K Number
K133933

Validate with FDA (Live)

Date Cleared
2014-04-02

(100 days)

Product Code
Regulation Number
876.4300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Injection Gold ProbeTM Bipolar Electronemostasis Catheter is indicated for use in endoscopic injection therapy (to deliver pharmacological injection agents, such as vasoconstrictors) and endoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract. These sites include the following:

  • Peptic Ulcers
  • Dieulafoy Lesions
  • Mallory-Weiss Tears
  • Bleeding Polyp Stalks
  • Arteriovenous Malformations (AVMs)
  • Angiomata

The Injection Gold Probe Bipolar Electrohemostasis Catheter also has irrigation capability. Any other use is not recommended.

Device Description

The Injection Gold Probe Catheter is an injection therapy and bipolar electrohemostasis catheter with irrigation capabilities. The Injection Gold Probe Catheter is available in a working length of 210 cm with outer diameters of 7F (2.3 mm) and 10F (3.3 mm), compatible with a minimum of 2.8 mm and 3.7 mm scope channels, respectively. The catheter shaft provides firmness for persendicular and tangential use, while the Mediglide coating on the catheter promotes minimal friction during endoscope passage.

I. Injection Therapy
Each Injection Gold Probe Catheter contains a 25 gauge injection needle for injection therapy. The catheter handle contains the injection needle hub. Injection is performed by attaching a luerlock syringe to the "Injection" hub and advancing the needle.

II. Bipolar Electrohemostasis
The gold-spiral electrode on the probe tip provides bipolar electrohemostasis. The Injection Gold Probe Catheter connects directly to the Endostat™ family of Bipolar Electrosurgical Generators. Or utilize the Bipolar Cable Adapter supplied by Boston Scientific for other bipolar electrosurgical generators.

III. Irrigation
Irrigation with saline or sterile water is performed by connecting the irrigation hub to the irrigation port of the bipolar generator or to a 30 cm3 (30 cc) or 60 cm3 (60 cc) luer-lock syringe or mechanical irrigation system for generators without irrigation capability.

AI/ML Overview

This document describes a Special 510(k) Premarket Notification for the Injection Gold Probe™ Bipolar Electrohemostasis Catheter. The submission focuses on demonstrating substantial equivalence to a predicate device (K110887) by addressing material changes made to the proposed device.

Here's an analysis of the acceptance criteria and supporting study information:

Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Biocompatibility: Device materials are biocompatible."All biocompatibility tests conducted on the proposed Injection Gold Probe™ Bipolar Electrohemostasis Catheter passed. Therefore, the Injection Gold ProbeTM Bipolar Electrohemostasis Catheter is considered biocompatible."
Bench Testing - Required specifications for the following tests:"All device bench tests results were acceptable. The data demonstrate that the Injection Gold Probe™ Bipolar Electrohemostasis Catheter meets design specifications and is suitable for its intended use."
- Combination Therapy DeviceMet required specifications.
- Device Irrigation Flow RateMet required specifications.
- Injection Needle Flow RateMet required specifications.
- Injection Needle Extension and RetractionMet required specifications.
- Injection Needle Retraction LockMet required specifications.
- Injection Hub to 22-gauge Hypotube Joint Tensile StrengthMet required specifications.
- 25-gauge Injection Needle to the 22-Gauge Hypotube Joint Tensile StrengthMet required specifications.
- Needle SpecificationMet required specifications.

Study Information

  1. Sample size used for the test set and the data provenance:

    • The document does not specify sample sizes for the individual bench tests. It only states that tests were performed.
    • The data provenance is not explicitly mentioned (e.g., country of origin). However, as this is a 510(k) submission to the FDA (USA), the testing would typically be conducted under US or internationally recognized standards. The studies appear to be prospective in nature, designed to evaluate the modified device against predetermined specifications.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This document describes engineering bench testing and biocompatibility testing, not clinical studies involving expert assessment of images or patient outcomes. Therefore, no "experts" in the context of clinical ground truth establishment were utilized or mentioned. The "ground truth" for these tests would be the established engineering specifications and biocompatibility standards.
  3. Adjudication method for the test set:

    • Not applicable as this is not a study requiring adjudication of expert opinions (e.g., image interpretation). Bench tests and biocompatibility tests are evaluated against predefined, objective pass/fail criteria.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This is a submission for a medical device (a catheter) and not an AI/software device that assists human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical medical device, not an algorithm.
  6. The type of ground truth used:

    • For biocompatibility: Adherence to established biocompatibility standards (e.g., ISO 10993 series).
    • For bench testing: Established engineering design specifications and performance requirements for the device's functions (e.g., flow rates, tensile strength, extension/retraction mechanisms).
  7. The sample size for the training set:

    • Not applicable. This is not a machine learning or AI-based device that requires a training set.
  8. How the ground truth for the training set was established:

    • Not applicable for the same reason as above.

