K Number
K191282

Validate with FDA (Live)

Date Cleared
2019-11-08

(179 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PENTAX Medical EPK-i7010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation.

The PENTAX Medical EPK-i7010 includes a digital post-processing imaging enhancement technology (PENTAX i-Scan™) and an optical imaging enhancement technology (OE). These imaging enhancement technologies are intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan and OE are compatible with PENTAX Medical video gastrointestinal endoscopes.

The PENTAX EPK-i5010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation.

The PENTAX EPK-i5010 includes PENTAX i-Scan™, a digital post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan is compatible with PENTAX k-series and i-series gastrointestinal videoscopes and colonovideoscopes.

Device Description

The PENTAX Medical EPK-i7010 Video Processor and PENTAX EPK-i5010 Video Processor each consist of a video system, integrated light source, monitor, and ancillary equipment. These processors are intended for gastrointestinal endoscopic diagnostic, treatment and video observation.

The PENTAX Medical EPK-i7010 Video Processor contains two types of contrast enhancement techniques: PENTAX i-Scan technology, and optical enhancement (OE) technology. The PENTAX EPK-i5010 Video Processor contains only PENTAX i-Scan. The PENTAX i-Scan technology is a digital post-processing image enhancement technique with three modes, i-Scan 1, 2 and 3.

AI/ML Overview

The provided text is a 510(k) summary for the PENTAX Medical EPK-i7010 and EPK-i5010 Video Processors, with a focus on a minor design change related to the Water Bottle Assembly OS-H5. It is NOT a study proving the device meets acceptance criteria for an AI/imaging diagnostic device.

The document discusses performance testing in a general sense, but this testing relates to the physical and functional characteristics of the water bottle accessory and its reprocessing, not to the diagnostic accuracy or clinical performance of the imaging enhancement technologies (i-Scan and OE) with human readers.

Therefore, many of the requested criteria for describing an AI/imaging diagnostic study (e.g., sample size for test set, number of experts for ground truth, MRMC study, effect size) cannot be found in this document because the document does not describe such a study.

The closest relevant information relates to the performance of the Water Bottle Assembly OS-H5.

Here's an attempt to answer the questions based only on the provided text, highlighting where information is absent:


Acceptance Criteria and Study for PENTAX Medical EPK-i7010/EPK-i5010 (Water Bottle Assembly OS-H5)

The provided document (510(k) summary) does not describe a study related to the diagnostic performance of the PENTAX i-Scan or Optical Enhancement (OE) technologies, especially in conjunction with human readers or AI algorithms. Instead, it focuses on demonstrating the substantial equivalence of the PENTAX Medical EPK-i7010 and EPK-i5010 Video Processors, specifically addressing a minor design change: the replacement of a disposable water bottle system with the reusable PENTAX Water Bottle Assembly OS-H5.

The "acceptance criteria" and "study" described pertain to the safety and functional performance of the reusable water bottle accessory and its reprocessing methods, not the diagnostic utility of the imaging processors themselves or any associated AI.

1. Table of Acceptance Criteria and Reported Device Performance (Specific to Water Bottle Assembly OS-H5)

Acceptance Criteria CategorySpecific Criteria (as implied)Reported Device Performance and Confirmation
Reprocessing Validation- Effective soil accumulation prevention (for reusability)- Soil Accumulation Study performed. "All acceptance criteria were satisfied."
- Effective cleaning efficacy- Cleaning Efficacy Study performed. "All acceptance criteria were satisfied."
- Effective rinsing- Rinsing Validation performed. "All acceptance criteria were satisfied."
- Effective sterilization (for reusability)- Sterilization Validation performed. "All acceptance criteria were satisfied."
Biocompatibility- No cytotoxicity- Tested for Cytotoxicity. "toxicity was either not detected or of negligible level." "test results satisfy the acceptance criteria specified." Acceptance criteria conform to ISO 10993-1, 5, and 10 and FDA Guidance Document for mucosal membranes, limited contact duration (< 24 hours).
- No sensitization- Tested for Sensitization. "toxicity was either not detected or of negligible level." "test results satisfy the acceptance criteria specified." Acceptance criteria conform to ISO 10993-1, 5, and 10 and FDA Guidance Document for mucosal membranes, limited contact duration (< 24 hours).
- No intracutaneous reactivity- Tested for Intracutaneous Reactivity. "toxicity was either not detected or of negligible level." "test results satisfy the acceptance criteria specified." Acceptance criteria conform to ISO 10993-1, 5, and 10 and FDA Guidance Document for mucosal membranes, limited contact duration (< 24 hours).
Functional Performance (Water Bottle)- Prevention of backflow of contaminants- Tested for backflow prevention. "The results concluded that acceptance criteria were met, and no backflow occurred in OS-H5 under the worst-case test conditions."
- Equivalent or greater flow rate and air delivery compared to predicate (Endo Smart Cap)- Flow rate and air delivery testing conducted. "The results of the testing confirmed that the performance of OS-H5 is equivalent to or greater than that of the predicate Endo Smart Cap." This demonstrates substantial equivalence for this specific characteristic.

