The IB-Stim is a percutaneous electrical nerve field stimulator (PENFS) system intended to be used in patients 8-21 years of age with functional abdominal pain associated with irritable bowel syndrome (IBS) and functional dyspepsia (FD). The IB-Stim is intended to be used for 120 hours per week, using 1 device per week for 4 consecutive weeks, through application to branches of Cranial Nerves V, VII, IX and X, and the occipital nerves identified by transillumination, as an aid in the reduction of pain when combined with other therapies for IBS and FD. FD patients for whom IB-Stim is indicated have also been shown to experience reduction of nausea symptoms.
Device Story
IB-Stim is a battery-operated, 5-gram, single-use, disposable percutaneous electrical nerve field stimulator (PENFS). Device placed behind patient's ear; connected to stimulation needles on auricle via multi-pin wire harness array (1-1-1-4 configuration). Clinician uses transilluminator to identify nerve bundles (Cranial Nerves V, VII, IX, X, and occipital nerves) for needle implantation. Device transmits low-frequency electric pulses for 120 hours per week over 4 consecutive weeks. Used in clinical settings to aid pain reduction in pediatric/adolescent patients with IBS or FD; also reduces nausea in FD patients. Output is electrical stimulation; healthcare provider uses device to provide adjunctive therapy, potentially improving patient quality of life by reducing abdominal pain and nausea.
Clinical Evidence
Post-hoc analysis of 23 subjects (Rome III criteria for FD) from a prospective study of 115 adolescents with pain-associated DGBIs. Treatment arm (n=13) received active IB-Stim; control arm (n=10) received sham. At 8-12 weeks post-treatment, 7/13 active subjects achieved ≥30% abdominal pain reduction vs 0/10 sham subjects. FD patients also reported reduced nausea. No serious adverse events reported. Supported by NASPGHAN recommendations.
Technological Characteristics
Battery-operated, 5g micro-stimulation appliance. Disposable, single-use. Includes multi-pin wire harness array (1-1-1-4 configuration), sterile needles, transilluminator, and adhesive components. Delivers low-frequency electrical pulses to auricular nerve branches. Sterilization and materials are identical to predicate K241533.
Indications for Use
Indicated for patients 8-21 years of age with functional abdominal pain associated with irritable bowel syndrome (IBS) or functional dyspepsia (FD). Used as an aid in pain reduction when combined with other therapies; also indicated for reduction of nausea symptoms in FD patients.
Regulatory Classification
Identification
A nonimplanted nerve stimulator for functional abdominal pain relief is a device that stimulates nerves remotely from the source of pain with the intent to relieve functional abdominal pain. This generic type of device does not include devices designed to relieve pelvic pain.
Special Controls
In combination with the general controls of the FD&C Act, the non-implanted nerve stimulator for functional abdominal pain relief is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed.
(3) Electrical performance testing of the device and electrodes must be conducted to validate the specified electrical output and duration of stimulation of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Sterility testing of the percutaneous components of the device must be performed.
(6) Shelf life testing must be performed to demonstrate continued sterility, package integrity, and device functionality over the labeled shelf life.
(7) Labeling must include the following:
(i) A detailed summary of the device technical parameters;
(ii) A warning stating that the device is only for use on clean, intact skin;
(iii) Instructions for use, including placement of the device on the patient; and
(iv) A shelf life.
DEN180057 — IB-Stim · Innovative Health Solutions (Ihs), Inc. · Jun 7, 2019
K202940 — First Relief v1 · Dyansys, Inc. · Dec 29, 2020
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 15, 2025
NeurAxis
% Dawn Norman
Partner
MRC Global, LLC
9085 E. Mineral Cir., Suite 110
Centennial, Colorado 80112
Re: K250451
Trade/Device Name: NeurAxis IB-Stim (01-1020)
Regulation Number: 21 CFR 876.5340
Regulation Name: Nonimplanted Nerve Stimulator For Functional Abdominal Pain Relief
Regulatory Class: Class II
Product Code: QHH
Dated: February 13, 2025
Received: February 18, 2025
Dear Dawn Norman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250451 - Dawn Norman
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250451 - Dawn Norman
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
SIVAKAMI VENKATACHALAM -S
for
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
K250451
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250451 | |
| Device Name IB-Stim | |
| Indications for Use (Describe) The IB-Stim is a percutaneous electrical nerve field stimulator (PENFS) system intended to be used in patients 8-21 years of age with functional abdominal pain associated with irritable bowel syndrome (IBS) and functional dyspepsia (FD). The IB-Stim is intended to be used for 120 hours per week, using 1 device per week for 4 consecutive weeks, through application to branches of Cranial Nerves V, VII, IX and X, and the occipital nerves identified by transillumination, as an aid in the reduction of pain when combined with other therapies for IBS and FD. FD patients for whom IB-Stim is indicated have also been shown to experience reduction of nausea symptoms. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
FORM FDA 3881 (8/23)
PEC Publishing Services (301) 443-6740
{4}
K250451
Page 1 of 2
510(k) Summary
IB-Stim
May 14, 2025
Company: NeurAxis, Inc.
