K Number
K202940

Validate with FDA (Live)

Device Name
First Relief v1
Manufacturer
Date Cleared
2020-12-29

(90 days)

Product Code
Regulation Number
876.5340
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The First Relief v1 is a percutaneous electrical nerve field stimulator (PENFS) system intended to be used in patients 11-18 years of age with functional abdominal pain associated with irritable bowel syndrome (IBS). The First Relief v1 is intended to be used for 120 hours per week up to 3 consecutive weeks, through application to Cranial Nerves V, VII, IX and X, and the occipital nerves identified by trans-illumination, as an aid in the reduction of pain when with other therapies for IBS.

Device Description

The First Relief v1 is designed to aid in the reduction of pain when combined with other therapies of IBS in patients 11 - 18 years of age with Functional Pain associated with the Irritable bowel syndrome (IBS) by the method of Cranial electrical stimulation at the auricular stimulation points. The First Relief v1 is a wearable, battery-operated device that is designed to administer Periodical low level electrical pulses to the ear over Five days / 120 hours from the time of activation of the device.

The electrical pulse from the device will be delivered to the branches of Cranial Nerves (V, VII, IX and X) on the ear through a set of wire assembly and Stimulation needles. Three zinc air batteries with 1.4 V each provides the required stimulation energy for a maximum of 120 hours. There are three Stimulation electrodes and one ground electrode - which constitute of a needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves. The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device.

This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the skin.

The stimulation pattern consists of Rectangular pulses with differing inter-pulse intervals.

AI/ML Overview

This FDA 510(k) summary for the "First Relief v1" device does not contain a study that establishes acceptance criteria and then proves the device meets those criteria. Instead, it focuses on demonstrating substantial equivalence to a predicate device ("IB-Stim", DEN180057) through non-clinical testing and comparison of technical characteristics.

Here's a breakdown of the information that is and is not available in the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not present a formal table of acceptance criteria derived from a clinical trial or performance study, nor does it report specific device performance metrics against such criteria. The comparison table provided (pages 4-5) details the technical characteristics of the applicant's device (First Relief v1), their own previous device (Drug Relief), and the predicate device (IB-Stim). It highlights similarities and minor differences in design and specifications, not performance outcomes against predefined acceptance criteria.

2. Sample size used for the test set and data provenance:

  • Sample size: Not applicable, as no clinical test set for performance against acceptance criteria is described. The document mentions "non-clinical testing" and "bench tests" but does not specify sample sizes for these.
  • Data provenance: Not applicable for a clinical test set. The data presented are technical specifications and test results from non-clinical (bench) testing.

3. Number of experts used to establish the ground truth for the test set and their qualifications:

Not applicable, as no clinical test set with human-established ground truth is described.

4. Adjudication method for the test set:

Not applicable, as no clinical test set requiring adjudication is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size:

No MRMC comparative effectiveness study is mentioned. The document focuses on demonstrating physical and technical equivalence to a predicate device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. The "First Relief v1" is a physical medical device (percutaneous electrical nerve field stimulator), not an algorithm or software requiring standalone performance evaluation in the context of AI.

7. The type of ground truth used:

Not applicable, as there is no clinical study that would require a ground truth determination for diagnostic or treatment outcomes. The "ground truth" in this context refers to engineering specifications and performance benchmarks established during non-clinical testing (e.g., electrical output measurements).

8. The sample size for the training set:

Not applicable. This device is not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established:

Not applicable.


Summary of what is presented in the document instead:

The document demonstrates substantial equivalence to the predicate device (IB-Stim) based on:

  • Intended Use: Nearly identical.
  • Technological Characteristics: Very similar, with minor differences outlined in the table (e.g., product dimensions, weight, battery type/capacity, wire assembly type, sterilization method, shelf life).
  • Non-Clinical Performance Testing: "The First Relief v1 has equivalent Performance specifications when compared to the predicate device." This includes functional testing for 120 hours to monitor continuous performance, capturing pulse width, pulse duration, amplitude, and current values.
  • Compliance with Standards: The device complies with a list of IEC and ISO standards related to electrical safety, EMC, biocompatibility, and sterilization.
  • Sterilization Testing: Detailed information on EtO sterilization validation for the needle package.
  • Biocompatibility Testing: Performed per ISO 10993.

