K043031 · Serim Research Corp. · NTZ · Jun 13, 2005 · Gastroenterology, Urology
Device Facts
Record ID
K043031
Device Name
SERIM DIALYSATE MONITOR TEST STRIP
Applicant
Serim Research Corp.
Product Code
NTZ · Gastroenterology, Urology
Decision Date
Jun 13, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Indications for Use
The Serim Dialysate Monitor Test Strips provide semi- quantitative indications of pH, bicarbonate ion, and glucose in the final bicarbonate dialysate. SERIM™ Dialysate Monitor™ Test Strips determine whether dialysate concentrates for hemodialysis were diluted properly to give a final dialysate with the proper proportions of bicarbonate and acetic acid/glucose. Significant misdilutions of the acid concentrate will give rise to excessively high and low levels of pH, alkalinity, and glucose in the dialysate The test is intended to supplement current methods of dialysate testing.
Device Story
Firm plastic test strips containing three reagent pads; used to monitor hemodialysis dialysate preparation. Input: final bicarbonate dialysate solution. Principle: reagent pads undergo color changes based on chemical reactions; pH pad measures acidity; bicarbonate pad uses acid-neutralization causing pH-dependent color shift; glucose pad uses glucose oxidase/peroxidase reaction as tracer for acid concentrate dilution. Output: visual color change on strip compared against reference chart. Used in clinical dialysis settings by healthcare staff. Provides semi-quantitative verification of proper proportioning of sodium bicarbonate, acid concentrate, and purified water. Helps detect significant misdilutions (e.g., 35% or greater) to ensure patient safety during dialysis.
Clinical Evidence
Bench testing only. Performance evaluated using samples from a dialysis clinic and intentionally misdiluted samples (2X or 0.5X concentration). Results showed strips successfully detected misdilutions of 35% or more and correlated with known glucose concentrations and conductivity measurements. No clinical patient outcome data presented.
Technological Characteristics
Firm plastic strip with three reagent pads. pH pad: glass electrode comparison. Bicarbonate pad: acid-neutralization/pH indicator system. Glucose pad: glucose oxidase, peroxidase, buffers, and indicator. Standalone, non-electronic, visual-read device.
Indications for Use
Indicated for semi-quantitative monitoring of pH, bicarbonate, and glucose concentrations in final bicarbonate dialysate solutions to verify proper dilution of concentrates during hemodialysis preparation. Intended as a supplement to existing dialysate testing methods.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
K031267 — STERICHEK BICARB PH REAGENT STRIPS · Environmental Test Systems, Inc. · Dec 23, 2003
K101750 — RPC E-Z CHEK 6.8-8.5 PH TEST STRIPS, MODEL K100-0117 AND E-Z CHEK 0-14 PH TEST STRIPS, MODEL K100-0104 · Reprocessing Products Corp · Jan 7, 2011
K253863 — GUARDIAN Quick Dip Residual Chlorine (5212) · Serim Research · Mar 17, 2026
K964264 — HIGH RANGE PEROXIDE TEST STRIP · Integrated Biomedical Technology, Inc. · Jun 16, 1997
K042322 — STERICHEK BLOOD LEAK REAGENT STRIPS · Hach Company · Dec 7, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
CONFIDENTIAL
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
SERIM Research Corporation 510(k) Premarket Notification Serim™ Dialysate Monitor™ Test Strip
3 of 49
# 510 (K) SUMMARY
JUN 1 3 2005
| Prepared: | May 20, 2005 | |
|--------------------------------|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | SERIM Research Corporation | |
| Address | P.O. Box 4002<br>Elkhart IN 46514 | |
| Phone: | 574-264-3440 | |
| Fax: | 574-266-6222 | |
| Contact: | Patricia A. Rupchock<br>Manager of Regulatory Affairs | |
| Device: | Trade/Proprietary Name: | SERIM™ Dialysate Monitor™ Test<br>Strip |
| | Common/ Usual Name: | Test for pH in dialysate<br>Test for glucose as tracer for acid<br>concentrate in dialysate<br>Test for bicarbonate in dialysate |
| Device Classification<br>Name: | Strip, Dialysate, pH, bicarbonate, glucose, acid, indicator | |
| Medical Specialty: | GU | |
| Product Code: | NTZ | |
| Class: | II | |
| Regulation Number: | 876.5820 | |
{1}------------------------------------------------
K043031 page 2 of 4
CONFIDENTIAL
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
SERIM Research Corporation 510(k) Premarket Notification Serim™ Dialysate Monitor™ Test Strip
4 of 49
| Predicate Devices: | SERIM™ Dialysate Monitor™ Test Pad |
|--------------------|------------------------------------|
|--------------------|------------------------------------|
| | Legally marketed device |
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| | Test for pH in dialysate Previously cleared – no significant modifications which affect safety or effectiveness or impacts the indications for use. |
| | K945491 |
| | Test for glucose as tracer for acid concentrate in dialysate |
| | Bayer Multistix® 10SG (glucose pad) K992257 |
| | Test for bicarbonate in dialysate |
| | IBP Gmbh Model HDM97 conductivity meter 510(k) number K020908 |
| Device Description: | SERIM Dialysate Monitor Test Strips are firm plastic strips that contain three reagent areas to test for pH, bicarbonate/carbonate ion, and glucose in dialysate. The glucose concentration is a tracer of the amount of acid concentrate added to the dialysate. |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | The Serim Dialysate Monitor Test Strips provide semi- quantitative indications of pH, bicarbonate ion, and glucose in the final bicarbonate dialysate.<br>SERIM™ Dialysate Monitor™ Test Strips determine whether dialysate concentrates for hemodialysis were diluted properly to give a final dialysate with the proper proportions of bicarbonate and acetic acid/glucose. Significant misdilutions of the acid concentrate will give rise to excessively high and low levels of pH, alkalinity, and glucose in the dialysate The test is intended to supplement current methods of dialysate testing. |
| Technological Characteristics: | The characteristics of each pad on the test strip will be compared with its individual predicate device. The pH strip is identical to that which has been previously cleared by the FDA (K945491). In this submission, the pH readings of the strip were compared to that of pH meters using glass electrodes.<br>The bicarbonate test pad is based on the neutralization of a weak |
{2}------------------------------------------------
CONFIDENTIAL
### SERIM Research Corporation 510(k) Premarket Notification Serim™ Dialysate Monitor™ Test Strip
#### 5 of 49
acid incorporated into the test pad. This process will cause a change in the ratio of acid and salt form of the acid that in turn leads to a pH change in the buffer system. This pH change produces a visual change in the color of a pH indicator in the pad. This color change can be correlated with the concentration of bicarbonate originally present in the sample. The predicate device is a conductivity meter. These are routinely used in dialysis as a measurement of the conductivity of the final dialysate solution. We have specifically used as the predicate device, an IBP HDM97 conductivity meter, (510(k) number K020908).
