K Number
K190492

Validate with FDA (Live)

Date Cleared
2019-07-18

(140 days)

Product Code
Regulation Number
876.4680
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Peralta Stone Removal Catheter is intended to be used endoscopically to grasp, manipulate, and remove calculi from the urinary tract, and/or for ureteral dilation.

Device Description

The Peralta Stone Removal Catheter (Peralta) is a 3 French (1 mm) catheter designed for removal of ureteral stones and fragments that are 5 mm or less in diameter in the urinary tract and ureteral dilation. The catheter uses a wire basket to encapsulate the stone, and a noncompliant balloon which can be inverted around the basket to cover the sharp edges of the stone during removal. The balloon can also be used for dilation of the urinary tract.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the Peralta Stone Removal Catheter. It details the device's indications for use, its comparison to predicate devices, and a summary of non-clinical testing. However, it explicitly states "No clinical data were provided in order to demonstrate substantial equivalence."

Therefore, based on the provided document, a study proving the device meets acceptance criteria related to clinical performance or human reader studies does not exist in this submission. The acceptance criteria and "study that proves the device meets the acceptance criteria" would primarily refer to the non-clinical testing performed.

Here's an analysis based on the information available:

1. Table of acceptance criteria and the reported device performance:

Since this is a 510(k) submission primarily relying on non-clinical testing and substantial equivalence, the "acceptance criteria" are derived from the performance standards and tests outlined. The document does not provide specific numerical acceptance criteria for each test, but rather states that the device "conforms to the performance specifications."

Acceptance Criteria (Implied from tests)Reported Device Performance (Summary)
Biocompatibility:
- Cytotoxicity (ISO 10993-5)Compliant
- Irritation & Skin Sensitization (ISOCompliant
10993-10)
Performance Testing:
- Dimensional VerificationPassed (ensuring device aligns with design specifications)
- Functional & Performance TestingPassed
- Usability TestingPassed
- Mechanical TestingPassed
Ex Vivo Testing (Porcine Ureters):
- Simulation of real-world usePassed (successfully simulated stone removal and dilation)
Animal Testing (In Vivo Porcine Model):
- Comparison to currently marketed devicePerformed, results indicate substantial equivalence
(with pre-placed 3mm and 4.3mm stones)
Conformity to FDA Guidance:Conforms to "Checklist for Mechanical Lithotripters and Stone Dislodgers used in Gastroenterology and Urology (November 01, 1994)"

2. Sample size used for the test set and the data provenance:

  • Biocompatibility: Not specified, but standard ISO testing involves specific sample sizes for tests like cytotoxicity and sensitization.
  • Dimensional, Functional, Usability, Mechanical Testing: Not specified, but generally a representative sample of manufactured devices would be tested.
  • Ex Vivo Testing: "Ex vivo testing in porcine ureters was performed." Sample size (number of ureters or experimental runs) is not specified.
  • Animal Testing (In Vivo): "In vivo testing on four ureters in a porcine model." This implies a sample size of four ureters.
  • Data Provenance: The studies were sponsored by Calcula Technologies, Inc., and performed as part of their 510(k) submission. The location of the studies is not explicitly stated in the document, but it is implied to be part of their regulatory compliance process for the US market. The studies are prospective in nature, as they were conducted specifically for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable in the traditional sense for this submission. For a medical device 510(k) that relies on non-clinical testing for substantial equivalence, "ground truth" isn't established by expert consensus on clinical outcomes or images in the same way it would be for an AI/CADe device. The ground truth for the non-clinical tests is established by direct measurement, physical performance, and biological responses according to established protocols.
  • The "experts" involved would be the engineers, scientists, and technicians conducting the performance, mechanical, and biocompatibility tests, following validated methods. Their qualifications are not specified but are implied to meet industry standards for such testing.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • None in the context of expert adjudication of clinical outcomes or images. The "adjudication" for non-clinical tests would involve review and approval of the test protocols and results by internal quality and regulatory personnel.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC study was NOT done. This type of study is relevant for AI-powered diagnostic or CADe devices where human readers' performance is augmented by AI. The Peralta Stone Removal Catheter is a physical medical device, not an AI/software device designed to assist human interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No. This question is not applicable. The device is a physical catheter, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the non-clinical tests, the "ground truth" is based on:
    • Direct measurements (for dimensional verification).
    • Defined physical parameters and failure points (for functional and mechanical testing).
    • Biological responses (for biocompatibility, e.g., cell viability, irritation response).
    • Observation of performance in simulated environments (ex vivo and in vivo animal models demonstrating the ability to grasp, manipulate, and remove stones, and dilate ureters).
  • In the animal study, the ground truth was the pre-placed urinary stones (3mm and 4.3mm) and the observed ability of the device to remove them.

8. The sample size for the training set:

  • Not applicable. This device is not an AI/machine learning model, so there is no "training set."

9. How the ground truth for the training set was established:

  • Not applicable. As there is no training set for an AI model, there is no ground truth established for one.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 18, 2019

Calcula Technologies, Inc % Allison Komiyama Principal Consultant AcKnowledge Regulatory Strategies, LLC 2251 San Diego Avenue, Suite B-257 San Diego, CA 92110

Re: K190492

Trade/Device Name: Peralta Stone Removal Catheter Regulation Number: 21 CFR 876.4680 Regulation Name: Ureteral Stone Dislodger Regulatory Class: Class II Product Code: FFL, EZN Dated: June 7, 2019 Received: June 10, 2019

Dear Allison Komiyama:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Glenn B. Bell, Ph.D. Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K190492

Device Name Peralta Stone Removal Catheter

Indications for Use (Describe)

The Peralta Stone Removal Catheter is intended to be used endoscopically to grasp, manipulate, and remove calculi from the urinary tract, and/or for ureteral dilation.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Calcuia Technologies. The logo features a stylized kidney shape on the left, with the text "Calcuia" in a bold, sans-serif font to the right of the kidney shape. Below "Calcuia" is the word "TECHNOLOGIES" in a smaller, sans-serif font. The image also contains the text "K190492" and "page 1 of 3" at the top right.