The key statement regarding performance is: "The First Relief v1 has equivalent Performance specifications when compared to the predicate device." This is a qualitative statement based on non-clinical (bench) testing, rather than a quantitative comparison against specific acceptance criteria from a clinical study. The FDA's clearance is based on the argument that these similarities and non-clinical test results demonstrate that the new device is as safe and effective as the predicate.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 29, 2020

DyAnsys, Inc. Srini Nageshwar CEO 300 North Bayshore Boulevard San Mateo. California 94401

Re: K202940

Trade/Device Name: First Relief v1 Regulation Number: 21 CFR 876.5340 Regulation Name: Non-implanted nerve stimulator for functional abdominal pain relief Regulatory Class: Class II Product Code: OHH Dated: July 5, 2020 Received: September 30, 2020

Dear Mr. Nageshwar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

First Relief v 1

Indications for Use (Describe)

The First Relief v1 is a percutaneous electrical nerve field stimulator (PENFS) system intended to be used in patients 11-18 years of age with functional abdominal pain associated with irritable bowel syndrome (IBS). The First Relief v1 is intended to be used for 120 hours per week up to 3 consecutive weeks, through application to Cranial Nerves V, VII, IX and X, and the occipital nerves identified by trans-illumination, as an aid in the reduction of pain when with other therapies for IBS.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red symbol to the left of the company name, "DyAnsys," which is written in black. The symbol appears to be a combination of a greater-than sign and a stylized lambda symbol.

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.87 and 21 CFR 807.92.

510(k) Number: K202940

1. Applicant Information:

Date Prepared:
Name:DyAnsys, Inc
Address:300, North Bayshore Boulevard, San Mateo, CA 94401, USA
Contact Person:Srini Nageshwar
Phone Number:408.480.4700
Facsimile Number:(650)556-1621

De vice Information

Classification :Class II Trade Name :First Relief v1 Classification Name :Non-Implanted Nerve Stimulator For Pain Associated With Irritable Bowel Syndrome Product Code :OHHH

2. Predicate Device:

DEN Number: DEN180057 Model Name: IB - Stim Manufacturer: Innovative Health Solutions.Inc.,

3. De vice Description:

The First Relief v1 is designed to aid in the reduction of pain when combined with other therapies of IBS in patients 11 - 18 years of age with Functional Pain associated with the Irritable bowel syndrome (IBS) by the method of Cranial electrical stimulation at the auricular stimulation points. The First Relief v1 is a wearable, battery-operated device that is designed to administer Periodical low level electrical pulses to the ear over Five days / 120 hours from the time of activation of the device.

The electrical pulse from the device will be delivered to the branches of Cranial Nerves (V, VII, IX and X) on the ear through a set of wire assembly and Stimulation needles. Three zinc air batteries with 1.4 V each provides the required stimulation energy for a maximum of 120 hours. There are three Stimulation electrodes and one ground electrode - which constitute of a needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves. The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device.

This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the skin.

The stimulation pattern consists of Rectangular pulses with differing inter-pulse intervals. 510(k) Number: K202940 Confidential

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red symbol resembling a greater-than sign combined with a lambda symbol, followed by the company name "DyAnsys" in black font. The font appears to be sans-serif and is relatively bold.

A 3-pin connector is provided, which is used to check the output voltage of the device once it is activated and before applying to the patient with any one of the voltage measuring devices available in the market with the appropriate regulatory compliance

4. Intended Use :

The First Relief v1 is a Percutaneous Electric Nerve Field Stimulator (PENFS) system intended to be used in patients 11 -18 years of age with functional abdominal pain associated with irritable bowel syndrome (IBS). The First Relief v1 is intended to be used for 120 hours per week up to 3 consecutive weeks, through application to branches of cranial nerves V, VII, IX and X and the occipital nerves identified by transillumination as an aid in the reduction of pain when combined with other therapies for IBS.

5. Comparison to Predicate Device:

The DyAnsys .Inc., First Relief v1 is substantially equivalent to legally marketed predicate device IB-Stim (DEN180057). It was evaluated through Non-Clinical testing.