{0}------------------------------------------------

APR 0 2 2014

K133933 Page 1 of 3

SECTION 6 510(k) SUMMARY

00000.9.5

510(k) SUMMARY

1, Submitter:

Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752

Contact: Ashley Pyle Regulatory Affairs Manager Telephone: 508-683-4359 Fax: 508-683-5939

Secondary Contact: Jennifer Kimball Director of Regulatory Affairs Telephone: 508-683-4178 Fax: 508-683-5939

Date Prepared: December 20, 2013

2. Proposed Device:

Trade Name: Injection Gold ProbeTM Bipolar Electrohemostasis Catheter Classification Name: Unit, Electrosurgical, Endoscopic (with accessories) Regulation Number: 876.4300 Product Code: KNS Classification: Class II

3. Predicate Device:

Trade Name: Injection Gold ProbeTM Binolar Electrohemostasis Catheter Manufacturer and Clearance Number: Boston Scientific Corporation, K 110887 Classification Name: Unit, Electrosurgical, Endoscopic (with accessories) Regulation Number: 876.4300 Product Codc: KNS Classification: Class II

4. Proposed Device Description:

The Injection Gold Probe Catheter is an injection therapy and bipolar electrohemostasis catheter with irrigation capabilities. The Injection Gold Probe Catheter is available in a working length of 210 cm with outer diameters of 7F (2.3 mm) and 10F (3.3 mm), compatible with a minimum of 2.8 mm and 3.7 mm scope channels, respectively. The catheter shaft provides firmness for persendicular and tangential use, while the Mediglide coating on the catheter promotes minimal friction during endoscope passage.

Special 510(k) Premarket Notification, Injection Gold ProbeTM Bipolar Electrohemostasis Catheter

{1}------------------------------------------------

  • I. Injection Therapy
    Each Injection Gold Probe Catheter contains a 25 gauge injection needle for injection therapy. The catheter handle contains the injection needle hub. Injection is performed by attaching a luerlock syringe to the "Injection" hub and advancing the needle.

  • II. Bipolar Electrohemostasis
    The gold-spiral electrode on the probe tip provides bipolar electrohemostasis. The Injection Gold Probe Catheter connects directly to the Endostat™ family of Bipolar Electrosurgical Generators. Or utilize the Bipolar Cable Adapter supplied by Boston Scientific for other bipolar electrosurgical generators.

  • III. Irrigation
    Irrigation with saline or sterile water is performed by connecting the irrigation hub to the irrigation port of the bipolar generator or to a 30 cm3 (30 cc) or 60 cm3 (60 cc) luer-lock syringe or mechanical irrigation system for generators without irrigation capability.

5. Intended Use/Indications for Use:

The Injection Gold ProbeTM Bipolar Electronemostasis Catheter is indicated for use in endoscopic injection therapy (to deliver pharmacological injection agents, such as vasoconstrictors) and endoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract. These sites include the following:

  • Peptic Ulcers �
  • o Dieulafoy Lesions
  • o Mallory-Weiss Tears
  • 0 Bleeding Polyp Stalks
  • Arteriovenous Malformations (AVMs) o
  • Angiomata �

The Injection Gold Probe Bipolar Electrohemostasis Catheter also has irrigation capability. Any other use is not recommended.

6. Description of Change for this Submission:

The following changes are being made to the proposed Injection Gold Probe Bipolar Electrohemostasis Catheter:

  • The current PTF : coating on the Injection Cold Probe (IGP) Needle Hub to Hypotube � Assemblies, is changing to a slightly modified PTFE coating.
  • The current chemical weld used to attach the hypotube to injection hub, is being replaced o with an adhesive bond.

