BARD PLAT. CLASS 3 FRENCH URETEROSCOPIC FLAT WIRE STONE, TORQUE HELICAL STONE BASKETS/GRASPING FORCEPS
K961248 · C.R. Bard, Inc. · FFL · May 28, 1996 · Gastroenterology, Urology
Device Facts
Record ID
K961248
Device Name
BARD PLAT. CLASS 3 FRENCH URETEROSCOPIC FLAT WIRE STONE, TORQUE HELICAL STONE BASKETS/GRASPING FORCEPS
Applicant
C.R. Bard, Inc.
Product Code
FFL · Gastroenterology, Urology
Decision Date
May 28, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4680
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Bard® Platinum Class™ Ureteroscopic 3 French Flat Wire Stone Basket is indicated for use in the endoscopic removal of renal and ureteral stones and calculi. Bard® Platinum Class™ Ureteroscopic 3 French High-Torque Helical Stone Basket is indicated for use in the endoscopic removal of renal and ureteral stones and calculi. Bard® Platinum Class™ Ureteroscopic 3 French Grasping Forceps is indicated for use in the endoscopic removal of renal and ureteral stones and calculi.
Device Story
Ureteroscopic instruments (3 Fr. diameter) designed for stone retrieval; includes flat wire basket, helical stone basket, and grasping forceps. Operation: device inserted through endoscope working channel to ureter/renal pelvis; tip advanced past stone; basket/forceps opened, manipulated, and closed around stone; device and endoscope withdrawn with captured stone. Used by urologists in clinical settings. Benefits: minimally invasive stone removal.
Indicated for endoscopic removal of renal and ureteral stones and calculi in patients requiring ureteroscopic intervention.
Regulatory Classification
Identification
A ureteral stone dislodger is a device that consists of a bougie or a catheter with an expandable wire basket near the tip, a special flexible tip, or other special construction. It is inserted through a cystoscope and used to entrap and remove stones from the ureter. This generic type of device includes the metal basket and the flexible ureteral stone dislodger.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Bard® Platinum Class™ 3 French Flat Wire Stone Basket (K961248)
Bard® Platinum Class™ 3 French High-Torque Helical Stone Basket (K961248)
Bard® Platinum Class™ 3 French Grasping Forceps (K961248)
Related Devices
K012581 — REPROCESSED STONE RETRIEVAL BASKETS · Sterilmed, Inc. · Feb 14, 2002
K970121 — STONE DISLODGER, BASKET · Boston Scientific Corp · Apr 2, 1997
K190492 — Peralta Stone Removal Catheter · Calcula Technologies, Inc. · Jul 18, 2019
K974300 — TELEMED SYSTEMS RETRIEVAL BASKET · Telemed Systems, Inc. · Dec 19, 1997
K170898 — Wittich Nitinol Stone Basket · Cook Incorporated · Dec 12, 2017
Submission Summary (Full Text)
{0}
K961248
MAY 28 1996
0000062
Bard Urological Division
C. R. Bard, Inc.
8195 Industrial Blvd.
Covington, GA 30209-2695
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
## A. Submitter's Information:
Submitter's Name: C. R. Bard, Inc., Urological Division
Address: 8195 Industrial Boulevard
Covington, Georgia 30209
Phone: 770-784-6135
Fax: 770-784-6419
Contact Person: Donna J. Wilson
Date of Preparation: March 29, 1996
## B. Device Names:
Trade Names:
Bard® Platinum Class™ 3 French Flat Wire Stone Basket
Bard® Platinum Class™ 3 French High-Torque Helical Stone Basket
Bard® Platinum Class™ 3 French Grasping Forceps
Common Name: Ureteral Stone Dislodger
Classification Names: Dislodger, stone, basket, ureteral, metal
[Flat Wire Stone Basket, Helical Stone Basket, Grasping Forceps]
## C. Predicate Device Names:
Trade Names:
Bard® Platinum Class™ 3 French Flat Wire Stone Basket
Bard® Platinum Class™ 3 French High-Torque Helical Stone Basket
Bard® Platinum Class™ 3 French Grasping Forceps
## D. Device Description:
The Bard® Platinum Class™ 3 French Ureteroscopic Instrument product line consists of three separate instruments or stone dislodgers: the Flat Wire Stone Basket, the High-Torque Helical Stone Basket, and the Grasping Forceps. These instruments are each constructed of three main assemblies: a handle assembly, shaft assembly, and stone capture mechanism consisting of an expandable basket or grasping forceps assembly. Stone dislodgers (baskets or grasping forceps) may be used in the removal of renal and ureteral stones and calculi by inserting the stone basket or grasping forceps (while in a closed position) through the working channel of an endoscope and into the ureter or renal pelvis. The tip of the basket or forceps is advanced past the stone
{1}
K961248 0000063
and the expandable basket or forceps is opened, manipulated and closed around the stone, thereby capturing it for removal. Once the stone has been captured, the basket or forceps should be partially closed to keep the stone from dislodging while the device and endoscope are being withdrawn.
The devices have a 3 Fr. outer diameter shaft size and each is available in three different lengths, 70cm, 90cm and 115cm.
## E. Intended Use:
Bard® Platinum Class™ Ureteroscopic 3 French Flat Wire Stone Basket is indicated for use in the endoscopic removal of renal and ureteral stones and calculi.
Bard® Platinum Class™ Ureteroscopic 3 French High-Torque Helical Stone Basket is indicated for use in the endoscopic removal of renal and ureteral stones and calculi.
Bard® Platinum Class™ Ureteroscopic 3 French Grasping Forceps is indicated for use in the endoscopic removal of renal and ureteral stones and calculi.
## F. Technical Characteristics:
The Bard® Platinum Class™ 3 French Ureteroscopic Instruments are constructed of biocompatible materials. Most of the materials used in the construction are typical for similar accessories. These include thermoplastics, 300 or 400 series stainless steel, solder, nickel titanium, fluoropolymers, polyimide, cyanoacrylate and epoxy.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.