessory to Haemonetics TEG 6s Hemostasis System. TEG Manager 2.0.0 is an application that provides remote viewing of current and historical patient tracing and test results created by the TEG 6s analyzers, and administration of all connected TEG 6s devices. TEG Manager interfaces with the TEG analyzers to obtain clinical data and retrieves patient information from external Hospital Information System (HIS). Users cannot manipulate the data that is stored and displayed within TEG Manager or input any additional clinical data in the software.
4. Comparison Information (similarities and differences) to the applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, operating principle, hardware and test parameters.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis.
Risk Analysis methods comprise the Device Hazard Analysis (HAs); Risk Management Plan and Report to include Initial Risk Assessment, Residual Risk Level Interpretation and Risk Management Activities as well as Verification and Validation Documentation.
b) Based on the Risk Analysis an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied.
The risk analyses for the TEG Manager were conducted in accordance with ISO 14971: 2007, The analyses concluded that the controls and mitigations are effective at reducing the likelihood of the hazard causes. All remaining risks associated with the identified hazards have been evaluated and deemed to be within acceptable levels in regards to the intended application and use of the device to ensure safety and effectiveness according to the Quality System. Cybersecurity has been included in the risk analysis. It was concluded that the proposed change does not pose any additional risks and there is no impact on the safety or effectiveness of the product.
{1}
The labeling for the modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared device.
2
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.