2. Sample Size Used for the Test Set and Data Provenance:

  • This information is not provided in the document for any diagnostic performance study, as such a study is not described.
  • For the water bottle assembly tests (reprocessing, biocompatibility, functional performance), specific sample sizes for units tested are not detailed, but standard validation practices would imply a sufficient number of samples were used to meet test specifications. Data provenance is implied to be from internal lab testing conducted for the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Not applicable/Not provided. This pertained to the water bottle accessory's physical and functional properties, not diagnostic ground truth.

4. Adjudication Method for the Test Set:

  • Not applicable/Not provided.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. An MRMC study was not described. The document pertains to the video processor and its water bottle accessory, not an AI or human-assisted diagnostic system. The "i-Scan" and "OE" are described as "imaging enhancement technologies" and "digital post-processing" or "optical," which are image processing features, not explicitly deep-learning AI algorithms in the context of this submission.
  • No effect size is discussed.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • No. This document does not describe a standalone algorithm performance study.

7. The Type of Ground Truth Used:

  • Not applicable for diagnostic ground truth. For the water bottle evaluation, the "ground truth" was laboratory measurements and validation results conforming to established engineering, cleaning/sterilization, and biocompatibility standards.

8. The Sample Size for the Training Set:

  • Not applicable. No training set for an AI algorithm is mentioned in this submission.

9. How the Ground Truth for the Training Set was Established:

  • Not applicable. No training set or associated ground truth establishment is mentioned.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in large, bold, blue letters, and the word "ADMINISTRATION" in smaller, blue letters below.

November 8, 2019

PENTAX of America, Inc. William Goeller Vice President, Quality/Regulatory Affairs 3 Paragon Drive Montvale, NJ 07645-1782

Re: K191282

Trade/Device Name: PENTAX Medical EPK-i7010 Video Processor with GI family and PENTAX EPK-i5010 Video Processor Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: PEA Dated: October 8, 2019 Received: October 9, 2019

Dear William Goeller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Martha W. Betz, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191282

Device Name

PENTAX Medical EPK-i7010 Video Processor with GI Family and PENTAX EPK-i5010 Video Processor

Indications for Use (Describe)

The PENTAX Medical EPK-i7010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation.

The PENTAX Medical EPK-i7010 includes a digital post-processing imaging enhancement technology (PENTAX i-Scan™) and an optical imaging enhancement technology (OE). These imaging enhancement technologies are intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan and OE are compatible with PENTAX Medical video gastrointestinal endoscopes.

The PENTAX EPK-i5010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation.

The PENTAX EPK-i5010 includes PENTAX i-Scan™, a digital post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan is compatible with PENTAX k-series and i-series gastrointestinal videoscopes and colonovideoscopes.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

PENTAX Medical EPK-i7010 and EPK-i5010 with GI Family of Endoscopes 510(k) Summary

Image /page/3/Picture/1 description: The image shows the Pentax Medical logo. The logo is red and gray. Above the logo is the text "K191282" and "Page 1 of 5".

I. SUBMITTER

PENTAX of America, Inc., HOYA Corporation PENTAX Lifecare Division 3 Paragon Drive Montvale, New Jersey 07645-1782 Phone: 201-571-2318 Ext 2318 Fax: 201-571-2340

Contact: William Goeller Date Prepared: April 26, 2019

II. DEVICE

The purpose of this 510(k) is to obtain clearance of the PENTAX Medical EPK-i7010 Video Processor with GI family and PENTAX EPK-i5010 Video Processor for use with the PENTAX Water Bottle Assembly OS-H5.

This 510(k) captures some minor design changes that have occurred during the evolution of the product line that includes expansion of the product line to include the reusable water bottle accessory. Although the changes are believed to be minor, the 510(k) is being submitted to account for technological advances in associated the compatible devices and to ensure that FDA has the most current information concerning the PENTAX Medical video processors with GI family of scopes.

