K251127 · Pentax of America, Inc. · PEA · Jun 3, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K251127
Device Name
PENTAX Medical Video Processor (EPK-i8020c)
Applicant
Pentax of America, Inc.
Product Code
PEA · Gastroenterology, Urology
Decision Date
Jun 3, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The PENTAX Medical Video Processor EPK-i8020c is intended to be used with the PENTAX Medical camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal and pulmonary endoscopic diagnosis, treatment and video observation. The PENTAX Medical EPK- i8020c includes a digital post- processing imaging enhancement technology (PENTAX i - ScanTM), and optical imaging enhancement technology (OE). These imaging enhancement technologies are intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i - Scan and OE are compatible with PENTAX Medical video gastrointestinal endoscopes and bronchoscopes.
Device Story
Video processor system (EPK-i8020c) integrates light source, video processing, and monitor connectivity for endoscopic procedures. Inputs: optical signals from PENTAX endoscopes/camera heads. Processing: digital post-processing (i-scan) and optical imaging enhancement (OE) to improve visualization of mucosal structures. Output: enhanced video feed displayed on monitors for physician observation. Used in clinical settings (GI/pulmonary suites) by physicians to assist in diagnosis and treatment. Device features include Light Limit Mode and automatic OE mode cancellation upon endoscope removal. Output aids real-time clinical decision-making during endoscopy; does not replace biopsy/histopathology. Benefits: improved visualization of tissue during diagnostic/therapeutic procedures.
Clinical Evidence
Bench testing only. No clinical data provided. System performance verified through software validation, cybersecurity assessment, and functional testing to confirm equivalence to the predicate device.
Technological Characteristics
Video processor with integrated light source. Software-based imaging enhancement (i-scan, OE). Connectivity: compatible with PENTAX endoscopes/monitors. Software developed per IEC 62304:2006 + A1:2015. Hardware unchanged from predicate.
Indications for Use
Indicated for gastrointestinal and pulmonary endoscopic diagnosis, treatment, and video observation in patients requiring endoscopic procedures. Imaging enhancement technologies (i-Scan, OE) are optional adjuncts to white light endoscopy; not for use as a replacement for histopathological sampling.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
PENTAX Medical Video Processor EPK-i8020c (K231249)
Reference Devices
PENTAX Medical Video Processor EPK-i8020c (K232860)
Related Devices
K150618 — PENTAX Medical EPK-i7010 Video Processor with GI Family · Pentax of America, Inc. · Nov 24, 2015
K232860 — PENTAX Medical Video Processor EPK-i8020c, PENTAX Medical Video Bronchoscope EB11-J10, PENTAX Medical Video Bronchoscope EB15-J10, PENTAX Medical Video Bronchoscope EB19-J10U, PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U · Pentax of America, Inc. · Jan 18, 2024
K191282 — PENTAX Medical EPK-i7010 and PENTAX EPK-i5010 Video Processors with GI Family · Pentax of America, Inc. · Nov 8, 2019
K173679 — PENTAX Medical EPK-i7010 Video Processor with EB Family of Scopes · Pentax of America, Inc. · Jul 3, 2018
K190805 — PENTAX Medical Video Processor EPK-i5500c, PENTAX Medical Video Upper GI Scope EG29-i10c, PENTAX Medical Video Colonoscope EC34-i10cL, PENTAX Medical Video Colonoscope EC38-i10cL · Pentax of America, Inc. · Dec 20, 2019
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 3, 2025
Pentax of America, Inc
Gurvinder Singh Nanda
Senior Director, Regulatory and Quality
3 Paragon Dr
Montvale, New Jersey 07645
Re: K251127
Trade/Device Name: PENTAX Medical Video Processor (EPK-i8020c)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: PEA
Dated: April 30, 2025
Received: May 1, 2025
Dear Gurvinder Singh Nanda:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251127 - Gurvinder Singh Nanda
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251127 - Gurvinder Singh Nanda
Page 3
Sincerely,
**Shanil P. Haugen -S**
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251127 | ? |
| Please provide the device trade name(s). | | ? |
| PENTAX Medical Video Processor (EPK-i8020c) | | |
| Please provide your Indications for Use below. | | ? |
| The PENTAX Medical Video Processor EPK- i8020c is intended to be used with the PENTAX Medical camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal and pulmonary endoscopic diagnosis, treatment and video observation.
The PENTAX Medical EPK- i8020c includes a digital post- processing imaging enhancement technology (PENTAX i - ScanTM), and optical imaging enhancement technology (OE). These imaging enhancement technologies are intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i - Scan and OE are compatible with PENTAX Medical video gastrointestinal endoscopes and bronchoscopes. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
PENTAX Medical Video Processor EPK-i8020c 510(k) Submission
K251127 Pg 1/4
PENTAX MEDICAL
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 221 CFR 807.92. All data included in this document is accurate and complete to the best of PENTAX Medical's knowledge.
# 1. SUBMITTER
Applicant: PENTAX Medical
HOYA Corporation PENTAX Division
3 Paragon Drive
Montvale, New Jersey 07645-1782
Contact: Gurvinder Singh Nanda, PhD
Sr. Director, Regulatory and Quality
PENTAX of America, Inc.