11611 N. Meridian St.
Suite 330
Carmel, IN 46032
Primary/Secondary Contact: Dawn N. Norman, MS
Partner MRC Global
9085 E. Mineral Cir., Suite 110
Centennial, CO 80112
Phone: (618) 604-3064
Email: dawn.norman@askmrcglobal.com
Kathy Remsen – Regulatory Consultant
(901) 606-4856
Email: kathy.remsen@askmrcglobal.com
Company Contact: Dr. Thomas J. Carrico, CRO
Trade Name: NeurAxis IB Stim (01-1020)
Common Name: Non-implanted nerve stimulator for functional abdominal pain relief
Classification: Class II
Regulation: 21 CFR 876.5340
Panel: Neurology
Product Code: QHH
Primary Predicate: Neuraxis IB Stim (K241533)
Device Description:
IB-Stim stimulator is a battery-operated micro-stimulation appliance weighing 5 grams designed as a disposable product for a single use. IB-Stim stimulator is placed behind the patient’s ear and connected to stimulation needles on the auricle. IB-Stim stimulator offers regular therapy over several days. The appliance transmits low-frequency electric pulses.
IB-Stim is a device system that consists of a (1) percutaneous electrical nerve field stimulator (PENFS); (2) sterile multi-pin wire harness array; (3) round bandages to fasten the appliance leads, stimulator skin adhesive, alcohol swab, transparent dressing, biohazard bag, and Patient ID card; (4) tweezers; (5) surgical marker; and (6) transilluminator for use in transillumination technique that aids in needle implantation. The wire harness array
Page 1 of 2
{5}
K250451
Page 2 of 2
consists of 4 leads. The 1-1-1-4 configuration consists of three single-needle leads, and one 4-needle array. The transilluminator emits a light for use in the trans-illumination technique to aid the clinician in identifying the nerve bundles in the ear.
## Indications for Use:
The IB-Stim is a percutaneous electrical nerve field stimulator (PENFS) system intended to be used in patients 8-21 years of age with functional abdominal pain associated with irritable bowel syndrome (IBS) and or functional dyspepsia (FD). The IB-Stim is intended to be used for 120 hours per week, using 1 device per week for 4 consecutive weeks, through application to branches of Cranial Nerves V, VII, IX and X, and the occipital nerves identified by transillumination, as an aid in the reduction of pain when combined with other therapies for IBS and FD. FD patients for whom IB-Stim is indicated have also been shown to experience reduction of nausea symptoms.
## Substantial Equivalence:
The subject NeurAxis IB-Stim is substantially equivalent to the predicate device, NeurAxis, Inc. IB-Stim – K241533.
The design, materials, packaging, sterilization, and package label of the subject device are identical to those of the predicate device. The indication for use is being expanded to include functional dyspepsia (FD). Real world evidence including clinical literature support these changes, indicating that they do not raise new questions about device safety and effectiveness.
## Performance Testing:
There are no changes to the device design. Therefore, previously conducted testing in DEN180057 remains applicable and no additional performance testing is required.
## Clinical Data:
A post-hoc analysis of a study was conducted in 23 subjects who met Rome III criteria for FD. Data from those 23 subjects were extracted from a prospective study that enrolled 115 adolescents with pain-associated DGBIs recruited from a tertiary care pediatric gastroenterology clinic. The treatment arm subjects were treated with IB-Stim and control arm subjects had a sham device. At extended follow-up (8-12 weeks after the end of treatment), 7 of 13 subjects treated with active IB-Stim continued to have at least a 30% reduction in abdominal pain compared to zero of 10 subjects receiving sham treatment, consistent with the results observed in the general study population of adolescents with pain associated DGBIs. Additionally, FD patients reported reduced nausea symptoms when comparing the IB-Stim and sham group at extended follow-up. Given the limitations of a post-hoc analysis in a small subgroup of FD subjects, real world evidence including clinical literature supports these findings.
In addition, no serious adverse events were reported for these patients. Recommendations from the North American Society for Pediatric Gastroenterology, Hepatology and Nutrition (NASPGHAN) based on clinical literature and current practice of medicine support use of PENFS for functional abdominal pain.
## Conclusion:
The subject device is determined to be substantially equivalent to the predicate device.
Page 2 of 2
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.