The subject devices have the same indications for use, composition of patient contact materials, viewing direction, image size, and reprocessing/sterilization method as the predicate.

Device NamesPENTAX Medical EPK-i7010 Video Processor withGI FamilyPENTAX EPK-i5010 Video Processor
Common NameEndoscope, Accessories, Video Post-processing forColor Enhancement
Classification NameEndoscope and accessories
Regulation No.876.1500
Device Class2
Product CodePEA
Classification PanelGastroenterology/ Urology

Table 5.1: Regulatory Classification of PENTAX Medical Video Processors.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the Pentax Medical logo. The logo has the word "PENTAX" in red, with the word "MEDICAL" in gray underneath. The text "K191282 Page 2 of 5" is at the top of the image.

III. PREDICATE DEVICES

The predicate device for this submission, PENTAX Medical EPK-i7010 Video Processor with GI Family (K150618), is virtually identical to the subject device. It's predicate, the PENTAX EPKi5010 Video Processor was the predicate for the EPK-i7010 submission, cleared through K122470, and it will be bundled in this subject devices differ from the predicate due to the replacement of accessory PENTAX Medical Dispo Cap Air Tubing (using a disposable water bottle) with the PENTAX Water Bottle Assembly OS-H5.

IV. DEVICE DESCRIPTION

The PENTAX Medical EPK-i7010 Video Processor and PENTAX EPK-i5010 Video Processor each consist of a video system, integrated light source, monitor, and ancillary equipment. These processors are intended for gastrointestinal endoscopic diagnostic, treatment and video observation.

The PENTAX Medical EPK-i7010 Video Processor contains two types of contrast enhancement techniques: PENTAX i-Scan technology, and optical enhancement (OE) technology. The PENTAX EPK-i5010 Video Processor contains only PENTAX i-Scan. The PENTAX i-Scan technology is a digital post-processing image enhancement technique with three modes, i-Scan 1, 2 and 3.

The table below is a list of accessories and compatible devices that are used with the EPKi7010 and EPK-i5010 Video Processors.

Accessories NameModel Number
PENTAX MedicalCondenser Earth CableOL-Z4
PENTAX Foot SwitchOS-A61
PENTAX Medical KeyboardOS-A83OS-A79
PENTAX White BalanceAdjusterOS-A43H
Table 1: List of Accessories
--------------------------------

P E N T A X M e d i c a l C o n f i d e n t i d e n t i a l

{5}------------------------------------------------

PENTAX Medical EPK-i7010 and EPK-i5010 with GI Family of Endoscopes 510(k) Summary

Image /page/5/Picture/1 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold, uppercase letters. Below "PENTAX" is a horizontal gray line. Below the line, the word "MEDICAL" is in gray, bold, uppercase letters.

K191287

PENTAX Medical WaterBottle AssemblyOS-H5
PENTAX Medical DispoCapAir (MEDIVATORS EndoSmartCap Tubing - K093665)100160P(DispoCap AirTubing)

Table 2: List of Compatible Devices

CompatibleDevicesManufacturerModel nameCompatible with
EPK-i7010EPK-i5010
PrinterSony Business SolutionsCorporationUP-55MDYY
UP-25MDY---
UP-D25MDY---
UP-21MD---Y
UP-21MDA---Y
UP-D23MD---Y
UP-D23MDA---Y
External Hard DriveHD Video RecorderWestern Digital CorporationWDBBGB0030HBKY---
Sony Business SolutionsCorporationHVO-1000MDYY
USB Flash MemoryTranscend Information, Inc.SanDiskTS32GJF600Y---
SDCZ6-1024-A10---Y
LCD Monitor(Any display devices whichsatisfy safety requirementsand other specifications)SafetyRequirementsUL 60601-1CAN/CSA C22.2 No.601.1EN 60601-1

The EPK-i7010 and EPK-i5010 are currently compatible with ENT and GI families of endoscopes.

P E N T A X M e d i c a l C o n f i d e n t i d e n t i d e n t i a l

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the Pentax Medical logo. The word "PENTAX" is in large, red, bold letters. Below "PENTAX" is a horizontal gray line. Below the line, the word "MEDICAL" is in gray, bold letters. The text "K191282 Page 4 of 5" is in the upper left corner of the image.

INDICATIONS FOR USE

The PENTAX Medical EPK-i7010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation.

