Gold Probe Bipolar Electrohemostasis Catheter is indicated for use in transendoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract. Injection Gold Probe Bipolar Electrohemostasis Catheter is indicated for use in endoscopic injection therapy (to deliver pharmacological injection agents, such as vasoconstrictors) and endoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract. The Injection Gold Probe Catheter also has irrigation capability. Any other use is not recommended.
Device Story
Bipolar electrohemostasis catheters for GI tract bleeding; used transendoscopically. Gold-spiral electrode tip delivers bipolar current for tissue cauterization and blood coagulation. Injection variant includes 25-gauge needle for pharmacological agent delivery. Both models feature irrigation capability via saline/sterile water. Catheter shaft provides structural firmness for perpendicular/tangential application; HemoGlide coating reduces friction during endoscope passage. Operated by physicians in clinical settings; connects to Endostat bipolar electrosurgical generators or other compatible generators via adapter. Output is thermal energy for hemostasis; aids in managing GI bleeding sites.
Clinical Evidence
No clinical data; bench testing only (based on substantial equivalence to predicate devices with unchanged technological characteristics).
Technological Characteristics
Bipolar electrohemostasis catheter; gold-spiral electrode tip; HemoGlide coating. Available in 7 Fr and 10 Fr diameters; lengths 210cm, 300cm, 350cm. Irrigation port for saline/water. Injection model includes 25-gauge needle. Connects to bipolar electrosurgical generators. No software or electronic algorithms.
Indications for Use
Indicated for patients requiring transendoscopic electrohemostasis of actual or potential bleeding sites in the gastrointestinal tract; Injection Gold Probe also indicated for endoscopic injection therapy of pharmacological agents.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
K961349 — INJECTION GOLD PROBE · Boston Scientific Corp · May 15, 1996
K092571 — GYRUS ACMI BIOCOAG PROBE · Gyrus Acmi, Inc. · Oct 20, 2010
K031026 — ETI BIPOLAR HEMOSTATIC PROBE · Endo-Therapeutics, Inc. · Jun 27, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 28, 2023
Boston Scientific Lingling Guo Senior Regulatory Affairs Specialist 100 Boston Scientific Way Marborough, Massachusetts 01752
Re: K232633
Trade/Device Name: Gold Probe Bipolar Electrohemostasis Catheter; Injection Gold Probe Bipolar Electrohemostasis Catheter Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit And Accessories Regulatory Class: Class II Product Code: KNS Dated: August 29, 2023 Received: August 30, 2023
Dear Lingling Guo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Sivakami Venkatachalam -S
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K232633
Device Name
Gold Probe Bipolar Electrohemostasis Catheter; Injection Gold Probe Bipolar Electrohemostasis Catheter
Indications for Use (Describe)
Gold Probe Bipolar Electrohemostasis Catheter is indicated for use in transendoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract.
Injection Gold Probe Bipolar Electrohemostasis Catheter is indicated for use in endoscopic injection therapy (to deliver pharmacological injection agents, such as vasoconstrictors) and endoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract. The Injection Gold Probe Catheter also has irrigation capability. Any other use is not recommended.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b>☑</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K232633 510(k) Summary
#### 1. Submitter
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752
| Primary Contact: | Lingling Guo |
|--------------------|-----------------------------------|
| | Sr. Regulatory Affairs Specialist |
| Telephone: | (508) 382-0456 |
| E-mail: | lingling.guo@bsci.com |
| | |
| Secondary Contact: | Karin Desjardins |
| | Director, Regulatory Affairs |
| Telephone: | 508-683-0236 |
| E-mail: | Karin.Desjardins@bsci.com |
Date Prepared: November 20, 2023
## 2. Proposed Devices
| Trade Name: | Gold Probe Bipolar Electrohemostasis Catheter |
|----------------------------|---------------------------------------------------------|
| Common Name: | Endoscopic electrosurgical unit and accessories |
| Product Code: | KNS |
| Device Class and Panel: | Class II, Gastroenterology/Urology |
| Classification Regulation: | 21 CFR 876.4300 |
| Trade Name: | Injection Gold Probe Bipolar Electrohemostasis Catheter |
| Common Name: | Endoscopic electrosurgical unit and accessories |
| Product Code: | KNS |
| Device Class and Panel: | Class II, Gastroenterology/Urology |
| Classification Regulation: | 21 CFR 876.4300 |
## 3. Predicate Devices (Primary predicate is listed first)
| Trade Name: | Gold Probe Bipolar Electrohemostasis Catheter |
|----------------------------|-------------------------------------------------|
| Clearance Number: | K970278 |
| Common Name: | Endoscopic electrosurgical unit and accessories |
| Product Code: | KNS |
| Device Class and Panel: | Class II, Gastroenterology/Urology |
| Classification Regulation: | 21 CFR 876.4300 |
{5}------------------------------------------------
| Trade Name: | Injection Gold Probe Bipolar Electrohemostasis Catheter |
|----------------------------|---------------------------------------------------------|
| Clearance Number: | K133933 |
| Common Name: | Endoscopic electrosurgical unit and accessories |
| Product Code: | KNS |
| Device Class and Panel: | Class II, Gastroenterology/Urology |
| Classification Regulation: | 21 CFR 876.4300 |
| Trade Name: | Gyrus BiCoAG Hemostasis Probe |
| Clearance Number: | K123319 |
| Common Name: | Endoscopic electrosurgical unit and accessories |
| Product Code: | KNS |
| Device Class and Panel: | Class II, Gastroenterology/Urology |
| Classification Regulation: | 21 CFR 876.4300 |
#### 4. Device Description
The Gold Probe is a bipolar electrohemostasis catheter with irrigation capabilities. When passed through an endoscope and activated, it will deliver a bipolar current to cauterize tissue. The Gold Probe is available in 7 Fr and 10 Fr with a working length of 300cm (7 and 10 Fr) and 350cm (7 Fr), compatible with a minimum of 2.8 mm and 3.7 mm scope channels, respectively. The catheter shaft provides firmness for perpendicular and tangential use, while the HemoGlide Coating on the catheter promotes minimal friction during endoscope passage.