MODEL NAMEDrug Relief(K173841)First Relief v1(Subject)IB-Stim(DEN180057)
MANUFACTURERDyAnsys.Inc.,DyAnsys.Inc.,Innovative Health Solutions. Inc.,
INTENDED USEThe Drug Relief is apercutaneous nerve fieldstimulator (PNFS) system, thatcan be used as an aid to reducethe symptoms of, opioidwithdrawal, through applicationto branches of cranial nerves V,VI, IX and X, and the occipitalnerves identified bytransillumination.The First Relief v 1 is a PercutaneousElectrical Nerve Field Stimulator(PENFS) systemintended to be usedin patients 11-18 years of age withfunctional abdominal painassociated with the irritable bowelsyndrome (IBS). The First Relief v 1is intended to be used for 120 hoursper week up to 3 consecutive weeks,through application to branches ofCranial nerves V, VI, IX and X theoccipital nerves identified bytransillumination as an aid in thereduction of pain when combinedwith other therapies for IBS.The IB-Stim is a percutaneousElectrical Nerve Field Stimulator(PENFS) systemintended to be usedin patients 11-18 years of age withfunctional abdominal pain associatedwith the irritable bowel syndrome(IBS). The IB-Stim is intended to beused for 120 hours per week up to 3consecutive weeks, throughapplication to branches of CranialNerves V, VI, IX and X and theoccipital nerves identified bytransillumination as an aid in thereduction of pain when combined withother therapies for IBS.
PATIENTPOPULATIONAdultsChildChild
TECHNOLOGICAL CHARACTERISTICS
SHAPERectangleRectangleRectangle
PRODUCTDIMENSION, mm50237 mm50237 mm36167 mm
WEIGHT, Kg6 gm (including battery)6 gm (including battery)5 gm
NEEDLEDIMENSIONS, mm0.42 mm (widthlength)0.42 mm (widthlength)0.52 mm (widthlength)
Wire Assembly4 units of wire with snap-fitring, where 3 nos constitute a4 units of wire with snap-fit ring,where 3 nos constitute a singleFour stainless steel wire with fourstimulation needles each. There are 3

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for DyAnsys. The logo consists of a stylized "A" in red, with a black greater-than sign to the left of it. To the right of the logo is the company name, "DyAnsys," in black sans-serif font.

single assembly for stimulationand the other one separate wirewill act as a ground electrode.assembly for stimulation and theother one separate wire will act as aground electrode.stimulation wires with 4 array ofneedles each and 1 ground wire with 1needle.
WIRE ASSEMBLYTYPEWire assembly withoutstimulation needles is solderedwith the pulse generatorWire assembly without stimulationneedles is soldered with the pulsegeneratorWire assembly is connected with thestimulation needles at one end and hasthe facility to connect with the pulsegenerator. It is sterilized and packedseparately.
VOLTAGEMEASUREMENTFEASIBILITYA 3-pin connector is provided tomeasure the output voltage ofthe device once it is activatedA 3-pin connector is provided tomeasure the output voltage of thedevice once it is activatedNot Feasible
POWER:
FREQUENCY1 Hz to 10 Hz1 Hz to 10 Hz1 Hz to 10 Hz
DUTY CYCLE2 Hours ON/1 minute OFF2 Hours ON/2 Hours OFF2 Hours ON/2 Hours Off
OPERATING TIME,HOURS120 hours (5 days)120 hours (5 days)120 hours (5 days)
WAVEFORMRectangle pulseRectangle PulseRectangle Pulse
(ENERGYSOURCE)BATTERYOPERATIONYesYesYes
BATTERY TYPEZinc Air, Battery P10Zinc Air Battery P10Lithium ion battery, CR1225
BATTERYCAPACITY100 mAh100 mAh50 mAh
VOLTAGE$3*1.4V$$3*1.4V$$1*3V$
PULSE WIDTH0.980 ms0.980 ms0.980 ms
ENVIRONMENT
OPERATINGTEMPERATURE5 °C to 45 °C5 °C to 45 °C5 °C to 45 °C
OPERATINGHUMIDITY40% to 80%40% to 80%20% to 80%
ENVIRONMENTOF USEClinics, Hospital, HomeenvironmentsClinics, Hospital, HomeenvironmentsClinics, Hospital, Home environments
STERILIZAITONEtO SterilizationEtO SterilizationIrradiation (Gamma)
RE-USESingle Use deviceSingle Use DeviceSingle Use Device
SHELF LIFE6 Months6 Months12 Months
PACKAGEThe drug relief device and itsaccessories are packed in ablister. This is then placed in acarton box. Each carton boxhasfive deviceThe First Relief v 1 device and itsaccessories are packed in a blister.This is then placed in a carton boxEach carton boxhas multipledevicesThe IB-Stim and its accessoriespacked in a blister
ACCESSORIES-Drug Relief with the nonactivated batteries- Cover of the Drug Relief- Sterile pack of needles for usewith the Drug Relief- Adhesive to fasten the needlesFirst Relief v1 with the nonactivated batteries- Cover of the First Relief v 1- Sterile pack of needles for use withthe First Relief v 1- Adhesive to fasten the needles- The IB-Stim device with insertedbatteries- Cover of the IB-Stim- Sterile pack of Needles for use withthe IB-Stim- Tegaderm has been used to fasten
- Adhesive for the Drug Reliefdevice- Instruction for use- Adhesive for the First Relief v 1device- Instruction for usethe device- Alcohol Swab-Instruction for use
PLANNING & PURCHASE
WARRANTYNANANA

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized symbol to the left, resembling an abstract letter 'A' with a combination of black and red lines. To the right of the symbol is the company name, 'DyAnsys', written in a clear, sans-serif font, with the letters in black.