The PENTAX Medical EPK-i7010 includes a digital post-processing imaging enhancement technology (PENTAX i-Scan™) and an optical imaging enhancement technology (OE). These imaging enhancement technologies are intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan and OE are compatible with PENTAX Medical video gastrointestinal endoscopes.

The PENTAX EPK-i5010 Video Processor is intended to be used with the PENTAX camera heads, endoscopes, light sources, monitors and other ancillary equipment for gastrointestinal endoscopic diagnosis, treatment and video observation.

The PENTAX EPK-i5010 includes PENTAX i-Scan™, a digital post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. i-Scan is compatible with PENTAX k-series and i-series gastrointestinal videoscopes and colonovideoscopes.

COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE

The subject devices are virtually identical to the predicate device. The subject devices have the same fundamental technology and operating principles of the predicate device, including the same intended use and design technological characteristics, such as Insertion Portion, Control Body and optical characteristics. The only substantive difference between the subject and predicate devices is the use of reusable PENTAX Water Bottle Assembly OS-H5 instead of Endo Smart Cap System (K093665). Since the PENTAX Water Bottle Assembly OS-H5 is initially supplied as non-sterile, an IFU for the device is provided with the EPK-i7010 and EPK-i5010 Video Processors. The IFU clearly instructs the user to clean and sterilize the water bottle daily or more frequently, depending on the patient and/or type of endoscopic procedure. However, these additional features did not affect the final product performance and did not raise any questions of safety or effectiveness as demonstrated by the biocompatibility and sterilization validation studies.

PERFORMANCE DATA

Reprocessing Validation

The PENTAX Water Bottle Assembly OS-H5 is a reusable semi-critical device (Spaulding Classification System). Since it is packaged non-sterile, it must be properly cleaned and sterilized BEFORE initial use. The OS-H5 must be subjected to an appropriate cleaning and sterilization process daily or more frequently, depending on the patient and/or type of endoscopic procedure.

The following tests have been performed for the PENTAX Water Bottle Assembly OS-H5:

  • . Soil Accumulation Study
  • Cleaning Efficacy Study

PENTAX Medical Confidential

{7}------------------------------------------------

PENTAX Medical EPK-i7010 and EPK-i5010 with GI Family of Endoscopes 510(k) Summary

Image /page/7/Picture/1 description: The image shows the Pentax Medical logo. The word "PENTAX" is in large, red, bold letters, and the word "MEDICAL" is in smaller, gray, bold letters below it. Above the logo, the text "K191282 Page 5 of 5" is visible.

  • Rinsing Validation
  • Sterilization Validation

All acceptance criteria were satisfied.

Biocompatibility

The biocompatibility of the PENTAX Water Bottle Assembly (OS-H5) (indirect patient contact materials) was confirmed by testing for Cytotoxicity, Sensitization, and Intracutaneous Reactivity. The acceptance criteria for the studies conform to those dealing with surface contact devices that contact mucosal membranes for limited contact duration (< 24 hours), as depicted in ISO 10993-1, 5, and 10, "Biological evaluation of medical devices," and the FDA Guidance Document, "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing." The results of the Cytotoxicity, Sensitization, and Intracutaneous Reactivity tests conducted for PENTAX Water Bottle Assembly OS-H5, including its components/accessories, indicated that toxicity was either not detected or of negligible level. Thus, it is determined that the test results satisfy the acceptance criteria specified.

Performance Testing

The PENTAX Water Bottle Assembly OS-H5 has been tested for backflow prevention. The testing provided evidence that the contaminants from the distal end of the endoscope do not backflow into Water Bottle Assembly OS-H5. The results concluded that acceptance criteria were met, and no backflow occurred in OS-H5 under the worst-case test conditions.

In addition, flow rate and air delivery testing has been conducted with a rigid-walled OS-H5 water bottle to demonstrate its substantial equivalence to the predicate Endo Smart Cap. The results of the testing confirmed that the performance of OS-H5 is equivalent to or greater than that of the predicate Endo Smart Cap.

VIII. CONCLUSION

The subject devices are virtually identical to the predicate device. The intended use of the subject and predicate devices is identical. Any differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness.

The data submitted to support the performance and reprocessing for the PENTAX Water Bottle Assembly OS-H5 demonstrate that the device performs as intended and can be reprocessed in a safe and effective manner. The data provided in this 510(k) Premarket Notification support the equivalence of the subject and predicate devices.

P E N T A X M e d i c a l C o n f i d e n t i d e n t i d e n t i a l

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.