7. Technological Characteristics:

The proposed Injection Gold Probe™ Bipolar Electrohemostasis Catheter has the same fundamental design, operating principal and intended use as the predicate Injection Gold Probe™ Binolar Electrohemostasis Catheter (K110887). The proposed device has material differences when compared to the currently marketed Injection Gold Probe™ Bipolar Electrohemostasis Catheter (K110887).

8, Performance Data:

Biocompatibility Testing has been performed to demonstrate that the material changes made to the proposed device are biocompatible.

0000116

Special 510(k) Premarket Notification, Injection Gold Probe131 Bijolar Electrohemostasis Catheter

{2}------------------------------------------------

Bench Testing has been performed on the proposed Injection Gold ProbeTM Bipolar Electrohemostasis Cather to demonstrate that the modified device meets the required specifications of the completed tests. The following bench tests were performed:

  • o Combination Therapy Device
  • o Device Irrigation Flow Rate
  • 0 Injection Needle Flow Rate
  • o Injection Needle Extension and Retraction
  • Injection Needle Retraction Lock 0
  • o Injection Hub to 22-gauge Hypotube Joint Tensile Strength
  • 25-gauge Injection Needle to the 22-Gauge Hypotube Joint Tensile Strength 0
  • o Needle Specification

9. Conclusion:

All biocompatibility tests conducted on the proposed Injection Gold Probe™ Bipolar Electrohemostasis Catheter passed. Therefore, the Injection Gold ProbeTM Bipolar Electrohemostasis Catheter is considered biocompatible.

All device bench tests results were acceptable. The data demonstrate that the Injection Gold Probetal Bipolar Electrohemostasis Catheter meets design specifications and is suitable for its intended use.

Boston Scientific Corporation has demonstrated that the proposed Injection Gold Probe™ Bipolar Electrohemostasis Catheter is substantially equivalent to Boston Scientific Corporation's currently marketed Injection Gold Probe™ Bipolar Electrohemostasis (K110887).

000017

Special 510(k) Premarket Notification, Injection Cold Probe131 Bipolar Electrohemostasis Catheter

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo for the Department of Health & Human Services - USA. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 2, 2014

Boston Scientific Corporation Ashley Santos Regulatory Affairs Manager 100 Boston Scientific Way Marlborough, MA 01752

Re: K133933

Trade/Device Name: Injection Gold Probe™ Bipolar Electrohemostasis Catheter Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: KNS Dated: February 28, 2014 Received: March 4, 2014

Dear Ashley Santos,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to connine. that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic Act (110) that to nevice, subject to the general controls provisions of the Act. The r ou may, therefore, mans of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability addition: "Fease note: "Obter ver, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elaboried (controls. Existing major regulations affecting your device can be may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please oe advised that I Dri 3 issuance of a sudenine complies with other requirements of the Act that I DA has made a and regulations administered by other Federal agencies.

{4}------------------------------------------------

Page 2 - Ashley Santos

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/4/Picture/6 description: The image shows the text "Benjamin Fisher -S". The words are written in a bold, sans-serif font. The letters "DA" are stylized with a geometric pattern. The text is black against a white background.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

SECTION 5 INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

Device Name:

Indications For Use:

To Be Determined K133933

Injection Gold ProbeTB1 Bipolar Electrohemostasis Catheter

The Injection Gold ProbeTM Bipolar Electrohemostasis Catheter is indicated for use in endoscopic injection therapy (to deliver pharmacological injection agents, such as vasoconstrictors) and endoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or notential bleeding sites in the gastrointestinal tract. These sites include the following:

  • Peptic Ulcers �
  • ◆ Dieulafoy Lesions
  • Mallory-Weiss Tears .
  • Bleeding Polyp Stalks ●
  • Arteriovenous Malformations (AVMs) .
  • Angiomata ●

The Injection Gold Probe Bipolar Blectrohemostasis Catheter also has irrigation capability. Any other use is not recommended.

Prescription Use X (Part 21 CPR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NBEDED)

Concurrence of CDRH, Office of Device Bvaluation (ODB)

Beniami 2014.04. 8 -04'00'

00014

Special 510(k) Premarket Nollfication, Injection Gold Probota Electrohenrostasis Catheter

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).