- 1. Irrigation
Irrigation with saline or sterile water is performed by connecting the irrigation hub to the irrigation port of the bipolar generator or to a 30 cm3 (30 cc) or 60 cm3 (60 cc) luer-lock syringe or mechanical irrigation system for generators without irrigation capability.
- 2. Bipolar Electrohemostasis
The gold-spiral electrode on the probe tip provides bipolar electrical continuity, necessary hemostasis depths, and minimum tissue adhesion. The Gold Probe connects directly to the Endostat family of Bipolar Electrosurgical Generators. Or utilize the Bipolar Cable Adapter supplied by Boston Scientific for other bipolar electrosurgical generators.
The Injection Gold Probe is an injection therapy and bipolar electrohemostasis catheter with irrigation capabilities. The Injection Gold Probe is available in a working length of 210 cm with outer diameters of 7F (2.3 mm) and 10F (3.3 mm), compatible with a minimum of 2.8 mm and 3.7 mm scope channels, respectively. The catheter shaft provides firmness for perpendicular and tangential use, while the HemoGlide Coating on the catheter promotes minimal friction during endoscope passage.
{6}------------------------------------------------
- 1. Injection Therapy
Each Injection Gold Probe contains a 25-gauge injection needle for injection therapy. The catheter handle contains the injection needle hub. Injection is performed by attaching a luer-lock syringe to the "Injection" hub and advancing the needle.
- 2. Bipolar Electrohemostasis
The gold-spiral electrode on the probe tip provides bipolar electrohemostasis. The Injection Gold Probe connects directly to the Endostat family of Bipolar Electrosurgical Generators. Or utilize the Bipolar Cable Adapter supplied by Boston Scientific for other bipolar electrosurgical generators.
- 3. Irrigation
Irrigation with saline or sterile water is performed by connecting the irrigation hub to the irrigation port of the bipolar generator or to a 30 cm3 (30 cc) or 60 cm3 (60 cc) luer-lock syringe or mechanical irrigation system for generators without irrigation capability.
# 5. Indications for Use
Gold Probe Bipolar Electrohemostasis Catheter is indicated for use in transendoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract.
Injection Gold Probe Bipolar Electrohemostasis Catheter is indicated for use in endoscopic injection therapy (to deliver pharmacological injection agents, such as vasoconstrictors) and endoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract. The Injection Gold Probe Catheter also has irrigation capability. Any other use is not recommended.
# 6. Technological Characteristics
The technological characteristics of the devices remain unchanged from the predicate devices Gold Probe Bipolar Electrohemostasis Catheter cleared under K970278, Injection Gold Probe Bipolar Electrohemostasis Catheter cleared under K133933 and similar as the predicate Gyrus BiCoAG Hemostasis Probe (K123319).
# 7. Performance Data
The proposed Gold Probe Bipolar Electrohemostasis Catheter and Injection Gold Probe Bipolar Electrohemostasis Cathter technological characteristics remain unchanged from predicate Gold Probe Bipolar Electrohemostasis Catheter (K970278) and Injection Gold Probe Bipolar Electrohemostasis Catheter (K133933) respectively; therefore, no further performance testing was required.
{7}------------------------------------------------
#### 8. Conclusion
Boston Scientific Corporation has determined that the proposed Gold Probe Bipolar Electrohemostasis Catheter is substantially equivalent to the predicate Gold Probe Bipolar Electrohemostasis Catheter (K970278) and Gyrus BiCoAG Hemostasis Probe (K123319), and that the proposed Injection Gold Probe Bipolar Electrohemostasis Catheter is substantially equivalent to the predicate Injection Gold Probe Bipolar Electrohemostasis Catheter (K133933) and Gyrus BiCoAG Hemostasis Probe (K123319).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.