510(k) Summary K190492

DATE PREPARED

July 18, 2019

MANUFACTURER AND 510(k) OWNER

Calcula Technologies, Inc. 959 Peralta Ave, San Francisco, CA 94110, USA Telephone: Official Contact: Raymond "Buzz" Bonneau, Chief Executive Officer

REPRESENTATIVE/CONSULTANT

Allison C. Komiyama, Ph.D., R.A.C. Lucie Dalet, Ph.D. AcKnowledge Regulatory Strategies, LLC Telephone: +1 (619) 458-9547 Email: akomiyama@acknowledge-rs.com, Idalet@acknowledge-rs.com Website: www.AcKnowledge-RS.com

DEVICE INFORMATION

Proprietary Name/Trade Name:Peralta Stone Removal Catheter
Common Name:Dislodger, Stone, Basket, Ureteral, Meta
Regulation Description:Ureteral stone dislodger
Regulation Number:21 CFR 876.4680
Regulatory Class:Class II
Product Code:FFL, EZN
Premarket Review:OPEQ/OHT3/DTH3B
Review Panel:Gastroenterology/Urology

PREDICATE DEVICE IDENTIFICATION

The Peralta Stone Removal Catheter is substantially equivalent to the following predicates:

510(k) NumberPredicate Device Name / ManufacturerPrimary Predicate
K831875Balloon Dilation Helical Stone Extractor / VPI
K941853Passport / Boston Scientific Corp
510(k) ExemptZero Tip Nitinol Stone Retrieval Basket / Boston Scientific Corp.Reference Device

The predicate devices have not been subject to a design related recall.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a logo for a company called "Calcula Technologies". The logo features a stylized image of a kidney with a flower-like design in the center. Above the logo, the text "K190492 page 2 of 3" is visible, likely indicating a document or page number.

DEVICE DESCRIPTION

The Peralta Stone Removal Catheter (Peralta) is a 3 French (1 mm) catheter designed for removal of ureteral stones and fragments that are 5 mm or less in diameter in the urinary tract and ureteral dilation. The catheter uses a wire basket to encapsulate the stone, and a noncompliant balloon which can be inverted around the basket to cover the sharp edges of the stone during removal. The balloon can also be used for dilation of the urinary tract.

INDICATIONS FOR USE

The Peralta Stone Removal Catheter is intended to be used endoscopically to grasp, manipulate, and remove calculi from the urinary tract, and/or for ureteral dilation.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

Calcula believes that the Peralta Stone Removal Catheter is substantially equivalent to the predicate devices based on the information summarized here:

The subject device has a similar intended use and a similar design as the predicate device cleared in K831875, and similar dimensions and materials as the device cleared in K941853. The main differences between the primary predicate and subject devices are (1) the device cleared in K831875 features a helical basket made of four stainless steel wires and the subject device features a basket made of four nitinol wires and (2) the balloon of the subject device is able to invert around the captured stone and cover the sharp edges of the stone during removal. The technological characteristics of the subject device are similar in design, dimensions and materials, to the 510(k) exempt Zero Tip Nitinol Stone Retrieval Basket device marketed by Boston Scientific (reference device). The subject device has undergone testing to ensure that the device is substantially equivalent to the predicate and reference devices.

SUMMARY OF NON-CLINICAL TESTING

The Peralta Stone Removal Catheter conforms to the performance specifications outlined in FDA guidance document Checklist for Mechanical Lithotripters and Stone Dislodgers used in Gastroenterology and Urology (Issued November 01, 1994)

The following tests were performed to demonstrate safety based on current industry standards:

Biocompatibility: The subject device was subjected to biocompatibility testing in compliance to ISO 10993-5 Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity and ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for "Calcula Technologies". The logo features a stylized kidney graphic on the left, followed by the word "Calcula" in large, bold font. Below "Calcula" is the word "TECHNOLOGIES" in a smaller font. In the upper right corner of the image, the text "K190492 page 3 of 3" is visible.

Performance: The performance testing of the subject device included dimensional verification, functional and performance testing, usability testing, and mechanical testing. Ex vivo testing in porcine ureters was performed to simulate real world use of the device.

Animal Testing: In vivo testing on four ureters in a porcine model that incorporated preplaced urinary stones (3 mm and 4.3 mm in size) was performed to simulate real world use of the device when compared to a currently marketed device.

The results of these tests indicate that the Peralta Stone Removal Catheter is substantially equivalent to the predicate devices.

SUMMARY OF CLINICAL TESTING

No clinical data were provided in order to demonstrate substantial equivalence.

CONCLUSION

Non-clinical testing (including biocompatibility, dimensional verification, functional and performance testing, usability and human factors testing, and mechanical testing) has been performed to demonstrate that any differences in technological characteristics do not raise new issues of safety or effectiveness. The similar indications for use, technological characteristics, and performance characteristics for the proposed Peralta Stone Removal Catheter are assessed to be substantially equivalent to the predicate devices and the reference device.

§ 876.4680 Ureteral stone dislodger.

(a)
Identification. A ureteral stone dislodger is a device that consists of a bougie or a catheter with an expandable wire basket near the tip, a special flexible tip, or other special construction. It is inserted through a cystoscope and used to entrap and remove stones from the ureter. This generic type of device includes the metal basket and the flexible ureteral stone dislodger.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.