Difference Between the Legally Marketed Predicate Device:

  • Difference in Components ●
  1. The energy source (battery) has been replaced with higher capacity for more effective treatment
    1. Device adhesive has been replaced with a biocompatible, stretchable adhesive exclusively used for wearables and the EtO sterilization has been made used for needles.
      These differences does not affect the safety and effectiveness of the device.

6. Performance Testing Summary

The First Relief v1 device and its components are subjected to performance testing to validate the effectiveness of each unit. The final product testing is performed to verify and compare the effectual output along with that of the predicate device. The functional test is performed for 120 hours to monitor the continuous performance. The pulse width, pulse duration, amplitude and current values are captured for the First Relief v1 device. The First Relief v1 has equivalent Performance specifications when compared to the predicate device.

The form factors, material for sterilization and some patient contacting materials of the First Relief v1 are similar to the 510(K) cleared device Drug Relief (K173861).

7. Compliance with Standards

The First Relief v1 complies with the following standards

  1. IEC 60601-1 2. IEC 60601-1-2 3. ISO 10993-1 4. ISO 10993-5 5. ISO 10993-6 6. ISO 10993-7 7. ISO 10993-10 8. ISO 10993-11 9. ISO 11135

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red symbol resembling an arrow or a stylized letter 'A' on the left, followed by the text "DyAnsys" in black. The text is in a simple, sans-serif font, with the 'D' and 'A' capitalized and the rest of the letters in lowercase.

8. Sterilization Testing Summary

The needle package was subjected to Bio-burden test. The Needle packs are exposed to EtO Sterilization to curtail the presence of microorganisms and to achieve the defined sterility assurance level (SAL). During the sterilization validation process the biological indicators are used to ensure the desired sterility assurance level. These Bls were placed at the appropriate location, where the sterilizing conditions are the most difficult to achieve. These needle packages carry a chemical indicator on the rear side which indicates the exposure to EtO. The sterility test performed on the needles indicates that there is no turbidity. The residual risk report carried out on the sterilized needle packs evidenced that the results are inline with standards requirement. All Sterlization testing was performed in accordance with ISO 11135:2014-Sterilization of healthcare products- ethylene oxide, ISO 11140-1:2005(R) 2010 Sterilization of healthcare products- chemical indicators, ISO 10993-7:2008/(R) 2012 Biological evaluation of medical devices- ethylene oxide sterilization residuals. ISO 11737-1:2006/R) 2011 Sterilization of medical devices- Microbiological methods-Part 1: Determination of a population of microorganisms on products, ISO 11737-2: 2009 Sterilization of medical devices- Microbiological methods- Part2: Tests of Sterility performed in the definition, validation and maintenance of a Sterilization process and ISO 11138-2:2006(R) 2010-Sterilization Of Healthcare Products-Biological Indicators- Part2: Biological Indicators for ethylene oxide Sterilization Processes.

9. Non-Clinical Testing Summary

The bench test has been performed and found that the First Relief v1 met all the requirements specifications and standards requirements. The testings includes the following:

    1. MEE testing as per IEC 60601-1
    1. EMI/EMC testing as per IEC 60601-1-2
    1. Biocompatibility testing as per ISO 10993
    1. Performance testing

10.Conclusion

Hence it is concluded that by demonstrating the Performance testing and with the Indications For Use, Environment for Use, Biocompatibility and compliance with the same harmonized standards, the product First Relief v1 is substantially equivalent to the predicate device IB-Stim (DEN180057) that was cleared by FDA.

§ 876.5340 Nonimplanted nerve stimulator for functional abdominal pain relief.

(a)
Identification. A nonimplanted nerve stimulator for functional abdominal pain relief is a device that stimulates nerves remotely from the source of pain with the intent to relieve functional abdominal pain. This generic type of device does not include devices designed to relieve pelvic pain.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed.
(3) Electrical performance testing of the device and electrodes must be conducted to validate the specified electrical output and duration of stimulation of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Sterility testing of the percutaneous components of the device must be performed.
(6) Shelf life testing must be performed to demonstrate continued sterility, package integrity, and device functionality over the labeled shelf life.
(7) Labeling must include the following:
(i) A detailed summary of the device technical parameters;
(ii) A warning stating that the device is only for use on clean, intact skin;
(iii) Instructions for use, including placement of the device on the patient; and
(iv) A shelf life.