Conductivity is a function of the concentration of all ionic material in the test solution and of the mobility of the individual ions, with hydrogen ion being especially mobile. As the concentration of bicarbonate in the final dialysate changes due to misdilution, the conductivity of the solution correlates with changing bicarbonate concentration. Therefore, we expect and we have found that the color of the test strip also correlates with bicarbonate concentration. Since both the conductivity and the test strip colors correlate with total ionic concentrations it was expected and was found experimentally that the test strips correlated well with the conductivity of the final dialysate assuming that the concentration of other ions remains about constant.
Generally, glucose (a.k.a. dextrose) is added routinely to dialysis solutions for all patients in quantities sufficient to insure that, when properly diluted, the resulting glucose concentration should be 200 mg/dl. (0.2%). The constant level of glucose in these solutions is included by the manufacturer with the acetic acid concentrate. Therefore, the glucose can act as a tracer for the acid concentrate.
The glucose test strip used in this test is similar in composition to the predicate device. It consists of glucose oxidase, peroxidase, buffers, and an indicator. We compared the performance of the Serim glucose test strip against the performance of the glucose pad on the Bayer MultiStix 10 SG test strip using standard glucose solutions prepared in dialysate. For simplicity, the glucose pad on the Bayer MultiStix 10 SG test strip will be referred to as the Bayer glucose strip. Each test strip separately correlated well with known glucose concentrations. Therefore, it was expected and found that the Serim test strip correlated well
{3}------------------------------------------------
SERIM Research Corporation
510(k) Premarket Notification
Serim™ Dialysate Monitor™ Test Strip
`K043031` page 4 of 4
CONFIDENTIAL
6 of 49
with the predicate Bayer test strip.
| Performance: | The performance of the SERIM Dialysate Monitor Test Strips<br>was studied using samples obtained from a dialysis clinic. In<br>addition, normal and intentionally misdiluted samples were<br>prepared. Significantly misdiluted samples had intentional<br>dilution errors of 2X or 0.5X the correct amount of bicarbonate<br>concentrate. All normal samples gave the expected values as read<br>by the test strips. Significantly misdiluted samples were readily<br>detected by the strips and also provided indications of which<br>component, acid or bicarbonate concentrate was added in the<br>improper amount. We have shown that when solutions were<br>misdiluted by 35% or more, there was a good probability of<br>detecting this misdilution by means of the test strip. Each of the<br>pads on the test strips was substantially equivalent to the results<br>of the corresponding predicate devices. |
|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | These studies indicate that SERIM Dialysate Monitor Test Strips<br>can be used to monitor the proportioning of sodium bicarbonate,<br>acid concentrate and purified water to determine that the final mix<br>is appropriately prepared. These strips are to be used to<br>supplement current dialysate testing regimens. |
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
Public Health Service
JUN 1 3 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Patricia Rupchock Manager of Regulatory Affairs Serim Research Corporation 23565 Reedy Drive ELKHART IN 46514
Re: K043031
Trade/Device Name: SERIM™ Dialysate™ Monitor " Test Strip Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Codes: NTZ Dated: January 19, 2005 Received: January 21, 2005
Dear Ms. Rupchock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
SERIM Research Corporation 510(k) Premarket Notification Serim™ Dialysate Monitor™ Test Strip
Ko43031 CONFIDENTIAL
11 of 49
## INDICATIONS FOR USE
510(k) Number (if known): K043031
Device Name: Serim™ Dialysate Monitor™ Test Strip
### Indications for Use:
The Serim™ Dialysate Monitor™ Test Strip provides semi-quantitative indications of pH, bicarbonate, and glucose (which acts as a tracer for the acid concentrate) concentrations in final bicarbonate dialysate solutions. They can be used to monitor the proper preparation of such dialysate solutions. The Serim™ Dialysate Monitor™ Test Strips are to be used to supplement current methods for testing dialysate.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
V
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
David K. Heyron
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device 510(k) Number_
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.