3 Paragon Drive
Montvale, New Jersey 07645-1782
Telephone: 650-722-4189
Email: gurvinder.nanda@pentaxmedical.com
Date Prepared: 04/11/2025
# 2. SUBJECT DEVICE
The objective of this 510(k) is to get clearance for PENTAX Medical Video Processor EPK-i8020c modified for software and instructions for use.
| Device Name | PENTAX Medical Video Processor EPK-i8020c |
| --- | --- |
| Common Name | Gastroscope and Accessories, Flexible/Rigid |
| Classification Name | Endoscope and accessories |
| Regulation No. | 876.1500 |
| Device Class | II |
| Product Code | PEA |
| Classification Panel | Gastroenterology / Urology |
PENTAX Medical CONFIDENTIAL
{5}
PENTAX Medical Video Processor EPK-i8020c 510(k) Submission
K251127 Pg 2/4
PENTAX MEDICAL
# 3. PREDICATE DEVICE and REFERENCE DEVICE
Previously cleared PENTAX Medical Video Processor EPK-i8020c has been chosen as predicate device:
| Subject Device | Predicate Device & Reference device |
| --- | --- |
| PENTAX Medical Video Processor EPK-i8020c | PENTAX Medical Video Processor EPK-i8020c (Predicate: K231249, Reference: K232860) |
The gastrointestinal use (product code: PEA) of the subject device has been cleared in K231249 and pulmonary use (product code: EOQ) has been cleared in K232860.
# 4. DEVICE DESCRIPTION
## PENTAX Medical Video Processor EPK-i8020c
The PENTAX Medical Video Processor EPK-i8020c is intended to be used with PENTAX Medical endoscopes, monitors and other peripheral devices for endoscopic diagnosis, treatment, and video observation. The video processor consists of a video system, integrated light source, monitor, and ancillary equipment.
The EPK-i8020c includes a digital post-processing imaging enhancement technology (PENTAX Medical i-scan™) and an optical imaging enhancement technology (OE). These post-imaging functions are not intended to replace histopathological sampling.
The brand name "INSPIRA™" is provided for the EPK-i8020c video processor and the name is found in the instructions for use (IFU) and/or in the commercial materials such as brochures.
# 5. INTENDED USE AND INDICATIONS FOR USE
## Intended use and Indications for use for PENTAX Medical Video Processor EPKi8020c
The PENTAX Medical Video Processor EPK-i8020c is intended to be used with the PENTAX Medical camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal and pulmonary endoscopic diagnosis, treatment and video observation.
The PENTAX Medical EPK-i8020c includes a digital post-processing imaging enhancement technology (PENTAX i-ScanTM), and optical imaging enhancement
PENTAX Medical CONFIDENTIAL
{6}
PENTAX Medical Video Processor EPK-i8020c 510(k) Submission
K251127 Pg 3/4 PENTAX MEDICAL
technology (OE). These imaging enhancement technologies are intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan and OE are compatible with PENTAX Medical video gastrointestinal endoscopes and bronchoscopes.
# 6. COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE/REFERENCE DEVICES
The subject device is functionally equivalent to the predicate device, and the differences between the two devices are minor technological changes such as the software modification to add the Light Limit Mode feature, and the automatic cancellation of the OE Mode when the endoscope is removed, and related changes to the instructions for use.
The changes to the subject device were evaluated through performance testing in design validation/verification. This testing did not raise any issues regarding the safety and effectiveness of the device, as these differences do not affect the performance, function, or general intended use of the device.
The components of the subject device have the same fundamental technology and principle of operation as the predicate device. Both the subject device and the predicate device are intended for illuminating and viewing the inside of the human body. The components of the subject device consist of the same components as the predicate device, including:
- A video processor
# 7. NON-CLINICAL PERFORMANCE DATA
The subject device has been successfully tested for their functions, performance, and safety. The device is not reprocessed, is not sold sterile does not contain any direct or indirect patient contact components.
The following performance data are provided in support of the substantial equivalence determination.
i. Software and Cybersecurity
Software verification and validation including cybersecurity assessments were conducted according to IEC 62304: 2006 + A1: 2015 and FDA Guidance for Industry and Staff "Content of Premarket Submissions for Device Software Functions.", "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions", and "Postmarket Management of Cybersecurity in Medical Devices".
PENTAX Medical CONFIDENTIAL
{7}
PENTAX Medical Video Processor EPK-i8020c 510(k) Submission
K251127 pg 4/4 PENTAX MEDICAL
ii. Electrical Safety and EMC
Additional Electrical Safety and EMC evaluation are not required, because the hardware of the subject device has not changed since past Premarket Notifications.
iii. System Performance
The system performance of the subject device demonstrated the equivalence to the predicate device.
## Substantial Equivalence Discussion:
After analyzing the intended use, indications for use, technological characteristics (including fundamental operating principle, energy source, scientific technology, functional characteristics, design features, performance characteristics, and constituent materials) and labeling, PENTAX Medical concludes that the subject device is as safe and effective as the predicate devices. There are no differences in indications for use and intended use between the subject and predicate devices and are therefore substantially equivalent. The technological differences in terms of design features, performance characteristics and constituent materials are not significant.
## 8. CONCLUSION
The modified PENTAX Medical Video Processor EPK-i8020c does not pose any unique questions regarding safety and effectiveness. It is deemed to be substantially equivalent to the identified predicate: the PENTAX Medical Video Processor EPK-i8020c (K231249).
PENTAX Medical CONFIDENTIAL
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.